← Product Code [ODC](/productcode/ODC) · K220884

# Disposable Endoscope Valve Sets (K220884)

_Sml Med-Tech Solutions Limited · ODC · Jan 17, 2023 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K220884

## Device Facts

- **Applicant:** Sml Med-Tech Solutions Limited
- **Product Code:** [ODC](/productcode/ODC.md)
- **Decision Date:** Jan 17, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Disposable Endoscope Valve Sets is a collection of several sterile units. It is intended to be fitted to multiple endoscope working channels/ports to control the flow of fluids, gases and other materials. It includes an Air/Water Valve, a Suction Valve, a Biopsy Valve and an Auxiliary Water Connector. - Air/Water Valve: This unit is intended to be fitted to an endoscope air/water channel to control the inflow of medical gases and water, whilst preventing back-flow. - Suction Valve: This unit is intended to be fitted to an endoscope suction channel to control the operations of suction, whilst preventing inflow of air. - Biopsy Valve: This unit is intended to be fitted to an endoscope biopsy port to prevent leakage of gases and body fluids during an endoscopic procedure. - Auxiliary Water Connector: This unit is intended to provide irrigation via sterile water supply during GI endoscopic procedures when used in conjunction with an irrigation pump.

## Device Story

Disposable Endoscope Valve Sets comprise sterile, single-use components (Air/Water Valve, Suction Valve, Biopsy Valve, Auxiliary Water Connector) for GI endoscopes. Valves attach to endoscope ports to manage fluid/gas flow, suction, and irrigation; prevent backflow; and maintain insufflation/seal during procedures. Used in hospitals/clinics by trained medical staff. Air/Water valve controls gas/water inflow; Suction valve enables debris/fluid aspiration; Biopsy valve prevents leakage during device passage; Auxiliary Water Connector facilitates irrigation via external pump. Benefits include maintaining sterile field, preventing cross-contamination, and ensuring proper endoscope function during diagnostic/therapeutic procedures.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including endoscope compatibility, flow rates, leakage tests, depression force, and backflow performance. Biocompatibility testing (cytotoxicity, sensitization, irritation) performed per ISO 10993-1. Package integrity validated via ASTM F88/F88M-2015 and ASTM F1929-2015.

## Technological Characteristics

Materials: ABS, Silicone Rubber, TPE, PC, Stainless steel 304. Components: Air/Water valve, Suction valve, Biopsy valve, Auxiliary Water Connector. Principle: Mechanical flow control/sealing. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Single-use. No software or electronic components.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Disposable Endoscope Valves Set, Disposable Endoscope Valves Set B ([K200769](/device/K200769.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" on the bottom.

January 17, 2023

SML Med-Tech Solutions Limited % Cassie Lee, Manager Share Info (Guangzhou) Medical Consultant Ltd. No. 1919-1920, Building D3, Minjie Plaza, Shuixi Road, Huangpu District Guangzhou, Guangdong CHINA

Re: K220884

Trade/Device Name: Disposable Endoscope Valve Sets Regulation Number: 21 CFR 876.1500 Name: Endoscope and accessories Regulatory Class: Class II Product Code: ODC Dated: October 13, 2022 Received: October 14, 2022

Dear Cassie Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Shanil P. Haugen -S

Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K220884

Device Name Disposable Endoscope Valve Sets

#### Indications for Use (Describe)

The Disposable Endoscope Valve Sets is a collection of several sterile units. It is intended to be fitted to multiple endoscope working channels/ports to control the flow of fluids, gases and other materials. It includes an Air/Water Valve, a Suction Valve, a Biopsy Valve and an Auxiliary Water Connector.

- Air/Water Valve: This unit is intended to be fitted to an endoscope air/water channel to control the inflow of medical gases and water, whilst preventing back-flow.

- Suction Valve: This unit is intended to be fitted to an endoscope suction channel to control the operations of suction. whilst preventing inflow of air.

- Biopsy Valve: This unit is intended to be fitted to an endoscope biopsy port to prevent leakage of gases and body fluids during an endoscopic procedure.

- Auxiliary Water Connector: This unit is intended to provide irrigation via sterile water supply during GI endoscopic procedures when used in conjunction with an irrigation pump.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary of K220884

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.

