K220739 · Zhejiang Gongdong Medical Technology Co., Ltd. · JSM · Nov 17, 2022 · Microbiology
Device Facts
Record ID
K220739
Device Name
Disposable Urine Collection Tube
Applicant
Zhejiang Gongdong Medical Technology Co., Ltd.
Product Code
JSM · Microbiology
Decision Date
Nov 17, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.2390
Device Class
Class 1
Indications for Use
Disposable Urine Collection Tube is intended for the collection, transport and preservation of urine specimens from the collection site to the testing laboratory. In the laboratory, urine specimens are processed using standard clinical laboratory operating procedures for the cultivation of uropathogenic bacteria and yeasts.
Device Story
Disposable Urine Collection Tube system facilitates safe collection and transport of urine specimens. Device consists of PET tube, butyl-rubber safety plug, and PE cap containing boric acid preservative. Vacuum-based mechanism enables automated filling of urine sample via puncture needle. Used by healthcare professionals in clinical settings to transport specimens to laboratories. Boric acid additive prevents overgrowth of uropathogenic bacteria and yeasts during transport. Healthcare providers process samples using standard clinical laboratory procedures for cultivation. System maintains specimen viability for up to 48 hours at 2-4°C and 22-25°C, supporting accurate diagnostic testing and clinical decision-making.
Clinical Evidence
Bench testing only. No clinical data. Microbial recovery studies evaluated viability of C. albicans, E. coli, E. faecalis, P. aeruginosa, P. mirabilis, and S. saprophyticus at 0, 24, and 48 hours at 2-4°C and 22-25°C. All results showed <1 log10 change. Fill volume impact study (90% and 110% fill) confirmed <1 log10 change for E. coli, P. aeruginosa, and P. mirabilis after 48 hours. Shelf-life validated for 18 months via real-time aging. Sterilization validated to SAL 10^-6 per ISO 11137-1/2.
Indicated for the collection, transport, and preservation of urine specimens from the collection site to the testing laboratory for the cultivation of uropathogenic bacteria and yeasts. Intended for use by healthcare professionals.
Regulatory Classification
Identification
A transport culture medium is a device that consists of a semisolid, usually non-nutrient, medium that maintains the viability of suspected pathogens contained in patient specimens while in transit from the specimen collection area to the laboratory. The device aids in the diagnosis of disease caused by pathogenic microorganisms and also provides epidemiological information on these diseases.
Predicate Devices
UriSwab™ - Urine Collection, Transport and Preservation System (K180052)
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K220739
B Applicant
Zhejiang Gongdong Medical Technology Co., Ltd.
C Proprietary and Established Names
Disposable Urine Collection Tube
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| JSM | Class I, reserved | 21 CFR 866.2390 - Transport Culture Medium | MI - Microbiology |
| LIO | Class I, reserved | 21 CFR 866.2900 - Microbiological specimen collection and transport device | MI - Microbiology |
## II Submission/Device Overview:
A Purpose for Submission:
To obtain a substantial equivalence determination for the Disposable Urine Collection Tubes for the collection, transport, and preservation of urine specimen from the collection site to the testing laboratory for the cultivation of uropathogenic bacteria and yeasts.
B Measurand:
Not applicable
C Type of Test:
Non-propagating Transport Device
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
K220739 - Page 2 of 8
# III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
B Indication(s) for Use:
Disposable Urine Collection Tube is intended for the collection, transport and preservation of urine specimens from the collection site to the testing laboratory. In the laboratory, urine specimens are processed using standard clinical laboratory operating procedures for the cultivation of uropathogenic bacteria and yeasts.
C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
D Special Instrument Requirements:
None
# IV Device/System Characteristics:
A Device Description:
The Disposable Urine Collection Tubes for the collection, transport and preservation system consist of a safety rubber plug, a plastic cap and boric acid as an additive inside the tube. The tube is made of Polyethylene terephthalate (PET) and the safety plug is made of butyl-rubber while the disposable cap is made of Polyethylene (PE). The disposable tube device sustains the viability of organisms during transport for up to 48 hrs. at 2-4°C and 22-25°C. There are two main tube types (round and conical bottom), and four configurations by volume of urine for the round bottom tubes (4 ml, 7 ml, 8 ml, 10 ml). The conical bottom tubes only have one configuration by volume (8 ml). Each volume is available in yellow and white caps. The different device configurations and information on the amount of preservative contained in each tube is provided in table 1.
