← Product Code [NRY](/productcode/NRY) · K220808

# Tigertriever 13 Revascularization Device (K220808)

_Rapid Medical Ltd. · NRY · Jul 25, 2022 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K220808

## Device Facts

- **Applicant:** Rapid Medical Ltd.
- **Product Code:** [NRY](/productcode/NRY.md)
- **Decision Date:** Jul 25, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Tigertriever Revascularization Device is intended to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA), or who fail IV t-PA therapy, are candidates for treatment.

## Device Story

Stentriever device for mechanical thrombectomy in large intracranial vessels; consists of adjustable nitinol braided mesh, stainless steel shaft, nitinol core wire, and handle. Physician operates device under fluoroscopic guidance; pulls slider on handle to manually expand mesh within vessel to conform to diameter; mesh encapsulates thrombus for retrieval. Used in neuro-interventional procedures by trained physicians. Device restores blood flow in ischemic stroke patients; benefits include effective clot removal and reperfusion. Supplied sterile for single-use.

## Clinical Evidence

No clinical data. Evidence consists of non-clinical bench testing and pre-clinical animal studies. Bench testing included simulated use in tortuous anatomical models, radial force, durability, torque strength, tip flexibility, kink resistance, tensile strength, and particulate evaluation. Pre-clinical animal studies (swine model) evaluated safety and clot retrieval performance at sub-acute and 30-day time points, demonstrating full reperfusion and no vessel damage.

## Technological Characteristics

Stentriever with adjustable nitinol braided mesh (8 wires, platinum core), 304 stainless steel shaft, and nitinol core wire. Distal OD 0.5mm (unexpanded) to 2.5mm (expanded). Sterilized via 100% ethylene oxide. Biocompatibility per ISO 10993. Bench testing per ISO 10555-1. Af temperature per ASTM F2082.

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- Tigertriever and Tigertriever 17 Revascularization Device ([K203592](/device/K203592.md))

## Reference Devices

- Comaneci Embolization Assist Device ([DEN170064](/device/DEN170064.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

July 25, 2022

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Rapid Medical Ltd. % Janice Hogan Partner Hogan Lovells US LLP 1735 Market Street, 23rd Floor Philadelphia, Pennsylvania 19103

Re: K220808

Trade/Device Name: Tigertriever 13 Revascularization Device Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NRY Dated: June 21, 2022 Received: June 21, 2022

Dear Janice Hogan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below

510(k) Number (if known) K220808

#### Device Name

Tigertriever 13 Revascularization Device

#### Indications for Use (Describe)

The Tigertriever Revascularization Device is intended to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA), or who fail IV t-PA therapy, are candidates for treatment.

Type of Use (Select one or both, as applicable) 図 Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA)Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image contains the logo for Rapid Medical. The logo consists of a stylized green symbol on the left, resembling three curved lines converging. To the right of the symbol, the words "Rapid Medical" are written in bold, black font. The logo appears to be for a medical company.

## 510(k) Summarv

## K220808

#### Submission Sponsor

Rapid Medical Ltd. Carmel Building, P.O. Box 337 Yokneam, 2069205 Israel Company Phone No.: +972-72-250-3331

### Contacts:

Dr. Orit Yaniv, VP of QA/RA Email: orit@rapid-medical.com

Ronen Eckhouse, CEO Email: ronen@rapid-medical.com

### Date Prepared

July 15, 2022

## Device Identification

Trade/Proprietary Name: Tigertriever 13 Revascularization Device Common/Usual Name: Catheter, Thrombus Retriever Classification Name: Percutaneous catheter Regulation Number: 21 CFR 870.1250 Product Code: NRY Device Class: II Classification Panel: Neurology

## Legally Marketed Predicate Device(s)

Tigertriever and Tigertriever 17 Revascularization Device (K203592)

## Indications for Use Statement

The Tigertriever Revascularization Device is intended to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA), or who fail IV t-PA therapy, are candidates for treatment.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image contains the logo for Rapid Medical. The logo consists of a stylized green symbol on the left, resembling three curved lines converging. To the right of the symbol is the text "Rapid Medical" in a bold, sans-serif font. The text is black, providing a strong contrast against the white background.