#### 1. Date of the summary prepared: January 11, 2023

#### 2. Submitter's Information

Sponsor Name: SML Med-Tech Solutions Limited Address: Rm 406, WAH YIU INDL CTR, 30-32 AU PUI WAN ST, FO TAN, N.T., Hong Kong Establishment Registration Number: Applying Post Code: 518119 Contact name: Mark KO (Marketing Director) Tel: (852) 2690 1113 Phone: +86 138 2302 2407 Fax: (852) 2601 2074 E-mail: mark@sml-medtech.com

#### Application Correspondent:

Contact Person: Ms. Cassie Lee Company: Share Info (Guangzhou) Medical Consultant Ltd. Address: No. 1919-1920, Building D3, Minjie Plaza, Shuixi Road, Huangpu District, Guangzhou, China Tel: +86 20 8200 6973 Email: regulatory@share-info.com

### 3. Subject Device Information

Type of 510(k): Traditional Common Name: Endoscope and accessories Classification Name: Endoscope Channel Accessory Trade Name: Disposable Endoscope Valve Sets Review Panel: Gastroenterology/Urology Product Code: ODC Regulation Number: 21 CFR 876.1500 Regulatory Class: Class II

### 4. Predicate Device Information

Sponsor: Wilson Instruments (SHA) Co., LTD

{4}------------------------------------------------

Common Name: Endoscope and accessories Classification Name: Endoscope Channel Accessory Trade Name: Disposable Endoscope Valves Set, Disposable Endoscope Valves Set B 510(k) Number: K200769 Review Panel: Gastroenterology/Urology Product Code: ODC Regulation Number: 21 CFR 876.1500 Regulatory Class: Class II

#### 5. Device Description

The Disposable Endoscope Valve Sets is a collection of several sterile units, it is intended for single-use and supplied sterile. The subject device is intended to multiple endoscope working channels/ports to control the flow of fluids, gases and other materials. It includes an Air/Water Valve, a Suction Valve, a Biopsy Valve and an Auxiliary Water Connector.

- The Suction Valve component of the Disposable Endoscope Valve Sets is designed to be attached । to the suction port of the endoscope, allowing the user to aspirate excess fluids or other debris obscuring the endoscopic image.
- । The Air/Water Valve component of the Disposable Endoscope Valve Sets is designed to be attached to the air/ water port of the endoscope, the activation of the air/ water valve allows the user to control air and water flow to assist in cleansing the lens during procedures.
- -The Biopsy Valve component of the Disposable Endoscope Valve Sets is intended to cover the endoscope biopsy port during an endoscopy procedure. In addition, the valve provides access for endoscopic device passage and exchange, helps maintain insufflation, and minimizes leakage of biomaterial from the biopsy port throughout the endoscopic procedure.
- The Auxiliary Water Connector component of the Disposable Endoscope Valve Sets is designed to be attached to the auxiliary water port of the endoscopes. The auxiliary water connector consists of a backflow valve that prevents the backflow of water or biomaterials from the endoscope to the sterile water bottle.

And there were no prior submissions for the subject devices.

#### 6. Intended Use / Indications for Use

The Disposable Endoscope Valve Sets is a collection of several sterile units. It is intended to be fitted to multiple endoscope working channels/ports to control the flow of fluids, gases and other materials. It includes an Air/Water Valve, a Suction Valve, a Biopsy Valve and an Auxiliary Water Connector.

- -Air/Water Valve: This unit is intended to be fitted to an endoscope air/water channel to control the inflow of medical gases and water, whilst preventing back-flow.

{5}------------------------------------------------

- -Suction Valve: This unit is intended to be fitted to an endoscope suction channel to control the operations of suction, whilst preventing inflow of air.
- Biopsy Valve: This unit is intended to be fitted to an endoscope biopsy port to prevent leakage of gases and body fluids during an endoscopic procedure.
- Auxiliary Water Connector: This unit is intended to provide irrigation via sterile water supply during -GI endoscopic procedures when used in conjunction with an irrigation pump.

