← Product Code [HAO](/productcode/HAO) · K220920

# Matreneu Percutaneous Balloon Compression Kit (K220920)

_Shenzhen Shineyard Medical Device Co. , Ltd. · HAO · Sep 7, 2022 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K220920

## Device Facts

- **Applicant:** Shenzhen Shineyard Medical Device Co. , Ltd.
- **Product Code:** [HAO](/productcode/HAO.md)
- **Decision Date:** Sep 7, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4535
- **Device Class:** Class 1
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The Matreneu® Percutaneous Balloon Compression Kit is intended to the Percutaneous Balloon Compression (PBC) procedure for primary trigeminal neuralgia.

## Device Story

Device consists of balloon catheter, puncture needle, syringe, connector, and pressure gauge. Used in sterile operating room by neurosurgeons. Under X-ray fluoroscopy, micro-balloon is introduced via sheath into meniscus of trigeminal nerve in Meckel's cavity. Contrast agent injected to fill balloon; expansion provides compression injury to nerve fibers to treat trigeminal neuralgia. Pressure gauge monitors injection pressure. Device is single-use and sterile.

## Clinical Evidence

No clinical data. Bench testing only, including biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, hemolysis, pyrogenicity), EO residuals, shelf-life, packaging, and electrical safety/EMC testing.

## Technological Characteristics

Non-powered neurosurgical instrument. Components: balloon catheter, puncture needle, syringe, connector, pressure gauge. Sterilization: ETO (SAL 10^-6). Standards: ISO 11135, ISO 10993-7, ISO 11607-1, ISO 10993-1, AAMI/ANSI ES 60601-1, IEC 60601-1-2.

## Regulatory Identification

A nonpowered neurosurgical instrument is a hand instrument or an accessory to a hand instrument used during neurosurgical procedures to cut, hold, or manipulate tissue. It includes specialized chisels, osteotomes, curettes, dissectors, elevators, forceps, gouges, hooks, surgical knives, rasps, scissors, separators, spatulas, spoons, blades, blade holders, blade breakers, probes, etc.

## Predicate Devices

- MULLAN GANGLION MICROCOMPRESSION SET ([K940973](/device/K940973.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

September 7, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size, also in blue.

Shenzhen Shineyard Medical Device Co. Ltd. % Joyce Yang Consultant Shenzhen Joyantech Consulting Co., Ltd. 1713A, 17th Floor, Block A, Zhongguan Times Square Nanshan District Shenzhen, Guangdong 518100 China

Re: K220920

Trade/Device Name: Matreneu Percutaneous Balloon Compression Kit Regulation Number: 21 CFR 882.4535 Regulation Name: Nonpowered Neurosurgical Instrument Regulatory Class: Class I Product Code: HAO Dated: August 8, 2022 Received: August 8, 2022

Dear Joyce Yang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Adam Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K220920

Device Name

Matreneu Percutaneous Balloon Compression Kit

Indications for Use (Describe)

The Matreneu Percutaneous Balloon Compression Kit is intended to the Percutaneous Balloon Compression (PBC) procedure for primary trigeminal neuralgia.

| Type of Use (Select one or both, as applicable)                                                                                                                                                       |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <input checked="true" type="checkbox"/> For Therapeutic Use in an OTP or Other Setting    </div> <div> <input type="checkbox"/> For Take-Home Use (if allowed by OTP's accreditation)    </div> |

X Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

Date of Summary prepared: September 7, 2022

## 1. Submission Sponsor

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

| Applicant Name | Shenzhen Shineyard Medical Device Co. Ltd.                                                                          |
|----------------|---------------------------------------------------------------------------------------------------------------------|
| Address        | 3F, Changfeng Industrial Block No.3 Liuxian Road,<br>Xin'an Bao'an District, Shenzhen, Guangdong,<br>518000, China. |
| Contact person | Yan Ping                                                                                                            |
| Phone          | +86-755-26912231                                                                                                    |

#### Submission correspondent 2.

| Name           | Shenzhen Joyantech Consulting Co., Ltd                                            |
|----------------|-----------------------------------------------------------------------------------|
| Address        | 1713A, 17th Floor, Block A, Zhongguan Times<br>Square, Nanshan District, Shenzhen |
| Post Code      | 518000                                                                            |
| Phone No.      | +86-755-86069197                                                                  |
| Contact Person | Joyce Yang                                                                        |
| Email          | joyce@cefda.com                                                                   |

