K220612 · Innovative Product Brands, Inc. · NHA · Jun 6, 2022 · Dental
Device Facts
Record ID
K220612
Device Name
PrimeLOC Attachment System
Applicant
Innovative Product Brands, Inc.
Product Code
NHA · Dental
Decision Date
Jun 6, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
The PrimeLOC Attachment System is designed to facilitate patient removal of a dental prosthesis for use with full arch overdentures or partial dentures retained in whole or in part by endosseous implants in the mandible or maxilla.
Device Story
PrimeLOC Attachment System consists of titanium alloy abutments, retention inserts, and denture housings; facilitates resilient attachment of dental prostheses to endosseous implants. Abutments feature coronal ridge retention design for interference (snap) fit with denture housing; threaded apical end connects to specific compatible implants. System accommodates path of insertion divergence up to 20° per implant and 40° between implants. Used by dental professionals to restore patient function; allows patient removal of prosthesis. Retention inserts fixed in metal housing embedded in overdenture; majority of loading supported by contact of denture with gingival tissue. Abutments provided with titanium nitride (TiN) coating. Includes instruments for seating/removing components and laboratory accessories for fixation.
Clinical Evidence
Bench testing only. Included sterilization validation (ISO 17665-1/2), biocompatibility (ISO 10993-5/12), mechanical fatigue testing (ISO 14801), and retention force testing over multiple insertion/removal cycles compared to predicate. MRI safety evaluated via scientific rationale and literature review.
Technological Characteristics
Materials: Ti 6Al-4V ELI (ASTM/ISO standards implied by biocompatibility testing). Coating: Titanium Nitride (TiN). Interface: Internal Hex. Dimensions: 3.5mm platform, cuff heights up to 6mm. Sterilization: Moist heat (ISO 17665). Connectivity: None (mechanical device).
Indications for Use
Indicated for patients requiring full arch overdentures or partial dentures retained by endosseous implants in the mandible or maxilla.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
LOCATOR Implant Anchor Attachment System (K083324)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The logos are displayed side-by-side.
Innovative Product Brands, Inc. % Melissa Burbage Senior Regulatory Specialist PaxMed International. LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K220612
Trade/Device Name: PrimeLOC Attachment System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: May 9, 2022 Received: May 9, 2022
Dear Melissa Burbage:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K220612
Device Name
PrimeLOC Attachment System
Indications for Use (Describe)
The PrimeLOC Attachment System is designed to facilitate patient removal of a dental prosthesis for use with full arch overdentures or partial dentures retained in whole or in part by endosseous implants in the mandible or maxilla.
| PrimeLOC Attachment Implant Compatibility | |
|----------------------------------------------------|----------------------------------------------------|
| Implant | Configurations |
| Zimmer Biomet Dental<br>Tapered Screw-Vent Implant | 3.5mm Platform (3.7mm and 4.1mm body<br>diameters) |
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
## K220612
## PrimeLOC Attachment System
## Innovative Product Brands, Inc.
June 6, 2022
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | Innovative Product Brands, Inc. |
|-------------------|---------------------------------|
| | 7045 Palm Avenue |
| | Highland, CA 92346 |
| | Telephone: +1 909-864-7477 |
| | Fax: +1 909-864-7232 |
| Official Contact | Chris Gervais, Chief Scientific Officer |
|---------------------------|----------------------------------------------------------------------|
| Representative/Consultant | Melissa Burbage |
| | Kevin A. Thomas, PhD; Floyd G. Larson, MS, MBA |
| | PaxMed International, LLC |
| | 12264 El Camino Real, Suite 400 |
| | San Diego, CA 92130 |
| | Telephone: +1 858-792-1235 |
| | Fax: +1 858-792-1236 |
| | Email: mburbage@paxmed.com<br>kthomas@paxmed.com; flarson@paxmed.com |
#### DEVICE NAME AND CLASSIFICATION
| Trade/Device Name | PrimeLOC Attachment System |
|----------------------|------------------------------------|
| Common Name | Dental implant abutment |
| Classification Name | Endosseous dental implant abutment |
| Regulation Number | 21 CFR 872.3630 |
| Regulation Name | Endosseous dental implant abutment |
| Regulatory Class | Class II |
| Product Code | NHA |
| Classification Panel | Dental Products Panel |
| Reviewing Branch | Dental Devices Branch |
## PREDICATE DEVICE INFORMATION
Primary Predicate Device K083324, LOCATOR Implant Anchor Attachment System, Zest Anchors, Inc.
Reference Device K133339, Tapered Screw Vent Implant, Zimmer Dental Inc.
{4}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT
The PrimeLOC Attachment system is designed to facilitate patient removal of a dental prosthesis for use with full arch overdentures or partial dentures retained in whole or in part by endosseous implants in the mandible or maxilla.
| PrimeLOC Attachment Implant Compatibility | |
|-----------------------------------------------------|----------------------------------------------------|
| Implant | Configurations |
| Zimmer Biomet Dental Tapered Screw-<br>Vent Implant | 3.5mm Platform (3.7mm and 4.1mm body<br>diameters) |
#### SUBJECT DEVICE DESCRIPTION
The PrimeLOC Attachment System consists of abutments and device-specific accessories (retention inserts and denture housings) to serve in a similar function to LOCATOR® abutments for resilient attachment of prostheses to endosseous dental implants. All PrimeLOC abutments are made of titanium alloy and have the same coronal ridge retention design that attaches to the overdenture component by an interference (snap) fit. The threaded apical end of the abutment connects to the implant and is specific to each compatible implant system and diameter. The PrimeLOC Attachment System is designed to accommodate a path of insertion divergence of up to 20° per implant and no more than 40° of divergence between implants. PrimeLOC abutments are provided with a titanium nitride (TiN) coating and are available in cuff heights up to 6 mm, with an implant interface diameter of 3.5 mm.
