K220738 · Cymedica Orthopedics, Inc. · IPF · May 12, 2022 · Physical Medicine
Device Facts
Record ID
K220738
Device Name
Motive Knee Wrap
Applicant
Cymedica Orthopedics, Inc.
Product Code
IPF · Physical Medicine
Decision Date
May 12, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Motive™ Knee Wrap device is intended to strengthen the quadriceps muscle stimulation to provide symptomatic temporary pain relief associated with knee arthritis. The Motive™ Knee Wrap device is indicated for adults of 22 years of age and older.
Device Story
Motive Knee Wrap is an OTC neuromuscular electrical stimulation (NMES) device for knee arthritis pain. It consists of a textile wrap with a modular 'Smart Panel' containing a pulse generator (Controller) and four skin-contacting electrodes. The user operates the device via a mobile app (iOS/Android) connected via Bluetooth Low Energy (BLE 4.2). The system delivers a proprietary, closed-loop, power-regulated monophasic electrical waveform to the quadriceps to induce muscle contractions. The app provides a virtual control panel for therapy initiation, intensity adjustment, and compliance tracking. The device is intended for home use. By strengthening the quadriceps, the device aims to reduce knee pain and improve joint function. The Controller includes a physical button for power and emergency stop, ensuring safety if the Bluetooth connection is lost.
Clinical Evidence
No new clinical trials were conducted for the Motive device. Substantial equivalence relies on non-clinical bench testing (electrical safety, EMC, biocompatibility, software validation) and clinical data previously submitted for the predicate IntelliHab System (K210604). Human factors and self-selection studies were performed to validate OTC usability.
Technological Characteristics
Materials: PC/ABS plastic (Bayblend FR3010). Energy: 3.7V Li-Po rechargeable battery. Connectivity: BLE 4.2. Waveform: Monophasic, asymmetrical, complex. Output: Regulated power, closed-loop feedback. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, IEC 60601-1-11, ISO 10993. Software: Embedded controller firmware, mobile app, and cloud-based web services.
Indications for Use
Indicated for adults 22+ years old for strengthening quadriceps muscles and providing symptomatic temporary pain relief associated with knee arthritis.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
RelieforMe TENS/EMS Device Model UPK-GE01 (K192733)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and name on the right. The FDA part of the logo is in blue and reads "FDA U.S. FOOD & DRUG ADMINISTRATION".
May 12, 2022
CyMedica Orthopedics, Inc. Kereshmeh Shahriari Vice President, Regulatory, Clinical, & Quality 2120 East 6th Street Tempe, Arizona 85281
Re: K220738
Trade/Device Name: Motive Knee Wrap Regulation Number: 21 CFR 890.5850 Regulation Name: Powered muscle stimulator Regulatory Class: Class II Product Code: IPF Dated: March 7, 2022 Received: March 14, 2022
### Dear Kereshmeh Shahriari:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For: CDR Jitendra Virani, MS. Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220738
Device Name Motive™ Knee Wrap Device
Indications for Use (Describe)
The Motive™ Knee Wrap device is intended to strengthen the quadriceps muscle stimulation to provide symptomatic temporary pain relief associated with knee arthritis.
The Motive™ Knee Wrap device is indicated for adults of 22 years of age and older.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| Prescription Use (Part 21 CFR 801 Subpart D) | |
| | |
| | |
| Over-The-Counter Use (21 CFR 801 Subpart C) | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary Motive™ Knee Wrap CyMedica Orthopedics, Inc.
| <b>1 Regulatory Information</b> | |
|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| <b>1.1 Trade/Proprietary Name:</b> | Motive™ Knee Wrap |
| <b>1.2 Common Name:</b> | Powered muscle stimulator |
| <b>1.3 Regulation Names & Numbers:</b> | Powered muscle stimulator, 21 CFR 890.5850 |
| | Product Codes:<br>Code- IPF; Powered muscle stimulator |
| <b>1.4 Classification:</b> | Powered muscle stimulator: II |
| <b>1.5 Manufacturer Name:</b> | CyMedica Orthopedics, Inc.<br>2120 East 6th Street, Suite 8<br>Tempe, AZ 85281<br>Telephone (480) 664-1282<br>FAX (866) 296-2772 |
These devices are reviewed by the Division of Neurological and Physical Medicine Devices.
