Venus Bulk Flow ONE is used for extended fissure sealing, first-layer lining for class I and class II cavities, Class I, II, III and V direct restorations, repair of direct and indirect restorations in combination with a suitable adhesive, splinting loosened teeth resulting from trauma or periodontal associated events.
Device Story
Venus Bulk Flow ONE is a light-curing, radio-opaque nano-hybrid dental composite. Applied by dentists in clinical settings to tooth structures or restorations; cured using dental light. Used for fissure sealing, cavity lining, direct restorations, repair of restorations, and tooth splinting. Provides restorative material for dental patients. Bulk-fill capability allows 4mm curing depth, simplifying and accelerating restoration procedures compared to traditional flowable composites. Clinical benefit achieved through structural restoration of tooth integrity.
Clinical Evidence
No clinical trial performed. Evidence based on non-clinical bench testing per ISO 4049:2019, including light sensitivity, depth of cure, flexural strength, water sorption/solubility, color stability, and radio-opacity. Results demonstrate equivalence or superiority to predicate. Clinical evaluation of predicate device remains valid for subject device.
Technological Characteristics
Nano-hybrid composite; light-cured; radio-opaque. Dimensions: 2g syringe. Curing depth: 4mm. Tested per ISO 4049:2019. Identical chemistry to predicate; modification of opacity/color to achieve increased curing depth.
Indications for Use
Indicated for dental patients requiring extended fissure sealing, cavity lining (Class I/II), direct restorations (Class I, II, III, V), repair of direct/indirect restorations, and splinting of teeth loosened by trauma or periodontal events.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 21, 2022
Kulzer, LLC Marc Henn Director QA/RA 4315 S. Lafayette Blvd South Bend, Indiana 46614
Re: K220605
Trade/Device Name: Venus Bulk Flow ONE Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF Dated: February 22, 2022 Received: March 2, 2022
Dear Marc Henn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220605
Device Name Venus Bulk Flow ONE
Indications for Use (Describe) Extended fissure sealing First-layer lining for class I and class II cavities Class I, II, III and V direct restorations Repair of direct and indirect restorations in combination with a suitable adhesive Splinting loosened teeth resulting from trauma or periodontal associated events
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Kulzer Mitsui Chemicals Group. On the left is a green circle with a white tooth in the center. To the right of the circle is the word "KULZER" in gray, with the words "MITSUI CHEMICALS GROUP" in a smaller font below it.
SUBJECT: Venus Bulk Flow ONE Special 510(k) Summary- K220605
Date: February 15, 2022
Owner: Kulzer, LLC 4315 S. Lafayette Blvd. South Bend, IN 46614
PRIMARY CORRESPONDENT: Marc Henn, Director of QA/RA Phone (574) 261-2555 Email Marc.Henn(@kulzer-dental.com
ADDITIONAL CORRESPONDENT: David Vincent, Director of QA/RA Phone (574) 298-5424 Email David.Vincent(@kulzer-dental.com
Trade Name (80792): Venus Bulk Flow ONE Common Name: Restorative Material Product Code: 76 E BF Classification Name (21 CFR 872.3690): Material, Tooth Shade Resin 807.92(a)3 Legally marketed device of which equivalence is claimed: VenusDiamond Flow (K091635)
#### [807.92(a)(4)] Description of Device:
Venus Bulk Flow ONE is a light curing, radio-opaque nano-hybrid composite.
The uncured dental composite is introduced into the oral cavity to be applied and cured to the tooth structure or restoration pre-treated by the dentist, thus achieving the clinicalbenefit.
Venus Bulk Flow ONE is used for extended fissure sealing, first-layer lining for class I and class II cavities, Class I, II, III and V direct restorations, repair of direct and indirectrestorations in combination with a suitable adhesive, splinting loosened teeth resulting from trauma or periodontal associated events.
#### Product Image:
Image /page/3/Picture/14 description: The image shows a collection of dental products. The products are packaged in white boxes with the brand name "Venus" prominently displayed. There are also syringes and other dental tools visible in the image, suggesting a comprehensive dental care kit.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Kulzer, a Mitsui Chemicals Group company. The logo consists of a green circle with a white tooth inside, followed by a vertical gray line. To the right of the line is the word "KULZER" in large, gray, sans-serif font, with the words "MITSUI CHEMICALS GROUP" in smaller font underneath.
| Parameter | Required Specification |
|----------------------------------------|-------------------------------------------|
| 5.2.7 sensitivityto<br>dental OP light | > 60 s |
| 5.2.8 depth of<br>cure | > 1,5 mm |
| 5.2.9 flexural<br>strength | > 80 MPa |
| 5.2.10 water<br>sorption | < 40 µg/mm3 |
| 5.2.10 water<br>solubility | < 7,5 µg/mm3 |
| 5.4. colour<br>stability | no more than<br>slight change in<br>color |
| 5.5 radio-opacity | > 100%-Al and<br>SD < 50%, if<br>claimed |
## [807.92(a)(5)| Intended Use:
Venus Bulk Flow ONE is used for extended fissure sealing, first-layer lining for class I and class II cavities, Class I, II, III and V direct restorations, repair of direct and indirectrestorations in combination with a suitable adhesive, splinting loosened teeth resulting from trauma or periodontal associated events.