#### 7. Comparison to predicate device

| Elements of<br>Comparison     | Subject Device                                                                                                                                                                                                                                                             | Predicate Device                                                                                                                                                | Result                                                                                                                                                                |                   |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| Company                       | SML Med-Tech Solutions Limited                                                                                                                                                                                                                                             | Wilson Instruments (SHA) Co.,<br>LTD                                                                                                                            | --                                                                                                                                                                    |                   |
| 510 (k) Number                | K220884                                                                                                                                                                                                                                                                    | K200769                                                                                                                                                         | --                                                                                                                                                                    |                   |
| Trade Name                    | Disposable Endoscope Valve Sets                                                                                                                                                                                                                                            | Disposable Endoscope Valves<br>Set, Disposable Endoscope<br>Valves Set B                                                                                        | --                                                                                                                                                                    |                   |
| Product Code                  | ODC                                                                                                                                                                                                                                                                        | ODC                                                                                                                                                             | Same                                                                                                                                                                  |                   |
| Common Name                   | Endoscope and accessories                                                                                                                                                                                                                                                  | Endoscope and accessories                                                                                                                                       | Same                                                                                                                                                                  |                   |
| Classification Name           | Endoscope Channel Accessory                                                                                                                                                                                                                                                | Endoscope Channel Accessory                                                                                                                                     | Same                                                                                                                                                                  |                   |
| Classification                | Class II                                                                                                                                                                                                                                                                   | Class II                                                                                                                                                        | Same                                                                                                                                                                  |                   |
| Regulation No.                | 21 CFR 876.1500                                                                                                                                                                                                                                                            | 21 CFR 876.1500                                                                                                                                                 | Same                                                                                                                                                                  |                   |
| Comparison of Air/Water Valve |                                                                                                                                                                                                                                                                            |                                                                                                                                                                 |                                                                                                                                                                       |                   |
| Indications for use           | This unit is intended to be fitted to<br>an endoscope air/water channel to<br>control the inflow of medical gases<br>and water, whilst preventing back-flow.                                                                                                               | This unit is intended to be fitted<br>to an endoscope air/water<br>channel to control the inflow of<br>medical gases and water, whilst<br>preventing back-flow. | Same                                                                                                                                                                  |                   |
| Compatible<br>endoscopes      | SML002_LP01_AW: Olympus® 140/<br>160/ 180/ 190/ 240/ 260/ 290 Series<br>GI Endoscope<br>SML001_PT01_AW: Pentax® 90/<br>i10 GI Endoscope<br>SML003_FJ01_AW: Fujifilm® 700<br>Series GI Endoscope                                                                            | WS-VO-01: Olympus<br>140/160/180/190/240/260/290<br>series endoscopes<br>WS-VP-01: PENTAX GI<br>Video Endoscope 90K/90i<br>Series; K10/i10 Series               | Similar<br>Note 1                                                                                                                                                     |                   |
| Sterile                       | EO sterilization                                                                                                                                                                                                                                                           | EO sterilization                                                                                                                                                | Same                                                                                                                                                                  |                   |
| Single Use                    | Yes                                                                                                                                                                                                                                                                        | Yes                                                                                                                                                             | Same                                                                                                                                                                  |                   |
| Elements of<br>Comparison     | Subject Device                                                                                                                                                                                                                                                             | Predicate Device                                                                                                                                                | Result                                                                                                                                                                |                   |
| Material                      | ABS, Silicone Rubber, TPE,<br>Stainless steel 304                                                                                                                                                                                                                          | WS-VO-01: SIR, ABS, TPE,<br>Stainless steel 304<br>WS-VP-01: SIR, ABS, TPE,<br>Stainless steel 304                                                              | Same                                                                                                                                                                  |                   |
| Environment<br>of use         | Hospitals/clinics                                                                                                                                                                                                                                                          | Hospitals/clinics                                                                                                                                               | Same                                                                                                                                                                  |                   |
| Comparison of Suction Valve   |                                                                                                                                                                                                                                                                            |                                                                                                                                                                 |                                                                                                                                                                       |                   |
| Indications for use           | This unit is intended to be fitted to<br>an endoscope suction channel to<br>control the operations of suction,<br>whilst preventing inflow of air.                                                                                                                         | This unit is intended to be fitted<br>to an endoscope suction channel<br>to control the operations of<br>suction, whilst preventing inflow<br>of air.           | Same                                                                                                                                                                  |                   |
| Compatible<br>endoscopes      | SML002_LP01_SU,<br>SML002_LP02_SU: Olympus® 140/<br>160/ 180/ 190/ 240/ 260/290 Series<br>GI Endoscope<br>SML001_PT01_SU,<br>SML001_PT02_SU,<br>SML001_PT03_SU: Pentax® 90/ i10<br>GI Endoscope<br>SML003_FJ01_SU,<br>SML003_FJ02_SU: Fujifilm® 700<br>Series GI Endoscope | WS-VO-02: Olympus<br>140/160/180/190/240/260/2<br>90 series endoscopes<br>WS-VP-02: PENTAX GI<br>Video Endoscope<br>90K/90i Series; K10/i10<br>Series           | Similar<br>Note 1                                                                                                                                                     |                   |
| Sterile                       | EO sterilization                                                                                                                                                                                                                                                           | EO sterilization                                                                                                                                                | Same                                                                                                                                                                  |                   |