#### 2 Submission Information
| Submission Number: | K220738 |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | March 31, 2022 |
| Contact: | Kereshmeh Shahriari<br>2120 East 6th Street, Suite 8<br>Tempe, AZ 85281<br>kereshmeh@cymedicaortho.com<br>Telephone (480) 664-1282<br>FAX (866) 296-2772 |
{4}------------------------------------------------
#### 3 Indications for Use
The Motive Knee Wrap device is intended to strengthen the quadriceps muscle using powered muscle stimulation to provide symptomatic temporary pain relief associated with knee arthritis. The Motive Knee Wrap device is indicated for adults of 22 years of age and older.
#### 4 Device Description
The Motive Knee Wrap device (Motive device) is an over-the-counter (OTC) electrical muscle stimulator (or Neuromuscular Electrical Stimulation (NMES) therapy) for relief of knee pain associated with knee arthritis. The Motive device is used to apply an electrical current through a power regulated output closed-loop feedback NMES system to provide relief of knee pain associated with arthritis. The stimulator continuously contracts quadricep muscle groups for strengthening of the muscles and knee pain relief. The stimulator is placed over the thigh muscle and just above the knee.
#### 4.1 Electrical Muscle Stimulator (NMES therapy)
The device is used to apply a patented unique waveform by means of an electrical current through a power regulated output and closed-loop feedback system to provide relief of knee pain associated with arthritis. The stimulator continuously contracts quadricep muscle groups for strengthening of the muscles and for knee pain relief. Electrical stimulation waveform is generated using a pulse generator or Controller. Electrical stimulation is delivered non-invasively to the treatment site (quadricep muscles of the knee joint) using connected electrodes placed directly on the skin.
#### 4.2 Mobile Application (App)
The electrical stimulation therapy is initiated and managed wirelessly by the User using a mobile application (App) developed for use on smart phones or tablets that interact with the system's pulse generator or Controller.
#### 4.3 Wrap with Smart Panel
The electrical muscle stimulator along with the cutaneous electrodes and wires are incorporated in a Wrap with a Smart Panel™ placed on the thigh and above the knee. Wrap with Smart Panel is worn on the thigh covering quadriceps muscles to apply the NMES therapy. The Wrap is comprised of a modular design including a textile-based wrap and a Smart Panel. The modular Smart Panel is attached to the Wrap by means of a hook and loop interface or other mechanisms. The Smart Panel is an assembly of a Controller (NMES pulse generator), cradle, plastics for hosting the electronics, and four skin contacting snap type electrodes. The Wrap with Smart Panel design is a single size for all size Users and intended for a bilateral use (right and left thigh use independently).
#### 4.4 Wireless Communications
The Motive device incorporates a Bluetooth Low Energy, BLE 4.2 connection module to enable wireless communication that can be paired with a Bluetooth enabled mobile device running the App, available from the App stores. The App implements a virtual control panel on the screen of the smart device where on-screen buttons are provided to the User.
The Controller is connected to the Smart Panel via a multi-pin interface. The Controller contains the primary safety controls for operation of the device and a push button is available for switching the unit on or off. The power button remains active and can be used in the event
{5}------------------------------------------------
of loss of Bluetooth connection to power on/off the device or start/stop stimulation while performing stimulation sessions. The Controller contains light emitting diodes (LED) which indicate status relating to battery charge, stimulation, and Bluetooth™ activity. Power is derived from a 3.7V Li- Po rechargeable battery pack and the unit can be recharged by using the supplied USB cable.
#### 4.5 Software
The Motive device utilizes an embedded Controller firmware, App software, and Web Services software.
The treatment program embedded software is a fundamental component of the system that supports the selection, adjustment, and delivery of the treatment program. The embedded software runs on the Controller and is interfaced with the stimulation pulse generator and App.
The App runs on a mobile platform operating system and communicates with the Controller over a BLE 4.2 connection. The free Motive App is downloaded by the Users from their respective App Stores. The Motive device allows the User to control the NMES therapy program in an easy and intuitive way. It allows the User to select, administer, and adjust the NMES therapy intensities, and manage their health and wellness conditions related to their therapy.
