K220813 · Jun 17, 2022
ART-PLAN
Therapanacea
Retrospective clinical patient image datasets (CT, MR, PET); Routine clinical cancer treatment records
Retrospective clinical data was used to train, validate, and test the performance of AI-based auto-segmentation and synthetic-CT generation algorithms. The data represents routine clinical practice across multiple centers and vendors.
AI-based contouring; Synthetic-CT generation; Retrospective clinical data; Algorithm training and validation
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
ART-Plan AI Model Training and Validation Cohort; Retrospective analysis of clinical patient data
Cancer patients undergoing radiotherapy; Sample Size: 8,736 patients (374,160 images); Number of Sites: Multiple centers (not specified)
Clinical performance evaluation comparing pseudo-CT to standard CT for pelvis and brain.
—
Indications for Use
ART-Plan is a software for multi-modal visualization, contouring and processing of 3D images of cancer patients for whom radiotherapy treatment has been prescribed. It allows the user to view, create and modify contours for the regions of interest. It also allows to generate automatically, and based on medical practices, the contours for the organs at risk and healthy lymph nodes and to register combinations of anatomical and functional images. Contours and images require verifications, potential modifications, and subsequently the validation of a trained user with professional qualifications in anatomy and radiotherapy before their export to a Treatment Planning System. ART-Plan offers the following visualization, contouring and manipulation tools to aid in the preparation of radiotherapy treatment: - Multi-modal visualization and rigid- and deformable registration of anatomical and functional images such as CT, MR, PET-CT, 4D-CT and CBCT - Display of fused and non-fused images to facilitate the comparison and delineation of image data by the user - Manual modification and semi-automatic generation of contours for the regions of interest - Automatic generation of contours for organs at risk and healthy lymph nodes, based on medical practices, on medical images such as CT and MR images. - Generation of pseudo-CT for supported anatomies The device is intended to be used in a radiation therapy clinical setting, by trained professionals only.
Device Story
ART-Plan is a software application for radiation therapy planning; it processes 3D multi-modal medical images (CT, PET-CT, CBCT, 4D-CT, MR). The device uses AI-based deep learning algorithms to automatically segment organs at risk and healthy lymph nodes, and to generate pseudo-CT images from MRI data. It also provides tools for rigid and deformable image registration, manual/semi-automatic contouring, and image fusion. Used in clinical settings by radiation oncologists, dosimetrists, and medical physicists; the software is accessed via web browser (cloud or server-based). Healthcare providers use the output to visualize anatomy, delineate regions of interest, and prepare data for export to treatment planning systems. By automating contouring and registration, the device aims to standardize workflows, reduce manual effort, and improve efficiency in radiotherapy treatment planning.
Clinical Evidence
Bench testing only. No clinical studies were conducted. Performance was validated using retrospective clinical data cohorts (374,160 samples from 8,736 patients). Metrics included Dice Similarity Coefficient (DSC) compared to AAPM thresholds (mean DSC >= 0.8) or inter-expert variability (mean DSC >= 0.54), and qualitative clinical evaluation (85% acceptance rate). Pseudo-CT generation was validated using gamma passing criteria (2%/2mm >= 95%; 3%/3mm >= 99%) and mean dose deviation (<= 2%).
Technological Characteristics
Software-based medical image management and processing system. Uses deep learning algorithms for automatic segmentation and synthetic CT generation. Supports DICOM 3.0. Connectivity via web-based platform (Chrome/Firefox). Deployment is server-based or cloud-based. No physical materials; software-only device.
Indications for Use
Indicated for cancer patients for whom radiation treatment has been planned. Intended for use by trained medical professionals including radiation oncologists, dosimetrists, and medical physicists. Supports display/visualization of 3D multi-modal medical image data (CT, PET-CT, CBCT, 4D-CT, MR).
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
TheraPanacea % Edwin Lindsay QA/RA consultant Pépinière Cochin Paris Santé, 29 rue du Faubourg Saint-Jacques Paris. 75014 FRANCE
Re: K220813
Trade/Device Name: ART-Plan Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management and Processing System Regulatory Class: Class II Product Code: QKB Dated: March 16, 2022 Received: March 21, 2022
Dear Edwin Lindsay:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Julie Sullivan, Ph.D. Director DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
for
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K220813
Device Name ART-Plan
#### Indications for Use (Describe)
ART-Plan is indicated for cancer patients for whom radiation treatment has been planned. It is intended to be used by tramed medical professionals including, but not limited to, radiation oncologists, dosimetrists, and medical physicists.
