← Product Code [NJM](/productcode/NJM) · K220689

# Vatech Clismile (K220689)

_Rayence Co., Ltd. · NJM · May 6, 2022 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K220689

## Device Facts

- **Applicant:** Rayence Co., Ltd.
- **Product Code:** [NJM](/productcode/NJM.md)
- **Decision Date:** May 6, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.5470
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

Vatech Clismile is indicated for orthodontic movement of natural teeth, excluding mandibular bicuspid teeth.

## Device Story

Vatech Clismile is a self-ligating orthodontic bracket system used to treat malocclusion. The device consists of a polycrystalline alumina ceramic body and a nickel-cobalt alloy metal cap. It is applied to teeth by an orthodontist or dental professional to transmit physical force via orthodontic wires to correct tooth alignment. The device functions by securing the archwire within the bracket slot using the integrated metal cap mechanism. It is available in MBT and ROTH designs. By providing a stable attachment point for orthodontic forces, the device facilitates controlled tooth movement, aiding in the correction of dental malocclusions.

## Clinical Evidence

Bench testing only. Performance evaluation included dimensional analysis per ISO 27020, biological stability testing per ISO 10993-5 (cytotoxicity), ISO 10993-10 (sensitization), and ISO 10993-11 (systemic toxicity), and corrosion/hazardous element testing per ISO 22674. Results confirmed the device meets design specifications and is equivalent to the predicate.

## Technological Characteristics

Materials: Polycrystalline Alumina (100%) body, Nickel-Cobalt Alloy (MP35N) cap. Design: Self-ligating bracket, MBT/ROTH configurations. Standards: ISO 27020 (dimensions), ISO 10993 (biocompatibility), ISO 22674 (corrosion). Form factor: Half-transparent ceramic body with metal cap. Non-sterile packaging.

## Regulatory Identification

An orthodontic plastic bracket is a plastic device intended to be bonded to a tooth to apply pressure to a tooth from a flexible orthodontic wire to alter its position.

## Predicate Devices

- In-Ovation C ([K060837](/device/K060837.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.

May 6, 2022

Rayence Co.,Ltd Dave Kim Medical Device Regulatory Affairs Mtech Group 7505 Fannin St. Ste 610 Houston, Texas 77054

Re: K220689

Trade/Device Name: Vatech Clismile Regulation Number: 21 CFR 872.5470 Regulation Name: Orthodontic Plastic Bracket Regulatory Class: Class II Product Code: NJM Dated: February 28, 2022 Received: March 9, 2022

Dear Dave Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

K220689

Device Name Vatech Clismile

Indications for Use (Describe)

Vatech Clismile is indicated for orthodontic movement of natural teeth, excluding mandibular bicuspid teeth.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### K220689

# 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date 510k summary prepared: January 28th, 2022

#### Submitter's Name, address, telephone number, a contact person:

| Submitter's Name :      | Rayence Co., Ltd.                                       |
|-------------------------|---------------------------------------------------------|
| Submitter's Address:    | 14, Samsung 1-ro 1-gil, Hwaseong-si, Gyeonggi-do, Korea |
| Submitter's Telephone:  | +82-31-8015-6459                                        |
| Contact person:         | Mr. Kee Dock Kim / RA Team Manager / +82-31-8015-6459   |
| Official Correspondent: | Dave Kim (davekim@mtech-inc.net)                        |
| Address:                | 7505 Fannin St. Ste 610-V111, Houston, TX 77054         |
| Telephone:              | +713-467-2607                                           |

Name of the device, including the trade or proprietary name if applicable, the common or usual name and the classification name, if known:

| Trade/proprietary name | : Vatech Clismile                  |
|------------------------|------------------------------------|
| Common Name            | : Orthodontic ceramic bracket      |
| Regulation Number      | : 21 CFR 872.5470                  |
| Regulation Name        | : Bracket, Ceramic, Orthodontic    |
| Regulatory Class       | : Class II                         |
| Product Code           | : NJM(Orthodontic plastic bracket) |

#### Predicate Device :

| Trade/Device Name | : In-Ovation C                  |
|-------------------|---------------------------------|
| Common Name       | : Orthodontic ceramic bracket   |
| 510(k) Number     | : K060837                       |
| Regulation Number | : 21 CFR 872.5470               |
| Regulation Name   | : Bracket, Ceramic, Orthodontic |
| Regulatory Class  | : Class II                      |
| Product Code      | : NJM                           |

{4}------------------------------------------------

### 2. Device Description

This product is an orthodontic bracket and is used with orthodontic wire materials and other orthodontic products to treat malocclusion of the upper and lower teeth.

It is used to correct teeth by transmitting physical force to each tooth, and is a selfligation bracket using a ceramic body and a metal cap made of a metal material. Shelf-life is not applicable because of the low likelihood of time-dependent product degradation.

### 3. Indication for use

Vatech Clismile is indicated for orthodontic movement of natural teeth, excluding mandibular bicuspid teeth.