Table 1. Disposable Tubes Description and Packaging configurations
| Ref No. | Draw volume (mL) | Tube size (Diameter × Length) in mm | Additive Boric acid in mg | Material | Type of tube | Cap color |
| --- | --- | --- | --- | --- | --- | --- |
| GD040UB | 4 | 13×75 | 40 | PET | Round bottom | yellow |
| GD040UBW | 4 | 13×75 | 40 | PET | Round bottom | white |
| GD070UB | 7 | 13×100 | 70 | PET | Round bottom | yellow |
| GD070UBW | 7 | 13×100 | 70 | PET | Round bottom | white |
| GD080UB | 8 | 16×100 | 80 | PET | Round bottom | yellow |
{2}
K220739 - Page 3 of 8
| GD080UBW | 8 | 16×100 | 80 | PET | Round bottom | white |
| --- | --- | --- | --- | --- | --- | --- |
| GD100UB | 10 | 16×100 | 100 | PET | Round bottom | yellow |
| GD100UBW | 10 | 16×100 | 100 | PET | Round bottom | white |
| GD080UAC | 8 | 16×100 | 80 | PET | Conical bottom | yellow |
| GD080UACW | 8 | 16×100 | 80 | PET | Conical bottom | white |
## B Principle of Operation:
The Disposable Urine Collection Tube system is used to safely collect and transport urine from collection sites to the testing laboratories. It is intended to be used by health care professionals. The vacuum in the collection tube allows autofill the urine sample through the puncture needle and the boric acid in the tubes prevents overgrowth of uropathogenic bacteria and yeasts.
## V Substantial Equivalence Information:
### A Predicate Device Name(s):
UriSwab-Urine Collection, Transport and Preservation System
### B Predicate 510(k) Number(s):
K180052
### C Comparison with Predicate(s):
| Device & Predicate Device(s): | Device: K220739 | Predicate: K180052 |
| --- | --- | --- |
| Device Trade Name | Disposable Urine Collection Tube | UriSwab-Urine Collection, Transport and Preservation System |
| Device Product Code and Classification | JSM, LIO, Class I | Same |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | Disposable Urine Collection Tube is intended for the collection, transport, and preservation of urine specimens from the collection site to the testing laboratory. In the laboratory, urine | Copan UriSwab - Urine Collection, Transport and Preservation System is intended for the collection, transport, and preservation of urine specimens from the collection site to the testing laboratory. In |
{3}
| | specimens are processed using standard clinical laboratory operating procedures for the cultivation of uropathogenic bacteria and yeasts. | the laboratory, UriSwab specimens are processed using standard clinical laboratory operating procedures for the cultivation of uropathogenic bacteria and yeasts. |
| --- | --- | --- |
| Specimen Stability | 2-4°C and 22-25°C; 48 hrs. | Same |
| Device Storage Temperature | 22-25°C | Same |
| Sterilization | Radiation | Same |
| Sterile | Yes; 10-6SAL | Same |
| General Device Characteristic Differences | | |
| Tube material | Polyethylene Terephthalate (PET) | Plastic |
| Tube type | Round and Conical bottom | Conical bottom |
| Tube size | 13 mm × 75 mm; 13 mm × 100 mm; 16 mm × 100 mm | 12 mm × 80 mm; 16 mm × 100 mm |
| Tube volume | 4 ml, 7 ml, 8 ml, 10 ml | 1.5 ml, 3.2 ml |
| Cap | Safety cap, yellow and white. | Screw cap, yellow |
| Additive | Boric acid | Boric acid and Sodium formate |
| Collection device | PET tube with additive | Plastic tube, stick with sponge |
| Tube filling | Predefined vacuum | Sponge applicator |
VI Standards/Guidance Documents Referenced:
Quality Control of Microbiological Transport Systems, Clinical and Laboratory Standards Institute (CLSI) guideline M40-A2, 2015.