## Device Description

The Tigertriever 13 device is a stentriever that is comprised of an adjustable nitinol braided mesh, stainless steel shaft, nitinol core wire and a handle. The shaft connects the mesh and the handle by the core wire that runs inside the shaft from the distal end of the slider activation element in the handle. The mesh is expanded when the physician pulls the slider, since the wires of the mesh are completely radiopaque, the physician sees the mesh under fluoroscopy and controls it until it conforms to the vessel diameter. The design of the wire mesh is optimized to penetrate the clot and encapsulate it during retrieval. The Tigertriever 13 Revascularization Device is supplied sterile and is intended for single-use only by physicians trained in neuro-interventional catheterization and the treatment of ischemic stroke.

## Comparison of Technological Characteristics with the Predicate Device

The Tigertriever 13 Revascularization Device is substantially equivalent to the Tigertriever 17 Revascularization Device (model: TRPP7166) previously cleared as part of K203592. The two devices have the same indications for use and principles of operation, and similar technological characteristics. There are minor differences in dimensions between the two device models; however, they provide physicians with a slightly broader range of lengths and diameters to accommodate the range of large vessel occlusion (LVO) dimensions. This difference in dimensions does not raise new questions of safety or effectiveness. A summary of the technological characteristics of the Tigertriever 13 Revascularization Device in comparison to those of the predicate device is presented in the table below.

