K220852 · Icare Finland OY · HKY · Oct 6, 2022 · Ophthalmic
Device Facts
Record ID
K220852
Device Name
iCare IC200
Applicant
Icare Finland OY
Product Code
HKY · Ophthalmic
Decision Date
Oct 6, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1930
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K220852 · Oct 6, 2022
iCare IC200
Icare Finland OY
Retrospective clinical data analysis of iCare IC200 IOP measurements
Retrospective analysis was used to compare the performance of the new 'Quick Measure' feature against clinical data from the predicate device to demonstrate substantial equivalence and adherence to ANSI Z80.10:2014 performance goals.
Retrospective clinical data analysis of Quick Measure function; Retrospective clinical data analysis
Human eyes (N=149 total); Sample Size: 149 eyes
GAT reference tonometer (Sitting) and Perkins reference (Supine)
Accuracy within ±5 mmHg vs. GAT/Perkins reference
Indications for Use
The iCare IC200 tonometer is intended to be used for the measurement of intraocular pressure of the human eye.
Device Story
Hand-held, battery-operated rebound tonometer; measures intraocular pressure (IOP) without topical anesthesia. Device uses sterile, single-use probe that makes brief contact with eye; measures probe deceleration and rebound time to calculate IOP. Deceleration is slower at low IOP; faster at high IOP. Operated by healthcare professionals in clinical settings. Features 'Quick Measure' mode (2-3 measurements) and 'Default' mode (6 measurements). Provides IOP readings to clinicians to assist in clinical decision-making. Benefits include rapid, non-invasive IOP assessment across various patient positions (horizontal to supine).
Clinical Evidence
Retrospective clinical data analysis compared 'Quick Measure' mode to predicate IC200 data (N=149 eyes). Performance evaluated against GAT (sitting) and Perkins (supine) reference tonometers. Results showed high agreement: 0.0% to 6.2% of measurements outside ±5 mmHg across low, medium, and high IOP subgroups. Bench testing using manometrically controlled artificial cornea confirmed high agreement with manometric pressure and default measurement mode, regardless of operator or angle.
Indicated for measurement of intraocular pressure in human eyes by healthcare professionals.
Regulatory Classification
Identification
A tonometer and accessories is a manual device intended to measure intraocular pressure by applying a known force on the globe of the eye and measuring the amount of indentation produced (Schiotz type) or to measure intraocular tension by applanation (applying a small flat disk to the cornea). Accessories for the device may include a tonometer calibrator or a tonograph recording system. The device is intended for use in the diagnosis of glaucoma.
{0}------------------------------------------------
October 6, 2022
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Icare Finland Oy Hannes Hyvönen Regulatory Affairs Manager Ayritie 22 Vantaa. 01510 Finland
Re: K220852
Trade/Device Name: iCare IC200 Regulation Number: 21 CFR 886.1930 Regulation Name: Tonometer And Accessories Regulatory Class: Class II Product Code: HKY Dated: September 5, 2022 Received: September 8, 2022
Dear Hannes Hyvönen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices
{1}------------------------------------------------
or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Elvin Ng Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia. Respiratory, ENT & Dental Devices Office of Product Evaluation and Ouality (OPEO) Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K220852
Device Name iCare IC200
Indications for Use (Describe)
The iCare IC200 tonometer is intended to be used for the measurement of intraocular pressure of the human eye.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# ical
iCare IC200 Traditional 510(k)
# 510(k) Summary
## 1.1 Submitter Name and Address
Icare Finland Oy Äyritie 22 Vantaa, Finland FI-01510
Contact: Hannes Hyvönen Phone: 358 9 8875 1150 Email: regulatory(at)icare-world.com Date prepared: October 5th, 2022
## 1.2 Device Name
| Trade Name: | iCare IC200 |
|--------------------------|---------------------------|
| Type/model: | TA031 |
| Common/Usual Name: | Tonometer |
| Classification Name: | Tonometer and Accessories |
| Regulation No: | 21 CFR 886.1930 |
| Device Regulatory Class: | II |
| Review Panel: | Ophthalmic |
| Product Code: | HKY |
Premarket Notification 510(k) Number: -
This premarket notification type is Traditional 510(k).
# 1.3 Predicate Device
Predicate device: iCare IC200 tonometer (type TA031), K190316.
The predicate device has not been subject to a design-related recall.
{4}------------------------------------------------
icare
## 1.4 Indications for Use / Intended Use
"The iCare IC200 tonometer is intended to be used for the measurement of intraocular pressure of the human eye."