#### Device Identification 3.

| Type of 510(k) submission: | Traditional                                    |
|----------------------------|------------------------------------------------|
| Trade Name:                | Matreneu® Percutaneous Balloon Compression Kit |
| Classification name:       | Instrument, Surgical, Non-Powered              |
| Review Panel:              | Neurology                                      |
| Product Code:              | HAO                                            |
| Device Class:              | 1                                              |
| Regulation Number:         | 21 CFR § 882.4535                              |

#### Legally Marketed Predicate Device 4.

| Trade Name        | MULLAN GANGLION MICROCOMPRESSION<br>SET |
|-------------------|-----------------------------------------|
| Regulation number | 21 CFR § 882.4535                       |
| Regulation class  | 1                                       |
| Regulation name   | Nonpowered neurosurgical instrument     |
| 510(k) Number     | K940973                                 |
| Product Code      | HAO                                     |
| Manufacturer      | COOK Incorporated                       |

{4}------------------------------------------------

#### Device Description 5.

The Matreneu® Percutaneous Balloon Compression Kit is intended to be used to the Percutaneous Balloon Compression (PBC) procedure for primary trigeminal neuralgia.

The subject device consist of five parts, such as the balloon catheter, puncture needle, svringe, connector, and pressure gauge. Through a sheath under X-ray fluoroscopy. a micro-balloon is introduced into the meniscus of the trigeminal nerve in Meckel's cavity to assist in retraction and establishment of a pathway. Then inject the contrast agent to fill the balloon to relieve the compression of the nerve fibers that cause trigeminal neuralgia. And through the expansion of the micro-balloon compression injury to treat trigeminal neuralgia.

The Kit is supplied sterile and is intended for single use.

### 6. Intended Use/ Indications for Use

The Matreneu® Percutaneous Balloon Compression Kit is intended to the Percutaneous Balloon Compression (PBC) procedure for primary trigeminal neuralgia.

| Comparison<br>item                           | Subject Device:  Matreneu®<br>Percutaneous Balloon<br>Compression Kit                                                                                                                                                                                                           | Predicate Device:  MULLAN<br>GANGLION<br>MICROCOMPRESSION<br>SET(K940973)                                                                                                                                                                                                                                                                                                                                                                                                                            | Comments               |
|----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Product<br>Code                              | HAO                                                                                                                                                                                                                                                                             | HAO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                   |
| Regulation<br>Number                         | 21 CFR § 882.4535                                                                                                                                                                                                                                                               | 21 CFR § 882.4535                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                   |
| Classificatio<br>n                           | Class I                                                                                                                                                                                                                                                                         | Class I                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                   |
| Type of use                                  | Prescription Use                                                                                                                                                                                                                                                                | Prescription Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                   |
| Intended<br>use &<br>Indication<br>s for Use | The Matreneu<br>Percutaneous Balloon<br>Compression Kit is<br>intended to be used to the<br>Percutaneous Balloon<br>Compression (PBC)<br>procedure for primary<br>trigeminal neuralgia.                                                                                         | The Mullan Percutaneous<br>Trigeminal Ganglion<br>Microcompression Set is<br>intended for use in the<br>percutaneous treatment of<br>trigeminal neuralgia. This<br>set is supplied sterile in a<br>peel - open package and is<br>intended for one - time use.                                                                                                                                                                                                                                        | Same                   |
| Applicable<br>user                           | The subject device is<br>indicated for patients with<br>trigeminal neuralgia treated<br>by compression                                                                                                                                                                          | The subject device is<br>indicated for patients with<br>trigeminal neuralgia treated<br>by compression                                                                                                                                                                                                                                                                                                                                                                                               | Same                   |
| Comparison<br>item                           | Subject Device: Matreneu®<br>Percutaneous Balloon<br>Compression Kit                                                                                                                                                                                                            | Predicate Device: MULLAN<br>GANGLION<br>MICROCOMPRESSION<br>SET(K940973)                                                                                                                                                                                                                                                                                                                                                                                                                             | Comments               |
| neurosurgery.                                | neurosurgery.                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                        |
| Environment<br>of use                        | The subject device<br>expected to be used in a<br>sterile operating room.                                                                                                                                                                                                       | The subject device<br>expected to be used in a<br>sterile operating room.                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                   |
| Single<br>/repeat use                        | Single use                                                                                                                                                                                                                                                                      | Single use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                   |
| Sterile /non-<br>sterile                     | Marketed as a sterile device                                                                                                                                                                                                                                                    | Marketed as a sterile device                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same                   |
| Sterilization<br>method and<br>SAL           | ETO sterile<br>SAL=10-6                                                                                                                                                                                                                                                         | ETO sterile<br>SAL=10-6                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                   |
| Components                                   | Balloon catheter<br>Puncture needle<br>Syringe<br>Connector<br>Pressure gauge                                                                                                                                                                                                   | Microcompression Balloon<br>Catheter<br>Needle with stylet<br>Syringe<br>Scalpel<br>Needle<br>Angled needle gauge<br>Gauze pad<br>Adhesive bandage                                                                                                                                                                                                                                                                                                                                                   | Different<br>(Issue 1) |
| Working<br>principle                         | Through a sheath under X-<br>ray fluoroscopy, a micro-<br>balloon is introduced into<br>the meniscus of the<br>trigeminal nerve in<br>Meckel's cavity to assist in<br>retraction and<br>establishment of a<br>pathway. Then slowly inject<br>the contrast agent to fill the<br> | Through a sheath under X-<br>ray fluoroscopy, a micro-<br>balloon is introduced into<br>the meniscus of the<br>trigeminal nerve in<br>Meckel's cavity to assist in<br>retraction and<br>establishment of a<br>pathway. Then slowly inject<br>the contrast agent to fill the<br>balloon to relieve the<br>compression of the nerve<br>fibers that cause trigeminal<br>neuralgia. And through the<br>expansion of the micro-<br>balloon compression injury,<br>so as to treat trigeminal<br>neuralgia. | Same                   |