|                                                                       | Tigertriever 17 Revascularization<br>Device (TRPP7166)<br>(Predicate device)                                                                                                                                                                                                                                                                                                          | Tigertriever 13 Revascularization Device<br>(TRPP7144)<br>(Subject device)                                                                                                                                                                                                                                                                                                      |
|-----------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number                                                         | K203592                                                                                                                                                                                                                                                                                                                                                                               | K220808                                                                                                                                                                                                                                                                                                                                                                         |
| Regulation No.                                                        | 21 CFR 870.1250                                                                                                                                                                                                                                                                                                                                                                       | 21 CFR 870.1250                                                                                                                                                                                                                                                                                                                                                                 |
| Regulation Name                                                       | Percutaneous catheter                                                                                                                                                                                                                                                                                                                                                                 | Percutaneous catheter                                                                                                                                                                                                                                                                                                                                                           |
| Classification                                                        | Class II                                                                                                                                                                                                                                                                                                                                                                              | Class II                                                                                                                                                                                                                                                                                                                                                                        |
| Product Code                                                          | NRY                                                                                                                                                                                                                                                                                                                                                                                   | NRY                                                                                                                                                                                                                                                                                                                                                                             |
|                                                                       | Tigertriever 17 Revascularization<br>Device (TRPP7166)<br>(Predicate device)                                                                                                                                                                                                                                                                                                          | Tigertriever 13 Revascularization Device<br>(TRPP7144)<br>(Subject device)                                                                                                                                                                                                                                                                                                      |
| Indications for Use                                                   | The Tigertriever Revascularization<br>Device is intended to restore blood flow<br>by removing thrombus from a large<br>intracranial vessel in patients<br>experiencing ischemic stroke within 8<br>hours of symptom onset. Patients who<br>are ineligible for intravenous tissue<br>plasminogen activator (IV t-PA), or who<br>fail IV t-PA therapy, are candidates for<br>treatment. | The Tigertriever Revascularization Device is<br>intended to restore blood flow by removing<br>thrombus from a large intracranial vessel in<br>patients experiencing ischemic stroke within<br>8 hours of symptom onset. Patients who are<br>ineligible for intravenous tissue plasminogen<br>activator (IV t-PA), or who fail IV t-PA<br>therapy, are candidates for treatment. |
| Anatomical Location                                                   | Neurovasculature                                                                                                                                                                                                                                                                                                                                                                      | Neurovasculature                                                                                                                                                                                                                                                                                                                                                                |
| Technological Characteristics                                         |                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                 |
| Mode of operation                                                     | Manual expansion of the braided distal<br>portion into the clot using the handle<br>component                                                                                                                                                                                                                                                                                         | Manual expansion of the braided distal<br>portion into the clot using the handle<br>component                                                                                                                                                                                                                                                                                   |
| Design of distal<br>portion                                           | Close end braided nitinol mesh,<br>manually expandable                                                                                                                                                                                                                                                                                                                                | Close end braided nitinol mesh, manually<br>expandable                                                                                                                                                                                                                                                                                                                          |
| Stent length (un-<br>expanded<br>configuration)                       | 23mm                                                                                                                                                                                                                                                                                                                                                                                  | 20.5mm                                                                                                                                                                                                                                                                                                                                                                          |
| Stent size Distal OD<br>(unexpanded and<br>expanded<br>configuration) | Unexpanded configuration 0.5mm<br>Expanded configuration 3mm                                                                                                                                                                                                                                                                                                                          | Unexpanded configuration 0.5mm<br>Expanded configuration 2.5mm                                                                                                                                                                                                                                                                                                                  |
| Stent structure                                                       | Braided from eight (8) Niti DFT 0.075<br>mm wires with a tantalum core                                                                                                                                                                                                                                                                                                                | Braided from eight (8) Niti DFT 0.05 mm<br>wires with a platinum core                                                                                                                                                                                                                                                                                                           |
| Overall length                                                        | 228 cm<br>(shaft+cable+mesh+tip)                                                                                                                                                                                                                                                                                                                                                      | 208 cm<br>(shaft+cable+mesh+tip)                                                                                                                                                                                                                                                                                                                                                |
| Compatibility                                                         | Microcatheter with an internal diameter<br>of 0.017 inches                                                                                                                                                                                                                                                                                                                            | Microcatheter with an internal diameter of<br>0.0165 inches                                                                                                                                                                                                                                                                                                                     |
| Materials                                                             |                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                 |
| Stent (mesh)                                                          | Nitinol                                                                                                                                                                                                                                                                                                                                                                               | Nitinol                                                                                                                                                                                                                                                                                                                                                                         |
| Markers                                                               | 90% Platinum/ 10% Iridium                                                                                                                                                                                                                                                                                                                                                             | 90% Platinum/ 10% Iridium                                                                                                                                                                                                                                                                                                                                                       |
| Core wire (shaft)                                                     | Nitinol core wire and 304 Stainless Steel<br>shaft                                                                                                                                                                                                                                                                                                                                    | Nitinol core wire and 304 Stainless Steel<br>shaft                                                                                                                                                                                                                                                                                                                              |
| Introducer sheath                                                     | PTFE                                                                                                                                                                                                                                                                                                                                                                                  | PTFE                                                                                                                                                                                                                                                                                                                                                                            |
| Packaging                                                             |                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                 |
| Sterilization Method                                                  | Ethylene oxide                                                                                                                                                                                                                                                                                                                                                                        | Ethylene oxide                                                                                                                                                                                                                                                                                                                                                                  |
|                                                                       | Tigertriever 17 Revascularization<br>Device (TRPP7166)<br>(Predicate device)                                                                                                                                                                                                                                                                                                          | Tigertriever 13 Revascularization Device<br>(TRPP7144)<br>(Subject device)                                                                                                                                                                                                                                                                                                      |
| Single Use                                                            | Yes                                                                                                                                                                                                                                                                                                                                                                                   | Yes                                                                                                                                                                                                                                                                                                                                                                             |
| Packaging                                                             | Placed into a Dispenser hoop, blister,<br>Tyvek pouch, and Carton box                                                                                                                                                                                                                                                                                                                 | Placed into a Dispenser hoop, Tyvek pouch,<br>and Carton box                                                                                                                                                                                                                                                                                                                    |

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Rapid Medical. The logo consists of a green abstract symbol on the left and the words "Rapid Medical" in bold, dark gray font on the right. The abstract symbol appears to be three curved lines that converge to a point.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for Rapid Medical. The logo consists of a stylized green symbol on the left, resembling three curved lines. To the right of the symbol is the text "Rapid Medical" in a bold, sans-serif font. The text is black.

