The CORE-SNARE is used endoscopically in the removal of diminutive polyps, sessile polyps, pedunculated polyps and tissue from within the GI tract.
Device Story
Disposable endoscopic electrosurgical snare; used for polyp/tissue removal in GI tract. Device comprises snare loop, catheter tube, and handle. Physician inserts catheter through endoscope operating channel; manipulates handle to deploy/rotate loop around lesion. Connected to external electrosurgical generator; delivers monopolar electrical current to cut tissue. Single-patient use; sterile. Benefits include effective tissue resection during endoscopic procedures.
Clinical Evidence
Bench testing only. No clinical data. Performance testing included dimensional analysis, tensile strength (electrode cable, inner rope, catheter-handle), withstand voltage, high-frequency leakage current, and electrical continuity per ANSI/AAMI HF18:2001 and IEC 60601-2-2:2017. Biocompatibility testing per ISO 10993-1 (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity). Accelerated aging per ASTM F1980.
Technological Characteristics
Materials: SUS 304 (loop), PTFE (catheter tube), ABS (handle). Energy: Monopolar electrosurgical current. Form factor: Disposable snare with various loop shapes (oval, crescent, diamond) and sizes (10-30mm). Connectivity: Connects to external electrosurgical generator. Sterilization: Ethylene Oxide (EO). Standards: IEC 60601-2-2:2017, ANSI/AAMI HF18:2001, ISO 10993, ISO 11135, ASTM F1980.
Indications for Use
Indicated for endoscopic removal of diminutive, sessile, and pedunculated polyps and tissue from the GI tract.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
November 21, 2022
Incore Co., Ltd. Jae-Hun Lee Department head of Regulatory Affairs 11, Hyeoksin-daero, 78-gil, Dong-gu Daegu, 41072 KOREA, SOUTH
Re: K220846
> Trade/Device Name: Core-Snare Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic electrosurgical unit and accessories Regulatory Class: Class II Product Code: FDI Dated: March 15, 2022 Received: March 23, 2022
Dear Jae-Hun Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Sivakami Venkatachalam -S
#### for
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K220846
Device Name CORE-SNARE
Indications for Use (Describe)
The CORE-SNARE is used endoscopically in the removal of diminutive polyps, pedunculated polyps and tissue from within the GI tract.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the words CORE-SNARE in large, bold, black font. Below that, the words Traditional 510(K) are written in a smaller, bold, black font. The image is simple and contains only text.
Image /page/3/Picture/2 description: The image shows the logo for INCORE. The logo consists of three overlapping circles on the left side, with the word "INCORE" in bold, sans-serif font to the right. Below the word "INCORE" is a thin line, and below that is the text "innovative core technologies of medical devices" in a smaller font size.
#### Section 5 - 510(K) Summary
In accordance with the Food and Drug Adminstration Rule to implement provisions of the Safe Medical Device Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Safety and Effectiveness for the use of the CORE-SNARE.
#### I. Submitted by
Company Name : INCORE CO.,LTD. Company Address : 11, Hyeoksin-daero 78-gil, Dong-gu, Daegu, Republic of Korea
Contact Person : Mr. Jae-Hun, Lee Department head of Regulatory Affairs
Contact Phone : (82) 2-866-3514 Contact Fax : (82) 2-6919-1346
Date of Preparation : March 15, 2022
#### II. Device
Trade of Device : CORE-SNARE
- Model : IC-SN0210, IC-SN0215, IC-SN0310, IC-SN0315, IC-SN0324, IC-SN0330, IC-SN0410, IC-SN0413, IC-SN0415, IC-SN0424, IC-SN0430, IC-SN3310, IC-SN3313, IC-SN3315, IC-SN3324, IC-SN3330, IC-SN0410CR, IC-SN0415CR, IC-SN0424CR, IC-SNR0210, IC-SNR0215, IC-SNR0310, IC-SNR0313, IC-SNR0315, IC-SNR0324, IC-SNR0410, IC-SNR0413, IC-SNR0415, IC-SNR0424, IC-SNR0330, IC-SNR0430, IC-SNR3310, IC-SNR3313, IC-SNR3315, IC-SNR3324, IC-SNR3330, IC-SNR0410CR, IC-SNR0415CR, IC-SNR0424CR
Common or Usual Name : Disposable electrically-operated medical snare Classification Name : Endoscopic Electrosurgical Unit and Accessories Device Product Code : FDI Review Panel : Gastroenterology/Urology Regulation Number: 21 CFR 876.4300 Endoscopic Electrosurgical Unit and Accessories. Regulatory Class : Class II
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for "INCORE", with the tagline "innovative core technologies of medical devices" underneath. To the left of the text is a graphic of three overlapping circles, with the intersection of the circles filled in with a darker color. The text is in a sans-serif font and is black in color. The tagline is in a smaller font size than the main logo.
