← Product Code [JDQ](/productcode/JDQ) · K220765

# HS Fiber Cerclage (K220765)

_Riverpoint Medical, LLC · JDQ · Jul 26, 2023 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K220765

## Device Facts

- **Applicant:** Riverpoint Medical, LLC
- **Product Code:** [JDQ](/productcode/JDQ.md)
- **Decision Date:** Jul 26, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3010
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

HS Fiber Cerclage sutures are indicated for use in general soft tissue approximation and/or ligation. These sutures may be used in cardiovascular surgeries, and orthopedic surgeries using allograft tissue. When used as a bone fixation cerclage, the sutures are intended for: - Trauma surgery indications including olecranon, ankle, patella, and some shoulder rewiring. - Repair of long bone fracture due to trauma or reconstruction.

## Device Story

Non-absorbable surgical suture composed of braided ultra-high molecular weight polyethylene (UHMWPE) strands; used for soft tissue approximation, ligation, and bone fixation cerclage. Intended for hospital/clinic/surgical settings; operated by surgeons. Device provides mechanical fixation for bone fractures or soft tissue; acts as a structural tether to maintain alignment during healing. Benefits include high tensile strength and fatigue resistance for orthopedic reconstruction and trauma repair. Supplied sterile for single use.

## Clinical Evidence

Bench testing only. No clinical data. Performance verified via mechanical testing (tensile, knot, fatigue, creep, wear debris/particle analysis), biocompatibility (ISO 10993-1), sterilization validation (EN ISO 14937), and simulated use usability validation (EN 62366-1). All acceptance criteria met.

## Technological Characteristics

Material: Braided ultra-high molecular weight polyethylene (UHMWPE). Form factor: Suture with/without pre-attached needles. Sterilization: Ethylene oxide gas. Standards: USP (except diameter), ISO 10993-1 (biocompatibility), ISO 11607-1 (packaging), EN ISO 14937 (sterilization). Non-electronic, mechanical implant.

## Regulatory Identification

A bone fixation cerclage is a device intended to be implanted that is made of alloys, such as cobalt-chromium-molybdenum, and that consists of a metallic ribbon or flat sheet or a wire. The device is wrapped around the shaft of a long bone, anchored to the bone with wire or screws, and used in the fixation of fractures.

## Predicate Devices

- Arthrex FiberTape Cerclage ([K170206](/device/K170206.md))

## Reference Devices

- Riverpoint Medical HS Fiber (Polyblend) ([K100006](/device/K100006.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a symbol of the agency's mission to protect and promote public health through the regulation and supervision of food safety, tobacco, pharmaceuticals, medical devices, and other related products.

July 26, 2023

Riverpoint Medical, LLC Edwin Anderson VP of Regulatory Affairs 815 NE 25th Ave Portland, Oregon 97232

Re: K220765

Trade/Device Name: HS Fiber® Cerclage Regulation Number: 21 CFR 888.3010 Regulation Name: Bone Fixation Cerclage Regulatory Class: Class II Product Code: JDQ, GAT Dated: July 13, 2023 Received: July 14, 2023

Dear Edwin Anderson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Brent Showalter -S

Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K220765

Device Name HS Fiber® Cerclage

Indications for Use (Describe)

HS Fiber Cerclage sutures are indicated for use in general soft tissue approximation and/or ligation. These sutures may be used in cardiovascular surgeries, and orthopedic surgeries using allograft tissue. When used as a bone fixation cerclage, the sutures are intended for:

- · Trauma surgery indications including olecranon, ankle, patella, and some shoulder rewiring.
- · Repair of long bone fracture due to trauma or reconstruction.

| Type of Use (Select one or both, as applicable)                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                            |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg class="bi bi-check-square-fill" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2zm10.03 4.97a.75.75 0 0 0-1.08.022L7.477 9.417 5.384 7.323a.75.75 0 0 0-1.06 1.06L6.97 11.03a.75.75 0 0 0 1.079-.02l3.992-4.99a.75.75 0 0 0-.01-1.05z" fill-rule="evenodd"></path> </svg>     Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span> <svg class="bi bi-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> </svg>     Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Riverpoint. The logo consists of the letters "RP" in a green rounded square, followed by the word "RIVERPOINT" in green. The font is sans-serif and bold.