#### 4.6 Controller and the App
The Controller's main function is the NMES therapy stimulation generator and communicating therapy metrics to the App. The Controller contains the primary safety controls for operation of the device. A push button is available for powering the unit on/off or pausing/resuming stimulation. The push button "Stop" control remains active at all times during stimulation and can be used in the event of loss of Bluetooth connection.
Once the App is installed and a wireless connection is established between the Controller and the mobile device (User's phone or tablet), the User can interact with all of the device features. The App downloaded on the User's mobile device provides the graphical user interface and governs the interactions between the controller and the User. The screens and user interface are designed to be intuitive and interactive. Improved User's engagement and compliance through the App are significant design features of the device.
Help and informational screens are provided throughout the App and on specific screens where a User needs to be aware of a feature or condition. Upon downloading the App, the User is guided through chronological screens on the App to perform the following tasks:
- Reading and accepting the Motive App Terms of Use and Conditions ●
- Creating a User profile
- Pairing the App with the Controller ●
- Starting the treatment
- Managing the treatment
- Reviewing User's related data and information on the App
#### 4.7 Web Services (API, Database)
A Medical Device Data System (MDDS), the CyMedica Web Services allows for the electronic transfer, storage, retrieval, and display of data sent or received by the Users' App
{6}------------------------------------------------
and Controller. Web Services includes an encrypted database hosted on secure cloud-based servers and an Application Programming Interface (API) allowing the App to send and receive data to read or write to the database and execute logic tasks. Web Services provides persistent storage and retrieval of data.
#### 5 General Electrical Muscle Stimulation Device Features and Output Characteristics
The following two Tables represent the Motive device electrical stimulation features and the neuromuscular electrical stimulation (NMES) therapy output specifications.
Table 1- Motive device Therapy Program General Electrical Stimulation Features
| No. of Output Modes | 1 |
|-------------------------------------------|-----------------|
| No. of Output Channels | 2 |
| Regulated Current, Voltage, or Power | Regulated Power |
| Software/Firmware/Microprocessor Control? | Yes |
| Automatic shut off? | Yes |
| Patient device control? | Yes |
| Indicator display- Low Battery? | Yes |
| Indicator display- Voltage/Current Level? | Yes |
| Timer range (minutes) | 20 minutes |
Table 2- Motive device NMES Therapy Program Output Specifications
| Waveform | Monophasic |
|------------------------------------------------------------------------------------|--------------------------------------|
| Shape | Asymmetrical, Complex |
| Pulse width | 5 ms |
| Frequency | 50 pps |
| Maximum current density @ 500Ω | 0.7 mA/cm² (rms) |
| Maximum power density (using smallest electrode conductive<br>surface area) @ 500Ω | 0.006 W/cm² |
| Maximum phase charge @ 500Ω | 359.8 μC |
| Maximum output voltage (VRMS) (±10%) @ 500Ω | 9.0 V |
| Maximum output current (IRMS) (±10%) @ 500Ω | 18 mA |
| Power source | Li-Polymer Battery<br>950mAh 3.7 VDC |
| Contraction time | 1.0 s |
| Relaxation time | 1.4 s |
| Treatment session | 20 minutes |
#### 6 NMES Waveform
The NMES treatment program pulse parameters are defined below, Table 3:
{7}------------------------------------------------
| Table 3- Motive NMES Waveform Pulse parameters | | |
|------------------------------------------------|---------------------------------------|-------|
| Pulse shape | Asymmetrical, monophasic, and complex | |
| Treatment duration | 20 minutes | |
| Frequency | 50 pps | |
| Pulse width | 5 ms | |
| Duty cycle | 25% | |
| Work cycle | 12 s | |
| Relaxation time | 10 s | |
| Contraction time | 5 cycles | 1 s |
| Rest time | | 1.4 s |
In the treatment program, the work cycle consists of the combination of five cycles of contraction and rest. Contraction time is the actual stimulation contraction period. Rest time is the period between contractions to wait to oscillate the stimulation between the two channels. Relaxation time is a period of no stimulation between the work cycles.