## Non-Clinical Performance Data

As part of demonstrating safety and effectiveness of the device and in showing substantial equivalence to the predicate device, Rapid Medical Ltd. completed a number of non-clinical performance tests. The device meets all the requirements of overall design, sterilization, and biocompatibility. Testing results confirm that the design output meets the design specification for the device.

# Biocompatibility

Biocompatibility of the Tigertriever 13 was based on the biocompatibility testing data for the Comaneci Embolization Assist Device (DEN170064). The two devices share the same manufacturing process and same manufacturing environment. In addition, the two devices are intended to be used in the same anatomical locations and are identical in terms of frequency and duration of exposure. Biocompatibility testing was completed for Comaneci device and consisted of the following tests: Cytotoxicity, Irritation (Intracutaeous Reactivity), Sensitization, Hemocompatibility, Pyrogenicity, Acute Systemic and Toxicity Testing. In addition, in vivo thrombogenicity was performed for the Tigertriever model. Results of the biocompatibility testing indicate that Tigertriever 13 Revascularization Device is biocompatible and is substantially equivalent for its intended use.

| Biological Endpoint                                | Test Results                                                                                                                                                                          | Conclusion                                |
|----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|
| Cytotoxicity – ISO Elution<br>Method               | Grade 0 reactivity observed 48 hours post<br>exposure to test article extract.                                                                                                        | Non-cytotoxic                             |
| Irritation – Intracutaeous<br>Reactivity in Rabbit | Difference of overall mean score between test<br>article and control was 0.                                                                                                           | Non-irritant                              |
| Sensitization – Guinea Pig<br>Maximization Test    | Grade 0, no evidence of causing delayed dermal<br>contact sensitization.                                                                                                              | Does not elicit<br>sensitization response |
| Hemocompatibility –<br>Complement Activation Assay | SC5b-9 concentration of the test article sample<br>was statistically lower than the activated NHS<br>control (p<0.05), and was not statistically higher<br>than the negative control. | Pass                                      |
| Hemocompatibility – In Vitro<br>Hemolysis          | Test article = 0.0% hemolysis.                                                                                                                                                        | Non-Hemolytic                             |

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image contains the logo for Rapid Medical. The logo features a green abstract symbol on the left, resembling three curved lines. To the right of the symbol, the words "Rapid Medical" are written in a bold, sans-serif font.

| Biological Endpoint                                                                                                                | Test Results                                                                                                       | Conclusion    |
|------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|---------------|
| Pyrogenicity – Material<br>Mediated Pyrogenicity in<br>Rabbit                                                                      | The total rise of rabbit temperatures during the<br>3-hour observation period was within acceptable<br>USP limits. | Non-pyrogenic |
| Systemic Toxicity –Systemic<br>Toxicity Study in Mice                                                                              | No mortality or evidence of systemic toxicity<br>from the extracts injected into mice.                             | Non-toxic     |
| Thrombogenicity – Acute Pre-<br>Clinical Evaluation of the<br>Safety of Tigertriever device in<br>a Swine Model<br>Thrombogenicity | Test device did not show higher<br>thrombogenicity rate compared to the predicate<br>device.                       | Pass          |

# Sterilization and Shelf life

The device and its accessories are sterilized by 100% Ethylene Oxide. The previously conducted shelf-life testing for the Tigertriever Revascularization Device supports a 2.5-year shelf life for the Tigertriever 13 Revascularization Device.

