← Product Code [FMI](/productcode/FMI) · K220890

# IO Needle Safety Sheath (K220890)

_Actuated Medical, Inc. · FMI · Oct 31, 2022 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K220890

## Device Facts

- **Applicant:** Actuated Medical, Inc.
- **Product Code:** [FMI](/productcode/FMI.md)
- **Decision Date:** Oct 31, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5570
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Pediatric

## Indications for Use

The IO Needle Safety Sheath (IOSS) is intended to assist in the safe removal of intraosseous needles.

## Device Story

IOSS is a single-use, non-sterile accessory designed to prevent accidental needlesticks during the removal of intraosseous (IO) needles. The device is placed around a BD Luer-Lok 10mL syringe attached to an EZ-IO needle prior to removal from the patient. During the removal process, the IOSS irreversibly captures the IO needle; tactile and visual confirmation (via clear device body) indicate the needle is in 'Safe Mode'. Used in pre-hospital, hospital, and acute care settings by clinicians. The device benefits patients and users by reducing the risk of sharps injuries during the extraction of IO access needles.

## Clinical Evidence

Bench testing only. Evidence includes biocompatibility (ISO 10993-5, -10, -11, -12, -23), usability testing (30 users, 500 uses), distribution testing (ISTA 3A 2018), and accelerated shelf-life testing (up to 5 years). Usability testing confirmed the device can be used without needle sticks or serious use errors.

## Technological Characteristics

Material: INEOS K-Resin KR01 styrene butadiene block copolymer. Form factor: Single-use, non-sterile sheath. Mechanism: Mechanical capture of IO needle hub. Biocompatibility: ISO 10993-1 compliant for limited duration surface contact. Shelf life: 5 years.

## Regulatory Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

## Predicate Devices

- EZ-IO Intraosseous Vascular Access System (EZ-IO) ([K180395](/device/K180395.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

October 31, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Actuated Medical, Inc Douglas Dillon Director, Quality Assurance & Regulatory Affairs 310 Rolling Ridge Drive Bellefonte, Pennsylvania 16823

#### Re: K220890

Trade/Device Name: IO Needle Safety Sheath Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: September 28, 2022 Received: September 29, 2022

#### Dear Douglas Dillon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

CAPT Alan M. Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K220890

Device Name IO Needle Safety Sheath (IOSS)

#### Indications for Use (Describe)

The IO Needle Safety Sheath (IOSS) is intended to assist in the safe removal of intraosseous needles.

IOSS is Indicated for Use with 25mm EZ-IO Vascular Access Needles and BD Luer-Lok 10mL Syringes.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

| <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------------------|
| <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)                                     |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Actuated Medical. The logo features a stylized medical cross in orange, followed by the company name in gray and orange. Below the name is the tagline "Innovative motion + Positive outcomes" in gray.

310 Rolling Ridge Drive Bellefonte, PA 16823 + p (814) 355-0003 + f (814) 355–1532

ActuatedMedical.com

#### Certified ...

- + ISO 13485:2016
- + Medical Device Single Audit Program (MDSAP)
- + Women's Business Enterprise (WBE)
- + Women-Owned Small Business (WOSB)

# K 220890 510(k) SUMMARY

# Applicant Information

Date Prepared: October 28, 2022

Name and Address: Actuated Medical, Inc. 320 Rolling Ridge Drive Bellefonte, PA 16823 Ph: (814) 355-0003 Fx: (814) 355-1523

Contact Person: Douglas R. Dillon Director, Quality Assurance and Regulatory Affairs Ph: (814) 355-0003 x107 Fx: (814) 355-1523 Email: Douglas.Dillon@actuatedmedical.com

### Device Information

| Trade Name           | IO Needle Safety Sheath (IOSS)   |
|----------------------|----------------------------------|
| Common Name:         | IO Needle Safety Sheath (IOSS)   |
| Classification:      | 21 C.F.R. §880.5570              |
| Classification Name: | Needle, Hypodermic, Single Lumen |
| Product Code:        | FMI                              |

Image /page/3/Picture/15 description: The image is an abstract design featuring a combination of solid and patterned areas. The left side of the image is a solid orange color, while the right side has a checkerboard pattern in two shades of orange. A wavy line runs across the image, overlapping both the solid and patterned areas.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image is an abstract design featuring shades of orange. The background transitions from a solid, lighter orange on the left to a pattern of darker orange squares on the right. A wavy line design overlays the background, adding a dynamic element to the composition.

