K220678 · Aspenstate, Inc. · IZL · Mar 31, 2022 · Radiology
Device Facts
Record ID
K220678
Device Name
APX-24 Portable X-ray System
Applicant
Aspenstate, Inc.
Product Code
IZL · Radiology
Decision Date
Mar 31, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1720
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The APX-24 Portable X-ray is intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic X-rays.
Device Story
Portable X-ray system; generates diagnostic X-ray images for adult and pediatric patients. Operates via high-frequency (102kHz) inverter; utilizes monoblock generator; controlled by dual-stage, deadman-type exposure switch. Used by physicians or technicians in clinical settings. Operator selects kVp and exposure time via pushbuttons; system uses software-based control with 2 CPUs to manage X-ray generation. Provides diagnostic images to assist healthcare providers in clinical decision-making; benefits patients by enabling portable, point-of-care diagnostic imaging.
Clinical Evidence
Bench testing only. Performance evaluation confirmed compliance with IEC 60601-1, IEC 60601-1-3, IEC 60601-2-54, IEC 60601-2-65, and related electromagnetic compatibility and safety standards.
Indicated for use by qualified/trained physicians or technicians on adult and pediatric subjects for diagnostic X-ray imaging.
Regulatory Classification
Identification
A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
March 31, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Aspenstate, Inc. % Mr. Dave Kim Regulatory Affairs Mtech Group 7505 Fannin St, Suite 610 HOUSTON TX 77054
Re: K220678
Trade/Device Name: APX-24 Portable X-ray System Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: Class II Product Code: IZL Dated: February 28, 2022 Received: March 8, 2022
## Dear Mr. Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel Burk, Ph.D. Assistant Director Diagnostic X-ray Systems Team Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220678
Device Name APX-24 Portable X-ray System
### Indications for Use (Describe)
The APX-24 Portable X-ray is intended for use by a qualified/trained physician on both adult and pediatric subjects for taking diagnostic X-rays.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <div style="display:inline-block;">☑</div> |
|----------------------------------------------|--------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <div style="display:inline-block;">☐</div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
This summary of 510(k) information is submitted following the requirements of 21 CFR Part 807.92.
| Date 510k summary prepared: | February 28, 2022 |
|-----------------------------|-------------------|
|-----------------------------|-------------------|
#### I. SUBMITTER
| Submitter's Name<br>Submitter's Address | Aspenstate, Inc<br>801 North Jupiter Rd, Suite 200<br>Plano TX 75074 |
|------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Submitter's Telephone<br>Contact person | Albert Kim/Managing Director<br>albert.kim@aspenstate.com<br>Tel: +1- 214-257-0113 |
| Official Correspondent<br>Address<br>Telephone | Dave Kim (davekim@mtech-inc.net)<br>7505 Fannin St, Suite 610, Houston, TX 77054<br>+713-467-2607 |
#### II. DEVICE
| Trade/Device Name: | APX-24 Portable X-ray System |
|--------------------|----------------------------------------|
| Regulation Number: | 21 CFR 892.1720 |
| Regulation Name: | Mobile x-ray system |
| Regulatory Class: | II |
| Product Codes: | IZL |
| Common/Usual Name: | Digital Mobile Diagnostic X-Ray System |
#### III. PREDICATE DEVICE
| Trade/Device Name: |
|--------------------|
| 510(k) Number: |
| Regulation Number: |
| Regulation Name: |
| Regulatory Class: |
| Product Codes: |
| Common/Usual Name: |
MinXray HF100H+™ K052721 21 CFR 892.1720 Mobile x-ray system II IZL Digital Mobile Diagnostic X-Ray System
{4}------------------------------------------------
#### DEVICE DESCRIPTION: IV.
The APX-24 Portable X-ray is a portable unit that operates from120 V / 220-240V, 50 / 60 Hz. The unit utilizes a newly designed high-frequency inverter and can bemounted to a tripod or support arm. The operator must observe the usual safety precautions regarding the use of x- rays.
#### V. Indications for Use: 21 CFR 807 92 (a) (5)
The APX-24 Portable X-ray is intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic X-rays.
