← Product Code [OMP](/productcode/OMP) · K220660

# Prevena Restor Adapti-Form Dressing (K220660)

_3M Healthcare Business Group · OMP · Jul 1, 2022 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K220660

## Device Facts

- **Applicant:** 3M Healthcare Business Group
- **Product Code:** [OMP](/productcode/OMP.md)
- **Decision Date:** Jul 1, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4780
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Prevena™ Restor™ Adapti-Form™ Dressing is part of the Prevena™ Restor™ Incision Management System and is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudate via the application of negative pressure wound therapy.

## Device Story

Prevena™ Restor™ Adapti-Form™ Dressing is a customizable dressing component of the Prevena™ Restor™ Incision Management System. It manages surgical incision environments by maintaining a closed system and removing exudate via negative pressure wound therapy (NPWT). The dressing consists of a foam interface, V.A.C. drape, SensaT.R.A.C. pad, and hydrocolloid sealing strips. It is applied in an operating room or surgery center and may transition home with the patient. The dressing is designed to be cut and customized by the clinician to fit linear, non-linear, or intersecting incisions across various anatomical locations. It connects to compatible NPWT therapy units (Prevena Plus, ActiV.A.C., V.A.C. ULTA, or V.A.C. Rx4) which provide the vacuum source. By maintaining a closed environment and removing fluid, the device aims to manage the incision site, potentially benefiting the patient by supporting the healing environment of the surgical wound.

## Clinical Evidence

No clinical data was required. Substantial equivalence was demonstrated through non-clinical bench testing, including negative pressure performance, package integrity/stability (ISO 11607-1), product performance stability post-sterilization, human factors evaluation, and biocompatibility (ISO 10993-1).

## Technological Characteristics

Customizable dressing kit including foam interface, V.A.C. drape, SensaT.R.A.C. pad, and hydrocolloid sealing strips. Operates via negative pressure wound therapy (NPWT) using external therapy units. Sterilized via gamma radiation (SAL 10^-6). Biocompatibility per ISO 10993-1. Packaging per ISO 11607-1.

## Regulatory Identification

A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.

## Predicate Devices

- Prevena™ Restor™ Incision Management System ([K181507](/device/K181507.md))

## Reference Devices

- Prevena Plus Incision Management System (Prevena Customizable Dressing components of system) ([K153199](/device/K153199.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 1, 2022

3M Health Care Business Group Teri Feeley Sr. Regulatory Affairs Associate 6203 Farinon Dr. San Antonio, Texas 78249

Re: K220660

Trade/Device Name: Prevena™ Restor™ Adapti-Form™ Dressing Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: OMP Dated: March 4, 2022 Received: March 7, 2022

Dear Teri Feeley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Julie Morabito, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K220660

Device Name Prevena™ Restor™ Adapti-Form™ Dressing

#### Indications for Use (Describe)

The Prevena™ Restor™ Adapti-Form™ Dressing is part of the Prevena™ Restor™ Incision Management System and is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudate via the application of negative pressure wound therapy.

| Type of Use (Select one or both, as applicable)                                                                         |                                                                                                         |
|-------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <div> <span> <input checked="true" type="checkbox"/>     Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> | <div> <span> <input type="checkbox"/>     Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div> |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the 3M logo. The logo is in red and features the number "3" followed by the letter "M". The font is bold and sans-serif. The logo is simple and recognizable.

510(k) Summary Prevena™ Restor™ Adapti-Form™ Dressing

3M Health Care Business Group 6203 Farinon San Antonio, TX 78249

Contact Person: Teri Feeley Email: tfeeley@mmm.com Phone: 210-459-1952 Facsimile: 210-255-6727 Date Prepared: 25 Feb 2022

Name of Subject Device: Prevena™ Restor™ Adapti-Form™ Dressing Predicate Device: Prevena™ Restor™ Incision Management System (K181507) Reference Device: Prevena Plus Incision Management System (Prevena Customizable Dressing components of system) (K153199) Common or Usual Name: Dressing component of Prevena Incision Management System Classification Name: Negative Pressure Wound Therapy Powered Suction Pump (and components) Requlation Number: 21 CFR 878.4780 Regulatory Class: Class II Product Code: OMP

## Device Description

The Prevena Restor™ Incision Management System consist of the following components for use together as a system:

- Prevena Plus Therapy Unit & canister
- Prevena Restor Adapti-Form Dressing: ●
	- Foam Dressing with a skin interface layer o
	- V.A.C. Drape o
	- o SensaT.R.A.C. Pad
	- Hydrocolloid Sealing strips o

The Prevena™ Restor™ Adapti-Form™ Dressing is designed to be compatible for use with previously cleared V.A.C. Negative Pressure Wound Therapy (NPWT) Units:

- · ActiV.A.C. Therapy Unit,
- · V.A.C. ULTA Therapy Unit, and
- · V.A.C. Rx4 Therapy Unit.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the 3M logo. The logo is red and features the number 3 followed by the letter M. The logo is simple and recognizable.