#### Ⅲ, Predicate Device
Device Name : ClearGrasp Snare Manufacturer : Finemedix Co. Ltd 510(K) Number : K183289 Classification Name : Endoscopic Electrosurgical Unit and Accessories Product Code : FDI Regulatory Class : Class II Regulation Number: 21 CFR 876.4300 Endoscopic Electrosurgical Unit and Accessories.
The predicates have not been subject to a design-related recall. No reference devices were used in this submission.
#### IV. Device Description
The CORE-SNARE is a disposable device for cuts out and removes the tissues such as polyps after entering a site to be treated during endoscopic procedures.
The device consists of Snare-loop connected with catheter tube and Catheter tube pass through operating channel of endoscope and Handle connected with snare through the catheter tube. It is available in various sizes and working lengths.
Users can choose an Oval or a Crescent or Diamond type based on their preference and the characteristic of the loop sizes can be chosen in accordance with the size of the lesion and the rotational type helps approaching and grasping the lesion.
When connected to an electrosurgical generator and activated, the loop delivers a monopolar electrical current to the surgical site.
The proposed product is packed in a sealed pouch following EO Sterilization (SAL 10t%). This device is supplied sterile for single-patient use and shall be not reused or re-sterilized.
#### V. Indications for Use
The CORE-SNARE is used endoscopically in the removal of diminutive polyps, sessile polyps, pedunculated polyps and tissue from within the GI tract.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for INCORE, a company that specializes in innovative core technologies for medical devices. The logo features a stylized graphic of overlapping circles in the upper left corner, followed by the company name in bold, sans-serif font. Below the name is a tagline that describes the company's focus on innovative core technologies for medical devices. The overall design is clean and modern, conveying a sense of innovation and expertise.
#### VI. Comparison of Technological Characteristics with predicate device
The CORE-SNARE has substantially equivalent device design, configuration, packaging fundamental technology, sterilization process and intended use as those featured in the predicate device Finemedix CO.,Ltd's ClearGrasp Snare, K183289. the differences between the proposed device and the predicated devices do not raise any questions regarding its safety and effectiveness. the differences are listed in the table below.
The following tests evaluate the substantial equivalence of the subject device through the performance test applying the same standard as the predicate device.
Performance testing such as appearance, dimension, continuity, tensile strength test of electrode cable, withstand voltage, high frequency leakage current test of electrode cable according to IEC 60601-2-2:2017 and ANSI/AAMI HF18:2001. The test results show that the subject device is substantially equivalent to the predicate device.
