K220680 · Shandong Huge Dental Material Corporation · EBI · Aug 12, 2022 · Dental
Device Facts
Record ID
K220680
Device Name
Denture Base Polymers
Applicant
Shandong Huge Dental Material Corporation
Product Code
EBI · Dental
Decision Date
Aug 12, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3760
Device Class
Class 2
Indications for Use
Denture Base Polymers is used for fabrication and repair of partial and full denture base for patients with missing teeth.
Device Story
Denture Base Polymers consists of a powder (polymethyl methacrylate) and a liquid (methyl methacrylate) component; mixed to form a resin for the fabrication and repair of partial and full denture bases. Used in dental clinics and hospitals by dental professionals. The device functions as a dental restorative material; physical and chemical properties are verified against ISO 20795-1 standards. The output is a cured denture base; clinical benefit is the restoration of dental function for patients with missing teeth.
Clinical Evidence
No clinical data. Bench testing only. Biocompatibility testing performed per ISO 10993, including cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity, subchronic toxicity, genotoxicity, and implantation. Physical/chemical properties verified against ISO 20795-1.
Technological Characteristics
Materials: Polymethyl methacrylate (PMMA), Methyl methacrylate (MMA), Ethyleneglycol dimethacrylate (EDMA). Curing methods: Heat Curing and Self Curing. Standards: ISO 20795-1 (dental base material), ISO 10993 (biocompatibility). Non-sterile.
Indications for Use
Indicated for fabrication and repair of partial and full denture bases in patients with missing teeth.
Regulatory Classification
Identification
A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 12, 2022
Shandong Huge Dental Material Corporation Maggie Zheng Regulatory Affairs Manager No. 68 Shanhai Road, Donggang District Rizhao City, Shandong 276800 China
Re: K220680
Trade/Device Name: Denture Base Polymers Regulation Number: 21 CFR 872.3760 Regulation Name: Denture Relining, Repairing, Or Rebasing Resin Regulatory Class: Class II Product Code: EBI Dated: April 27, 2022 Received: May 2, 2022
Dear Maggie Zheng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K220680
Device Name Denture Base Polymers
Indications for Use (Describe)
Denture Base Polymers is used for fabrication and repair of partial and full denture base for patients with missing teeth.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 005 510 (k) Summary K220680
This summary of 510(k) substantial equivalence information is being submitted in accordance with the requirements of 21 C.F.R. 807.92.
### 1. Date Summary Prepared: July 20, 2022
### 2. Submitter Information:
| Name | Shandong Huge Dental Material Corporation |
|----------------|-----------------------------------------------------------------------------------------------|
| Address | No. 68 Shanhai Road, Donggang District, Rizhao City, Shandong<br>Province, 276800, P.R. China |
| Telephone | 086-633-2277268 |
| Contact Person | Ms. Maggie Zheng |
| Contact Title | Regulatory Affairs Manager |
| E-mail | zhengxy@hugedent.com |
### 3. Device Name
Trade name: Denture Base Polymers Common name: Dental base material Classification name: resin, denture, relining, repairing, rebasing (21 C.F.R. 872.3760) Regulatory Class: II Product Code: EBI
## 4. Substantially Equivalent Device
| Company Name | Device Name | 510 (k)<br>NO. | Substantially<br>Equivalent (SESE)<br>Decision Date | Product code | Remarks |
|---------------------|----------------------------------------------------------------|----------------|-----------------------------------------------------|--------------|------------------------|
| VERTEX-DENTAL<br>BV | Vertex Rapid<br>Simplified | K102654 | 12/03/2010 | EBI | Primary<br>Predicate |
| VERTEX-DENTAL<br>BV | Vertex Self Curing,<br>Vertex Castavaria,<br>Vertex Castapress | K102640 | 03/25/2011 | EBI | Secondary<br>Predicate |
These predicate devices have not been subject to a design-related recall.
### 5. Description of Device
Denture Base Polymers contains powder and liquid, wherein the power is mainly composed of polymethyl methacrylate, and the main composition of the liquid is methyl methacrylate.
{4}------------------------------------------------
Denture Base Polymers is used for fabrication and repair of partial and full denture base for patients with missing teeth, and its extensive indications can meet the needs of the production of complete denture.
### 6. Indications for use
Denture Base Polymers is used for fabrication and repair of partial and full denture base for patients with missing teeth.
## 7. Summary of Physical and Chemical Properties Tests
- Chemical Composition:
The device has similar chemical composition as the predicate devices. All of them are mainly composed of Polymethyl methacrylate and Pigments (Powder) and Methyl methacrylate, Ethyleneglycol dimethacrylate (Liquid).
- Technological characteristics:
The device has the same technological characteristics as the predicate devices (Heat Curing and Self Curing). And the device is similar in specifications and shades as the predicate devices.
- Properties:
The device has comparable physical and chemical properties as the predicate devices.
(Meeting the requirements of ISO standards for the Dental base material, ISO 20795-1)
- Applications:
The device has similar indications for use as the sum of the predicate devices: fabrication and repair of partial and full denture base.
