Powder-Free Nitrile Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl (White, Blue, Black, Orange and Green Color)
K220609 · Encompass Industries Sdn. Bhd. · LZA · May 27, 2022 · General Hospital
Device Facts
Record ID
K220609
Device Name
Powder-Free Nitrile Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl (White, Blue, Black, Orange and Green Color)
Applicant
Encompass Industries Sdn. Bhd.
Product Code
LZA · General Hospital
Decision Date
May 27, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, the glove was tested for chemotherapy drugs permeation.
Device Story
Disposable, non-sterile, powder-free nitrile examination glove; available in white, blue, black, orange, and green. Used by healthcare personnel to prevent cross-contamination between patient and examiner. Tested for permeation resistance against specific chemotherapy drugs and Fentanyl Citrate per ASTM D6978-05. Provides physical barrier protection. Not made with natural rubber latex. Used in clinical settings; operator is healthcare professional or patient (OTC). Output is physical protection; healthcare provider visually inspects glove integrity before use. Benefits include reduced risk of disease transmission and protection from hazardous drug exposure.
Clinical Evidence
No clinical data. Bench testing only, including dimension, freedom from holes (ASTM D5151-19), physical properties (ASTM D412, D573), residual powder (ASTM D6124-06), and chemotherapy drug permeation (ASTM D6978-05). Biocompatibility testing performed per ISO 10993-5, 10, and 11.
Technological Characteristics
Material: Nitrile. Form: Powder-free, non-sterile, disposable. Standards: ASTM D6319-19 (dimensions, physical properties, freedom from holes), ASTM D6978-05 (chemotherapy permeation). Dimensions: 230mm+ length, 0.05mm+ thickness. Biocompatibility: ISO 10993-5, 10, 11. No software or electronic components.
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hand to prevent contamination between patient and examiner, and to provide protection against exposure to chemotherapy drugs and Fentanyl. Contraindicated for use with Carmustine and ThioTepa.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Black Colored, Powder Free Nitrile Examination Gloves, Non-sterile, Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (K200181)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol is a stylized representation of a caduceus, a traditional symbol of medicine, with intertwined snakes and wings. To the right of the symbol is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue, with "FDA" in a larger font size than the rest of the text.
May 27, 2022
Encompass Industries Sdn. Bhd. Nur Muhyiddin OA Manager Lot 18256, Kawasan Perindustrian Lot Q Kertih BioPolymer Park Kemaman, Terengganu 24300 Malaysia
Re: K220609
Trade/Device Name: Powder-Free Nitrile Examination Glove (White, Blue, Black, Orange and Green Color) Tested for Use with Chemotherapy Drugs and Fentanyl Regulation Number: 21 CFR 880.6250 Regulation Name: Non-powdered patient examination glove Regulatory Class: Class I, reserved Product Code: LZA, LZC Dated: February 24, 2022 Received: March 2, 2022
Dear Nur Muhyiddin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Bifeng Qian, M.D., Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K220609
#### Device Name
Powder-Free Nitrile Examination Glove (White, Blue, Black, Orange and Green Color) Tested for Use with Chemotherapy Drugs and Fentanyl
#### Indications for Use (Describe)
A powder-free patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, the glove was tested for chemotherapy drugs permeation.