| Item | Proposed Device | Predicate Device | Substantial<br>Equivalence |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Classfication<br>regulation | CORE-SNARE | ClearGrasp Snare | |
| | 21 CFR 876.4300 | 21 CFR 876.4300 | Same |
| Classification<br>and Code | Class II , FDI | Class II , FDI | Same |
| Device<br>Classfication<br>Name | Endoscopic Electrosurgical<br>Unit and Accessories | Endoscopic Electrosurgical<br>Unit and Accessories | Same |
| 510(K) number | K220846 | K183289 | - |
| Indications for<br>Use | The CORE-SNARE is<br>used endoscopically in the<br>removal of diminutive<br>polyps, sessile polyps,<br>pedunculated polyps and<br>tissue from within the GI<br>tract. | The ClearGrasp Snare is<br>used endoscopically in the<br>removal of diminutive<br>polyps, sessile polyps,<br>pedunculated polyps and<br>tissue from within the GI<br>tract. | Same |
| Configuration | Loop, Catheter Tube,<br>Handle(Slider, Body) | Loop, Catheter Tube,<br>Handle, Slider, Plug | Similar |
| Table 1 : Comparison to Predicate Device | |
|------------------------------------------|--|
|------------------------------------------|--|
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for INCORE. The logo consists of a circular graphic to the left of the word "INCORE". The graphic is made up of three overlapping circles, with the intersection of the circles colored in blue. Below the word "INCORE" is a thin black line, and below that is some smaller text that is difficult to read.
| | Loop | SUS 304 | SUS 304 | |
|----------|--------------------------------------|------------------|----------------------------------|-------------------|
| Material | Catheter<br>Tube | PTFE | PTFE | Same |
| | Handle | ABS | ABS | |
| | | Shape of Loop | Oval and Crescent and<br>Diamond | Oval and Crescent |
| | Working Length | 2200, 2400mm | 1600, 1800, 2200, 2400mm | Similar |
| | Loop Size | 10, 15, 24, 30mm | 6,10,13,15,24,30 mm | Similar |
| | Rotational/Non-<br>Rotational | Both | Both | Same |
| | Sterilization | EO Sterilization | EO Sterilization | Same |
| | Single Use | Yes | Yes | Same |
| | Used with<br>Electrosurgical<br>Unit | Yes | Yes | Same |
## VII. Non-clinical testing data
#### 1) Sterility
A sterility validation was completed following ISO 11135 requirements to demonstrate a 10° SAL. And the proposed device meets the requirements of ISO 10993-7 "Biological evaluation of medical devices - Part 7: ethylene oxide sterilization residuals"
The Sterility test is tested with a direct method.
No evidence of microbial growth is found, the test article to be examined complies with sterility test.
## 2) Biocompatibility Testing
Biocompatibility testing has been performed in accordance with ISO 10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing"
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for INCORE. The logo consists of three overlapping circles on the left side, with the word "INCORE" in bold, sans-serif font to the right. Below the word "INCORE" is a tagline in a smaller font that reads "Innovative core technologies of medical devices."
The CORE-Snare is considered tissue contacting for a duration of less than 24 hours.
The proposed device passed all biocompatibility tests.
In accordance to ISO10993-1, the following biocompatibility tests were performed : Cytotoxicity, Maximization, Intractutaneous Reactivity, Acute Systemic Toxicity, Pvrogen test.
#### 3) Performance Testing
#### (1) Demension test
When measured by vernier calipers, the tolerance of the nominal size should be ±10 % of the stndard. For the dimensional test, all dimensions of the core-snare non-rotation type product were measured. As a result of the measurement, it was confirmed that all dimensional values do not exceed the tolerance of ±10%. Therefore, the CORE-SNARE meets the performance required by demension test.
#### (2) Tensile strength test
#### (1) Electrode Cable
The tensile strength test method for Electrode Cable was performed in accordance with ANSI/AAMI HF 18 (2001) 4.2.5.5, and according to method, Electrode cables and connecting parts shall not be damaged from 40 N pulling and 0.64 J impact. The proposed device was tested, and there was no damaged from 40 N pulling and 0.64 J impact.
Therefore, the CORE-SNARE meets the performance required by tensile strength of electrode cable.
#### (2) Inner rope-cutting wire
The tensile strength (inner rope-cutting wire) was tested as follows.
