K220869 · Vapotherm, Inc. · CAF · Oct 28, 2022 · Anesthesiology
Device Facts
Record ID
K220869
Device Name
Vapotherm Aerosol Adapter AAA-2
Applicant
Vapotherm, Inc.
Product Code
CAF · Anesthesiology
Decision Date
Oct 28, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5630
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Vapotherm Aerosol Adapter AAA-2 is intended to facilitate the connection between Vapotherm High Velocity Therapy HVT 2.0 and the Aerogen Solo Nebulizer.
Device Story
The Vapotherm Aerosol Adapter AAA-2 is a passive accessory designed to connect the Vapotherm High Velocity Therapy (HVT) 2.0 system to the Aerogen Solo Nebulizer. It allows for the delivery of aerosolized medication through the HVT 2.0 breathing circuit. The device is used in hospital and sub-acute settings by clinicians. It functions as a mechanical interface, enabling the integration of the nebulizer into the high-velocity therapy flow path without interfering with the delivery of supplemental oxygen or the aerosolization performance of the Aerogen system. By facilitating this connection, the device allows patients receiving high-velocity respiratory support to simultaneously receive nebulized medications, potentially improving clinical efficiency and patient comfort.
Clinical Evidence
No clinical data. Bench testing only. Testing included aerosolized solution output (laser verification), flow rate verification (8-45 L/min), occlusion alarm testing, rain-out volume assessment, connection integrity (ISO 5356-1), and continuous use durability. Aerosolization performance was validated using a Next Generation Cascade Impactor (NGI) to confirm particle size distribution and respirable dose consistency compared to the predicate. Human factors/usability testing was conducted to identify potential use errors.
Technological Characteristics
Passive mechanical adapter; polymer construction; non-sterile; single-patient use (up to 30 days). Interfaces with 22mm ports per ISO 5356-1. Biocompatibility per ISO 10993-1 and ISO 18562-1. No energy source or software.
Indications for Use
Indicated for spontaneously breathing adult and pediatric patients (5 kg and up, 29 days or older) requiring aerosolized medication delivery via the Aerogen Solo Nebulizer in conjunction with Vapotherm HVT 2.0 therapy in hospital or sub-acute settings.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the words "U.S. FOOD & DRUG" on one line and "ADMINISTRATION" on the line below. The FDA logo is in blue.
Vapotherm Inc. Michael Webb VP of Regulatory Affairs 100 Domain Drive Exeter, New Hampshire 03833
Re: K220869
Trade/Device Name: Vapotherm Aerosol Adapter AAA-2 Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: Class II Product Code: CAF Dated: September 29, 2022 Received: September 30, 2022
Dear Michael Webb:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
James Lee Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220869
Device Name Vapotherm Aerosol Adapter AAA-2
Indications for Use (Describe)
The Vapotherm Aerosol Adapter AAA-2 is intended to facilitate the connection between Vapotherm High Velocity Therapy HVT 2.0 and the Aerogen Solo Nebulizer.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Vapotherm logo. The logo consists of a blue sphere with white lines on the left and the word "VAPOTHERM" in blue on the right. The sphere has a design of three curved lines that converge at the bottom. The word "VAPOTHERM" is in a sans-serif font, and there is a registered trademark symbol after the word.
Image /page/3/Picture/1 description: The image is a logo for Hi-VNI Technology. The logo is blue and green. The word "Hi-VNI" is in blue, and the word "TECHNOLOGY" is in green. There is a green swoosh under the word "Hi-VNI".