| Characteristic | MinXray HF100H (K052721) | Aspenstate APX-24 |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: | Intended for use by a<br>qualified/trained physician or<br>technician on both adult and<br>pediatric subjects for taking<br>diagnostic x-rays. | The APX-24 Portable X-ray is<br>intended for use by a<br>qualified/trained physician or<br>technician on both adult and<br>pediatric subjects for taking<br>diagnostic X-rays. |
| Size/weight | 406 x 222 x 241 mm18.6kgs | 414 x 230 x 215.5mm, 12.7kg |
| Energy Source | 100-140V 50-60 HzAC3 0kVA | 120VAC/220-240VAC, 50/60Hz |
| Mounting method | Unit is usually mounted to a<br>MinXray XGS MKIII Portable<br>Stand | This unit is not mounted to a<br>Portable Stand |
| User Interface | Up-Down pushbuttons for kVp<br>selections and exposure time<br>selections with LED indictors<br>and mAs indicators | Up/Down pushbuttons for kVp<br>selections and exposure time<br>selections with LED indicators |
| Exposure switch | Dual-stage, deadman type | Dual-stage, deadman type |
| Controls | Software-based, 2 CPUs. | Software-based, 2 CPUs. |
| Construction | Monobloc HF a generator,<br>Medical full-bridge inverter<br>system | Monoblock HF a generator, Medical<br>full-bridge inverter system |
| High Voltage<br>Energy Source | High frequency (60kHz)<br>inverter | High frequency (102kHz) inverter |
| Line Voltage<br>adjustment | Automatic, dynamic | Automatic, dynamic |
| | | |
| Exposure times | 0.03-0.2 sec(in 0.01 sec. Steps)<br>0.2-04 sec(in 0.02 sec. Steps)<br>0.4-1.0 sec(in 0.05 sec. Steps)<br>1.0-4.0 sec(m 0.1 sec. Steps) | 0.03-0.2 sec(in 0.01 sec. Steps)<br>0.2-04 sec(in 0.02 sec. Steps)<br>0.4-1.0 sec(in 0.05 sec. Steps)<br>1.0-4.0 sec(m 0.1 sec. Steps) |
| Tube potential(kV) | 40 - 100kV 2kVstep | 40-100kV (1kV step) |
| kV steps | 31(2kV-step) | 31(2kV-step) |
| Tube current(mA) | 30mA(40-60 kV)<br>25mA(62-80 kV)<br>20mA(82-100 kV) | 30 mA (40-50 kV)<br>30, 35, 40 mA (51-60 kV)<br>25, 30 mA (61-70 kV)<br>15, 20, 30 mA (71-80 kV)<br>13, 25 mA (81-90 kV)<br>10, 20 mA (91-100 kV) |
| mA steps | Constant | Constant |
| X-ray tube | Canon D-124S | Canon D-125SB |
| Anode heatStorage | 20,000HU | 50,000 HU |
| Focal SpotSize | 1.2 mm | 1.2 mm |
| mAs | 0.6-120mAs | 0.32-100mAs |
| Total filtration | 3.2mm AL equivalent | Min 2.8 mm AL |
| Collimator | Advantech R72<br>Continuously adjustable<br>light beam type with<br>central x-ray indicator | Continuously adjustable light beam type with central x-ray indicator |
| Source to Skin<br>Distance (SSD) | 300 mm | 300 mm |
| Performance<br>Standard | 21CFR 1020.30 | IEC60601-1-3, IEC60601-1-6,<br>IEC60601-2-28, IEC60601-2-54<br>21CFR 1020.30, 21CFR 1020.31 |
| Electrical safety | UL2601, IEC60601-1 | IEC60601-1 |
Technological characteristics: 21 CFR 807 92 (a) (6) Comparison Table
{5}------------------------------------------------
### ASPENSTATE, INC
801 North JupiterRd, Suite 200, Plano, TX 70574 USA
Tel.: +214-257-0113
#### VI. Discussion of differences
The subject device is similar to the predicate device regarding the indications for use and technological application. The subject and predicate devices are portable X-ray systems for taking diagnostic X-rays of human anatomy using a fixed tube current and voltage (kVp).
The differences are the device appearance, size, and user interface.
{6}------------------------------------------------
ASPENSTATE, INC 801 North JupiterRd, Suite 200, Plano, TX 70574 USA Tel.: +214-257-0113
#### VII. Nonclinical testing
Testing was performed successfully according to the following standards:
- > IEC 60601-1-3:2008+A1:2013
- > IEC 60601-2-65:2012
- > EN 60601-2-65:2013
- > EN 60601-1-2:2015
- > EN 61000-3-2:2014
- > EN 61000-3-3:2013
- > IEC 62133:2012
- > EN 62133:2013
- > EN 60601-1:2006/A1:2013
- > IEC 60601-2-54:2009 (First edition) + A1:2014 for use in conjunction with IEC 60601-1:2015 (third edition) + A1:2012
Furthermore, the following Specific Guidance Document was utilized in the device development to ensure the safety of this device for both the operators and patients: "The Content of Premarket Submissions for Software Contained in Medical Devices"
#### VIII. Conclusion:
By the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided above comparison table. the APX-24 Portable X-ray System has little difference with its size and user interface as the information in the table. However, the system is substantially equivalent to the predicate devices with its design, mechanical and electrical performance as described.
Performance evaluation (test) reports and the device inspection report confirmed that the APX-24 Portable X-ray System is suitable for the device's intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.