Selecta Duet LED, Digital Duet, Selecta LED Trio, Digital Trio Laser Systems
K220877 · Lumenis Be, Inc. · HQF · Jul 15, 2022 · Ophthalmic
Device Facts
Record ID
K220877
Device Name
Selecta Duet LED, Digital Duet, Selecta LED Trio, Digital Trio Laser Systems
Applicant
Lumenis Be, Inc.
Product Code
HQF · Ophthalmic
Decision Date
Jul 15, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4390
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Selecta Duet LED or Digital Duet: Photodisruption of ocular tissue using light energy emitted by a Nd:YAG laser, including discission of the posterior capsule of the eye (posterior capsulotomy), iridotomy/ iridectomy, and selective laser trabeculoplasty. Selecta LED Trio or Digital Trio: Photodisruption of ocular tissue using light energy emitted by a Nd: Y AG laser, including discission of the posterior capsule of the eye (posterior capsulotomy), iridotomy/ iridectomy, selective laser trabeculoplasty, and retinal photocoagulation.
Device Story
Integrated diagnostic slit lamp and therapeutic Nd:YAG laser delivery system; used in clinical settings by ophthalmologists. Inputs: manual user control via joystick button or footswitch. Operation: 1064 nm Nd:YAG laser for photodisruption; 532 nm laser (integrated or via external LaserLink S connection) for selective laser trabeculoplasty and retinal photocoagulation. Outputs: laser energy delivered to ocular tissue. Clinical utility: enables precise anterior segment and retinal procedures; improves patient outcomes through non-invasive tissue disruption and coagulation. System includes LED illumination for examination.
Clinical Evidence
Bench testing only. Verification of software and hardware modifications performed per recognized standards including ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1, ISO 15004-2, and ISO 14971. Software verified per IEC 62304.
Technological Characteristics
Integrated slit lamp and Nd:YAG laser. Wavelengths: 1064 nm (PD), 532 nm (SLT/photocoagulation). Pulse duration: 3 ns. Energy: 30 mJ max (1064 nm), 2 mJ max (532 nm). Illumination: 5W LED. Connectivity: LaserLink S for external photocoagulator. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1, ISO 15004-2, ISO 14971.
Indications for Use
Indicated for photodisruption of ocular tissue (posterior capsulotomy, iridotomy/iridectomy, selective laser trabeculoplasty) and retinal photocoagulation (Trio models only) in patients requiring ophthalmic laser treatment.
Regulatory Classification
Identification
An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 15, 2022
Lumenis Be, Inc. Shlomit Segman Sr. Director, Regulatory Affairs 1870 South Milestone Drive Salt Lake City, UT 84104
Re: K220877
Trade/Device Name: Selecta Duet LED, Digital Duet, Selecta LED Trio, Digital Trio Laser Systems Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: Class II Product Code: HOF Dated: March 14, 2022 Received: March 25, 2022
Dear Shlomit Segman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
542 of the Act); 21 CFR 1000-1050.
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Anjana Jain, PhD Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K220877
Device Name
Selecta Duet LED, Digital Duet, Selecta LED Trio, and Digital Trio Laser Systems
Indications for Use (Describe)
· Selecta Duet LED or Digital Duet: Photodisruption of ocular tissue using light energy emitted by a Nd: Y AG laser, including discission of the posterior capsule of the eye (posterior capsulotomy) iridectorny, and selective laser trabeculoplasty.
· Selecta LED Trio or Digital Trio: Photodisruption of ocular tissue using light energy emitted by a Nd: Y AG laser, including discission of the posterior capsule of the eye (posterior capsulotomy) iridectorny, selective laser trabeculoplasty, and retinal photocoagulation.
| Type of Use (Select one or both, as applicable) |
|-----------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size:10pt"> <span style="font-family:Wingdings">þ</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
| <span style="font-size:10pt"> <span style="font-family:Wingdings">□</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Lumenis logo. The logo consists of a blue circle with a white "U" inside, followed by the word "Lumenis" in blue. Below the word "Lumenis" is the phrase "Energy to Healthcare" in a smaller font size.
K220877
# 510(k) SUMMARY
| PREPARATION DATE: | July 14, 2022 |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT: | Lumenis Be, Inc.<br>1870 South Milestone Drive<br>Salt Lake City, Utah 84104 |
| CONTACT PERSON: | Shlomit Segman, PhD<br>Sr. Director, Regulatory Affairs<br>Lumenis Be Ltd.<br>Tel: +972-4-9599230 or +972-52-3667233 |
| DEVICE TRADE NAME: | Selecta Duet LED Laser System<br>Digital Duet Laser System<br>Selecta LED Trio Laser System<br>Digital Trio Laser System |
| CLASSIFICATION NAME: | Powered Laser Surgical Instrument |
| DEVICE CLASSIFICATION: | Class II (21CFR§886.4390) |
| PRODUCT CODE: | HQF |
| PREDICATE DEVICE: | Selecta Duet and Trio Laser Systems<br>Applicant: Lumenis, Inc.<br>510(k) Number: K081704<br>Classification Name: Powered Laser Surgical Instrument<br>Device Classification: Class II (21CFR§878.4810)<br>Product Code: GEX, HQF |
The Selecta Duet and Trio Laser Systems have not been subject to a design related recall.### DEVICE DESCRIPTION:
The Selecta Duet LED and Digital Duet subject devices are each a fully integrated, highperformance diagnostic slit lamp and therapeutic laser delivery system. The diagnostic slit lamp is for examining the anterior segment of the eye, from the corneal epithelium to the posterior capsule. The integrated Nd:YAG Q-Switch laser module provides short pulses at the wavelength corresponding with the selected operating mode. Two modes are available:
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Lumenis logo. The logo consists of a blue circle with a white "U" shape inside, followed by the word "Lumenis" in blue. Below the word "Lumenis" is the phrase "Energy to Healthcare" in a smaller font, also in blue.