#### 7 Recommended Usage
Motive device NMES therapy is recommended to be used twice a day for a minimum of 5 days per week. Each session of NMES therapy is 20 minutes. The therapy is recommended to be applied as long as the knee pain relief is sustained.
#### 8 Summary of Non-Clinical/ Bench Studies
To demonstrate the safety, the Motive device was tested for electrical safety, electromagnetic compatibility, usability, biocompatibility, and risk management requirements.
To demonstrate the safety, the Motive device was tested per the following standards:
- IEC 60601-1: 2005, 3rd Edition, Medical electrical equipment- General requirements for ● basic safety and essential performance + CORR. 1:2006 + CORR. 2:2007 + A1:2012
- IEC 60601-1-2: 2014-02, 4th Edition, Medical electrical equipment-Part 1-2: General . requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility requirements
- IEC 60601-2-10: 2012, 2nd Edition, Medical electrical equipment- Part 2-10: Particular ● requirements for the basic safety and essential performance of nerve and muscle stimulator
- IEC 60601-1-11: 2015, 2nd Edition, Medical electrical equipment- Part 1-11: General requirements for basic safety and essential performance- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
{8}------------------------------------------------
- IEC 60601-1-6: 2010+ A1:2013, 3rd Edition, Medical electrical equipment- Part 1-6: . General requirements for basic safety and essential performance- Collateral standard: Usability including IEC 62366: Application of usability engineering to medical devices
- IEC 62366: 2007+ A1: 2014, 1st Edition, Medical devices -- Application ofusability . engineering to medical devices
- IEC 62304: 2006, 15T Edition, Medical device software Software life cycleprocesses ●
- ISO 10993-1: 2009, Biological evaluation of medical devices- Part 1: Evaluation and ● testing within a risk management process
- . ISO 10993-5: 2009, Biological evaluation of medical devices- Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10: 2010, Biological evaluation of medical devices- Part 10: Tests for irritation and skin sensitization
In addition, the performance of BLE module for wireless co-existence was tested in an environment with equipment operating in the ISM band i.e., Bluetooth and Wi-Fi devices, cellphones, cordless phones, etc. The device met all specified requirements.
BLE module testing was conducted in accordance with the following standards:
- FCC CFR47 Part 15, Subpart C, July 2019, Intentional Radiator, §15.247, Operation within the bands 902-928 MHz, 2400-2483.5 MHz, and 5725-5850MHz
- FCC 47CFR PT 15 SPT B, August 2018 Title 47 CFR Part 15 Subpart B: Unintentional ● Radiators, FCC Part 15, Subpart B [FCC \$15.107 & FCC §15.109]
Battery testing was conducted in accordance with IEC 62133:2012 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells and for batteries made from them, for use in portable applications.
#### 9 Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC testing were conducted on the Motive device and complies with the IEC 60601-1, IEC 60601-2-10, and 60601-11 standards for safety and the IEC 60601-1-2 standard for EMC.
# 10 Software Verification & Validation Testing
The device's software has been validated in accordance with the requirements set forth in the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 11, 2005) and IEC 62304: 2006, 15" Edition, Medical device software - Software life cycle processes. The software validation tests demonstrated that the software version meets its design requirements.
#### Cybersecurity Controls 11
Cybersecurity related activities included system assessment and mitigations for potential
{9}------------------------------------------------
cybersecurity hazards and risks on the Motive device performance and data security. Information related to system design features, processes, testing, and controls to manage and mitigate the risks are included in this submission.
Cybersecurity information and supporting documents created and submitted according to the requirements of FDA guidance documents, Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, October 2, 2014 and Postmarket Management of Cybersecurity in Medical Devices, December 28, 2016.
# 12 Human Factors and Usability
A self-selection study and human factors study including usability assessment was conducted to validate the usability of the Motive device as an over-the-counter device and for home use. The results of the study support the labeling content and User's Manual instructions for successfully using the device as intended. The results of the human factors and usability study substantiates the acceptability of the risks identified during the risk assessment activities.
Human Factors information and supporting documents created and submitted according to the requirements of FDA guidance documents, FDA guidance document, Applying Human Factors and Usability Engineering to Medical Devices, February 3, 2016.
Additionally, the Motive device complies with the IEC 60601-1-6: 2010 for usability and IEC 62366: Application of usability engineering to medical devices.