## 4. Summary of Design Control Risk management

Vatech Clismile orthodontic brackets were developed with the same structure, material / operation method, and purpose as In-Ovation C.

The risks and the hazardous impact of the device modification were analyzed with FMEA method. The specific risk control and protective measures to mitigate the risks from the modification were reviewed and implemented in the new product design phase. The overall assessment concluded that all risks and hazardous conditions identified arising from the design change were successfully mitigated and accepted.

{5}------------------------------------------------

# 5. Summary of the technological characteristics of the device compared to the predicate device:

Vatech Clismile described in this 510(k) has the same indications for use and similar technical characteristics as its predicate devices, IN-OVATION C (K060837).

| Characteristic                        | Proposed Device                                                                                                                            | Predicate Device                                                                                                                           | Similarity |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Manufacturer                          | Rayence CO., LTD                                                                                                                           | DENTSPLY International                                                                                                                     | -          |
| Product Name                          | Vatech Clismile                                                                                                                            | IN-OVATION C                                                                                                                               | -          |
| Picture                               | Image: Vatech Clismile                                                                                                                     | Image: IN-OVATION C                                                                                                                        | Similar    |
| 510(k) number                         | K220689                                                                                                                                    | K060837                                                                                                                                    | -          |
| Indications for<br>use                | Vatech Clismile is indicated for<br>orthodontic movement of natural<br>teeth, excluding mandibular<br>bicuspid teeth.                      | In-Ovation C is indicated for<br>orthodontic movement of natural<br>teeth, excluding mandibular<br>bicuspid teeth.                         | Same       |
| Material<br>composition<br>of Bracket | Polycrystalline Alumina(100%)                                                                                                              | Polycrystalline Alumina(100%)                                                                                                              | Same       |
| Material<br>composition<br>of Door    | Nikel-Cobalt Alloy(100%)                                                                                                                   | Nikel-Cobalt Alloy(100%)                                                                                                                   | Same       |
| Transparency                          | Half-transparency                                                                                                                          | Half-transparency                                                                                                                          | Same       |
| Bracket design                        | MBT, ROTH designs with<br>and without hook,<br>conforming to ISO<br>27020:2010 Dentistry –<br>Brackets and tube for Use in<br>Orthodontics | MBT, ROTH designs with<br>and without hook,<br>conforming to ISO<br>27020:2010 Dentistry –<br>Brackets and tube for Use in<br>Orthodontics | Same       |
| Self-ligating<br>mechanism            | Yes                                                                                                                                        | Yes                                                                                                                                        | Same       |
| Orientation<br>marking                | Yes                                                                                                                                        | Yes                                                                                                                                        | Same       |
| Non-sterile<br>packaging              | Yes                                                                                                                                        | Yes                                                                                                                                        | Same       |
| In-out(mm)                            | 1.05 to 1.54                                                                                                                               | 1.05 to 1.55                                                                                                                               | Same       |
| Torque(°)                             | -11 to +17                                                                                                                                 | -11 to +17                                                                                                                                 | Same       |
| Angulation(°)                         | 0 to 13                                                                                                                                    | 0 to 13                                                                                                                                    | Same       |
| slot sizes                            | 0.022 inch                                                                                                                                 | 0.018" / 0.022 inch                                                                                                                        | Similar    |

#### Comparison table

{6}------------------------------------------------

### 6. Summary of Performance Testing

IN-OVATION C and Vatech Clismile have the same composition and raw material of the components constituting the device, and the purpose and principle of operation of the calibration bracket are the same. Bracket generally uses a one-kit product consisting of 20 brackets. The K100-552-81 model of In-Ovation-C is composed of 16 ceramic and 4 metal brackets, and the Clismile is also composed of the same number of sets.

The ceramic bracket is composed of alumina (Body) and Ni-Co alloy (Cap, MP35N)

The design characteristics of the product were evaluated for seven items specified in ISO 27020, which is a bracket standard. As representative products, dimensional analysis of "Clismile 12-1223" and "IN-OVATION C H100-132-81" corresponding to the maximum canine was conducted, and it was confirmed that the dimensions were the same within the error range.

### 7. Summary for any testing and reference guidance:

> Dimension evaluation of prescription of orthodontic bracket according to standard ISO27020 (Brackets and tubes for use in orthodontics)

> Evaluation report on biological stability according to standard ISO10993-5 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity, ISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for skin sensitization, ), ISO 10993-11 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity

> Evaluation of hazardous element content and corrosion test for metal cap bracket parts according to standard ISO 22674(Metallic materials for fixed and removable restorations and appliances)

#### 8. Conclusions:

In accordance with the performance outcomes. Vatech Clismile demonstrated equivalent performance compared to In-Ovation-C. Therefore, Rayence claims the substantial equivalency between the proposed device and predicate device in terms of safety and effectiveness.

---

**Source:** [https://fda-staging.innolitics.com/device/K220689](https://fda-staging.innolitics.com/device/K220689)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com