### Predicate Device

The legally marketed device to which substantial equivalence is being claimed:

| 510(k)<br>Number | Trade Name                                              | Submitter                                             |
|------------------|---------------------------------------------------------|-------------------------------------------------------|
| K180395          | EZ-IO Intraosseous<br>Vascular Access System<br>(EZ-IO) | Arrow International, Inc.<br>(Subsidiary of Teleflex) |

### Device Description

The IO Needle Safety Sheath (IOSS) is a single-use, non-sterile, antineedlestick accessory to assist in the removal of intraosseous (IO) needles. IOSS is placed around the syringe and IO needle prior to removal from the patient. It irreversibly captures the IO needle within the device upon removal from the patient; helping to prevent accidental needle sticks to the patient or user. There is tactile confirmation of capture, and the clear body of IOSS allows for visual confirmation that the IO needle is in Safe Mode.

# Intended Use

IOSS is intended to assist in the safe removal of intraosseous needles.

### Indications for Use

The IO Needle Safety Sheath (IOSS) is intended to assist in the safe removal of intraosseous needles. IOSS is indicated for use with 25mm and 45mm EZ-10® Vascular Access Needles and BD Luer-Lok™ 10 mL Syringe.

### Technological Characteristics

Because IOSS is only used to remove the parent predicate device (EZ-IO) from the patient, substantial equivalence considerations are primarily focused on the removal of EZ-IO as compared to the removal of EZ-IO with IOSS. The contraindications, target population, and anatomical sites used are identical to the predicate device. Key differences between the predicate and IOSS include needle removal technique, required accessories, sterility, patient contact, and materials of construction.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image is an abstract background with a pattern of orange squares and a wavy line design. The squares are arranged in a grid-like fashion, with alternating shades of orange to create a subtle contrast. The wavy lines are also orange and appear to be overlaid on top of the squares, adding a dynamic element to the design. The overall effect is a visually interesting and textured background.