### 7. Technological characteristics comparison

{5}------------------------------------------------

Issue 1: The components of the subject device are not exactly the same as the predicate device. Compared with the predicate device, the subject device contains the same key components as the predicate device, such as Balloon Catheter, Puncture needle, and Syringe. The difference is that the predicate device contains some accessories, such as Scalpel, Angled needle gauge,

{6}------------------------------------------------

Gauze pad and Adhesive bandage. However, these accessories do not affect the intended use, performance and clinical operation of the device. In addition, the subject device includes connector and pressure gauge. The connector is used to connect the balloon catheter to the pressure gauge. The pressure gauge is only used to show the injection pressure, not to determine the amount of contrast agent to be injected. The connector and the pressure gauge also do not affect the intended use and performance of the device. Therefore, the differences dot not raise new risk of safety and effectiveness.

### 8. Summary of non-clinical testing

Performance testing was conducted on physical and chemical properties to support the marketing claims and to confirm that the safety and effectiveness of the Matreneu® Percutaneous Balloon Compression Kit is at least equivalent to the predicate device.

The EO residual was measured after sterilization of the device to meet the criteria defined in ISO 11135 Second edition 2014 and ISO 10993-7:2008.

The shelf-life for three years had been validated in real-time aging test and the requirements on packaging for terminally sterilized medical device per ISO 11607-1:2006 are also met. The testing successfully demonstrated essential performance is achieved before and after the shelf life test.

The biocompatibility evaluations were conducted in accordance with the 2020 FDA Guidance document Use of International Standard ISO-10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The cytotoxicity, sensitization, intracutaneous irritation, system toxicity, In Vitro Hemolytic Properties Test, and pyrogen tests were performed to demonstrate the biocompatibility of the device.

Tests on Electromagnetic Compatibility and Electrical Safety were performed in accordance to requirements per AAMI/ANSI ES 60601-1:2005/(R)2012 and A1:2012 and IEC 60601-1-2 Edition 3: 2007-03.

#### 9. Brief discussion of clinical tests

No clinical tests were performed.

### 10. Conclusions

The conclusion drawn from the nonclinical tests demonstrates that the subject device, the Matreneu® Percutaneous Balloon Compression Kit are as safe and effective as the legally marketed predicate device cleared under K940973.

---

**Source:** [https://fda-staging.innolitics.com/device/K220920](https://fda-staging.innolitics.com/device/K220920)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