# 510(k) Summary Prevena™ Restor™ Adapti-Form™ Dressing

# Intended Use / Indications for Use

The Prevena Restor Adapti-Form Dressing is part of the Prevena Restor Incision Management System and is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudate via the application of negative pressure wound therapy.

## Summary of Technological Characteristics

The Prevena™ Restor™ Adapti-Form™ Dressing is a customizable dressing, designed for use over linear, non-linear, intersecting incisions. The dressings are for use over various anatomical locations as determined at the discretion of the healthcare provider. The dressing is intended to be used with the Prevena Plus™ Therapy Unit or compatible V.A.C. Therapy Unit to achieve its intended purpose as the Prevena Restor Incision Management system.

The subject device indications for use, technological characteristics and principles of operation are substantially equivalent to the predicate. The difference between the subject and predicate device is the subject device dressing components are provided separately to offer the ability to cut/customize the dressing to address application needs that may not be met with the predicate (Prevena Restor Dressing, K181507)).

A table comparing the key features of the subject and predicate devices is provided below.

| Characteristic                                  | Subject Device:        | Predicate Device: Prevena Restor Incision<br>Management System K181507<br>Reference Device: Prevena Plus Incision<br>Management system with Prevena<br>Customizable Dressing, K153199                                                                                                                     |
|-------------------------------------------------|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use                                    | Identical to predicate | The Prevena Restor™ Incision Management<br>System is intended to manage the<br>environment of surgical incisions that continue<br>to drain following sutured or stapled closure<br>by maintaining a closed environment and<br>removing exudate via the application of<br>negative pressure wound therapy. |
| Indicated Wound Types                           | Identical to predicate | • Surgically closed incisions                                                                                                                                                                                                                                                                             |
| V.A.C. Negative Pressure<br>Wound Therapy Units | Identical to predicate | • Prevena Plus Therapy Unit 7 day<br>• Prevena Plus Therapy Unit 14-day<br>• ACTIV.A.C.™ Therapy Unit*<br>• V.A.C.ULTA™ Therapy Unit*<br>• V.A.C.RX4™ Therapy Unit*                                                                                                                                       |

Summary of the technological characteristics of the device compared to the predicate device [21 CFR 807.92(a)(6)]

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the 3M logo. The logo is red and features the number "3" followed by the letter "M". The logo is simple and recognizable.

# 510(k) Summary Prevena™ Restor™ Adapti-Form™ Dressing

|                                                 |                               | *and associated canisters                                                                                                            |
|-------------------------------------------------|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Use environment/Care<br>Setting of dressing kit | Identical to predicate        | The Dressing will be initially applied in the<br>operating room/surgery center and then may<br>transition home with the patient      |
| Dressing Components                             | Identical to reference device | 1) Foam dressing with skin interface<br>layer<br>2) Hydrocolloid sealing strips<br>3) Adhesive drape<br>4) Tubing with interface pad |
| Sterilization Method &<br>SAL                   | Identical to predicate        | Gamma radiation<br>10-6                                                                                                              |

# Performance Data

Summary of non-clinical tests conducted for determination of substantial equivalence:

- Prevena Restor Incision Management System negative pressure test .
- Package Integrity/Stability testing was done in accordance with ISO 11607-1 ●
- Product performance stability testing of dressing components after sterilization ●
- . Human factors evaluation
- Biocompatibility testing to ISO 10993-1 .

In all instances, the Prevena™ Restor™ Adapti-Form™ Dressing which is a part of the Prevena™ Restor™ Incision Management System functioned as intended and all test results passed.

Clinical and Pre-clinical testing were not necessary to demonstrate substantial equivalence.

## Conclusions

The subject device's Intended Use, indications for use, fundamental technology and principles of operation are unchanged compared to the predicate and reference devices cleared under K181507 & K153199.

The performance data demonstrates that the Prevena Restor Adapti-Form Dressing is substantially equivalent to the predicate product in terms of safety and effectiveness.

---

**Source:** [https://fda-staging.innolitics.com/device/K220660](https://fda-staging.innolitics.com/device/K220660)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).