The intended use of the iCare IC200 tonometer remains unchanged by this design modification.
## 1.5 Device Description
The iCare IC200 tonometer (model TA031) is a hand-held, battery-operated device which measures intraocular pressure (IOP) without the need for topical anaesthesia by rebound tonometry. The tonometer is to be used by a healthcare professional.
The tonometer uses the rebound method. A small and light, sterile, single-use probe makes brief contact with the eye. The device measures the deceleration of the probe and the rebound time and calculates the IOP from these parameters. Deceleration of the probe is slower at low IOP compared to high IOP. The measurement method, the IOP measurement algorithm and rebound technology (including disposable probe) are identical with the predicate device.
iCare IC200 tonometer can obtain IOP measurements in all angles between horizontal and supine directions. External design has not been modified. The user interface menu has been updated to contain a Quick Measure mode. The new Quick Measure feature is used to measure patient's IOP with fewer rebound measurements and faster measurement cycle. Quick Measure takes two or three rebound measurements; two if both results are within 2 mmHg and third if the difference between the first two measurements is greater than 2 mmHg.
## 510(k) Summary
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "icare" in a bold, sans-serif font. The letters are black, and there is a thin blue line underneath the word. The word appears to be a logo or brand name.
# 1.6 Comparison of Technological Characteristics with the Predicate Device
Similarities of the subject device and the predicate device is presented in the Table 1 below:
| | Table 1 Similarities of the subject device and the predicate device | | | | |
|--|---------------------------------------------------------------------|--|--|--|--|
|--|---------------------------------------------------------------------|--|--|--|--|
| Characteristic | Modified device (subject of this 510k) | Cleared Device (K190316) |
|----------------------------------------------------|--------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product<br>Appearance and<br>materials | Image: Modified device | Image: Cleared device |
| | Same | Same |
| Product/Device<br>Identification | Same | iCare IC200 Tonometer<br>(Type: TA031) |
| Intended Use /<br>Indications for<br>Use Statement | Same | IOP Measurement<br>“The iCare IC200 tonometer is intended<br>to be used for the measurement of<br>intraocular pressure of the human eye.” |
| Intended users | Same | Healthcare professionals |
| Measurement<br>method | Same | Rebound tonometry |
| Measurement<br>range | Same | 7-50 mmHg |
| Default<br>measurement<br>sequence | Same | Default measurement mode: 6<br>measurements, calculation of median IOP<br>and standard deviation based on 4<br>measurements (highest and lowest<br>excluded) |
| Versatility of<br>Measurement<br>position | Same | Tonometer can be used in any angle<br>between 0° (sitting, standing) and 90°<br>(patient in supine position) |
| Device Display | Same | 1.50" OLED display |
| Automatic eye<br>recognition<br>system | Same | Not applicable |
| Design | Same | Handheld microprocessor based |
| Calibration | Same | No maintenance calibration required |
| Contact tip<br>(probe) | Same | Lightweight, disposable, single use,<br>plastic probe<br>(TP01s) |
| Contact tip<br>sterilization | Same | Gamma-sterilized |
| Anaesthesia<br>required | Same | No |
| Power supply | Same | 4 x 1,5V AA Alkaline LR6 batteries |
| Device<br>dimensions and<br>weight | Same | Dimensions:<br>43mm x 104mm x 214mm<br>Weight: 165g (without batteries), 267g<br>(with batteries) |
| Connectivity<br>interface | Same | Bluetooth (Microchip RN4678 Module) |
| Electronics | Same | Inclination sensor, External Flash<br>memory, Real time clock |
| Mechanics | Same | As described in K190316 |
| User interface | Audio indications:<br>Same | Audio indications:<br>Beeps for device too near or too far<br>situation, and for other errors. |
| | Graphical User Interface:<br>Same | Graphical User Interface:<br>OLED display |
| | Probe base light:<br>Same | Probe base light:<br>Green: angle correct, device readiness |
| | Red: angle incorrect, measurement not possible<br>Blinking red: measurement error messages | |
<sup>510(</sup>k) Summary
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "icare" in a bold, sans-serif font. The word is black and is positioned on a white background. A thin blue line is located directly underneath the word.