ART-Plan is a software application intended to display and visualize 3D multi-modal medical image data. The user may import, define, display, transform and store DICOM3.0 compliant datasets (including regions of interest structures). These images, contours and objects can subsequently be exported/distributed within the system, across computer networks and/or to radiation treatment planning systems. Supported modalities include CT. PET-CT. CBCT, 4D-CT and MR images.
ART-Plan supports Al-based contouring on CT and MR images and offers semi-automatic and manual tools for segmentation.
To help the user assess changes in image data and to obtain combined multi-modal image information, ART-Plan allows the registration of anatomical and functional images and display of fused and non-fused images to facilitate the comparison of patient image data by the user.
With ART-Plan, users are also able to generate, visualize, evaluate and modify pseudo-CT from MRI images.
Type of Use (Select one or both, as applicable)
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
This 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92.
#### Submitter's Name:
TheraPanacea
#### Submitter's Address:
Pépinière Paris Santé Cochin 29 rue du Faubourg Saint-Jacques 75014 Paris France
Telephone: +33 9 62 52 78 19
#### Establishment Registration Number:
3019834893
#### Contact Person:
Edwin Lindsay
Telephone +44 (0) 7917134922
#### Date Prepared:
16 Mar 2022
#### Below summaries the Device Classification Information regarding the TheraPanacea ART-Plan:
#### Primary Product Code:
| Regulation<br>Number | Device | Device<br>Class | Product<br>Code | Classification<br>Panel |
|----------------------|------------------------------------------------------|-----------------|-----------------|-------------------------|
| 892.2050 | Medical image<br>management and<br>processing system | Class II | QKB | Radiology |
#### Device Trade Name:
ART-Plan
Device Common Name:
ART-Plan
#### Intended Use:
Page 1 of 28
{4}------------------------------------------------
ART-Plan is a software for multi-modal visualization, contouring and processing of 3D images of cancer patients for whom radiotherapy treatment has been prescribed.
It allows the user to view, create and modify contours for the regions of interest. It also allows to generate automatically, and based on medical practices, the contours for the organs at risk and healthy lymph nodes and to register combinations of anatomical and functional images. Contours and images require verifications, potential modifications, and subsequently the validation of a trained user with professional qualifications in anatomy and radiotherapy before their export to a Treatment Planning System.
ART-Plan offers the following visualization, contouring and manipulation tools to aid in the preparation of radiotherapy treatment:
- Multi-modal visualization and rigid- and deformable registration of anatomical and O functional images such as CT, MR, PET-CT, 4D-CT and CBCT
- O Display of fused and non-fused images to facilitate the comparison and delineation of image data by the user
- O Manual modification and semi-automatic generation of contours for the regions of interest
- Automatic generation of contours for organs at risk and healthy lymph nodes, based 0 on medical practices, on medical images such as CT and MR images.
- Generation of pseudo-CT for supported anatomies 0
The device is intended to be used in a radiation therapy clinical setting, by trained professionals only.
#### Indications for Use:
ART-Plan is indicated for cancer patients for whom radiation treatment has been planned. It is intended to be used by trained medical professionals including, but not limited to, radiologists, radiation oncologists, dosimetrists, and medical physicists.
ART-Plan is a software application intended to display and visualize 3D multi-modal medical image data. The user may import, define, display, transform and store DICOM3.0 compliant datasets (including regions of interest structures). These images, contours and objects can subsequently be exported/distributed within the system, across computer networks and/or to radiation treatment planning systems. Supported modalities include CT, PET-CT, CBCT, 4D-CT and MR images.
ART-Plan supports Al-based contouring on CT and MR images and offers semi-automatic and manual tools for seqmentation.
To help the user assess changes in image data and to obtain combined multi-modal image information, ART-Plan allows the registration of anatomical and functional images and display of fused and non-fused images to facilitate the comparison of patient image data by the user.
With ART-Plan, users are also able to generate, visualize, evaluate and modify pseudo-CT from MRI images.