# SECTION 5: 510(k) SUMMARY Date: 26-October-2022 Company: Vapotherm, Inc. 100 Domain Drive Exeter, NH 03833 Official Contact: Michael J. Webb - Vice President of Regulatory Affairs Tel - (844) 381-8276 Proprietary or Trade Name: AAA-2 Common/Usual Name: Nebulizer, Accessory to Classification Name: 21 CFR 868.5630, Class II Product Code: CAF Predicate Device: K133360 Aerogen Limited, Aeroneb Solo Adapter (Primary) Reference Device Predicate: K162753 Instrumentation Industries, Inc., RTC 26-C Inline Aerosol Adapter
### Device Description:
The Vapotherm Aerosol Adapter AAA-2 is an accessory that is intended to facilitate the connection between Vapotherm High Velocity Therapy HVT® 2.0, cleared under K203357, and the Aerogen Aeroneb® Solo Nebulizer System, cleared under K133360, and more recently with K143719. which is intended to aerosolize solutions for inhalation. The AAA-2 facilitates the connection between the noted devices, and therefore aerosolization use is dictated by the Aerogen Aeroneb Solo Nebulizer System labeling.
#### Indications for Use:
The Vapotherm Aerosol Adapter AAA-2 is intended to facilitate the connection between Vapotherm High Velocity Therapy HVT 2.0 and the Aerogen Solo Nebulizer.
#### Patient Population:
The AAA-2 shares the same intended patient populations as the Vapotherm HVT 2.0 which are spontaneously breathing adult and pediatric patients (5 kg and up), and the Aeroneb Solo Adapter which includes pediatric patients (29 days or older) as well as adult patients.
#### Environment of Use:
Hospital, sub-acute facility
#### Contraindications:
None
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Vapotherm. The logo consists of a blue circle on the left with white lines forming a stylized "V" shape inside. To the right of the circle is the word "VAPOTHERM" in blue, with a registered trademark symbol next to it. The logo is clean and modern, with a focus on the company name.
Image /page/4/Picture/1 description: The image shows the logo for Hi-VNI Technology. The logo is blue and green, with the words "Hi-VNI" in large, bold letters. The word "TECHNOLOGY" is in smaller letters below the main text. A green swoosh design is underneath the "Hi-VNI" text, adding a dynamic element to the logo. The logo appears clean and professional.
#### Substantial Equivalence:
The AAA-2 is substantially equivalent to the predicate device, K133360 Aeroneb Solo Adapter. The reference predicate device K162753 RTC 26-C Inline Aerosol Adapter, also supports substantial equivalence.
A Benefit-Risk Assessment was completed and concluded that when assessing the minor differences between the subject and predicate device, there is no decrease in the expected benefit or increase in potential risk with the subject device. This assessment supports the conclusion that the minor differences in technological characteristics between the subject device AAA-2 and predicate Aeroneb Solo Adapter do not raise new questions of safety and effectiveness. Additionally, this assessment when considered with the Performance Testing-Bench, supports that the AAA-2 is "as safe and effective" and performs as well as the predicate device. The probable benefit to health from of the use of the AAA-2 outweighs any potential risk of injury or illness from such use.
The tables below present the similarities and differences between the products for substantial equivalence purposes. The differences between the subject device and the predicate device do not raise new questions of safety and effectiveness. Performance data support a substantial equivalence determination.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Vapotherm. The logo consists of a blue sphere with white lines forming a "V" shape inside of it. To the right of the sphere is the word "VAPOTHERM" in blue, with a registered trademark symbol next to it.
-
Image /page/5/Picture/1 description: The image is a logo for Hi-VNI Technology. The logo is blue and green. The words "Hi-VNI" are in blue, with a green swoosh underneath. The word "TECHNOLOGY" is in smaller blue letters below the swoosh. The logo is simple and modern.