- PD (YAG) mode: 1064 nm for photodisruption (PD) of ocular tissues in procedures such ● as posterior capsulotomy, iridotomy, and iridectomy, and
- SLT mode: 532 nm for Selective Laser Trabeculoplasty (SLT). ●
The Selecta LED Trio and Digital Trio subject devices are each created by connecting the Selecta LED Duet and Digital Duet, respectively, to a compatible, FDA-cleared, Lumenis continuous 532 nm wavelength photocoagulator laser, via the LaserLink S. When the user connects the Duet to the LaserLink S, the Duet-part acts only as a Slit Lamp since the integrated Duet-laser is disabled and the laser in the external photocoagulator takes over. This combination adds the ability to perform retinal photocoagulation, as well as other indications cleared for the continuous 532 nm wavelength photocoagulator laser. The user disconnects the Duet from the LaserLink S for the system to perform as a Duet.
### INTENDED USE/INDICATIONS FOR USE:
- · Selecta Duet LED or Digital Duet: Photodisruption of ocular tissue using light energy emitted by a Nd:YAG laser, including discission of the posterior capsule of the eye (posterior capsulotomy), iridotomy/ iridectomy, and selective laser trabeculoplasty.
- · Selecta LED Trio or Digital Trio: Photodisruption of ocular tissue using light energy emitted by a Nd: Y AG laser, including discission of the posterior capsule of the eye (posterior capsulotomy), iridotomy/ iridectomy, selective laser trabeculoplasty, and retinal photocoagulation.
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS:
The subject laser systems are substantially equivalent to their respective predicate, predecessor K081704 laser systems. This is a catch-up 510(k) submission intended to update the Selecta Family of Laser Systems record at FDA with the changes that have been implemented in the subject devices. As such, the intended use/indications remain substantial equivalent to that for the cleared devices. Further, the implemented changes were primarily to be responsive to user feedback, upgrade to the latest user interface, or reflect a need to change a supplier or manufacturing contractor. The key changes are captured in the comparison table below.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Lumenis logo. The logo consists of a blue circle with a white "L" inside, followed by the word "Lumenis" in blue. Below the word "Lumenis" is the phrase "Energy to Healthcare".
# Table: Comparison of Technological Characteristics
| | Cleared Devices | Subject Devices: | |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | Selecta Duet and Trio Laser<br>Systems (K081704) | Selecta Duet LED, Digital Duet,<br>LED Trio, and Digital Trio Laser<br>Systems | |
| | GENERAL | | |
| Manufacturer | Lumenis, Inc. | Lumenis Be, Inc. | |
| Regulation<br>Number | 21CFR§878.4810 | Same | |
| Regulation<br>Name | Powered Laser Surgical Instrument | Same | |
| Procodes | GEX and HQF | Same | |
| Indications<br>for Use | • Selecta Duet: Photodisruption of<br>ocular tissue using light energy<br>emitted by a Nd:YAG laser,<br>including discission of the posterior<br>capsule of the eye (posterior<br>capsulotomy), and discission of<br>pupillary membranes (pupillary<br>membranectomy) in aphakic and<br>pseudophakic patients, and<br>iridotomy/iridectomy and selective<br>laser trabeculoplasty.<br>• Selecta Trio: Photodisruption of<br>ocular tissue using light energy<br>emitted by a Nd:YAG laser,<br>including discission of the posterior<br>capsule of the eye (posterior<br>capsulotomy), and discission of<br>pupillary membranes (pupillary<br>membranectomy) in aphakic and<br>pseudophakic patients, and<br>iridotomy/iridectomy, retinal<br>photocoagulation and selective<br>laser trabeculoplasty. | • Selecta Duet LED or Digital<br>Duet: Photodisruption of ocular<br>tissue using light energy emitted<br>by a Nd:YAG laser, including<br>discission of the posterior capsule<br>of the eye (posterior<br>capsulotomy), iridotomy/<br>iridectomy, and selective laser<br>trabeculoplasty.<br>• Selecta LED Trio or Digital<br>Trio: Photodisruption of ocular<br>tissue using light energy emitted<br>by a Nd:YAG laser, including<br>discission of the posterior capsule<br>of the eye (posterior<br>capsulotomy), iridotomy/<br>iridectomy, retinal<br>photocoagulation, and selective<br>laser trabeculoplasty. | |
| LASER SPECIFICATIONS | | | |
| Type | Frequency-doubled/Q-switched<br>Nd:YAG | Identical | |
| | Cleared Devices | Subject Devices: | |
| | Selecta Duet and Trio Laser<br>Systems (K081704) | Selecta Duet LED, Digital Duet,<br>LED Trio, and Digital Trio Laser<br>Systems | |