# 13 Biocompatibility/ Material
The patient contacting materials used in the Motive device components include the Wrap with Smart Panel, electrodes, and electrode gel. The biocompatibility of the Wrap, electrodes, and electrode gel were previously evaluated and cleared under 510(K) submission K163067. The biocompatibility testing had been conducted according to the requirements of ISO 10993: 2009 and tested as a surface contact/ >24 hour to 30 days prolonged exposure. From the evaluation and submitted information, the components of the Motive device were found to be biocompatible for its use.
# 14 Shelf life/ Sterility
The non-invasive nature of the device obviates the need for sterile components; however, patientcontacting surfaces should be capable of being cleaned as needed. The Motive device is provided for single person use and does not require any of the components to be sterilized by the end user. It is intended for external use only. The electrodes are disposable and can be replaced as needed.
# 15 Performance Testing- Bench Testing of Electrical Muscle Stimulation (NMES) Waveform
All features and output specifications of the device, including those identified in Tables 1 and 2, were verified by individual pulse output waveform tracings for loads of 500, 2k, and 10k ohms, to simulate conditions that the device could counter during use.
{10}------------------------------------------------
# 16 Comparison of Technological Characteristics with the Predicate Device
CyMedica Orthopedics, Inc. demonstrated that, for the purposes of FDA's regulation of medical devices, the Motive device is substantially equivalent in design principles to the predicate device, CyMedica Orthopedics, Inc. IntelliHab System, K210604, the primary predicate device. The data included in this submission demonstrates design substantial equivalence to the predicate device, IntelliHab System. Both devices utilize the same NMES therapy electrical stimulation waveform and electrodes to deliver the therapy to the quadriceps muscles of the knee. Both devices are indicated for the treatment of knee pain and strengthening of the quadriceps muscle.
The primary predicate IntelliHab system is intended for prescription use and the subject Motive device is intended for over-the-counter (OTC) use. However, both devices are intended for home-use. Data from a self-selection study and human factors study were provided to support the Motive device safety for the OTC use. Additionally, a second predicate device is included, RelieforMe TENS/EMS Device Model UPK-GE01 (K192733) as an OTC device.
The intended use, design, materials, and functional characteristics of the Motive device and the primary predicate device are substantially the same. The subject device and primary predicate device are both intended as home-based use devices. Both devices have the same general purpose and function. Both devices are intended for the same general health population. Both devices utilize the same electrical stimulation therapy to strengthen the quadricep muscle to provide symptomatic temporary knee pain relief. Similar to the primary predicate device, Motive device provides the same NMES waveform, protocol, and pulse amplitude for the purpose of muscle strengthening and knee pain relief. Similar to the primary predicate device, Motive device includes a Wrap. Controller (pulse generator), RF communications, electrodes, and an App for delivery of the NMES therapy.
The Motive device is substantially equivalent in design and labeling to the primary predicate device and secondary predicate device. Nonclinical testing including self-selection study and human factors study performed on the Motive device are sufficient to demonstrate that the subject device is as safe and effective as the legally marketed predicate devices. The technological and labeling differences do not raise new or different questions about safety or effectiveness. Motive device is substantially equivalent to the predicate devices.
The following table demonstrates the similarities and differences between the subject Motive device and the predicate devices, IntelliHab System and RelieforMe TENS/EMS Device.