# 510(k) SUMMARY

### Riverpoint Medical HS Fiber® Cerclage

## Submitter Information

| Submitter's Name:    | Riverpoint Medical                     |
|----------------------|----------------------------------------|
| Address:             | 825 NE 25th Ave.<br>Portland, OR 97232 |
| Phone Number:        | (503) 517-8001                         |
| Fax Number:          | (503) 517-8002                         |
| Registration Number: | 3006981798                             |
| Contact Person:      | Becca DeFrancia<br>(503) 517-8001      |
| Date of Preparation: | July 13, 2023                          |

# Device Name

| Trade Name:            | HS Fiber® Cerclage                                                                                                        |
|------------------------|---------------------------------------------------------------------------------------------------------------------------|
| Common or Usual Names: | Bone Fixation Cerclage, Suture                                                                                            |
| Classification Name:   | 21 CFR 888.3010: Bone Fixation Cerclage<br>21 CFR 878.5000: Nonabsorbable Poly(ethylene) Terephthalate<br>Surgical Suture |
| Device Classification  |                                                                                                                           |

### Devi

| FDA Class:              | II                 |
|-------------------------|--------------------|
| Product Classification: | 888.3010, 878.5000 |
| Classification Code:    | JDQ, GAT           |

# Predicate Device(s)

K170206 (Primary) – Arthrex FiberTape Cerclage

K100006 (Reference) – Riverpoint Medical HS Fiber (Polyblend)

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Riverpoint. The logo consists of the letters "RP" inside of a green rounded square. To the right of the square is the word "RIVERPOINT" in all capital letters and in the same shade of green as the square.

### Device Description

The Riverpoint Medical HS Fiber® Cerclage sutures are non-absorbable, sterile, surgical sutures composed of multiple single strands of ultra-high molecular weight polyethylene (UHMWPE) braided together to form the implant. HS Fiber Cerclage sutures are available in common sizes and lengths with or without pre-attached needles. Suture supplied meet United States Pharmacopeia (USP) requirements for non-absorbable suture except for diameter. The device is sterilized by ethylene oxide gas, and is provided sterile for single use. The device is intended for use in a hospital/clinic/surgical setting.

The classification for the HS Fiber Cerclage is FDA Class II device with product classification 21 CFR §878.5000, Product Code GAT and subsequent 21 CFR §888.3010, Product Code IDQ.

### Indications for Use:

HS Fiber Cerclage sutures are indicated for use in general soft tissue approximation and/or ligation. These sutures may be used in cardiovascular surgeries, and orthopedic surgeries using allograft or autograft tissue. When used as a bone fixation cerclage, the sutures are intended for:

- Trauma surgery indications including olecranon, ankle, patella, and some shoulder rewiring. ●
- Repair of long bone fracture due to trauma or reconstruction. ●

### Performance Data

The sutures used to construct the HS Fiber Cerclage meet requirements established by the United States Pharmacopeia (USP), except for diameter. The UHMWPE sutures are tested per USP performance requirements for needle attachment and tensile strength. FDA Guidance "Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA" were followed during the preparation of this submission.

Non-clinical performance testing for the Riverpoint Medical HS Fiber Cerclage included sterilization validation per EN ISO14937:2009 - Sterilization of health care products - General requirements for characterization of a sterilizing agent and the development, validation and routine control of a sterilization process for medical devices, biocompatibility testing per ISO 10993-1:2018 - Biological Evaluation of Medical Devices, stability testing on the product and packaging per ISO 11607-1:2019 - Packaging for terminally sterilized medical devices --Part 1: Requirements for materials, sterile barrier systems and packaging systems, and a simulated use Usability Validation performed per EN62366-1: 2015- Medical devices -Application of usability engineering to medical devices. All acceptance criteria were met, and the Riverpoint Medical HS Fiber Cerclage Suture performed as intended. Non-clinical mechanical testing was performed to verify the fixation performance for the HS Fiber Cerclage using tensile strength, knot strength, fatigue strength, creep, and wear debris with particle analysis. Results of the performance testing for the HS Fiber Cerclage device concluded that the device performed comparably to the predicate device in tensile strength, fatigue strength, creep, and wear debris. The tests performed demonstrated that the HS Fiber Cerclage met all requirements for its intended use.

---

**Source:** [https://fda-staging.innolitics.com/device/K220765](https://fda-staging.innolitics.com/device/K220765)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