| Principal<br>output -<br>wavelength | 532 and 1064 nm | Identical | |
| Operating<br>mode | Frequency doubled/Fundamental,<br>pulsed | Identical | |
| Pulse<br>duration | 3 nanoseconds | Identical | |
| Energy (mJ) | For 532 nm laser: 2 maximum<br>For 1064 nm laser: 30 maximum | Identical | |
| Laser beam<br>spot size | For 532 nm laser:<br>400 um at visual focal plane<br>For 1064 nm laser:<br><8 um full width half maximum | For 532 nm laser:<br>400 um at visual focal plane<br>For 1064 nm laser:<br><11 µm at visual focal plane, 8 µm<br>full width half maximum calculated | |
| CDRH Rad.<br>Health<br>classification | Class IIIb Lasers | Identical | |
| IEC 60825<br>classification | Class 3B | Identical | |
| Laser<br>Actuation | Button or footswitch | Joystick button or Footswitch | |
| | AIMING BEAM SPECIFICATIONS | | |
| Type | CW Diode laser | Identical | |
| Principal<br>output | 635 nm wavelength | Identical | |
| CDRH Rad.<br>Health<br>classification | Class II | Identical | |
| IEC 60825<br>classification | Class 2 | Identical | |
| | SLIT LAMP | | |
| Field of view | 11.3 mm diameter at 16X<br>magnification | Identical | |
| Working<br>distance | 70 mm | Identical | |
| | Cleared Devices<br>Selecta Duet and Trio Laser<br>Systems (K081704) | Subject Devices:<br>Selecta Duet LED, Digital Duet,<br>LED Trio, and Digital Trio Laser<br>Systems | |
| Slit rotation<br>(°) | -90 (horizontal) to + 90 (horizontal) | Identical | |
| Bulb | Halogen, 30 W | Light Emitting Diodes, 5W | |
| COMPATIBLE 532 nm PHOTOCOAGULATOR LASER | | | |
| | • Novus Spectra Laser System | • Smart532 (K151109) | |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Lumenis. The logo consists of a blue circle with a white "L" inside, followed by the word "Lumenis" in blue. Below the word "Lumenis" is the tagline "Energy to Healthcare".
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the Lumenis logo. The logo consists of a blue circle with a white "L" inside, followed by the word "Lumenis" in blue. Below the word "Lumenis" is the phrase "Energy to Healthcare" in a smaller font.
The differences between the subject devices and their respective, predecessor, predicated devices do not raise different questions of safety and effectiveness and have been confirmed through testing.
# PERFORMANCE TESTING:
Performance testing demonstrates that the subject devices are as safe and as effective as their respective, predicate, predecessor devices. These tests included:
- . Evaluating the implemented changes using the same recognized standards that were used for the cleared devices:
| Standard Number | Standard Name |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ANSI/AAMI ES60601-<br>1:2005/ A2:2010 | Medical electrical equipment - Part 1: General<br>requirements for basic safety and essential performance<br>(IEC 60601-1:2005, MOD) |
| IEC 60601-1-2 Edition<br>4.0:2014 | Medical electrical equipment - Part 1-2: General<br>requirements for basic safety and essential performance -<br>Collateral Standard: Electromagnetic disturbances -<br>Requirements and tests |
| IEC 60601-2-22 Ed. 3.1<br>b:2012 | Medical electrical equipment - Part 2-22: Particular<br>requirements for basic safety and essential performance<br>of surgical, cosmetic, therapeutic and diagnostic laser<br>equipment |
| IEC 60825-1:2014 | Safety of laser products - Part 1: Equipment classification<br>and requirements |
| ISO 15004-2 | Ophthalmic instruments — Fundamental requirements<br>and test methods — Part 2: Light hazard protection |
| ISO 14971 Third Edition<br>2019-12 | Medical devices - Application of risk management to<br>medical devices |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Lumenis, a company that specializes in energy to healthcare. The logo features a stylized letter "U" inside of a circle, followed by the company name "Lumenis" in a sans-serif font. Below the company name is the tagline "Energy to Healthcare" in a smaller font. The logo is clean and modern, with a focus on simplicity and clarity.
- Software/Bench Testing ●
Lumenis developed and verified the software in accordance with a major level of concern described in the FDA "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and also per the IEC 62304:2006 and A1:2015 Medical Device Software - Software Life Cycle Processes standard.
The software/bench tests conducted on the modifications confirmed that they met their requirements.
### CONCLUSIONS:
The provided performance testing supports a substantial equivalence determination. The subject devices are substantially equivalent to their respective, predicate, predecessor devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.