{5}------------------------------------------------
#### Summary of Substantial Equivalence:
The following predicate devices have been that the ART-Plan can claim equivalence with and these are detailed below
#### General Comparison
| | General Information | | | | | | | |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Property | Proposed Device<br>ART-Plan v1.10.0 | Primary<br>Predicate<br>ART-Plan<br>v1.6.1 | Reference<br>device<br>Contour<br>ProtégéAl | Reference<br>device<br>MIM 4.1 | Reference<br>device<br>MRCAT Pelvis | Reference<br>device<br>MRCAT Brain | Reference<br>device<br>Syngo.via RT<br>Image Suite | Comment |
| Common<br>Name | Radiological image<br>processing software for<br>radiation therapy | Radiological image<br>processing software<br>for radiation therapy | Radiological<br>Image<br>Processing<br>Software For<br>Radiation<br>Therapy | System, image<br>processing,<br>radiological | System,<br>Planning,<br>Radiation<br>Therapy<br>Treatment | System,<br>Planning,<br>Radiation<br>Therapy<br>Treatment | System,<br>Planning,<br>Radiation<br>Therapy<br>Treatment | N/A |
| Device<br>Manufacturer | TheraPanacea | TheraPanacea | MIM Software,<br>Inc | MIMvista Corp<br>(now MIM<br>Software Inc) | Philips Medical<br>Systems | Philips Medical<br>Systems | Siemens<br>Medical<br>Solutions USA,<br>Inc. | N/A |
| 510k | N/A | K202700 | K210632 | K071964 | K182888 | K193109 | K173635 | N/A |
| Device<br>Classification | II | II | II | II | II | II | II | N/A |
| Primary<br>Product Code | QKB | QKB | QKB | LLZ | MUJ | MUJ | MUJ | The primary product code<br>is QKB "Radiological<br>Image Processing<br>Software For Radiation<br>Therapy" as the software<br>uses Al algorithms and is<br>intended for radiation<br>therapy, just like the<br>primary predicate device |
| Secondary<br>Product Code | LLZ, MUJ | LLZ | - | - | - | - | LLZ | As secondary product<br>code:<br>LLZ (System, Image<br>Processing,<br>Radiological) was<br>included as the<br>software is used in |
| | | | | | | | | image processing and<br>some predicates use it<br>as primary or<br>secondary product<br>code:<br>MUJ (System,<br>Planning, Radiation<br>Therapy Treatment)<br>was includes as it is a<br>software used in the<br>planning of<br>radiotherapy<br>treatment and some of<br>the reference devices<br>use it as their primary<br>code |
| Target<br>Population | Any patient type for<br>whom relevant modality<br>scan image data is<br>available | Any patient type for<br>whom relevant<br>modality scan data is<br>available | Not stated | Not stated | Any patient with<br>soft tissue<br>cancers in the<br>pelvic region for<br>whom<br>radiotherapy<br>treatment has<br>been planned | Any patient with<br>primary and<br>metastatic brain<br>tumor for whom<br>radiotherapy<br>treatment has<br>been planned | Not stated | The proposed device has<br>identical target<br>populations to the primary<br>and reference devices. |
| Environment | Hospital | Hospital | Hospital | Hospital | Hospital | Hospital | Hospital | The proposed device<br>and predicates have<br>identical target<br>environments |
| Intended Use/<br>Indication for<br>Use | Intended Use<br>ART-Plan is a software<br>for multi-modal<br>visualization, contouring<br>and processing of 3D<br>images of cancer<br>patients for whom<br>radiotherapy treatment<br>has been prescribed.<br>It allows the user to<br>view, create and modify<br>contours for the regions<br>of interest. It also allows<br>to generate<br>automatically, and<br>based on medical | Intended Use<br>ART-Plan is a<br>software designed to<br>assist the contouring<br>process of the target<br>anatomical regions<br>on 3D-images of<br>cancer patients for<br>whom radiotherapy<br>treatment has been<br>planned.<br>The SmartFuse<br>module allows the<br>user to register<br>combinations of<br>anatomical and | Intended Use<br>Contour<br>ProtégéAl is an<br>accessory to<br>MIM software<br>used for the<br>contouring of<br>anatomical<br>structures in<br>imaging data<br>using<br>machine-learnin<br>g-based<br>algorithms<br>automatically.