| Single Use                    | Yes                                                                                                                                                                                                                                                                        | Yes                                                                                                                                                             | Same                                                                                                                                                                  |                   |
| Material                      | ABS, PC, Silicone Rubber, Stainless<br>steel 304                                                                                                                                                                                                                           | WS-VO-02: SIR, ABS, TPE,<br>Stainless steel 304, Stainless<br>steel 17-4PH<br>WS-VP-02: SIR, ABS, TPE,<br>Stainless steel 304                                   | Similar<br>Note 2                                                                                                                                                     |                   |
| Environment<br>of use         | Hospitals/clinics                                                                                                                                                                                                                                                          | Hospitals/clinics                                                                                                                                               | Same                                                                                                                                                                  |                   |
| Comparison of Biopsy Valve    |                                                                                                                                                                                                                                                                            |                                                                                                                                                                 |                                                                                                                                                                       |                   |
| Indications for use           | This unit is intended to be fitted to                                                                                                                                                                                                                                      | This unit is intended to be fitted                                                                                                                              | Same                                                                                                                                                                  |                   |
| Elements of<br>Comparison     | Subject Device                                                                                                                                                                                                                                                             |                                                                                                                                                                 | Predicate Device                                                                                                                                                      | Result            |
|                               |                                                                                                                                                                                                                                                                            |                                                                                                                                                                 |                                                                                                                                                                       |                   |
|                               | an endoscope biopsy port to prevent<br>leakage of gases and body fluids<br>during an endoscopic procedure.                                                                                                                                                                 |                                                                                                                                                                 | to an endoscope biopsy port to<br>prevent leakage of gases and<br>body fluids during an endoscopic<br>procedure.                                                      |                   |
| Compatible<br>endoscopes      | SML002_LP01_BP: Olympus® 140/<br>160/ 180/ 190/ 240/ 260/290 Series<br>GI Endoscope<br>SML001_PT01_BP: Pentax® 90/ i10<br>GI Endoscope<br>SML003_FJ01_BP: Fujifilm® 700<br>Series GI Endoscope                                                                             |                                                                                                                                                                 | Olympus<br>140/160/180/190/240/260/2<br>90 series endoscopes                                                                                                          | Similar<br>Note 1 |
| Sterile                       | EO sterilization                                                                                                                                                                                                                                                           |                                                                                                                                                                 | EO sterilization                                                                                                                                                      | Same              |
| Single Use                    | Yes                                                                                                                                                                                                                                                                        |                                                                                                                                                                 | Yes                                                                                                                                                                   | Same              |
| Material                      | Silicone Rubber                                                                                                                                                                                                                                                            |                                                                                                                                                                 | Silicon rubber                                                                                                                                                        | Similar<br>Note 2 |
| Environment<br>of use         | Hospitals/clinics                                                                                                                                                                                                                                                          |                                                                                                                                                                 | Hospitals/clinics                                                                                                                                                     | Same              |
| Comparison of Water Connector |                                                                                                                                                                                                                                                                            |                                                                                                                                                                 |                                                                                                                                                                       |                   |
| Indications for use           | This unit is intended to provide<br>irrigation via sterile water supply<br>during GI endoscopic procedures<br>when used in conjunction with an<br>irrigation pump.                                                                                                         |                                                                                                                                                                 | This unit is intended to provide<br>irrigation via sterile water supply<br>during Gl endoscopic<br>procedures when used in<br>conjunction with an irrigation<br>pump. | Same              |
| Compatible<br>endoscopes      | SML002_LP01_AU: Olympus® 140/<br>160/ 180/ 190/ 240/ 260/290 Series<br>GI Endoscope<br>SML001_PT01_AU: Pentax® 90/ i10<br>GI Endoscope<br>SML003_FJ01_AU: Fujifilm® 700<br>Series GI Endoscope                                                                             |                                                                                                                                                                 | Olympus<br>140/160/180/190/240/260/2<br>90 series endoscopes                                                                                                          | Similar<br>Note 1 |
| Sterile                       | EO sterilization                                                                                                                                                                                                                                                           |                                                                                                                                                                 | EO sterilization                                                                                                                                                      | Same              |
| Single Use                    | Yes                                                                                                                                                                                                                                                                        |                                                                                                                                                                 | Yes                                                                                                                                                                   | Same              |
| Material                      | PC, Silicone Rubber                                                                                                                                                                                                                                                        |                                                                                                                                                                 | SIR, PVC, PC, Stainless                                                                                                                                               | Similar           |
| Elements of<br>Comparison     | Subject Device                                                                                                                                                                                                                                                             | Predicate Device                                                                                                                                                | Result                                                                                                                                                                |                   |
|                               |                                                                                                                                                                                                                                                                            | steel 303                                                                                                                                                       | Note 2                                                                                                                                                                |                   |
| Environment<br>of use         | Hospitals/clinics                                                                                                                                                                                                                                                          | Hospitals/clinics                                                                                                                                               | Same                                                                                                                                                                  |                   |