## Bench Tests

The device passed all performance bench testing in accordance with internal requirements, national standards and international standards as shown in the table below. Results of the performance bench testing indicate that Tigertriever 13 Revascularization device meets established performance requirements and is substantially equivalent for its intended use.

| Performance Bench Testing             |                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                             |                                                                                                             |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| Test                                  | Test Method Summary                                                                                                                                                                                                                                                                                                                   | Conclusions                                                                                                                                                                                                                                                                                                 |                                                                                                             |
| Simulated use test                    | Simulated use testing of the Tigertriever 13<br>Revascularization Device was performed in<br>an anatomical model which simulated the<br>tortuosity of the neurovasculature. Devices<br>were delivered through the tortuous<br>anatomical model to evaluate the<br>effectiveness of the device at retrieval of<br>firm and soft clots. | The device was tested for handling and<br>clot retrieval in an in vitro tortuous<br>path anatomical model, which has<br>been used in the evaluation of the<br>predicate device. The subject device<br>effectively retrieved clot and restored<br>flow in the test model.                                    |                                                                                                             |
| Radial force                          | The radial force of the subject device was<br>measured within a range of diameters from<br>1.5mm and 2.5mm, and compared with the<br>radial forces measured for the predicate<br>devices.                                                                                                                                             | The radial force of the subject device<br>when tested in applicable lumen sizes<br>is comparable to the predicate<br>device.                                                                                                                                                                                |                                                                                                             |
| Durability                            | Damage was evaluated after delivery and<br>withdrawal of the device beyond the<br>recommended number of passes and                                                                                                                                                                                                                    | Devices tested demonstrated no<br>damage after delivery and<br>withdrawal testing. Durability<br>established acceptable performance                                                                                                                                                                         |                                                                                                             |
| Performance Bench Testing             |                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                             |                                                                                                             |
| Test                                  | Test Method Summary                                                                                                                                                                                                                                                                                                                   | Conclusions                                                                                                                                                                                                                                                                                                 |                                                                                                             |
|                                       | resheathings recommended in the<br>instructions for use.                                                                                                                                                                                                                                                                              | for 3 passes, which is at least<br>equivalent to the number of passes<br>specified in the predicate labeling (2<br>passes per device).                                                                                                                                                                      |                                                                                                             |
| Delivery, deployment<br>and retrieval | The delivery, deployment and retrieval<br>forces were measured during simulated use<br>of the subject device.                                                                                                                                                                                                                         | The device was tested for delivery,<br>deployment, and retrieval in an in<br>vitro tortuous path anatomical model,<br>which has been used in the<br>evaluation of the predicate device.<br>The subject device demonstrated<br>acceptable performance with respect<br>to delivery, deployment and retrieval. |                                                                                                             |
| Torque strength                       | Devices were tracked through a<br>microcatheter in a tortuous path anatomical<br>model and evaluated for damage following a<br>number of rotations with the distal end<br>constrained.                                                                                                                                                | The device demonstrated the ability<br>to withstand 5 rotations without<br>damage. Like the predicate device,<br>rotational maneuvers are not<br>expected under the intended<br>conditions of clinical use. Therefore,<br>the results demonstrate acceptable<br>torque strength.                            |                                                                                                             |
| Dimensions test                       | Dimensional conformance to specifications<br>was confirmed.                                                                                                                                                                                                                                                                           | The subject device dimensions are<br>within the range of existing predicate<br>device dimensions. The minor<br>differences in dimensions do not<br>affect performance, safety or<br>effectiveness.                                                                                                          |                                                                                                             |
| Tip flexibility                       | Tip Flexibility testing was performed to<br>measure the force required to deflect<br>Tigertriever 13 tips to 90 degrees.                                                                                                                                                                                                              | The subject device met acceptance<br>criteria.                                                                                                                                                                                                                                                              |                                                                                                             |
| Kink resistance                       | Kink evaluation was performed for the distal<br>section and for the proximal section of the<br>device, in simulated bends representative<br>of the intended clinical anatomy.                                                                                                                                                         | Kink resistance was evaluated under<br>conditions simulating anatomic<br>tortuosity and demonstrated<br>acceptable performance.                                                                                                                                                                             |                                                                                                             |
| Tensile test                          | The minimum force to break the subject<br>device was tested for all joints.                                                                                                                                                                                                                                                           | The tensile strength of the device met<br>acceptance criteria based on<br>recognized standards (ISO 10555-1).                                                                                                                                                                                               |                                                                                                             |
| Particulates                          | Particulate test was performed according to<br>the light obscuration test method. Simulated<br>use testing of the subject device was<br>performed in an anatomical model. Devices<br>were flushed and the fluid was evaluated for<br>particle sizes of ≥10, ≥25 and ≥50 µm.                                                           | The particulate generated by the<br>subject device was similar to the<br>particulate generated by the predicate<br>device.                                                                                                                                                                                  |                                                                                                             |
| Austenite Finish (Af)<br>Temperature  | The Active Af temperature was determined<br>from a graph of displacement as a function<br>of temperature (Bend and Free Recovery per<br>ASTM F2082)                                                                                                                                                                                   | The Af temperature of the device met<br>acceptance criteria.                                                                                                                                                                                                                                                |                                                                                                             |
| Performance Bench Testing             |                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                             |                                                                                                             |
| Test                                  | Test Method Summary                                                                                                                                                                                                                                                                                                                   | Conclusions                                                                                                                                                                                                                                                                                                 |                                                                                                             |
| Coating integrity<br>assessment       | The test was performed with the Comaneci device, as the design of the two devices is the same in terms of the core wire mechanism. Damage to the PTFE core wire coating was evaluated following simulated use.                                                                                                                        | Results demonstrated no damage to the coating following simulated use.                                                                                                                                                                                                                                      |                                                                                                             |
|                                       | Corrosion                                                                                                                                                                                                                                                                                                                             | The test was performed with the Comaneci device and not with the subject device. The device was immersed in saline for 5 hours at room temperature, boiled in distilled water for 30 min, and maintained in 37°C for 48 hours. After the above treatment, device is tested for corrosion.                   | No corrosion was observed, which meets the acceptance criteria based on recognized standards (ISO 10555-1). |