{7}------------------------------------------------
# ical
iCare IC200 Traditional 510(k)
Changes listed in Table 2 below.
| Characteristic | Modified device (subject of this 510k) | Cleared Device (K190316) |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|
| Quick Measure<br>Feature | Difference: a new Quick Measure feature<br>added to measure patient's IOP with fewer<br>rebound measurements and faster<br>measurement cycle.<br>-Quick Measure takes two or three rebound<br>measurements; two if both results are<br>within 2mmHg and third if the difference<br>between the first two measurements is<br>greater than 2mmHg. | Not applicable |
| User interface<br>color | Graphical User Interface Colors:<br>UI in Default measurement mode remains<br>blue<br>Difference: measurement sequence<br>information and results in Quick Measure<br>mode are shown in magenta color to<br>differentiate from default measurement<br>mode | Graphical User Interface Colors:<br>Blue color |
| Alignment error<br>indication | Alignment error remains enabled in default<br>measurement mode.<br>Difference: alignment error is disabled in<br>Quick Measure mode. | Enabled in default measure mode |
| Device Software<br>Accessories | Same during default measurement<br>Difference: iCare CLINIC/EXPORT is not<br>usable for Quick Measure results | iCare CLINIC, iCare EXPORT |
## Table 2 Differences of the subject device to the predicate device
## 510(k) Summary
{8}------------------------------------------------
| Labelling | Difference: User manual is updated to<br>include Quick Measure instructions and<br>applicable warnings.<br>Minor edits to quick guide and type label. | User manual, quick guide, type label |
|-----------|-------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|
|-----------|-------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|
## 1.7 Performance and Safety Data
The device has been tested according to relevant FDA recognized consensus standards. The following performance and safety data are provided in support of the substantial equivalence:
- ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012 C1:2009/(R)2012 and . A2:2010/(R)2012 Medical electrical equipment -- Part 1: General requirements for basic safety and essential performance
- ISO 15004-1:2006 Ophthalmic instruments Fundamental requirements and test . methods - Part 1: General requirements
- IEC 60601-1-6:2010 + A1:2013 Medical electrical equipment Part 1-6: General ● requirements for basic safety and essential performance - Collateral standard: Usability
- IEC 62304:2006 + A1:2015 Medical device software Software life-cycle . processes (Device firmware and software level of concern: Class B)
- . ANSI Z80.10:2014 Ophthalmic Instruments - Tonometers
There were no changes related to sterilization, biocompatibility, EMC, or light hazard protection. Device and software risk analysis has been performed in accordance with ISO 14971 risk management standard.
Quick Measure performance data was obtained from a retrospective clinical data analysis in which Quick Measure function was retrospectively compared to the clinical data of iCare IC200. In the analysis Quick Measure met ANSI Z80.10:2014 (in accordance with
## 510(k) Summary
{9}------------------------------------------------
FDA's extent of recognition) performance goals per GAT reference tonometer (see Table 3).
| Group | N (eyes) | Outside ±5 mmHg<br>vs. GAT reference<br>(Sitting) | Outside ±5 mmHg<br>vs. Perkins reference<br>(Supine) |
|-----------------------------------------|----------|---------------------------------------------------|------------------------------------------------------|
| GAT group = low<br>≤ 16 mmHg | 44 | 0/44<br>(0.0 %) | 0/44<br>(0.0 %) |
| GAT group = medium<br>> 16 to < 23 mmHg | 65 | 1/65<br>(1.5 %) | 4/65<br>(6.2 %) |
| GAT group = high<br>≤ 23 mmHg | 40 | 1/40<br>(2.5 %) | 0/40<br>(0.0 %) |
## Table 3 Comparability Chart between IC200 Quick Measure and GAT, and Perkins, within IOP subgroups
Accuracy of the iCare IC200 quick measure mode was assessed in a bench test using a manometrically controlled artificial cornea. Testing compared the new Quick Measure mode against the existing default measurement mode (same as in predicate device IC200), clinically validated subnormal reference tonometer (iCare IC200), and manometric pressure.
Repeatability and accuracy testing completed with iCare IC200 quick measure mode demonstrated high agreement with manometric pressure and reference tonometer used in default measurement mode (iCare IC200).
Reproducibility test completed with iCare IC200 quick measure mode demonstrated high agreement with manometric pressure regardless of the device, operator, or the measurement angle.
{10}------------------------------------------------
## Software Verification and Validation
Testing Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Moderate" level of concern, since a failure or latent flaw in the software could indirectly result in minor injury to the patient or operator through incorrect or delayed information or through the action of a care provider.
## 1.8 Substantial Equivalence
The modified iCare IC200 including the Quick measure mode is substantially equivalent to the predicate device. The devices have the same or substantially equivalent technological characteristics and the iCare IC200 raises no new issues of safety and effectiveness. Performance and safety data demonstrate that the modified iCare IC200 is as safe and effective as the predicate device.
<sup>510(</sup>k) Summary
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.