{11}------------------------------------------------
| TABLE 4-<br>Substantial<br>Equivalence<br>Comparison | Proposed Subject Device, Motive™ Knee<br>Wrap (Motive device), K220738<br>CyMedica Orthopedics, Inc. | Primary Predicate Device, IntelliHab™<br>System, K210604<br>CyMedica Orthopedics, Inc. | 2nd Predicate Device, RelieforMe TENS/EMS<br>Device Model UPK-GE01, K192733<br>UMEHEAL Ltd. |
|-----------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>use | The Motive Knee Wrap device is intended to<br>strengthen the quadriceps muscle using powered<br>muscle stimulation to provide symptomatic<br>temporary pain relief associated with knee<br>arthritis.<br>The Motive Knee Wrap device is indicated for<br>adults of 22 years of age and older. | The IntelliHab system is intended to strengthen<br>the quadricep muscle using powered muscle<br>stimulation to provide symptomatic temporary<br>pain relief associated with knee osteoarthritis and<br>improvement of the knee joint mobility when the<br>recommended treatment regimens are followed.<br>In addition, the Intellihab System is indicated for<br>the following:<br>• Retardation or prevention of disuse atrophy<br>• Evaluation of joint mobility by measuring and<br>recording range of motion.<br>The IntelliHab System is indicated for adults of<br>22 years of age and older. | TENS:<br>1. It is intended for temporary relief of<br>pain associated with sore and aching<br>muscles in the shoulder, neck, back,<br>waist, abdomen, lower extremities (legs),<br>upper extremities (arms) due to strain<br>from exercise or normal household and<br>work activities.<br>2. It is also intended for symptomatic relief<br>and management of chronic, intractable pain<br>and relief of pain associated with arthritis.<br>EMS:<br>3. To stimulate healthy muscles to<br>improve and facilitate muscle performance.<br>To be used for the improvement of muscle<br>tone and firmness, and for strengthening<br>muscles in the intended areas onthe body.<br>4. To be used for relaxation of muscle<br>spasm, increase of blood flow circulation,<br>prevention or retardation of disuse<br>atrophy, muscle re-education, maintaining<br>or increasing range of motion, and<br>immediate post-surgical stimulation of<br>calf muscles to prevent venous thrombosis |
| Intended Use | Over-the-Counter (OTC)<br>Home-use | Prescription<br>Home-use | OTC<br>Home-use |
| Regulation<br>names and<br>numbers | Powered muscle stimulator, 21 CFR 890.5850 | Powered muscle stimulator, 21 CFR 890.5850<br>Goniometer, AC-powered, 21 CFR 888.1500 | Powered muscle stimulator, 21 CFR 890.5850<br>TENS, 21 CFR 882.5890 |
| Product codes<br>and<br>classifications | Powered muscle stimulator, IPF, Class II | Powered muscle stimulator, IPF, Class II<br>Goniometer, AC-powered, KQX, Class I | Powered muscle stimulator, IPF, Class II<br>TENS, GZJ and NUH, Class II |
| TABLE 4-<br>Substantial<br>Equivalence<br>Comparison | Proposed Subject Device, Motive™ Knee<br>Wrap (Motive device), K220738<br>CyMedica Orthopedics, Inc. | Primary Predicate Device, IntelliHab™<br>System, K210604<br>CyMedica Orthopedics, Inc. | 2nd Predicate Device, RelieforMe TENS/EMS<br>Device Model UPK-GE01, K192733<br>UMEHEAL Ltd. |
| Product<br>identifications | Powered muscle stimulator, A powered muscle<br>stimulator is an electrically powered device<br>intended for medical purposes that repeatedly<br>contracts muscles by passing electrical currents<br>through electrodes contacting the affected body<br>area. | Powered muscle stimulator, A powered muscle<br>stimulator is an electrically powered device<br>intended for medical purposes that repeatedly<br>contracts muscles by passing electrical currents<br>through electrodes contacting the affected body<br>area.<br>Goniometer, AC-powered, A goniometer is an<br>AC-powered or battery powered device intended<br>to evaluate joint function by measuring and<br>recording ranges of motion, acceleration, or<br>forces exerted by a joint. | Powered muscle stimulator, A powered<br>muscle stimulator is an electrically powered<br>device intended for medical purposes that<br>repeatedly contracts muscles by passing<br>electrical currents through electrodes contacting<br>the affected body area.<br>TENS, A transcutaneous electrical nerve<br>stimulator for pain relief is a device used to<br>apply an electrical current to electrodes on a<br>patient's skin to treat pain. |
| Electrical<br>stimulation<br>modalities | NMES "Strength" program therapy | NMES "Strength" program therapy | NMES and TENS therapies |
