OptiVu™ ROSA® MxR is indicated for displaying surgical workflow images from the ROSA® RECON platform in Mixed Reality. It includes functions for viewing the same surgical workflow steps and 2D visualizations as presented on the existing ROSA RECON platform user interface. When accessing ROSA® MxR from a stereoscopic head mounted display (HMD), images viewed are for informational purposes only and are not intended for diagnostic use.
Device Story
OptiVu ROSA MxR is a software-based accessory for the ROSA RECON platform; streams surgical workflow images and 2D visualizations from the ROSA RECON platform to a wireless, off-the-shelf head-mounted display (HMD) (e.g., HoloLens 2). Used intra-operatively in orthopedic surgery; allows surgeons to view duplicated user interface in a Mixed Reality environment. Surgeon interacts with the interface via hand gestures (touch, move, resize). Device provides informational display only; does not perform surgical planning or 3D visualization. Benefits include improved surgeon access to workflow data during procedures by placing the interface in a convenient location.
Clinical Evidence
Bench testing only. Verification and validation testing included physical/performance tests, engineering analysis, and usability engineering. Software verification and validation conducted per IEC 62304 standards.
Technological Characteristics
Software-based accessory; utilizes off-the-shelf HMD (e.g., HoloLens 2) for Mixed Reality visualization. Wireless connectivity (Wi-Fi) to ROSA RECON platform. Software level of concern: Major. No new hardware components; relies on existing ROSA RECON robotic/optical units.
Indications for Use
Indicated for displaying surgical workflow images and 2D visualizations from the ROSA RECON platform in a Mixed Reality environment during orthopedic surgeries. For use by surgeons intra-operatively. Not for diagnostic use.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
July 29, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Orthosoft, Inc. (d/b/a Zimmer CAS) Paul Hardy Associate Director 75 Queen St., Suite 3300 Montreal, QC H3C 2N6 Canada
# Re: K220733
Trade/Device Name: OptiVu ROSA MxR Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO, LLZ Dated: April 27, 2022 Received: May 2, 2022
# Dear Paul Hardy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For: Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220733
Device Name OptiVu™ ROSA® MxR
#### Indications for Use (Describe)
OptiVu™ ROSA® MxR is indicated for displaying surgical workflow images from the ROSA® RECON platform in Mixed Reality. It includes functions for viewing the same surgical workflow steps and 2D visualizations as presented on the existing ROSA RECON platform user interface. When accessing ROSA® MxR from a stereoscopic head mounted display (HMD), images viewed are for informational purposes only and are not intended for diagnostic use.
| Type of Use (Select one or both, as applicable) | | | | |
|-------------------------------------------------|---------------------------------------------|--|--|--|
| > Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | | | |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### K220733 510(k) Summary
In accordance with 21 CFR \$807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the OptiVu™ ROSA® MxR 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on September 13, 2019.
| Sponsor: | Orthosoft, Inc d/b/a Zimmer CAS<br>75 Queen St., Suite 3300<br>Montreal, QC, H3C 2N6, CANADA<br>Establishment Registration Number: 9617840 |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Paul Hardy<br>Regulatory Affairs Director<br>Telephone: 574-453-6739<br>Paul.Hardy@zimmerbiomet.com |
| Date: | July 29th, 2022 |
| Subject Device: | Trade Name: OptiVu™ ROSA® MxR<br>Common Name: OptiVu™ ROSA® MxR, OptiVu™ ROSA®<br>Mixed Reality<br>Classification Name:<br>• OLO-Stereotaxic Instrument (21 CFR 882.4560)<br>• LLZ- Medical Image Management and Processing System |
#### Predicate Device(s):
| 510(k) Number | Device Name | Manufacturer | Predicate |
|---------------|-------------------|------------------------------------|-----------|
| K182964 | ROSA® Knee System | Zimmer CAS | Primary |
| K200384 | HipXpert System | Surgical Planning Associates, Inc. | Reference |
(21 CFR 892.2050)
#### Purpose and Device Description:
OptiVu™ ROSA® Mixed Reality (ROSA® MxR) is indicated for displaying surgical workflow images from the ROSA® RECON Platform and the corresponding ROSA® clinical applications intra-operatively during orthopedic surgeries. It is intended to provide an additional means of display where the ROSA® RECON Platform user interface is duplicated into a Mixed Reality see-through environment. OptiVu™ ROSA® MxR allows the ROSA® RECON Platform user interface and corresponding ROSA® clinical applications (e.g. ROSA® Knee,
{4}------------------------------------------------
ROSA® Partial, or ROSA® Hip) to be streamed through a compatible head-mounted display (HMD) (e.g. HoloLens 2).