The following chemotherapy drugs have been tested with the glove:
| Chemotherapy Drugs | Concentration | Breakthrough Detection Time in Minutes | | | | |
|---------------------|---------------|----------------------------------------|------|-------|--------|-------|
| | | White | Blue | Black | Orange | Green |
| Carmustine | 3.3 mg/ml | 11.2 | 10.7 | 11.0 | 67.3 | 67.1 |
| Cisplatin | 1.0 mg/ml | >240 | >240 | >240 | >240 | >240 |
| Cyclophosphamide | 20.0 mg/ml | >240 | >240 | >240 | >240 | >240 |
| Dacarbazine | 10.0 mg/ml | >240 | >240 | >240 | >240 | >240 |
| Doxorubicin HCL | 2.0 mg/ml | >240 | >240 | >240 | >240 | >240 |
| Etoposide | 20.0 mg/ml | >240 | >240 | >240 | >240 | >240 |
| Fluorouracil | 50.0 mg/ml | >240 | >240 | >240 | >240 | >240 |
| Methotrexate | 25.0 mg/ml | >240 | >240 | >240 | >240 | >240 |
| Paclitaxel | 6.0 mg/ml | >240 | >240 | >240 | >240 | >240 |
| ThioTepa | 10.0 mg/ml | 25.2 | 21.5 | 11.1 | 47.8 | 169.1 |
| Mitomycin C | 0.5 mg/ml | >240 | >240 | >240 | >240 | >240 |
| Vincristine Sulfate | 1.0 mg/ml | >240 | >240 | >240 | >240 | >240 |
| Fentanyl Citrate | 100 mcg/2mL | >240 | >240 | >240 | >240 | >240 |
*Please note that the following 2 drugs have extremely low permeation times:
- 1. Carmustine
- 2. Thiotepa
Warning: Do not use with Carmustine and ThioTepa.
Type of Use (Select one or both, as applicable)
] Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
# K220609
Preparation date: May 27, 2022
| 1. Submitter | | |
|----------------|---|-------------------------------------------------------------------------------------------------------------|
| Name | : | Encompass Industries Sdn. Bhd. |
| Address | : | Lot 18256, Kawasan Perindustrian Lot Q,<br>Kertih Bio-Polymer Park, 24300 Kemaman,<br>Terengganu, Malaysia. |
| Telephone No. | : | +609 831 8866 |
| Contact Person | : | Nur Atikah binti Muhammad Muhyiddin |
| E-mail | : | regulatory@encompass-medical.com atikah@ems-inc.com |
## 2. Identification of Device
| Trade/Proprietary Name(s) | : 1. Powder-Free Nitrile Examination Glove (White, Blue, Black, Orange and Green Color) Tested for use with Chemotherapy Drugs and Fentanyl<br>2. Other clients' trade name and private labeling |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name(s) | : Powder-Free Nitrile Examination Glove |
| Classification Name: | : 1. Non-powdered Patient Examination Glove (21 CFR 880.6250)<br>2. Patient Examination Glove, Specialty (Product Code: LZC)<br>3. Polymer Patient Examination Glove (Product Code: LZA) |
| Device Classification | : Class I |
# 3. Identification of Legally Marketed Device as Predicate
| Predicate name | : Black Colored, Powder Free Nitrile Examination Gloves, Non-sterile,<br>Tested for Use with Chemotherapy Drugs and Fentanyl Citrate |
|----------------|--------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number | : K200181 |
| Company | : Comfort Rubber Glove Industries Sdn Bhd |
## 4. Description of Device
The proposed Powder-Free Nitrile Examination Glove (White, Blue, Black, Orange and Green Color) Tested for use with Chemotherapy Drugs and Fentanyl is a disposable device intended for over-the-counter use and is provided powder-free and non-sterile. These patient examination gloves are formulated using Nitrile. The patient examination gloves are not made with natural rubber latex. General specifications of the glove are as below:
| 1. Overall Length | : | 230 mm minimum |
|---------------------|---|--------------------------------------|
| 2. Width | : | 95 ± 5 mm minimum (for medium glove) |
| 3. Palm Thickness | : | 0.05 mm minimum |
| 4. Finger Thickness | : | 0.05 mm minimum |
| 5. Tensile Strength | | |
| a. Before Aging | : | 14 MPa minimum |
{5}------------------------------------------------
| b. After Aging | : | 14 MPa minimum |
|------------------------|---|----------------|
| 6. Ultimate Elongation | | |
| a. Before Aging | : | 500 % minimum |
| b. After Aging | : | 400 % minimum |
| 7. Pinhole AQL | : | 2.5 |
## 5. Intended Use / Indication for Use:
A Powder-Free Nitrile Examination Glove (White, Blue, Black, Orange and Green Color) Tested for use with Chemotherapy Drugs and Fentanyl is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. In addition, these gloves were tested for chemotherapy drugs permeation test.
