AccuCTP is an image processing software package to be used by trained professionals, including but not limited to physicians and medical technicians. The software runs on a standard off-the-shelf computer, and can be used to perform image viewing, processing and analysis of brain images. Data and images are acquired through DICOM compliant imaging devices. AccuCTP provides both viewing and analysis capabilities for functional and dynamic imaging datasets acquired with CT Perfusion (CTP), which can visualize and analyze dynamic imaging data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume.
Device Story
AccuCTP is standalone software for processing DICOM-compliant CT perfusion brain images; operates on standard off-the-shelf PC workstations. Inputs: dynamic CT perfusion datasets. Processing: performs motion correction, cranial removal, and time-intensity analysis to calculate perfusion parameters (CBF, CBV, MTT, Tmax, AIF/VOF). Outputs: threshold maps, time-intensity plots, and volumetric measurements (including rCBF/Tmax mismatch). Used by physicians and technicians in clinical settings to assist in visualizing tissue perfusion changes. Output aids clinical decision-making by providing quantitative perfusion data; benefits patient by facilitating assessment of brain tissue blood flow and volume.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including phantom testing to validate parameter maps and volumes against ground truth, and a comparative validation study against the RAPID predicate. The validation study evaluated agreement in parameter maps and volume results at rCBF<30% and Tmax>6s thresholds. All results met pre-defined pass/fail criteria.
Technological Characteristics
Standalone software; runs on standard off-the-shelf PC workstations. Processes DICOM-compliant CT perfusion data. Core functions: motion correction, cranial removal, volumetry of threshold maps, time-intensity plotting, and perfusion parameter calculation (CBF, CBV, MTT, Tmax).
Indications for Use
Indicated for trained professionals, including physicians and medical technicians, to perform viewing, processing, and analysis of brain images acquired via DICOM-compliant CT perfusion imaging. Used to visualize and analyze dynamic contrast changes to calculate tissue perfusion and blood volume parameters.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
ArteryFlow Technology Co., Ltd. % Ashley Fu RA Specialist 459 Qianmo Road, Suite C1-501, Binjiang District, Hangzhou, Zhejiang 310051 CHINA
## Re: K220663
November 22, 2022
Trade/Device Name: AccuCTP Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ Dated: October 18, 2022 Received: October 18, 2022
Dear Ashley Fu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jessica Lamb
Jessica Lamb, Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K220663
Device Name AccuCTP
Indications for Use (Describe)
AccuCTP is an image processing software package to be used by trained professionals, including but not limited to physicians and medical technicians. The software runs on a standard off-the-shelf computer, and can be used to perform image viewing, processing and analysis of brain images. Data and images are acquired through DICOM compliant imaging devices.
AccuCTP provides both viewing and analysis capabilities for functional and dynamic imaging datasets acquired with CT Perfusion (CT-P), which can visualize and analyze dynamic imaging data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue blood volume.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
PSC Publishing Services (301) 443-6740 BR
{3}------------------------------------------------
## K220663: 510(k) Summary
This 510(k) Summary of safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.
### 1. Submitter's Information
Submitter: ArteryFlow Technology Co., Ltd.
Address: 459 Qianmo Road, Suite C1-501, Binjiang District, 310051 Hangzhou City, Zhejiang Province, China
Phone Number: +86-571-86772567
Primary correspondent: Xiang Jianping, PhD, General Manager
Email: jianping.xiang@arteryflow.com
Secondary correspondent: Ashley Fu, RA Specialist
Email: fang.fu@arteryflow.com
Date of preparation: October 18, 2022
### 2. Device Information
Trade/ Device Name: AccuCTP
Common Name: Radiological Image Processing Software
Regulatory Class: Class II
Regulation Description: Medical image management and processing system
Regulation number: 892.2050
Classification Product Code: LLZ
#### 3. Predicate Device Information
Manufacturer: iSchemaView, Inc.
Device Name: RAPID
Regulatory Class: Class II
Regulation Number: 892.2050
Classification Product Code: LLZ
510(k) number: K121447
#### 4. Device Description
AccuCTP is a standalone software package that provides visualization and study of changes of tissue perfusion in digital images captured by CT (Computed Tomography). The software provides viewing, quantification, analysis and reporting capabilities, and it allows
{4}------------------------------------------------
repeated use and continuous processing of data and can be deployed on a supportive customer's PC that meets the minimum system requirements.
AccuCTP works with the DICOM compliant medical image data. AccuCTP provides tools for performing the following types of analysis:
- volumetry of threshold maps .
- time intensity plots for dynamic time courses .
- . measurement of mismatch between rCBF and Tmax threshold volumes obtained from the same scan.
#### 5. Indications for Use
AccuCTP is an image processing software package to be used by trained professionals, including but not limited to physicians and medical technicians. The software runs on a standard off-the-shelf computer, and can be used to perform image viewing, processing and analysis of brain images. Data and images are acquired through DICOM compliant imaging devices.
AccuCTP provides both viewing and analysis capabilities for functional and dynamic imaging datasets acquired with CT Perfusion (CTP), which can visualize and analyze dynamic imaging data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume.
#### 6. Performance Data
Software verification and validation testing have been performed in support that AccuCTP satisfies all design specifications and provides accurate computation of output parameters. During the development, potential hazards were controlled by the risk management report, including risk analysis, risk mitigation, verification and validation.
AccuCTP have conducted a phantom test and a validation study of the performance to validate the core algorithm through parameter maps and volumes validation. Besides, filtering and cranial removal have also been validated.