The subject device's main purpose is to place the duplicated virtual ROSA® user interface at a convenient location to provide the following functionalities:
- Mixed Reality visualization solution
- Register hand gestures as inputs such as to touch, move ● and resize views
- . Wireless connectivity between a HMD (e.g. HoloLens 2) and the ROSA® RECON Platform
OptiVu™ ROSA® MxR is indicated for displaying surgical workflow images from the ROSA® RECON platform in Mixed Reality. It includes functions for viewing the same surgical workflow steps and 2D visualizations as presented on the existing ROSA RECON platform user interface. When accessing ROSA® MxR from a stereoscopic head mounted display (HMD), images viewed are for informational purposes only and are not intended for diagnostic use.
#### Summary of Technological Characteristics:
The rationale for substantial equivalence is based on consideration of the following characteristics:
- . The proposed device and the primary predicate share the same previously cleared ROSA® RECON Platform hardware (robotic and optical unit) and the same previously cleared ROSA® RECON Platform software (OS, SDK Software, and Case Management software).
- The proposed device has similar principles of operation . as the predicate devices. OptiVu™ ROSA® MxR and the secondary predicate both utilize an off-the-shelf HMD (e.g. HoloLens 2) to stream images pertaining to surgical procedures for intra-operative use. The encrypted images streamed through the HMD (e.g HoloLens 2) allows the surgeon to view the display in a Mixed Reality environment. OptiVuTM ROSA® MxR and the secondary predicate device both display surgical workflow images. The ROSA® Knee System displays the workflow through its hardware component, the ROSA® RECON Platform whereas OptiVuTM ROSA® MxR connects through wifi to the ROSA®
# Indications for Use:
{5}------------------------------------------------
RECON Platform to stream a duplicate of the ROSA® RECON Platform user interface in Mixed Reality in real-time. The proposed device does not perform any pre-operative surgical planning as compared to the primary predicate device. The proposed device does not provide the viewing of images in 3D as compared to the secondary predicate.
#### Summary of Performance Data (Nonclinical and/or Clinical)
The following performance data was provided in support of the substantial equivalence determination:
#### Device Performance Testing
Verification and Validation Testing for OptiVu™ ROSA® MxR was conducted with the following aspects:
- Physical/Performance Tests- to ensure the performance of the implemented features and verify related design inputs
- . Engineering Analysis- to ensure the performance of the implemented features and verify related design inputs
- . Usability Engineering- addressed user interactions with OptiVu™ ROSA® MxR
#### Software Verification and Validation Testing
Software tests were conducted to satisfy requirements of the FDA Guidance for the Content Premarket Submissions for Software Contained in Medical Devices and IEC 62304 (Medical Device Software- Life Cycle Process). The software was considered a "major" level of concern, since OptiVu™ ROSA® MxR is an accessory to a medical device that has a major level of concern. The testing demonstrates that OptiVuTM ROSA® MxR does not raise any new issues of safety and effectiveness as compared to the predicate device(s).
#### Substantial Equivalence Conclusion
The proposed device and the primary predicate device share the same previously cleared ROSA® RECON Platform hardware (robotic and optical unit) and the same previously cleared ROSA® RECON Platform software (OS, SDK Software, and Case Management software).
{6}------------------------------------------------
In addition, the proposed device has similar principles of operation as the predicate devices. The proposed device utilizes an off-the-shelf HMD (e.g. HoloLens 2) to stream images pertaining to surgical procedures for intra-operative use which is the same as the secondary predicate device. The proposed device does not perform any pre-operative surgical planning as compared to the primary predicate. The proposed device does not provide the viewing of images in 3D as compared to the secondary predicate.
In summary, differences between the devices do not raise new questions of safety and effectiveness and the proposed device is at least as safe and effective as the legally marketed predicate device(s).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.