Retention inserts are fixed within a metal denture housing which is embedded in an overdenture prosthesis. The retention inserts allow for varying levels of retention. This connection allows the denture to be retained on the abutments while the majority of loading is supported by the contact of the denture with the gingival tissue surrounding the mandibular and maxillary ridges.
The PrimeLOC Attachment System are compatible with OEM implants, specifically Tapered Screw-Vent 3.5 mm platform dental implants manufactured by Zimmer Dental, Inc. Compatibility is based on agreements with the OEM to ensure that the abutments are designed to fit the corresponding implants. All interfaces of the subject device have been designed and verified by the OEM implant manufacturer. This ensures that the interface of the subject device with the identified compatible implants is appropriate.
A thin titanium nitride coating is applied to the PrimeLOC abutments in the area that contacts gingival tissue and in the area that contacts the retention inserts. The denture attachment housing, which sits within the denture, may be colored through a titanium anodization process.
This system includes instruments used for seating and removing abutments and retention inserts, and processing inserts, impression copings, and laboratory components used to facilitate the fixation of the retention insert and denture housing in the appropriate position within the denture. These components are endosseous dental implant accessories considered Class I 510(k)-exempt devices as defined in 21 CFR 872.3980.
{5}------------------------------------------------
### PERFORMANCE DATA
Non-clinical testing data submitted to demonstrate substantial equivalence included:
- sterilization validation according to ISO 17665-1 Sterilization of health care products -. Moist heat – Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices and ISO 17665-2 Sterilization of health care products - Moist heat - Part 2: Guidance on the application of ANSI/AAMI/ISO 17665-1
- . biocompatibility testing according to ISO 10993-5 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity and ISO 10993-12 Biological evaluation of medical devices – Part 12: Sample preparation and reference materials
- mechanical testing according to ISO 14801 Dentistry Implants Dynamic loading ● test for endosseous dental implants
- testing of each different type of retention insert in a denture housing to measure removal ● force from an abutment over multiple sets of insertion/removal cycles compared to the predicate device
- . non-clinical worst-case MRI review to evaluate the metallic devices in the MRI environment using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan. "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795), based on the subject device components and material composition. Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque
The subject abutment-implant interface of the PrimeLOC Attachment System is verified to be compatible through either OEM signed abutment drawings, OEM implant drawings, and/or established compatibility demonstrated through mutual agreements with the OEM. Once the specifications and tolerances have been identified, the subject device have been designed and verified by the OEM implant manufacturer. This ensures that the interface of the subject device with the identified compatible implants is appropriate.
#### EQUIVALENCE TO MARKETED DEVICES
The indications for use statement of the subject device is similar to that of the primary predicate device. The subject device has the added text, "designed to facilitate patient removal of a dental prosthesis for use with full arch overdenture ... , " whereas the primary predicate states, "appropriate for use with overdentures…." This added text only provides more information on how the device is used and does not affect the intended use of the device.
The subject device abutment diameter and abutment cuff height are similar to those of the predicate in that they are within the range of the predicate device. The subject device connection is similar to the internal hex connection of the predicate device, but the subject device is not offered with the additional connections offered by the predicate device. Any differences in technological characteristics were supported by fatigue testing and retention testing to demonstrate substantial equivalence.
Reference Device K133339 is included for the compatible implant body.
{6}------------------------------------------------
| Comparison | Subject Device | Primary Predicate Device | Conclusion |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| | PrimeLOC Attachment<br>Innovative Product Brands, Inc. | K083324<br>LOCATOR Implant Attachment<br>System<br>Zest Anchors, Inc. | |
| Indications | The PrimeLOC Attachment system is<br>designed to facilitate patient removal<br>of a dental prosthesis for use with full<br>arch overdentures or partial dentures<br>retained in whole or in part by<br>endosseous implants in the mandible<br>or maxilla. | The LOCATOR Implant Anchor<br>Abutment for Endosseous Dental<br>Implants is appropriate for use with<br>overdentures or partial dentures<br>retained in whole or in part by<br>endosseous implants in the mandible<br>or maxilla. | Similar |
| Design | | | |
| Abutment Diameter,<br>mm | 3.5 | 3.25 to 6.5 | Similar |
| Abutment Cuff<br>Height, mm | 1 to 6 | 0 to 8 | Similar |
| Abutment<br>Angulation | Straight | Straight | Same |
| Abutment/Implant<br>Interface | Internal Hex | Conical, External Hex, Internal Hex,<br>Internal Multi Lobe | Similar |
| Divergence<br>Allowance | 20°/40° | 20°/40° | Same |
| Prosthesis<br>Attachment Type | Nylon Insert retained in Denture<br>Attachment Housing | Nylon Insert retained in Denture<br>Attachment Housing | Same |
| Material | | | |
| Abutment | Ti 6Al-4V ELI | Ti 6Al-4V ELI | Same |
| Abutment Coronal<br>Surface Coating | Titanium Nitride | Titanium Nitride | Same |
| Prosthetic Retention<br>Component | Nylon | Nylon | Same |
## CONCLUSION
The subject devices and the primary predicate device have the same intended use and similar technological characteristics. They encompass a similar range of physical dimensions, are manufactured from the same materials, and are to be sterilized using identical methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.