| Category                                 | Predicate EZ-IO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Proposed IOSS Used as<br>Indicated with Predicate<br>Device                                                                                                                                                                                                                                                                                                                                                                                            | Comparison                                                                                                                                                                              |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use                   | For intraosseous access<br>anytime in which vascular<br>access is difficult to obtain<br>in emergent, urgent, or<br>medically necessary cases<br>for up to 24 hours. Insertion<br>sites: ADULTS (≥22 years<br>old): proximal humerus,<br>proximal tibia, distal tibia.<br>PEDIATRICS (≤21 years<br>old): proximal humerus,<br>proximal tibia, distal tibia,<br>distal femur.<br>Use of the device may be<br>extended for up to 48 hours<br>when alternate intravenous<br>access is not available or<br>reliably established.<br>Insertion sites:<br>ADULTS (≥22 years):<br>proximal humerus, proximal<br>tibia, distal tibia.<br>PEDIATRICS (≥12 years<br>through 21 years old):<br>proximal humerus, proximal<br>tibia, distal tibia, distal<br>femur. | The IO Needle Safety<br>Sheath (IOSS) is intended<br>to assist in the safe removal<br>of intraosseous needles.<br>IOSS is Indicated for Use<br>with 25 mm and 45 mm EZ-<br>IO® Vascular Access<br>Needles and BD Luer-<br>Lok™ 10 mL Syringes.                                                                                                                                                                                                         | Different -<br>Indications for<br>the same<br>intended use<br>and a subset of<br>the available<br>needle sets<br>does not raise<br>new questions<br>of safety or<br>efficacy.           |
| Category                                 | Predicate EZ-IO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Proposed IOSS Used as<br>Indicated with Predicate<br>Device                                                                                                                                                                                                                                                                                                                                                                                            | Comparison                                                                                                                                                                              |
| Needle<br>removal<br>technique           | Remove any extension sets<br>and dressings.<br>Attach a luer-lock syringe to<br>the hub of the catheter.<br>While maintaining<br>alignment of the needle and<br>the syringe, the syringe is<br>rotated clockwise while<br>pulling straight up. Rocking<br>or bending the catheter<br>upon removal should be<br>avoided.                                                                                                                                                                                                                                                                                                                                                                                                                                 | Remove any extension sets<br>and dressings.<br>Attach a BD Luer-Lok 10<br>mL Syringe to the hub of<br>the catheter.<br>Latch IOSS around the<br>syringe. Hold IOSS in place<br>with one hand.<br>While maintaining<br>alignment of the needle and<br>the syringe, the syringe is<br>rotated clockwise while<br>pulling straight up until the<br>EZ-IO hub locks in Safe<br>Mode. Rocking or bending<br>the catheter upon removal<br>should be avoided. | Different -<br>Adding the use<br>of IOSS into<br>the removal<br>process is a<br>difference. It<br>does not raise<br>new questions<br>of safety or<br>efficacy.                          |
| Disposal                                 | Dispose of the catheter with<br>the syringe attached in an<br>approved sharps container.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Dispose of catheter with the<br>syringe attached and<br>covered by IOSS<br>immediately after use in an<br>approved sharps container.                                                                                                                                                                                                                                                                                                                   | Same                                                                                                                                                                                    |
| Required<br>accessories                  | Luer-lock syringe (not<br>included) to use as handle<br>for removal of needle.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | IOSS indicated only for one<br>specific syringe: BD Luer-<br>Lok™ 10 mL Syringe (not<br>included) to use as handle<br>for removal of needle.                                                                                                                                                                                                                                                                                                           | Same                                                                                                                                                                                    |
| Contra-<br>indications                   | · Fracture in target bone.<br>· Previous, significant<br>orthopedic procedure at the<br>site, prosthetic limb or joint.<br>· IO access (or attempted<br>IO access) in targeted bone<br>within past 48 hours.<br>• Infection at the area of<br>insertion.<br>· Excessive tissue (severe<br>obesity) and/or absence of<br>adequate anatomical<br>landmarks.                                                                                                                                                                                                                                                                                                                                                                                               | Unchanged                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                                                                                    |
| Category                                 | Predicate EZ-IO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Proposed IOSS Used as<br>Indicated with Predicate<br>Device                                                                                                                                                                                                                                                                                                                                                                                            | Comparison                                                                                                                                                                              |
| Target<br>population                     | Adult and pediatric patients<br>who are in need of vascular<br>access.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Unchanged                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                                                                                    |
| Where used                               | Pre-hospital<br>In hospital<br>Acute care                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Unchanged                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                                                                                    |
| Anatomical<br>sites used                 | Proximal Humerus<br>Proximal Tibia<br>Distal Tibia<br>Distal Femur (pediatrics<br>only)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Unchanged                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                                                                                    |
| Needle/<br>Cannula<br>design             | Sterile, single use<br>Hubs: colored<br>polycarbonate<br>Stylet/catheter: Stainless<br>Steel<br>Faceted tip<br>Standard luer connection<br>Three lengths:<br>15 mm; 25 mm; 45 mm<br>Size: 15 gauge (0.071", 1.8<br>mm)<br>Needle Cover:<br>Polypropylene                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Device has no<br>needle/cannula.                                                                                                                                                                                                                                                                                                                                                                                                                       | Different -<br>IOSS assists in<br>the safe<br>removal<br>intraosseous<br>needles.                                                                                                       |
| Needle set<br>quidelines                 | Available Needle Sets:<br>15 mm: 3-39 kg<br>25 mm: 3 kg or over<br>45 mm: 40 kg or over                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Indicated for Needle Sets:<br>IOSS only indicated for the<br>25 mm and 45 mm Needle<br>Sets                                                                                                                                                                                                                                                                                                                                                            | Different -<br>IOSS only<br>indicated for<br>the 25 mm and<br>45 mm Needle<br>Sets due to 15<br>mm Needle<br>design.                                                                    |
| Depth control                            | Positioning marks at 5 cm<br>and 10 cm apart to provide<br>visual reference points<br>Tactile feedback for change<br>of pressure                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Unchanged.                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                    |
| Single use<br>components/<br>accessories | EZ-IO Needle Sets<br>EZ-Connect Extension Set<br>EZ-IO Patient Wristband<br>EZ-Stabilizer Dressing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | BD Luer-Lok™ 10 mL<br>Syringe                                                                                                                                                                                                                                                                                                                                                                                                                          | Different -<br>IOSS only<br>requires a<br>syringe for use.                                                                                                                              |
| Sterility                                | Sterile: Ethylene Oxide                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | IOSS is non-sterile.                                                                                                                                                                                                                                                                                                                                                                                                                                   | Different -<br>Patient contact<br>for IOSS does<br>not require<br>sterilization.                                                                                                        |
| Category                                 | Predicate EZ-IO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Proposed IOSS Used as<br>Indicated with Predicate<br>Device                                                                                                                                                                                                                                                                                                                                                                                            | Comparison                                                                                                                                                                              |
| Shelf life                               | 4 Years                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 5 years                                                                                                                                                                                                                                                                                                                                                                                                                                                | Different -<br>Additional year<br>of shelf life<br>does not raise<br>new questions<br>of safety or<br>efficacy.                                                                         |
| Bio-<br>compatibility                    | Biocompatible materials<br>used (per ISO 10993-1<br>prolonged contact<br>duration).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Biocompatibility of IOSS<br>established per ISO 10993-<br>1 for limited duration<br>surface contact:<br>Cytotoxicity, Sensitization,<br>Irritation, Pyrogenicity, and<br>Systemic Toxicity                                                                                                                                                                                                                                                             | Different -<br>IOSS and<br>Predicate both<br>meet<br>biocompatibility<br>endpoints<br>appropriate for<br>their patient<br>contact.                                                      |
| Materials of<br>construction             | Stainless steel catheter,<br>polycarbonate and color<br>additive hubs, and<br>polypropylene needle<br>cover.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Clear, colorless, INEOS K-<br>Resin KR01 styrene<br>butadiene block copolymer                                                                                                                                                                                                                                                                                                                                                                          | Different -<br>different<br>materials meet<br>all<br>requirements<br>for<br>biocompatibility<br>and physical<br>properties:<br>raising no new<br>questions of<br>safety or<br>efficacy. |

{6}------------------------------------------------

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image is an abstract design featuring a combination of geometric and organic elements. The background is a gradient of orange, transitioning from a lighter shade to a darker one. Overlaid on this background is a pattern of plus signs, arranged in a grid-like fashion, with the color of the plus signs alternating between the two shades of orange. A series of wavy lines are superimposed on the plus sign pattern, adding a sense of movement and fluidity to the design.