Sterilization of health care products - Radiation - Part 1: Requirements for development, validation, and routine control of a sterilization process for medical devices, 1st edition, 11137-1, International Organization for Standardization (ISO), 2006.
Sterilization of health care products - Radiation - Part 2: Establishing the sterilization dose, 3rd edition, 11137-2, International Organization for Standardization (ISO), 2013.
VII Performance Characteristics (if/when applicable):
K220739 - Page 4 of 8
{4}
K220739 - Page 5 of 8
# A Analytical Performance:
1. Precision/Reproducibility:
Not Applicable
2. Linearity:
Not Applicable
3. Analytical Specificity/Interference:
Not Applicable
4. Assay Reportable Range:
Not Applicable
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
## Stability Verification Report:
The shelf life for the Disposable Urine Collection Tube was determined to be 18 months from the date of manufacture when stored in well-ventilated clean room with the environment of relative humidity not more than 80% and at temperature 22 – 25°C. The shelf life of the Disposable Urine Collection Tube was evaluated using real-time aging performance test at time points T = 0, 6, 12 and 18 months. At each time point, the following performance tests were conducted as described below:
- Appearance test was carried out by inspecting visually for clear label, transparent PET tube and any sign of deformation or breakage.
- Vacuum test was performed by measuring volume drawn into the Disposable Urine Collection Tube between 90 to 110% of the nominal liquid capacity allowed.
- Physical and mechanical performance tests included leak test, strength test, and a 50 KPa pressure test.
- Mechanical stress test was conducted by checking the tubes could withstand centrifugation.
- The strength test, and urine collection volume capacity test were carried out as part of the evaluation.
- Sterility test and the standard of boric acid content used were also assessed.
- The package integrity was evaluated by visually checking for any changes or damage to the package. Six lots of varying production were used for the studies and results showed that all samples passed the real time aging test for 18 months at 22-25°C.
All the results were acceptable and support the claim that the Disposable Urine Collection Tube is physically or visually stable for 18 months.
{5}
# Sterilization:
The Disposable Urine Collection Tubes were sterilized by electron beam radiation with an average dose of $20\mathrm{kGy}$ to achieve a sterility assurance level (SAL) of $10^{-6}$ with a dosimeter attached to monitor dosage through the sterilization process. Sterilization was validated following ISO 11137-1: 2006 and ISO 11137-2: 2013 Sterilization of Health Care Products Radiation guidelines. Sterility test was carried out according to ISO11137-2:2009 Medical Devices Radiation Sterilization-Microbiological Methods-Part 2. In this sterility test, 100 samples after irradiated at the verification dose were inoculated with $200\mathrm{ml}$ soybean casein liquid medium and incubated at $30\pm 2^{\circ}\mathrm{C}$ for 14 days. The results were observed with no growth. Sterilization studies of the tube were acceptable.
# 6. Detection Limit:
# Microbial Recovery Studies:
Microbial recovery (viability) studies were conducted to support the claim of the Disposable Urine Collection Tube for the collection, transport and preservation of urine specimens from the collection site to the testing laboratory. Sample panels were prepared from the listed bacteria and yeast (below) that were commonly associated with urinary tract infections. Cultures from $1.5 \times 10^{8}$ (equivalent 0.5 McFarland standard) were diluted into the pooled human negative clinical urine matrix to give a final concentration of $1.5 \times 10^{2} \mathrm{CFU} / \mathrm{ml}$ . The appropriate volume of urine samples spiked with each organism was introduced into the Disposable Urine Collection Tubes. Three lots of the Disposable Urine Collection Tubes with three samples per lot were evaluated for the microbial recovery (viability) assay. The inoculated sample tubes were initially tested to establish a time point zero then stored at $2-4^{\circ} \mathrm{C}$ and $22-25^{\circ} \mathrm{C}$ for 24 hrs., and 48 hrs. After each storage time and temperature, the samples were plated to grow colonies which were counted to determine CFU/ml for each time points. According to CLSI document M40-A2, the acceptance criteria for microbial recovery method was set to be within $1 \log_{10}$ of the initial microorganism count. Results of recovery studies are presented in table 2.