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image shows the logo for Rapid Medical. The logo consists of a green abstract symbol on the left and the words "Rapid Medical" in bold, dark gray font on the right. The symbol appears to be three curved lines that converge to a point.

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image contains the logo for Rapid Medical. The logo consists of a stylized green symbol on the left, resembling three curved lines converging. To the right of the symbol, the words "Rapid Medical" are written in a bold, sans-serif font, with "Rapid" and "Medical" on the same line.

# Pre-Clinical Animal Testing Data

The safety of the Tigertriever 13 was assessed in two pre-clinical studies at sub-acute and 30-day (chronic) time points. After procedures, animals were assessed in vivo until their scheduled subchronic or chronic timepoints. In each tested vessel three passes were performed, one pass with clot and two passes without clots. The simulation included clots of variable hardness and consistency. Treated vessel diameters ranged from 1.1-3.2mm; this range complies with the target clinical indication of the large intracranial vessel occlusion, which includes vessels size of 1.5 mm and larger. Full reperfusion was demonstrated in all procedures, and there was no evidence of vessel damage (i.e., dissection, perforation, clot formation) related to device deployment. Histopathological assessment of treated vessels was considered acceptable. Tigertriever 13 preclinical studies thus demonstrate that the device has acceptable safety and clot retrieval performance.

# Conclusion

The Tigertriever 13 Revascularization Device has the same indications for use and principles of operation, and similar technological characteristics as the predicate device. The differences in technological characteristics between the subject and predicate devices do not raise different questions regarding the safety and effectiveness. Non-clinical performance testing as discussed above demonstrates similar performance and safety characteristics between the Tigertriever 13 Revascularization Device and the predicate device, thus supporting the conclusion that the Tigertriever 13 Revascularization Device is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K220808](https://fda-staging.innolitics.com/device/K220808)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