In the phantom test, a group of phantoms was designed to validate the parameter maps and volume compared with ground truth. Parameter map and Volume results were quantitatively analysed and met the pre-defined pass/fail criteria.
In the validation study, a calculation performance validation was conducted to evaluate the agreement between AccuCTP and RAPID CTP in calculating the parameter maps as well as the volume results at the rCBF<30% and Tmax>6s thresholds. Parameter map and Volume results were quantitatively analysed and met the pre-defined pass/fail criteria.
Performance data demonstrates that AccuCTP is as safe and effective as iSchemaView's RAPID for performing CTP analysis.
#### 7. Substantial Equivalence
AccuCTP is a generic tool without being disease-specific and it enables the users to
{5}------------------------------------------------
manually select the outputted threshold values. AccuCTP does not have diagnostic implications and has same intended use compared with predicate device.
AccuCTP shares the same principle of operation and technological characteristics as RAPID (K121447) with respects to CTP functionalities. The minor differences in indications do not alter the intended use of the device, and the minor technological differences raise no new issues of safety or effectiveness. A comparison of the technological characteristics of the predicate and subject device is given in the table below.
| Parameter | Subject device | Predicate device |
|-----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | AccuCTP<br>K220663 | RAPID<br>K121447 |
| Regulation | LLZ<br>21 CFR §892.2050 | LLZ<br>21 CFR §892.2050 |
| Indications for Use | AccuCTP is an image processing<br>software package to be used by<br>trained professionals, including but<br>not limited to physicians and<br>medical technicians. The software<br>runs on a standard off-the-shelf<br>computer, and can be used to<br>perform image viewing, processing<br>and analysis of brain images. Data<br>and images are acquired through<br>DICOM compliant imaging devices.<br><br>AccuCTP provides both viewing<br>and analysis capabilities for<br>functional and dynamic imaging<br>datasets acquired with CT<br>Perfusion (CTP), which can<br>visualize and analyse dynamic<br>imaging data, showing properties of<br>changes in contrast over time. This<br>functionality includes calculation of<br>parameters related to tissue flow<br>(perfusion) and tissue blood<br>volume. | iSchemaView's RAPID is an image<br>processing software package to be<br>used by trained professionals,<br>including but not limited to<br>physicians and medical technicians.<br>The software runs on a standard<br>"off-the-shelf" computer or a virtual<br>platform, such as VMware, and can<br>be used to perform image viewing,<br>processing and analysis of brain<br>images. Data and images are<br>acquired through DICOM compliant<br>imaging devices.<br><br>iSchemaView's RAPID provides<br>both viewing and analysis<br>capabilities for functional and<br>dynamic imaging datasets acquired<br>with CT Perfusion (CT-P) and MRI<br>including a Diffusion Weighted MRI<br>(DWI) Module and a Dynamic<br>Analysis Module (dynamic<br>contrast-enhanced imaging data for<br>MRI and CT).<br><br>The DWI Module is used to<br>visualize local water diffusion<br>properties from the analysis of<br>diffusion - weighted MRI data.<br>The Dynamic Analysis Module is |
| Computer<br>Platform | Standard off-the-shelf PC<br>workstation | used for visualization and analysis<br>of dynamic imaging data, showing<br>properties of changes in contrast<br>over time. This functionality<br>includes calculation of parameters<br>related to tissue flow (perfusion)<br>and tissue blood volume. |
| | | Standard off-the-shelf PC workstation/server Virtual platform such as VMware |
| Functional<br>Overview | AccuCTP is a software package<br>that provides for the visualization<br>and study of changes of tissue<br>perfusion in digital images captured<br>by CT. AccuCTP provides viewing<br>and quantification. | RAPID is a software package that<br>provides for the visualization and<br>study of changes of tissue perfusion<br>and diffusion in digital images<br>captured by CT and MRI. RAPID<br>provides viewing and quantification. |
| Data Acquisition | Acquires medical image data from<br>DICOM compliant imaging devices<br>and modalities | Acquires medical image data from<br>DICOM compliant imaging devices<br>and modalities |
| Data/Image Types | DICOM Format | DICOM Format |
| Acquisition and<br>Modalities<br>Features | CT (CT Perfusion) | CT (CT Perfusion) MRI (Perfusion Weighted Images and Diffusion Weighted Images) |
| Computed<br>Parameter Maps<br>of Perfusion CT | Cerebral blood flow (CBF) | Yes |
| | Cerebral blood volume (CBV) | Yes |
| | Mean transit time (MTT) | Yes |
| | Tissue residue function time to<br>peak (Tmax) | Yes |
| | Arterial input function (AIF)/Venous<br>output function (VOF) | Yes |
| | Time-course | Yes |
| | Brain mask | Yes |
| CT Tools | Region of interest (ROI) and<br>Volumetry | Yes |
| | Motion correction | Yes |
| | Export perfusion files to local<br>address. | Export perfusion and diffusion files<br>to PACS and DICOM file systems |
| | Acquire, transmit, process, and | Yes |
### Table 1 General Comparison
{6}------------------------------------------------
{7}------------------------------------------------
| | | store medical images | | |
|--|--|----------------------|--|--|
|--|--|----------------------|--|--|
# 8. Conclusion
AccuCTP has same intended use and is as safe and effective as the predicate device RAPID (K121447). The differences between the subject and predicate devices do not raise new questions of safety and effectiveness. Thus, the AccuCTP is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.