| Table 5.1: Primary Predicate Device Comparison Table and Discussion | | |
|---------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | Primary Predicate:<br>Aerogen Aeroneb Solo<br>Adapter (K133360) | Subject Device:<br>Vapotherm Aerosol Adapter<br>AAA-2 |
| Indications for<br>Use | The Aeroneb® Solo Adapter is<br>an accessory specific to the<br>Aeroneb® Solo Nebulizer. It<br>facilitates intermittent and<br>continuous nebulization and<br>optional supply of supplemental<br>Oxygen to pediatric (29 days or<br>older) and adult patients in<br>hospital use environments via a<br>mouthpiece or aerosol mask. If<br>supplemental oxygen is used, for<br>pediatric patients under 18 years<br>of age, a maximum flow rate of<br>2 LPM should be used. Note:<br>The mouthpiece should not be<br>used for children under 5 years<br>of age. | SIMILAR TEXT, SAME<br>INTENDED USE<br>The Vapotherm Aerosol Adapter AAA-<br>2 is intended to facilitate the connection<br>between Vapotherm High Velocity<br>Therapy HVT®2.0 and the Aerogen<br>Solo Nebulizer.<br><br>Both are aerosol adapters used with the<br>Aeroneb Solo Nebulizer System, that<br>allow connection to a larger breathing<br>circuit.<br><br>Because the subject device is only<br>indicated for use with the Aeroneb Solo<br>and HVT 2.0, it carries the same<br>intended patient populations as the two<br>parent devices which are pediatric and<br>adult patients.<br><br>The pediatric use is limited to 29 days<br>or older and 5kg and up, again,<br>consistent with the parent devices. |
| Intended Use | To allow use of the Aeroneb<br>Solo Nebulizer System with<br>supplemental oxygen. | SAME<br>To allow use of the Aeroneb Solo<br>Nebulizer System with the Vapotherm<br>HVT 2.0, which delivers supplemental<br>oxygen. |
| Patient<br>Population | Pediatric (29 days or older) and<br>adult patients | SIMILAR<br>Pediatric (29 days or older) and adult<br>patients.<br>Pediatric Patients 5kg and up and adult<br>patients.<br><br>The subject device is only for use with<br>the Aerogen Solo Nebulizer and HVT<br>2.0 and therefore is subject to their<br>cleared patient populations, which<br>would be pediatric patients 5 kg and up<br>and 29 days or older, as well as adult<br>patients. |
| Environment of<br>Use | Hospital | SIMILAR<br>Hospital, sub-acute facility<br><br>Environments are similar clinical<br>environments. Sub-acute facility is<br>included given the HVT 2.0 clearance<br>and is not a significant difference. |
| Table 5.1: Primary Predicate Device Comparison Table and Discussion | | |
| Characteristic | Primary Predicate:<br>Aerogen Aeroneb Solo<br>Adapter (K133360) | Subject Device:<br>Vapotherm Aerosol Adapter<br>AAA-2 |
| Type & Duration<br>of Use | Single patient use for up to 28 days | SIMILAR<br>Single patient use for up to 30 days<br><br>AAA-2 30 day use is consistent with<br>the HVT 2.0 device it is used with. |
| Prescriptive | Rx Only | SAME<br>Rx Only |
| Patient Interface –<br>for parent device | Mouthpiece, aerosol mask, nasal<br>cannula, or via ventilator | SIMILAR<br>Nasal cannula or TA-22<br><br>Both share the same patient interface of<br>nasal cannula or via trach access. |
| Patient contacting<br>materials | Externally communicating,<br>tissue, prolonged duration | SAME<br>Externally communicating, tissue,<br>prolonged duration<br><br>Both devices are comprised of<br>polymers; Same ISO 10993-1<br>classification |
| Sterility | Non-sterile | SAME<br>Non-sterile |
| Operating<br>Principle | Function is to allow access to a<br>breathing circuit for the purpose<br>of the introduction of the<br>aerosolized solution using<br>Aeroneb Solo Nebulizer System. | SAME<br>Function is to allow access to a<br>breathing circuit for the purpose of the<br>introduction of the aerosolized solution<br>using Aeroneb Solo Nebulizer System.<br><br>Both are passive adapters that facilitate<br>the connection of the Aeroneb Solo<br>Nebulizer with a larger breathing<br>circuit that delivers supplemental<br>oxygen. |
| Compatible<br>Device | Breathing circuit, multiple, not<br>specified<br>Aerogen Aeroneb SOLO<br>Nebulizer | SAME<br>Vapotherm HVT 2.0<br>Aerogen Aeroneb SOLO Nebulizer<br><br>Aerogen Solo Adapter IFU shows<br>multiple different configurations for<br>attaching the adapter to various<br>breathing circuits. The Subject Device<br>is more specific to HVT 2.0 |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Vapotherm. The logo consists of a blue circle with white lines inside, followed by the word "VAPOTHERM" in blue, with a registered trademark symbol next to it. The logo is simple and modern, and the colors are calming and professional.