Table 5.1 Summary Test Results for Resistance of Powder-Free Nitrile Examination Gloves (White, Blue, Black, Orange and Green Color) to Permeation by Chemotherapy Drugs
| Chemotherapy Drug | Concentration | Minimum Breakthrough Detection Time (minutes) | | | | |
|---------------------------------------------------------------------------|---------------|-----------------------------------------------|------|-------|--------|-------|
| | | White | Blue | Black | Orange | Green |
| Carmustine (BCNU)* | 3.3 mg/ml | 11.2 | 10.7 | 11.0 | 67.3 | 67.1 |
| Cisplatin | 1.0 mg/mL | >240 | >240 | >240 | >240 | >240 |
| Cyclophosphamide | 20.0 mg/mL | >240 | >240 | >240 | >240 | >240 |
| Dacarbazine | 10.0 mg/mL | >240 | >240 | >240 | >240 | >240 |
| Doxorubicin HCL | 2.0 mg/mL | >240 | >240 | >240 | >240 | >240 |
| Etoposide | 20.0 mg/mL | >240 | >240 | >240 | >240 | >240 |
| Fluorouracil | 50.0 mg/mL | >240 | >240 | >240 | >240 | >240 |
| Methotrexate | 25.0 mg/mL | >240 | >240 | >240 | >240 | >240 |
| Paclitaxel | 6.0 mg/mL | >240 | >240 | >240 | >240 | >240 |
| ThioTepa* | 10.0 mg/mL | 25.2 | 21.5 | 11.1 | 47.8 | 169.1 |
| Mitomycin C | 0.5 mg/mL | >240 | >240 | >240 | >240 | >240 |
| Vincristine Sulfate | 1.0 mg/mL | >240 | >240 | >240 | >240 | >240 |
| Fentanyl Citrate | 100.0 mcg/2mL | >240 | >240 | >240 | >240 | >240 |
| *Please note that the following drugs have extremely low permeation time: | | | | | | |
| 1. Carmustine (3.3 mg/mL) | | | | | | |
2. Thiotepa (10 mg/mL)
Warning: Do not use with Carmustine and ThioTepa.
# 6. Comparative Technological Characteristics & Performance Information Summary
Table 6.1 Summary of Technological Characteristics Comparison between Proposed Device and Predicate Device
| | | Device Performance / Characteristic | | Comparison Analysis | |
|---------------------------------------------|-------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|---------------------------------------------|
| Character-<br>istics | Requirement /<br>Standard | Predicate Device<br>(K200181) | Subject Device | Comparison | Any<br>Safety &<br>Effective-<br>ness Issue |
| Material | - | Nitrile | Nitrile | Similar | No |
| Color | - | Black | White<br>Blue<br>Black<br>Orange<br>Green | Minor<br>Different<br>(Supported by<br>Biocompatibility Study) | No |
| Product<br>Code | | LZA<br>LZC | LZA<br>LZC | Similar | No |
| Type of<br>Use | - | Single Use | Single Use | Similar | No |
| Material | - | Nitrile | Nitrile | Similar | No |
| Powder-<br>Free | - | Yes | Yes | Similar | No |
| Sterility | - | Non-sterile | Non-Sterile | Similar | No |
| Intended<br>Use /<br>Indications<br>for Use | - | The Black Colored,<br>Powder Free Nitrile<br>Examination Gloves,<br>Non-Sterile, and Tested<br>for Use with<br>Chemotherapy Drugs and<br>Fentanyl Citrate is a<br>specialty medical glove<br>which is a disposable<br>device intended for<br>medical purposes that is<br>worn on the examiner's<br>hand or finger to prevent<br>contamination between<br>examiner and patient. In<br>addition, these gloves are<br>worn to protect the<br>wearer against exposure<br>to chemotherapy