The cutting knife is cut out from the sheath by manipulating the handle. The inner rope and the cutting wire are cut at a position of about 15cm from the end to make a specimen, and the specimen is pulled at a rate of 1.0 in/min.
Under the conditions of the above method, the strength between the inner rope and the cutting wire should be not less than 15 lbs.
The proposed device was tested, and It was confirmed the strength is more than 15 lbs.
Therefore, the CORE-SNARE meets the performance required by tensile strength of
{8}------------------------------------------------
Image /page/8/Picture/2 description: The image shows the logo for INCORE. The logo features a graphic of three overlapping circles on the left side. To the right of the graphic is the word "INCORE" in large, bold, sans-serif font. Underneath the word "INCORE" is a thin line, and below that is some smaller text that is difficult to read.
inner rope-cutting wire.
#### 3 Catheter sheath-handle
The tensile strength (catheter sheath-handle) was tested as follows. After manipulating the handle, pull out the cutting knife out of the sheath, cut the sheath and cable at 2 inches of the distal end of the strain relief to make a specimen, and pulled at a rate of 1.0 in/min. Under the conditions of the above method, the strength between the catheter sheath and the handle should be not less than 2 lbs. The proposed device was tested, and there was more than 2 bs. Therefore, the CORE-SNARE meets the performance required by tensile strength of catheter sheath-handle.
- (3) Cable Withstand voltage
The main frequency withstand voltage meets the test criteria that the rate accessory voltage specified by the manufacturer of the high frequency surgical accessory must withstand a peak voltage of 1000V, large DC or Mains frequency for 5 minutes.
And the high-frequency withstand voltage must withstand for 30 seconds at high frequency voltage of 120% of the rated accessory voltage presented by the manufacturer. Maximum voltage of accessory is 920 Vp.
For the test, refer to the test method with the following IEC 60601-2-2:2011. The proposed device was tested according to the above criteria and passed. Therefore, the CORE-SNARE meets the performance required by Cable Withstand voltage.
- (4) High Frequency leakage current test
High frequency leakage current testing has been performed in accordance with ANSI / AAMI HF 18(2001) 4.2.5.2.
The monopolar high-frequency leakage current of the electrode cable shall be not more than 3.6 dfL mA and, the connection cord shall not have worked loose nor shall it show any damage.
For multiconductor cables there shall be no short circuits betwwen individual conductors.
The proposed device was tested according to the above criteria and passed.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for INCORE. The logo consists of a circular design on the left and the word "INCORE" in bold, sans-serif font on the right. Below the word "INCORE" is a thin line, and below that is some smaller text that is difficult to read. The circular design on the left appears to be three overlapping circles with a dark blue fill in the intersection.
Therefore, the CORE-SNARE meets the performance required by High frequency leakage current test
(5) Conductive Inspection (Continuity test)
This test is performed using electrical conduction circuit system, and tips for connection with electrode tip must be electrically connected to each other. The proposed device was tested according to the above criteria and passed. Therefore, the CORE-SNARE meets the performance required by Conductive inspection.
#### 4) Physical/Chemical
Extraction Test is tested accordance with In-house hold.
Test items are Appearance, pH, KMnO4 Consumption, Evaporating residue, Ultraviolet absorption, Heavy metal are acceptable level.
#### 5) Shelf Life Test
The CORE-SNARE has a maximum shelf life of 3 years from the date of sterilization. In compliance with the standard of ASTM F1980 ; Standard Guide For Accelerated Aging of Sterile Barrier System for Medical Device, accelerated aging studies for start and three(3) years were performed to determine product integrity over its lifespan, with acceptable results. A real-time aging study is currently in process to verify the result found in the accelerated aging studies.
## VIII. Conclusion
The conclusion drawn from the technological characteristics is that the CORE-SNARE has been found to have a safety and efficacy profile that is substantially equivalent to the predicate device ClearGrasp Snare which is marketed for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.