The predicate device (K091635) is restricted to areas not subjected to masticatory forces. This is the primary difference between the subject device Venus Bulk Flow ONE and the predicate device which is Venus Diamond Flow (K091635). The intended population for use is dental patients needing restorations. The expanded indication of not being limited to areas that are notsubjected to masticatory surfaces allows for more broad use of this product in dental restorations for said indications. Internal studies per ISO 4049 yielded results that demonstrate this product is appropriate for use in those areas safely.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Kulzer, a company that is part of the Mitsui Chemicals Group. The logo features a green circle with a white tooth shape inside, followed by the word "KULZER" in bold, gray letters. Below "KULZER" is the text "MITSUI CHEMICALS GROUP" in a smaller font, also in gray. The logo is clean and modern, with a focus on dental health and chemical innovation.
## [807.92(a)(6)] Technological Characteristics:
The basic fundamental scientific technology of the subject device Venus Bulk Flow ONE versusthe predicate device Venus Diamond Flow (K091635) are identical. The chemistries are identical. Conditions of use, time of patient contact, physical properties, biocompatibility, tissue contact, patient population, users are all identical.
The differences are different composite amount in the packaging, depth of cure, and less restrictive indications for use. The subject device Venus Bulk Flow ONE syringe contains 2g of material while the predicate device Venus Diamond Flow (K091635) contains 1.8g. This has no effect on safety or efficacy. The depth of cure for the subject device Venus Bulk Flow ONE is 4mm while for the predicate device Venus Diamond Flow (K091635) it is 2mm. The 4mm curing depth is achieved by modification of the opacity (color) only. This has no effect on safety or efficacy. The remaining difference which is a less restrictive indications for use due to the restriction not to use on masticatory surfaces also has no effect on safety and efficacy. The predicate device Venus Diamond Flow (K091635) same as subject device Venus Bulk Flow ONE was tested perISO 4049 and met the requirements therein. Since both devices meet the ISO 4049 requirements, both could have had this restriction removed. At the time Venus DiamondFlow (K091635) was cleared and released for sale, it simply was not thought of.
## [807.92(b)(1)]: Non-clinical bench testing:
The following testing was performed:
- light sensitivity O
- 0 depth of cure
- 0 flexural strength
- 0 water sorption
- 0 water solubility
- 0 color stability
- o radio-opacity
In all cases, results were either equal or better than the predicate Venus Diamond Flow(K091635) and exceeded specifications as required by ISO 4049:2019. With these results in mind, the subject device Venus Bulk Flow ONE is better than or equivalent tothe predicate device Venus Diamond Flow (K091635) in all cases.
## [807.92(b)(2)] Clinical Performance Data:
A formal clinical trial was not performed.
Venus Bulk Flow ONE is the latest product development of Kulzer's dental flowable composites. It is based on a modification of Venus Diamond Flow, marketed since 2009. According to the performed Equivalence Evaluation, the devices Venus Bulk FlowONE and Venus Diamond Flow can be considered equal. Venus Bulk Flow ONE was developed to further simplify and accelerate the process of direct composite restorationswith good aesthetics.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Kulzer, a company that is part of the Mitsui Chemicals Group. The logo consists of a green circle with a white tooth inside, followed by the word "KULZER" in gray, with the words "MITSUI CHEMICALS GROUP" underneath in a smaller font. The logo is simple and modern, and the use of green suggests a focus on health or nature.
The translucency of Venus Bulk Flow ONE is reduced to the possible minimum, while still allowing 4 mm of curing depth. In this way, the bulk-fill technique is possible without compromise in aesthetics. Both Venus Diamond Flow and Venus Bulk Flow ONE meet the performance requirements of ISO 4049:2019. There is no clinically significant difference in the safety and clinical performance of the devices Venus Diamond Flow and Venus Bulk Flow ONE. Therefore, the current clinical evaluation "Dental Composites" remains valid and applies to Venus Bulk Flow ONE. This statement verifies that Venus Bulk Flow ONE, when used under the conditions andfor the purposes intended, will perform only with known and foreseeable risks and side effects (inclusively from foreseeable misuse). No new risks or other previously unknownsafety issues or side effects were identified.
## 807.92(b)(3) Conclusions from Non-clinical and Clinical Tests:
Based on the above non-clinical and clinical evaluation data/tests, the subject device Venus Bulk Flow ONE is as safe and effective as the predicate device Venus DiamondFlow(K091635).
#### 807.92(d) Information reasonably deemed necessary by FDA:
Any additional information reasonably deemed necessary by the FDA will be provided ina timely a fashion as possible.
## 807.92(e) Indications for Use:
| | Subject Device Venus<br>Bulk Flow ONE<br>(K220605) | Predicate Device Venus<br>Diamond Flow (K091635) |
|--------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication<br>for<br>Use | • Extended fissure sealing<br>• First-layer lining for class<br>I and II cavities<br>• Class-I, -II, -III and -V<br>direct restorations<br>• Repair of direct and<br>indirect restorations in<br>combination with a suitable<br>adhesive<br><br>Splinting loosened teeth<br>resulting from trauma or<br>periodontal associated<br>events | • Enlarged fissure sealing<br>•Cavity lining – as the first<br>layer for Class I and II<br>cavities<br>•Class V fillings<br>•Minimally invasive Class I<br>and II fillings in areas not<br>subjected to masticatory<br>forces<br>•Minimally invasive Class III<br>fillings<br>• Small repairs of direct and<br>indirect restorations<br>combined with a suitable<br>bonding agent<br><br>Interlocking of loosened teeth |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.