drugs<br>and Fentanyl Citrate. | Powder-Free Nitrile<br>Examination Glove (White,<br>Blue, Black, Orange and<br>Green Color) Tested for<br>Use with Chemotherapy<br>Drugs and Fentanyl is a<br>disposable device intended<br>for medical purposes that is<br>worn on the examiner's<br>hand, in order to prevent<br>contamination between<br>patient and examiner, and<br>also prevent transmission of<br>a wide variety of diseases to<br>both patient and health care<br>personnel. In addition,<br>these gloves were tested for<br>chemotherapy drug<br>permeation test. | Similar | No |
| Dimension | ASTM D6319-10 | Meets requirement | Meets requirement | Similar | No |
| Physical<br>Properties | ASTM D6319-10 | Meets requirement | Meets requirement | Similar | No |
| Freedom<br>from Holes | ASTM D6319-10<br>(Tested according<br>to ASTM D5151-<br>06) | Meets requirements of<br>ASTM D6319-10<br>(ASTM D5151) | Meets ASTM D6319-10,<br>21 CFR 800.20 (Tested in<br>accordance with ASTM<br>D5151-06) | Similar | No |
| Powder<br>Residual | ASTM D6124-06 | Meets requirement | Meets requirement | Similar | No |
| Biocompati<br>bility | .In Vitro<br>Cytotoxicity-<br>ISO 10993-5 | Under the conditions of<br>the study, the device is<br>non-cytotoxic | Under the condition of<br>study, the device extract<br>shows potential toxicity. | Different<br>(Supported by<br>Acute<br>Systemic<br>Toxicity<br>Study) | No |
| | Acute Systemic<br>Toxicity - ISO<br>10993-11 | - | Not an acute systemic toxic<br>under the condition of the<br>study. | Different<br>(Device is<br>proven safe to<br>used) | No |
| | Skin Sensitization<br>- ISO 10993-10 | Under the condition of<br>the study, the device is<br>non-sensitization | Not a sensitizer under the<br>conditions of the study. | Similar | No |
| | Skin Irritation -<br>ISO 10993-10 | Under the conditions of<br>the study, the device is<br>non irritating | Not a primary skin irritant<br>under the conditions of the<br>study. | Similar | No |
{6}------------------------------------------------
{7}------------------------------------------------
| Characteristic | Predicate Device<br>(K200181) | Proposed Device | Comparison Analysis | Justification |
|--------------------------------------------------|----------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Color | Black | White<br>Blue<br>Black<br>Orange<br>Green | Minor Different<br>(Supported by<br>Biocompatibility<br>Study) | Device in different color has<br>the same non-sensitization,<br>non-irritating and non-toxic<br>characteristics. |
| Biocompatibility<br>(In Vitro<br>Cytotoxicity) | Meets ISO 10993:5<br>Under the<br>conditions of the<br>study, the device is<br>non-cytotoxic | ISO 10993:5<br>Under the condition of<br>the study, the<br>device extract shows potential<br>cytotoxicity. | Different<br>(Supported by<br>Acute Systemic<br>Toxicity Study) | Device has potential cytotoxic<br>effect. Therefore, Acute<br>Systemic Toxicity test is<br>conducted to prove that<br>device does not have potential<br>toxicity. |