| System<br>components | Controller, Controller changing cable, App,<br>Wrap with Smart Panel, electrodes, and<br>electrode gel | Controller, Controller changing cable, App,<br>conductive wrap, electrodes, and electrode gel | Controller and electrodes |
| Controller and<br>PCBA | Controller (pulse generator) consists of a<br>printed-circuit board within a plastic enclosure<br>that docks in a wrap docking receptacle via a 16-<br>pin interface. The Controller is the stimulation<br>generator and the tactile control device for the<br>system. The User can establish a wireless<br>connection between the Controller and their<br>mobile device. | Same | Unknown |
| Wrap | The Wrap design includes a modular design of a<br>basic textile-based wrap and a removable Smart<br>Panel where all the electronics, wires, and<br>electrodes are included in the Smart Panel.<br>Wrap with Smart Panel is single size.A single Wrap with Smart Panel can<br>provide treatment for both right and left<br>knees (not simultaneously). | The Wrap is one piece only and includes<br>integrated electronics, wires, and accelerometers<br>Wrap is not washable.<br>Wrap comes in different sizes.Two designated right and left wraps are<br>required to provide the treatment for the<br>right knee or left knee (not<br>simultaneously). | No wrap |
| TABLE 4-<br>Substantial<br>Equivalence<br>Comparison | Proposed Subject Device, Motive™ Knee<br>Wrap (Motive device), K220738<br>CyMedica Orthopedics, Inc. | Primary Predicate Device, IntelliHabTM<br>System, K210604<br>CyMedica Orthopedics, Inc. | 2nd Predicate Device, RelieforMe TENS/EMS<br>Device Model UPK-GE01, K192733<br>UMEHEAL Ltd. |
| Electrodes | Three snap type connected electrodes integrated<br>with the Smart Panel of the Wrap. One 2"<br>diameter electrode is required to activate the<br>Vastous Medialis Oblique (VMO) muscle and<br>two 2" x 3.5" electrodes for Rectus Femoris (RF)<br>muscle.<br>Although only three electrodes are required to<br>provide the therapy, in the Motive device setup<br>steps, Users are instructed to install 4 electrodes<br>(2 VMO electrodes) so they can apply the right<br>and left knee therapies when desired. | Three wire connected or pigtail electrodes<br>integrated with the conductive wrap. One 2x2"<br>electrode for VMO muscle and two 2" x 3.5"<br>electrodes for RF muscle. | Unknown |
| User Interface | The User's App serves as the primary user<br>interface and communication link between the<br>Controller and UI. The App includes screens for<br>application of NMES therapy, changing<br>intensities, starting, pausing, and stopping a<br>treatment. The App includes screens for creating<br>a profile, initial device setup guide, Help, and<br>reminders related to critical tasks such as<br>changing electrodes.<br>Additionally, the App improves therapy<br>compliance through a multi-faceted approach of<br>incentivizing User engagement through a<br>motivational User Dashboard, therapy reminders<br>and a rewards system. | Same UI design | Unknown |
| Stimulation<br>pulse<br>characteristics | The patented waveform is a unique,<br>asymmetrical, complex, and monophasic shaped<br>pulse which is designed to provide optimized<br>therapeutic benefits while maximizing comfort<br>and compliance.<br>The unique design features including a longer<br>pulse width (5 ms), monophasic polarity, work<br>cycles, and regulated power output provide a<br>longer duration of muscle contraction within a<br>20 minute treatment session. These unique<br>waveform characteristics lead to the therapeutic<br>benefits of the device reducing the knee pain and | Same | Unknown |
| TABLE 4-<br>Substantial<br>Equivalence<br>Comparison | Proposed Subject Device, Motive™ Knee<br>Wrap (Motive device), K220738<br>CyMedica Orthopedics, Inc. | Primary Predicate Device, IntelliHab™<br>System, K210604<br>CyMedica Orthopedics, Inc. | 2nd Predicate Device, RelieforMe TENS/EMS<br>Device Model UPK-GE01, K192733<br>UMEHEAL Ltd. |
| | improving knee joint functionality.<br>The Moive NMES stimulation pulses have three<br>distinct phases:<br>Phase 1: Pulse Spike (Rise Time, Peak,<br>and Decay)<br>Phase 2: Pulse Mesa<br>Phase 3: Recharge (off period) | | |
| Regulated<br>power output | Using a proprietary technology, the output of<br>stimulation circuit delivers energy to the User at<br>a constant power, independent of the load<br>impedance, hence power regulation during the<br>mesa portion of the pulse. | Same | Unknown…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.