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

#### Comparison in Detail(s):

#### Note 1:

Although the "Compatible endoscopes" of the subject device is a little different from the predicate device, the subject device is designed for intended use with the corresponding endoscopes mentioned and all the components of the subject device have good compatibility with the compatible endoscopes. So, the differences between the subject device and predicate device will not affect the safety and effectiveness.

#### Note 2:

Although the "Material" of the subject device is a little different from the predicate device, they all met the requirements of the ISO 10993 series biocompatibility standards. So, the differences between the subject device and predicate device will not affect the safety and effectiveness.

#### 8. Test Summary

### 8.1 Summary of Non-Clinical Performance Testing

#### 1)

The SML Med-Tech Solutions Limited has performed bench testing to support substantial equivalence. The following tests were performed on the subject devices and predicate devices.

For Suction Valve SML001_PT01_SU, SML001_PT02_SU, SML001_PT03_SU, SML002_LP01_SU, SML002_LP02_SU, SML003_FJ01_SU, SML003_FJ02_SU:

- -Endoscope Compatibility
- -Depression Force
- Leakage Test -
- । Suction Flow

For Air/Water Valve SML001_PT01_AW, SML002_LP01_AW, SML003_FJ01_AW:

- । Endoscope Compatibility
- -Air flow rate
- -Water flow rate
- Leakage Test -
- Depression Force -
- Backflow Performance Test י

{9}------------------------------------------------

For Biopsy Valve SML001_PT01_BP, SML002_LP01_BP / SML003_FJ01_BP:

- Endoscope Compatibility -
- । Leakage Test

For Auxiliary Water Connector SML001_PT01_AU, SML002_LP01_AU, SML003_FJ01_AU:

- -Endoscope Compatibility
- -Leakage Test
- Water flow rate -
- -Backflow Performance Test

#### Sterilization 2)

All the subject devices are sold in sterile packages, like the predicate devices have been sterilized in a validated EO sterilization cycle. The EO sterilization cycle has a Sterility Assurance Level (SAL) of 10°. EO residuals on the components are below the maximum levels defined in ISO 10993-7:2008 Biological Evaluation of Medical Devices - Part 7: Ethylene Oxide sterilization residuals.

#### 3) Biocompatibility

The biocompatibility of the subject devices was conducted in accordance with the FDA guideline "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process". Biocompatibility testing is conducted on the subject device in accordance with the ISO 10993 standards. It included the following tests:

- -In Vitro Cytotoxicity Test
- । Skin sensitization Test
- -Irritation or intracutaneous reactivity Test

The subject devices are classified as surface devices with mucosal membrane contact for a limited duration (not more than 24 hours). The test result shows that the subject devices are biocompatible.

#### 4) Package integrity

The package integrity test of the subject devices has been established to study whether the packaging will maintain a sterile barrier for the entirety of the proposed shelf life, after simulated distribution. And visual inspection, seal strength test (ASTM F88/F88M-2015), and dye penetration testing (ASTM F1929-2015) have been conducted on the subject device before aging products, and products have undergone accelerated aging followed by simulated distribution. The test results show that the subject devices maintain a sterile barrier for the entirety of the proposed shelf life, after simulated distribution.

### 8.2 Summary of Clinical Performance Test

{10}------------------------------------------------

No clinical study is included in this submission.

## 9. Final Conclusion:

The subject device Disposable Endoscope Valve Sets is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K220884](https://fda-staging.innolitics.com/device/K220884)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).