| Biocompatibility<br>(Acute Systemic<br>Toxicity) | - | Meets ISO 10993-11<br>Not an acute systemic<br>toxic under the<br>condition of the study. | Different<br>(Device is proven<br>safe to use) | Device is does not have<br>potential toxicity effect.<br>Therefore, similar with the<br>predicate. |
Table 6.2 Summary of Difference in Technological Characteristics
## Table 6.3: Test Results for Resistance of Medical Gloves to Permeation by Chemotherapy Drugs (ASTM D6978-05)
| Chemotherapy<br>Drug | Concentration | Minimum Breakthrough Detection Time (minutes) | | | | | |
|----------------------|---------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|------|-------|--------|-------|
| | | Predicate Device<br>(K200181) | Proposed Device | | | | |
| | | | White | Blue | Black | Orange | Green |
| Carmustine (BCNU)* | 3.3 mg/ml | 54.1 | 11.2 | 10.7 | 11.0 | 67.3 | 67.1 |
| Cisplatin | 1.0 mg/mL | >240 | >240 | >240 | >240 | >240 | >240 |
| Cyclophosphamide | 20.0 mg/mL | >240 | >240 | >240 | >240 | >240 | >240 |
| Dacarbazine | 10.0 mg/mL | >240 | >240 | >240 | >240 | >240 | >240 |
| Doxorubicin HCL | 2.0 mg/mL | >240 | >240 | >240 | >240 | >240 | >240 |
| Etoposide | 20.0 mg/mL | >240 | >240 | >240 | >240 | >240 | >240 |
| Fluorouracil | 50.0 mg/mL | >240 | >240 | >240 | >240 | >240 | >240 |
| Methotrexate | 25.0 mg/mL | - | >240 | >240 | >240 | >240 | >240 |
| Paclitaxel | 6.0 mg/mL | >240 | >240 | >240 | >240 | >240 | >240 |
| ThioTepa* | 10.0 mg/mL | 1.3 | 25.2 | 21.5 | 11.1 | 47.8 | 169.1 |
| Mitomycin C | 0.5 mg/mL | - | >240 | >240 | >240 | >240 | >240 |
| Vincristine Sulfate | 1.0 mg/mL | - | >240 | >240 | >240 | >240 | >240 |
| Fentanyl Citrate | 100.0 mcg/2mL | >240 | >240 | >240 | >240 | >240 | >240 |
| *Warning Statement: | | The maximum testing time is 240 minutes. Please note that the following drugs have extremely low permeation time:<br>1. Carmustine (3.3 mg/mL)<br>2. Thiotepa (10 mg/mL) | | | | | |
{8}------------------------------------------------
# 7. Summary of Non-Clinical Test:
Below are the non-clinical tests that was conducted and the purposes:
| Test | Purpose |
|-------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| Dimension Test | To evaluate whether the device meets the current ASTM D6319-19 |
| Freedom From Hole | To evaluate whether the subject meets current 21 CFR 800.20, ASTM<br>D6319-19 and test according to ASTM D5151-19 |
| Physical Property Test | To evaluate whether the subjects meet current ASTM D6319-19 and test<br>according to ASTM D412 and ASTM D573 |
| Residual Powder Test | To evaluate whether the subjects meet current ASTM D6319-19 and test<br>according to ASTM D6124-06 |
| Biocompatibility Test<br>(Skin Irritation) | To determine whether skin irritation potential of device meets ISO 10993-<br>10. |
| Biocompatibility Test<br>(Skin Sensitization) | To determine whether skin sensitization potential of device meets ISO<br>10993-10. |
| Biocompatibility Test<br>(Acute Systemic<br>Toxicity) | To determine whether leachable extracted from the device would cause<br>acute systemic toxicity and meet ISO 10993-11. |
Table 7.1 Summary of Dimension Test
| | Standard | | Minimum | Device Performance (Minimum) | | | | |