Table 2. Performance of Disposable Urine Collection Tube in Microbial Recovery Studies
| Strain tested | Storage temperature | Average CFU/ml recovered (N=9) at time points | | | Change in Log10(0 - 48 hrs) |
| --- | --- | --- | --- | --- | --- |
| | | 0 hr | 24 hrs | 48 hrs | |
| Candida albicans(ATCC 24433) | 2-4°C | 140 | 107 | 85 | -0.22 |
| | 22-25°C | 140 | 74 | 69 | -0.31 |
| Escherichia coli(ATCC 25922) | 2-4°C | 137 | 104 | 92 | -0.17 |
| | 22-25°C | 137 | 85 | 42 | -0.51 |
| Enterococcus faecalis(ATCC 29212) | 2-4°C | 145 | 102 | 91 | -0.20 |
| | 22-25°C | 145 | 80 | 54 | -0.43 |
| Pseudomonas aeruginosa(ATCC 27853) | 2-4°C | 143 | 103 | 69 | -0.32 |
| | 22-25°C | 143 | 106 | 72 | -0.30 |
| Proteus mirabilis(ATCC 7002) | 2-4°C | 140 | 111 | 72 | -0.29 |
| | 22-25°C | 140 | 105 | 71 | -0.29 |
| Staphylococcus saprophyticus | 2-4°C | 139 | 109 | 69 | -0.30 |
K220739 - Page 6 of 8
{6}
The microbial recovery studies demonstrated acceptable results.
## Fill Volume Impact Recovery Study:
A fill volume impact recovery study was performed to demonstrate that underfilling of Disposable Urine Collection Tube containing preservative does not negatively impact the viability of urine microorganism. Sample panels were prepared by listed microorganisms (below) diluted from $1.5 \times 10^{8}$ CFU/mL (equivalent to a $0.5$ McFarland standard) into the pooled human urine matrix to give a final concentration of $1.5 \times 10^{2}$ CFU/ml. Three production lots were tested in three replicates with each urine samples spiked with $E.$ coli, $P.$ aeruginosa or $P.$ mirabilis at the sample fill volume of $90\%$ and $110\%$. The inoculated sample tubes were initially tested to establish a time point zero then stored at $22-25^{\circ}\mathrm{C}$ for 48 hrs. After the storage time, the samples were plated to grow colonies which were counted to determine CFU/ml for each time points. The acceptance criteria for the study is less than 1 $\log_{10}$ change when the fill volume is above or below the line. The results were shown in table 3.
Table 3. Performance of Disposable Urine Collection Tube in fill volume impact recovery study.
| Strain tested | Fill volume | Average CFU/ml recovered (N=9) at time points | | Change in Log_{10} (0 - 48 hrs) |
| --- | --- | --- | --- | --- |
| | | 0 hr | 48 hrs | |
| E. coli (ATCC 25922) | 90% | 125 | 48 | -0.42 |
| | 110% | 127 | 46 | -0.44 |
| P. aeruginosa (ATCC 27853) | 90% | 143 | 49 | -0.47 |
| | 110% | 142 | 50 | -0.45 |
| P. mirabilis (ATCC 7002) | 90% | 140 | 40 | -0.54 |
| | 110% | 139 | 39 | -0.55 |
Fill volume recovery study results were acceptable.
## 7. Assay Cut-Off:
Not applicable.
K220739 - Page 7 of 8
{7}
B Comparison Studies:
1. Method Comparison with Predicate Device:
Method comparison is not applicable for a bacterial transport medium. The device itself does not provide a result that can be used in making a clinical decision. Bench testing studies were conducted to determine the ability of the Disposable Urine Collection Tube for urine collection, transport & preservation to maintain viability of microorganisms commonly found as the etiological agents of urinary tract infections. The analytical studies showed recovery of representative bacteria are within the acceptance criteria recommended in CLSI-M40-A2.
2. Matrix Comparison:
Not applicable.
C Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable.
D Clinical Cut-Off:
Not applicable.
E Expected Values/Reference Range:
Not applicable.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K220739 - Page 8 of 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.