K220880 · Arthrex, Inc. · GAT · Oct 27, 2022 · General, Plastic Surgery
Device Facts
Record ID
K220880
Device Name
Arthrex BioSuture
Applicant
Arthrex, Inc.
Product Code
GAT · General, Plastic Surgery
Decision Date
Oct 27, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.5000
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Arthrex BioSuture is intended for soft tissue approximation and or ligation. These sutures may be incorporated, as components, into surgeries where constructs, including those with allograft tissues, are used for repair.
Device Story
Arthrex BioSuture is a braided surgical suture composed of Ultra High Molecular Weight Polyethylene (UHMWPE) and polyester, featuring a collagen coating. Designed as a line extension to existing predicate sutures, it is available in various precut lengths with straight or loop configurations. The device is intended for use by surgeons for soft tissue approximation and ligation, including incorporation into surgical constructs involving allograft or autograft tissues. It is supplied sterile for single-patient use. Clinical benefit is derived from its mechanical strength, which facilitates secure tissue repair. The device is evaluated for MR safety.
Clinical Evidence
Bench testing only. Mechanical testing, including straight pull and knot pull strength, was conducted to verify that the device meets established acceptance criteria. Bacterial Endotoxins Testing (BET) was performed using the Kinetic Chromogenic Method per ANSI/AAMI ST72:2011/(R)2016, USP <161>, USP <85>, and EP 2.6.14 to confirm compliance with pyrogen limit specifications.
Technological Characteristics
Braided construct of UHMWPE and polyester with collagen coating. Available in various lengths and configurations (straight/loop). Sterilized for single use. Evaluated for MR safety. Testing performed per ANSI/AAMI ST72:2011/(R)2016, USP <161>, USP <85>, and EP 2.6.14 for endotoxins.
Indications for Use
Indicated for soft tissue approximation and ligation in patients requiring surgical repair, including procedures utilizing allograft or autograft tissue constructs.
Regulatory Classification
Identification
Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
10/27/2022
Arthrex Inc. Stacy Valdez Senior Regulatory Affairs Specialist 1370 Creekside Boulevard Naples. Florida 34108-1945
### Re: K220880
Trade/Device Name: Arthrex BioSuture Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable Poly(Ethylene Terephthalate) Surgical Suture Regulatory Class: Class II Product Code: GAT Dated: September 2, 2022 Received: September 7, 2022
Dear Stacy Valdez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Deborah Fellhauer RN, BSN Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K220880
Device Name Arthrex BioSuture
Indications for Use (Describe)
The Arthrex BioSuture is intended for soft tissue approximation and or ligation. These sutures may be incorporated, as components, into surgeries where constructs, including those with allograft tissues, are used for repair.
Type of Use (Select one or both, as applicable)
| <span></span> | Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|----------------------------------------------|
| <span></span> | Over-The-Counter Use (21 CFR 801 Subpart C) |
escription 303 (Part 21 CFR 301 Subpart B)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Date Prepared | October 17, 2022 |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter | Arthrex Inc.<br>1370 Creekside Boulevard<br>Naples, FL 34108-1945 |
| Contact Person | Stacy Valdez<br>Senior Regulatory Affairs Specialist<br>1-239-643-5553, ext. 72010<br>stacy.valdez@arthrex.com |
| Name of Device | Arthrex BioSuture |
| Common Name | Suture |
| Product Code | GAT |
| Classification Name | 21 CFR 878.5000: Nonabsorbable poly(ethylene terephthalate) surgical suture |
| Regulatory Class | II |
| Primary Predicate Device | K112899: Arthrex Bio-Suture |
| Additional Predicate Device | K140019: Arthrex BioSuture |
| Reference Devices | K193575: Arthrex SutureTape<br>K122374: Arthrex Suture<br>K041553: Arthrex Suture Grafting Kit<br>K032245: Arthrex FiberTape Family<br>K021434: Arthrex FiberWire Family, USP Size Sutures |
| Purpose of Submission | This Special 510(k) premarket notification is submitted to obtain<br>clearance for the Arthrex BioSuture. |
| Device Description | The proposed Arthrex BioSuture is a braided construct made of Ultra High<br>Molecular Weight Polyethylene (UHMWPE) and polyester and coated<br>with collagen coating. The proposed suture is braided flat with round<br>ends and may be available in precut lengths in straight and loop<br>configurations. The Arthrex BioSuture is packaged sterile for single use.<br>The Arthrex BioSuture is a line extension to the Arthrex Bio-Suture<br>consisting of a new size. |
| Indications for Use | The Arthrex BioSuture is intended for soft tissue approximation and or<br>ligation. These sutures may be incorporated, as components, into<br>surgeries where constructs, including those with allograft or autograft<br>tissues, are used for repair. |
| Performance Data | Mechanical testing (straight pull, knot pull) was conducted to<br>demonstrate that the strength of the proposed Arthrex BioSuture met<br>the established acceptance criteria. |
| Technological Comparison | Bacterial Endotoxins Test (BET) was performed on the representative<br>samples utilizing the Kinetic Chromogenic Method in accordance with<br>ANSI/AAMI ST72:2011/(R)2016, USP <161>, USP <85>, EP 2.6.14 to<br>demonstrate that the proposed device meets pyrogen limit specifications.<br>The Arthrex BioSuture is substantially equivalent to the predicate devices<br>cleared under K112899 and K140019 in which the overall design and<br>configuration of the suture, intended use/indications, surgical technique,<br>fundamental scientific technology, sterility, materials, packaging and<br>manufacturing process are identical. |
| | |
| | diameter than the predicate devices cleared under K112899. |
| | The Arthrex BioSuture is labeled with a 4-year shelf life; whereas the<br>predicate devices cleared under K112899 are labeled with a 2-year shelf<br>life. However, the additional predicate devices cleared under K140019<br>are labeled with a 4-year shelf life. |
| | The Arthrex BioSuture has been evaluated for MR Safe labeling; whereas<br>the predicate devices cleared under K112899 were not evaluated for MR<br>Safe labeling. The needles are not implantable and therefore have not<br>been evaluated for MR Safety. |
| | The Arthrex BioSuture is a line extension to the predicate devices, which<br>include minor dimensional modifications with no change to intended use<br>or function. Any differences between the Arthrex BioSuture and the<br>predicate devices are considered minor and do not raise different<br>questions of safety or effectiveness. |
| Conclusion | The Arthrex BioSuture is substantially equivalent to the predicate device<br>in which the overall design and configuration of the suture, the intended<br>use/indications, surgical technique, fundamental scientific technology,<br>sterility, materials, packaging, and manufacturing process remain<br>identical to the primary predicate Arthrex Bio-Suture (K112899) and<br>additional predicate Arthrex BioSuture (K140019). Any differences<br>between the proposed device and the predicate device are considered<br>minor and do not raise different questions concerning safety or<br>effectiveness. |
| | Mechanical testing (straight pull, knot pull) demonstrated that the<br>strength of the proposed Arthrex BioSuture is substantially equivalent to<br>that of the predicate device for the desired indications. |
| | Based on the indications for use, technological characteristics, and the<br>summary of data submitted, Arthrex Inc. has determined that the<br>proposed device is substantially equivalent to the currently marketed<br>predicate device. |
{4}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.