|-----------------------|----------------------|------|-----------------|------------------------------|---------|---------|---------|---------|
| Characteristics | | Size | Requirement | White | Blue | Black | Orange | Green |
| Length | ASTM<br>D6319-<br>19 | XS | 220 mm | 239 mm | 235 mm | 242 mm | 243 mm | 244 mm |
| | | S | 220 mm | 236 mm | 237 mm | 242 mm | 244 mm | 241 mm |
| | | M | 230 mm | 230 mm | 230 mm | 240 mm | 245 mm | 245 mm |
| | | L | 230 mm | 234 mm | 237 mm | 255 mm | 243 mm | 242 mm |
| | | XL | 230 mm | 232 mm | 246 mm | 240 mm | 240 mm | 245 mm |
| | | XXL | 230 mm | 238 mm | 238 mm | 242 mm | 240 mm | 245 mm |
| Width | ASTM<br>D6319-<br>19 | XS | $70 \pm 10$ mm | 75 mm | 77 mm | 77 mm | 77 mm | 77 mm |
| | | S | $80 \pm 10$ mm | 85 mm | 87 mm | 86 mm | 87 mm | 85 mm |
| | | M | $95 \pm 10$ mm | 93 mm | 86 mm | 92 mm | 97 mm | 97 mm |
| | | L | $110 \pm 10$ mm | 104 mm | 107 mm | 104 mm | 108 mm | 107 mm |
| | | XL | $120 \pm 10$ mm | 112 mm | 111 mm | 114 mm | 118 mm | 117 mm |
| | | XXL | $130 \pm 10$ mm | 128 mm | 128 mm | 127 mm | 124 mm | 126 mm |
| Thickness<br>(Finger) | ASTM<br>D6319-<br>19 | XS | 0.05 mm | 0.08 mm | 0.08 mm | 0.09 mm | 0.27 mm | 0.25 mm |
| | | S | | 0.08 mm | 0.09 mm | 0.05 mm | 0.27 mm | 0.27 mm |
| | | M | | 0.10 mm | 0.11 mm | 0.10 mm | 0.28 mm | 0.27 mm |
| | | L | | 0.09 mm | 0.08 mm | 0.10 mm | 0.27 mm | 0.27 mm |
| | | XL | | 0.07 mm | 0.06 mm | 0.10 mm | 0.28 mm | 0.27 mm |
| | | XXL | | 0.07 mm | 0.08 mm | 0.07 mm | 0.28 mm | 0.26 mm |
| Thickness<br>(Palm) | ASTM<br>D6319-<br>19 | XS | 0.05 mm | 0.05 mm | 0.05 mm | 0.06 mm | 0.19 mm | 0.19 mm |
| | | S | | 0.05 mm | 0.05 mm | 0.06 mm | 0.19 mm | 0.19 mm |
| | | M | | 0.06 mm | 0.06 mm | 0.05 mm | 0.19 mm | 0.18 mm |
| | | L | | 0.05 mm | 0.05 mm | 0.06 mm | 0.22 mm | 0.20 mm |
| | | XL | | 0.05 mm | 0.05 mm | 0.06 mm | 0.22 mm | 0.18 mm |
| | | XXL | | 0.05 mm | 0.05 mm | 0.06 mm | 0.20 mm | 0.19 mm |
{9}------------------------------------------------
| Characteristics | Standard | Size | Minimum<br>Requirement | Device Performance (Minimum) | | | | |
|-----------------------|-------------------------------------------------------------------------------------|------|------------------------|------------------------------|--------|--------|--------|--------|
| | | | | White | Blue | Black | Orange | Green |
| Freedom from<br>Holes | A;STM<br>D6319-19<br>21 CFR<br>800.20<br>(Tested<br>as per<br>ASTM<br>D5151-<br>19) | XS | AQL 2.5 | Passed | Passed | Passed | Passed | Passed |
| | | S | | Passed | Passed | Passed | Passed | Passed |
| | | M | | Passed | Passed | Passed | Passed | Passed |
| | | L | | Passed | Passed | Passed | Passed | Passed |
| | | XL | | Passed | Passed | Passed | Passed | Passed |
| | | XXL | | Passed | Passed | Passed | Passed | Passed |
Table 7.2 Summary of Freedom of Hole Test
Table 7.3 Summary of Physical Property Test
| Characteristics | Standard | Size | Minimum<br>Requirement | Device Performance (Minimum) | | | | |
|-------------------------------------|---------------------------------------|------|------------------------|------------------------------|------|-------|--------|-------|