Image /page/6/Picture/1 description: The image is a logo for Hi-VNI Technology. The logo is primarily blue, with the words "Hi-VNI" in a stylized font. A green swoosh underlines the "VNI" portion of the logo. Below the swoosh, the word "TECHNOLOGY" is written in a smaller, sans-serif font.
## Non-clinical performance testing:
## Biocompatibility / Materials -
Biocompatibility testing was conducted in accordance with FDA guidance, Use of International Standard ISO 10993-1, Biological evaluation of medical devices – Part 1: Evaluation and testing
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Vapotherm. The logo consists of a blue circle on the left with a white stylized "V" inside. To the right of the circle is the word "VAPOTHERM" in blue, with a registered trademark symbol after the word.
Image /page/7/Picture/1 description: The image shows the logo for Hi-VNI Technology. The logo is primarily blue, with a green swoosh underneath the "Hi" portion of the name. The word "Technology" is written in a smaller, sans-serif font below the main logo. A registered trademark symbol is present next to the "I" in "VNI".
within a risk management process" and ISO 18562-1 - Biocompatibility evaluation of breathing gas pathways in healthcare applications. The AAA-2 was determined to possess acceptable biocompatibility risk.
### Bench / Performance Testing -
Testing was performed to assure that the proposed design does not raise new questions of safety and effectiveness based on the proposed device intended use. Testing included:
Aerosolized solution output testing was performed utilizing a 500mW green laser to confirm delivery, with a nasal cannula patient and trach adapter interface.
Achievable flow rate consisted of testing the AAA-2 with HVT 2.0 while varying flow rates set to 8, 20, and 45 L/min, with varying cannula sizes which included Infant, Pediatric Small, and Adult XL cannulas, to confirm the set flows were achieved.
Occlusion testing with associated alarm included running the AAA-2 with HVT 2.0 at varying flow rates of 3 and 8 L/min for the infant cannula, 3 and 20 L/min for the Pediatric Small, and 3 and 45 L/min for the Adult XL.
The testing described as 'Volume of liquid exiting the adapter outlet' is to demonstrate that excessive rain-out does not occur. The system passed as the maximum of 5 ml of rainout was not exceeded with the AAA-2 placed in a worst-case orientation.
The physical connections testing included challenging all connection points under varying flow rates and worst-case simulated conditions. The 22mm female port that accommodates the connection to the Aerogen Solo Nebulizer was also demonstrated to be compliant to ISO 5356-1.
Continuous use testing was performed while running the HVT 2.0 at varying flow rates and temperatures and there were no signs of component or material degradation.
Aerosolization performance comparison testing for commonly nebulized solutions was completed to demonstrate similar performance for both the AAA-2 Aerosol Adapter and the Aeroneb Solo Adapter when used with an Aeroneb Solo Nebulizer System. To determine particle size distributions and respirable dose of the nebulized solution, nebulized solution respules were collected using the Next Generation Cascade Impactor (NGI) and demonstrated that the use of HVT 2.0 with AAA-2 did not interfere with the expected performance of the Aeroneb Solo device.
Usability/Human Factors testing was performed with results analyzed to identify potential use errors, close calls, and difficulties associated with the critical tasks. The results of this testing supports a substantial equivalence determination.
#### Substantial Equivalence Conclusion
The performance testing demonstrates that the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.