{8}------------------------------------------------

# Non-Clinical Performance Data

The methods and performance data for evaluating the technological differences, the questions of safety and effectiveness, as well as IOSS integrity and compatibility with the predicate device (EZ-IO) include bench performance testing, usability testing, and biocompatibility testing.

Bench Performance Testing: Verification of Product Specification testing confirmed that the IO Needle Safety Sheath (IOSS) met all product specifications high temperature/humidity, and low temperature environmental conditions. Distribution testing conforming to ISTA 3A 2018 (FDA Recognition Number

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image is an abstract design featuring shades of orange. The background consists of a pattern of squares in two slightly different shades of orange, creating a checkerboard effect. A wavy line, also in orange, meanders across the pattern, adding a dynamic element to the design.

5-126) confirmed that IOSS met all product specifications and acceptance criteria after being exposed to shipping and transportation conditions. Accelerated Shelf Life testing confirmed that IOSS met all product specifications and acceptance criteria after exposure to conditions simulating a shelf life of up to 5 years.

| Test Performed                              | Test Description                                                                                                                                                                                                                               | Standards<br>Organization and<br>Designation                                                              | Results        |
|---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|----------------|
| Biocompatibility                            | Final, finished devices tested for<br>cytotoxicity, sensitization, irritation or<br>intracutaneous reactivity, acute<br>systemic toxicity, and material<br>mediated pyrogenicity.                                                              | ISO 10993-12, 2021<br>ISO 10993-5, 2009<br>ISO 10993-10, 2010<br>ISO 10993-23, 2021<br>ISO 10993-11, 2017 | Pass           |
| Verification of<br>Product<br>Specification | Verification of Product Specifications<br>after exposure to ambient, high<br>temp/humidity, and low temperature<br>environmental conditions.                                                                                                   | NA                                                                                                        | Pass           |
| Usability                                   | A total of 500 uses by thirty (30)<br>Users to assess whether the<br>intended user population can use<br>IOSS without encountering a needle<br>stick, serious use errors, or problems<br>for the intended use and expected<br>use environment. | NA                                                                                                        | Pass           |
| Distribution                                | Confirmation of Product<br>Specifications following exposure to<br>simulated distribution stress and<br>conditions.                                                                                                                            | ISTA 3A 2018                                                                                              | Pass           |
| Accelerated<br>Shelf Life                   | Confirmation of Product<br>Specifications following exposure<br>accelerated conditions simulating a<br>shelf life of one (1), three (3), and five<br>(5) years.                                                                                | NA                                                                                                        | Pass           |
| Shelf Life                                  | Confirmation of Product<br>Specifications following real-time<br>ambient condition exposure for one<br>(1), three (3), and five (5) years.                                                                                                     | NA                                                                                                        | Test on-going. |

Usability Testing: Human factors validation testing confirmed that the user population can safely and effectively use IOSS for the intended use in the expected use environment without encountering a needle stick, serious use errors, or other problems.

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image is an abstract design featuring a combination of geometric shapes and wavy lines. The color palette is dominated by shades of orange, with a lighter orange used for the wavy lines and some of the geometric shapes, creating a contrast against the darker orange background. The geometric shapes appear to be arranged in a grid-like pattern, adding a structured element to the design. The wavy lines overlay the geometric shapes, introducing a sense of movement and fluidity to the overall composition.

Biocompatibility Testing: IOSS is categorized as a limited duration surface device expected to normally contact intact skin, but which may incidentally contact breached or compromised skin. Cytotoxicity, sensitization, irritation, acute systemic toxicity, and material mediated pyrogenicity testing was conducted on IOSS in its final finished form. All tests confirmed the suitability of IOSS.

### Conclusions

After evaluating IOSS for its intended use with EZ-IO (the predicate device), then identifying, evaluating, and mitigating the risks associated with both use and foreseeable misuse where practicable, and confirming the safety and effectiveness of the device was through worst-case testing, it is concluded that IOSS is as safe and effective as EZ-IO when used as indicated to remove EZ-IO 25 mm and 45 mm intraosseous needles.

---

**Source:** [https://fda-staging.innolitics.com/device/K220890](https://fda-staging.innolitics.com/device/K220890)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).