| | | | | White | Blue | Black | Orange | Green |
| Before Aging<br>(Tensile [MPa]) | | XS | 14 MPa | 23.2 | 21.0 | 21.3 | 20.7 | 21.1 |
| | | S | | 21.2 | 24.5 | 23.6 | 20.0 | 20.1 |
| | | M | | 22.8 | 21.6 | 22.3 | 21.0 | 28.8 |
| | | L | | 20.7 | 23.7 | 21.6 | 21.4 | 20.9 |
| | | XL | | 20.5 | 20.2 | 24.3 | 20.9 | 22.0 |
| | | XXL | | 22.8 | 23.6 | 22.2 | 23.0 | 22.1 |
| Before Aging<br>(Elongation<br>[%]) | ASTM<br>D6319-<br>19 | XS | 500 % | 528 | 520 | 515 | 527 | 533 |
| | | S | | 527 | 549 | 518 | 515 | 556 |
| | | M | | 520 | 522 | 530 | 549 | 548 |
| | | L | | 528 | 521 | 527 | 531 | 523 |
| | | XL | | 536 | 521 | 528 | 533 | 524 |
| | | XXL | | 526 | 524 | 528 | 525 | 525 |
| After Aging<br>(Tensile [MPa]) | (Tested<br>as per<br>ASTM<br>D412-16) | XS | 14 MPa | 23.0 | 20.2 | 23.0 | 24.5 | 21.4 |
| | | S | | 20.2 | 24.6 | 26.6 | 22.7 | 20.4 |
| | | M | | 26.8 | 19.8 | 22.7 | 22.2 | 26.4 |
| | | L | | 24.9 | 26.1 | 21.4 | 25.2 | 21.7 |
| | | XL | | 26.3 | 23.1 | 28.8 | 22.2 | 24.6 |
| | | XXL | | 23.3 | 21.0 | 20.0 | 27.4 | 20.1 |
| After Aging<br>(Elongation<br>[%]) | | XS | 400 % | 432 | 461 | 414 | 430 | 465 |
| | | S | | 427 | 466 | 421 | 412 | 446 |
| | | M | | 444 | 457 | 425 | 466 | 440 |
| | | L | | 488 | 423 | 436 | 447 | 455 |
| | | XL | | 458 | 424 | 434 | 432 | 431 |
| | | XXL | | 437 | 435 | 433 | 425 | 425 |
{10}------------------------------------------------
| Characteristics | Standard | Size | Minimum<br>Requirement | Device Performance (Minimum) | | | | |
|----------------------------|--------------------------------------------------------------------|------|------------------------------------------------|------------------------------|------|-------|--------|-------|
| | | | | White | Blue | Black | Orange | Green |
| Residual<br>Powder<br>(mg) | ASTM<br>D6319-<br>19<br>(Tested<br>as per<br>ASTM<br>D6124-<br>06) | XS | ≤ 2.0 mg of<br>residual<br>powder per<br>glove | 0.98 | 0.68 | 0.80 | 0.82 | 0.96 |
| | | S | | 0.86 | 0.72 | 0.86 | 0.98 | 0.82 |
| | | M | | 0.92 | 0.88 | 0.94 | 0.92 | 0.92 |
| | | L | | 0.78 | 0.76 | 0.92 | 0.78 | 0.90 |
| | | XL | | 0.84 | 0.98 | 0.84 | 0.68 | 0.86 |
| | | XXL | | 0.74 | 0.82 | 0.92 | 0.84 | 0.98 |
Table 7.4 Summary of Residual Powder Test
Table 7.5 Summary of Biocompatibility Test
| Characteristics | Standard | Minimum<br>Requirement | Device Performance | | | | |
|----------------------------|--------------|------------------------|------------------------------------------------------------------|------|-------|--------|-------|
| | | | White | Blue | Black | Orange | Green |
| Skin Irritation | ISO 10993-10 | N/A | Non-irritant under the conditions of the study. | | | | |
| Skin<br>Sensitization | ISO 10993-10 | N/A | Non-sensitizer under the conditions of the study. | | | | |
| Acute Systemic<br>Toxicity | ISO 10993-11 | N/A | Not an acute systemic toxic under the condition of<br>the study. | | | | |
# 8. Summary of Clinical Test:
Clinical data is not required.
# 9. Conclusion:
The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.