<br>Appropriate | Intended Use<br>MIM 4.1<br>(SEASTAR)<br>software is<br>intended for<br>trained medical<br>professionals<br>including, but not<br>limited to,<br>radiologists,<br>oncologists,<br>physicians,<br>medical<br>technologists,<br>dosimetrists and<br>physicists. | Intended Use:<br>MRCAT imaging<br>is<br>intended to<br>provide the<br>operator with<br>information of<br>tissue<br>properties<br>for radiation<br>attenuation<br>estimation<br>purposes<br>in photon<br>external beam<br>radiotherapy | Intended Use:<br>MRCAT imaging<br>is intended to<br>provide the<br>operator with<br>information of<br>tissue<br>properties for<br>radiation<br>attenuation<br>estimation<br>purposes in<br>photon external<br>beam<br>radiotherapy | Intended use:<br>Not available in<br>the summary<br>Indication for<br>use:<br>syngo.via RT<br>Image Suite is a<br>3D and 4D<br>image<br>visualization,<br>multimodality<br>manipulation<br>and<br>contouring tool<br>that helps the | The intended use and<br>indications for use of<br>the proposed device,<br>ART-Plan v1.10.0 and<br>the primary predicate<br>ART-Plan v1.6.1 are<br>similar as they are<br>both intended for<br>medical image<br>registration and<br>segmentation in the<br>context of radiotherapy<br>treatment planning:<br>they allow multi-modal<br>and mono-modal rigid |
| practices, the contours | functional images and | image | MIM 4.1<br>(SEASTAR) is a<br>medical image<br>and information<br>management<br>system that is<br>intended to<br>receive, transmit,<br>store, retrieve,<br>display, print and<br>process digital<br>medical images,<br>as well as create,<br>display and print<br>reports from those<br>images. The<br>medical<br>modalities of<br>these medical<br>imaging systems<br>include, but are<br>not limited to, CT,<br>MRI, CR, DX,<br>MG, US, SPECT,<br>PET and XA as<br>supported by<br>ACR/NEMA<br>DICOM 3.0. | treatment<br>planning. | treatment<br>planning. | preparation and<br>response<br>assessment of<br>treatments such<br>as, but not<br>limited to those<br>performed with<br>radiation (for<br>example,<br>Brachytherapy,<br>Particle<br>Therapy,<br>External<br>Beam Radiation<br>Therapy).<br>It provides tools<br>to efficiently<br>view existing<br>contours,<br>create, edit,<br>modify, copy<br>contours of<br>regions of<br>the body, such<br>as but not<br>limited to, skin<br>outline, targets<br>and<br>organs-at-risk. It<br>also provides<br>functionalities to<br>create and<br>modify simple<br>treatment plans.<br>Contours,<br>images and<br>treatment plans<br>can<br>subsequently be<br>exported to a<br>Treatment<br>Planning<br>System.<br>The software<br>combines<br>following digital | deformable registration<br>for the same modalities<br>of images (CT, MR, PET<br>they allow automatic<br>segmentation of<br>organs-at-risk and lymph<br>nodes on injected and<br>non-injected CT images<br>using deep learning<br>algorithms<br>they allow the import,<br>manipulation,<br>visualisation, generation<br>and the export of DICOM<br>images<br>The intended for the<br>proposed device has<br>been adapted to<br>provide a more specific<br>description of the<br>proposed device but<br>does not represent a<br>new intended use,<br>except for the<br>additional claims for<br>the proposed device<br>as compared to the<br>primary predicate as:<br>- it includes an improved<br>version of the existing<br>automatic segmentation<br>tool as compared to the<br>one of ART-Plan v1.6.1.<br>- it allows automatic<br>segmentation on more<br>anatomies and<br>organ-at-risk | |
| for the organs at risk | display them with | visualization<br>software must<br>be used to<br>review and, if<br>necessary, edit<br>results<br>automatically<br>generated by<br>Contour<br>ProtégéAI.<br>Contour<br>ProtégéAI is not<br>intended to<br>detect or<br>contour lesions. | MIM 4.1<br>(SEASTAR) provides the user<br>with the means to<br>display, register<br>and fuse medical<br>images from<br>multiple<br>modalities.<br>Additionally, it<br>evaluates cardiac<br>left ventricular<br>function and<br>perfusion,<br>including left<br>ventricular | Indications for<br>Use: | Indications for<br>use: | | | |
| and healthy lymph | fused and non-fused | | | | | | | |
| nodes and to register | displays to facilitate | | | | | | | |
| combinations of | the comparison and | | | | | | | |
| anatomical and | delineation of image | | | | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.