## 8. Technological Characteristics:
The new device, Denture Base Polymers, has the same design, main materials and chemical composition as the predicate device.
| Comparison Items | New Device | Primary Predicate | Secondary Predicate |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Denture Base Polymers | Vertex Rapid Simplified | Vertex Self Curing, Vertex Castavaria, Vertex Castapress |
| | K220680 | K102654 | K102640 |
| 1) Regulatory Classifications | same | same | same |
| 2) Indications for use | This product is used for fabrication and repair of partial and full denture base for patients with missing teeth. | The Vertex Rapid Simplified is indicated for:<br>1. Fabrication of full dentures<br>2. Fabrication of partial dentures | Vertex cold-curing denture base materials are indicated for:<br>1. Manufacture of full and partial dentures<br>2. Repair of full and partial dentures<br>3. Rebasing of full and partial dentures<br>4. Relining of full and partial |
{5}------------------------------------------------
# HUGE
| | Comparison Items | New Device | Primary Predicate | Secondary Predicate | | | | |
|-----|-----------------------------|--------------------------------------------------------------------|--------------------------------------------------------|------------------------------------------------------------------------|--------------------------------------|--------------------------------------------------------|-----------------------------------------------|------------------------------------------|
| | | Denture Base Polymers<br>K220680 | Vertex Rapid Simplified<br>K102654 | Vertex Self Curing, Vertex<br>Castavaria, Vertex Castapress<br>K102640 | | | | |
| 3) | Contraindications | NA | NA | dentures<br>NA | | | | |
| 4) | Composition<br>of Materials | PMMA, MMA, EDMA | PMMA, MMA, EDMA | PMMA, MMA, EDMA | | | | |
| 5) | Physical<br>Properties | New<br>Device | Physical<br>parameters | Flexural<br>strength | Water<br>absorption | Solubility | Residual<br>methyl<br>methacrylate<br>monomer | Flexural<br>modulus |
| | | | | Type 1: ≥ 65 MPa<br>Type 2: ≥ 60 MPa | ≤ 32 ug/mm³ | Type 1: ≤ 1.6 ug/mm³<br>Type 2: ≤ 8.0 ug/mm³ | Type 1: ≤ 2.2%<br>Type 2: ≤ 4.5% | Type 1: ≥ 2000 MPa<br>Type 2: ≥ 1500 MPa |
| | | Type 1 | Meet the criteria<br>per ISO 20795-1 | Meet the criteria<br>per ISO 20795-1 | Meet the criteria<br>per ISO 20795-1 | Meet the criteria<br>per ISO 20795-1 | Meet the criteria<br>per ISO 20795-1 | |
| | | Type 2 | Meet the criteria<br>per ISO 20795-1 | Meet the criteria<br>per ISO 20795-1 | Meet the criteria<br>per ISO 20795-1 | Meet the criteria<br>per ISO 20795-1 | Meet the criteria<br>per ISO 20795-1 | |
| | | Primary Predicate<br>(K102654)<br>Secondary Predicate<br>(K102640) | Meet the criteria<br>per ISO 20795-1 | Meet the criteria<br>per ISO 20795-1 | Meet the criteria<br>per ISO 20795-1 | Meet the criteria<br>per ISO 20795-1 | Meet the criteria<br>per ISO 20795-1 | |
| 6) | Labeling | | similar | similar | | similar | | |
| 7) | Target Population | | dental patients | dental patients | | dental patients | | |
| 8) | Anatomical Site | | on teeth | on teeth | | on teeth | | |
| 9) | Where Used | | used in hospital, dental clinic<br>and relevant places | used in hospital, dental clinic<br>and relevant places | | used in hospital, dental clinic<br>and relevant places | | |
| 10) | Human Factors | | dental professional | dental professional | | dental professional | | |
| 11) | Design | | same | same | | same | | |
| 12) | Cautions | | similar | | similar | | similar | |
| 13) | Standards Met | | same | | same | | same | |
| 14) | Biocompatibility | | biocompatible | biocompatible | | biocompatible | | |
| 15) | Sterility | | Non-sterile | | Non-sterile | | Non-sterile | |
| 16) | Chemical Safety | | similar | | similar | | similar | |
## 9. Summary of Biocompatibility
The new device, Denture Base Polymers, is substantially equivalent to the predicate devices that have been on the market for years and with no clinical adverse events. The formulation of new device does not contain any new or non-conventional chemicals compared to the legally marketed predicate device.
{6}------------------------------------------------
We selected our Denture Base Polymers (Model: Type 2 Class 1) as the representative in biocompatibility tests and those biocompatibility test reports can be used in the biological evaluation of Denture Base Polymers.
Biocompatibility tests were performed fully following the ISO 10993 standards. The test items include Cytotoxicity; Sensitization; Irritation or Intracutaneous Reactivity; Acute Systemic Toxicity; Material-Mediated Pyrogenicity; Subchronic Toxicity; Genotoxicity and Implantation.
### 10. Clinical Performance Data
Not applicable. Clinical performance testing has not been performed on the subject device.
### 11. Summary of Substantial Equivalence
As with the comparison shown in substantial equivalence discussion, these devices are same or similar in almost all aspects. The details of indications for use, composition of materials, physical properties, labeling, stability/shelf life and cautions are slightly different, but the minor differences between the new device and the predicate devices are out of significance and do not affect neither the general intended use nor substantial equivalence.
We conclude that our Denture Base Polymers is substantially equivalent to the predicate devices described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.