The NeoNavia biopsy system is intended for obtaining tissue samples from both breast lesions and axillary lymph nodes for diagnostic analysis of breast abnormalities.
Device Story
NeoNavia Biopsy System obtains tissue samples from breast lesions and axillary lymph nodes for histological evaluation. System comprises base unit, driver, and sterile single-use probes (FlexiPulse and VacuPulse). Operated under ultrasound imaging guidance by healthcare professionals in clinical settings. Probes utilize pulses to improve precision and control during needle insertion and positioning. VacuPulse uses vacuum-assisted side-cut technology; FlexiPulse uses front-loaded vacuum-assisted needle with rotational severing. Manual positioning with optional short longitudinal movement aids millimeter-by-millimeter advancement. Output is tissue sample for diagnostic analysis. Benefits include controlled, precise tissue acquisition for breast abnormality diagnosis.
Clinical Evidence
No human clinical data. Evidence consists of bench testing (performance verification, usability validation) and animal studies (sheep and pigs) comparing NeoNavia probes to the predicate. Animal study endpoints included biopsy sample quality (gross/histopathology), acute bleeding rate, and hematoma formation.
Technological Characteristics
System includes base unit, driver, and single-use probes. Materials: stainless steel. Sensing/Actuation: vacuum-assisted, rotational severing, pulse-based needle advancement. Connectivity: standalone. Sterilization: validated for single-use probes. Software: Moderate Level of Concern.
Indications for Use
Indicated for obtaining tissue samples from breast lesions and axillary lymph nodes for histologic examination in patients with breast abnormalities. To be used by healthcare professionals in hospitals or healthcare facilities.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
NeoDynamics AB % Leann Latham Regulatory Consultant M Squared Associates,Inc. 127 West 30th Street. 9th Floor New York, New York 10001
September 8, 2022
Re: K220595
Trade/Device Name: NeoNavia Biopsy System Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology Biopsy Instrument Regulatory Class: Class II Product Code: KNW, FCG Dated: August 1, 2022 Received: August 3, 2022
Dear Leann Latham:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin Chen Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220595
Device Name NeoNavia Biopsy System
Indications for Use (Describe)
The NeoNavia biopsy system is intended for obtaining tissue samples from both breast lesions and axillary lymph nodes for diagnostic analysis of breast abnormalities.
The FlexiPulse and VacuPulse probes are intended to provide tissue from breast lesions and axillary lymph nodes for histologic examination.
The NeoNavia biopsy system is to be used only by healthcare professionals in hospitals or healthcare facilities.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### Section 005: 510k Summary
The following information is provided as required by 21 CFR § 807.87 for NeoNavia Biopsy System 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the information upon which the substantial equivalence determination is based.
- Sponsor: NeoDynamics AB Anna Forsberg Medical Affairs Manager and Country Manager, US Lejonvägen 14 181 32 Lidingö SWEDEN Phone: 973-953-9885 Email: anna.forsberg@neodynamics.com
- LeAnn Latham Contact: M Squared Associates, Inc. 127 West 30th Street New York, NY 10001 E-mail: llatham@msquaredassociates.com Phone: 347-954-9482 Fax (703) 562-9797
| Date of Submission: | February 28, 2022 |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary Name: | NeoNavia Biopsy System |
| Common Name: | Instrument, Biopsy |
| Regulatory Class: | Class II |
| Regulation: | 21 CFR 876.1075 |
| Product Codes: | KNW, FCG |
| Predicate Devices: | Primary Device Trade Name: Mammotome elite® Biopsy System<br>Device Company: Devicor® Medical Products, Inc.<br>510(k) Number: K153709 |
### Device Description :
NeoNavia® biopsy system is designed to obtain tissue samples from breast lesions or axillary lymph nodes for histological evaluation. The system is composed of a base unit, a driver and two different types of sterile single use probes. Each probe type (also referred to as needle) utilizes pulses with the intention to improve precision and control when inserting and positioning the biopsy needle in a suspicious lesion. NeoNavia® must be operated together with ultrasound imaging guidance.
{4}------------------------------------------------
### Intended Use
The NeoNavia biopsy system is intended for obtaining tissue samples from both breast lesions and axillary lymph nodes for diagnostic analysis of breast abnormalities.
### Indications for Use:
The FlexiPulse and VacuPulse probes are intended to provide tissue from breast lesions and axillary lymph nodes for histologic examination.
The NeoNavia biopsy system is to be used only by healthcare professionals in hospitals or healthcare facilities.
# Technological characteristics, comparison to predicate device:
| | | Table 1: Comparison table of NeoNavia and the predicates | | | | | | |
|--|--|----------------------------------------------------------|--|--|--|--|--|--|
|--|--|----------------------------------------------------------|--|--|--|--|--|--|
| Characteristics | NeoNavia® Biopsy | Mammotome elite® | Differences and |
|--------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | System (subject device) | Biopsy System | Comments |
| | | (Primary predicate) | |
| Description | NeoNavia® biopsy<br>system is designed to<br>obtain tissue samples<br>from breast lesions or<br>axillary lymph nodes for<br>histological evaluation.<br>The system is composed<br>of a base unit, a driver<br>and two different types<br>of sterile single use<br>probes. | Mammotome elite®<br>Biopsy System consists<br>of a reusable Holster<br>and a single-patient use,<br>sterile Probe that is<br>used with ultrasound<br>imaging guidance to<br>excise and collect<br>diagnostic samples with<br>a single insertion of the<br>Probe. | |
| 510(k) number | TBD | K153709 | |
| Product Code | KNW, FCG (proposed) | KNW | |
| Intended Use | The NeoNavia biopsy<br>system is intended for<br>obtaining tissue<br>samples from both<br>breast lesions and<br>axillary lymph nodes for<br>diagnostic analysis of<br>breast abnormalities. | The Mammotome elite®<br>Biopsy System is<br>intended to provide<br>breast or axillary lymph<br>node tissue samples for<br>diagnostic analysis of<br>imaged or palpated<br>breast abnormalities. | Same, to provide breast<br>tissue and lymph node<br>samples for diagnostic<br>analysis. |
| Characteristics | NeoNavia® Biopsy<br>System (subject device) | Mammotome elite®<br>Biopsy System | Differences and<br>Comments |
| | | (Primary predicate) | |
| Indications for Use | The FlexiPulse and<br>VacuPulse probes are<br>intended to provide<br>tissue from breast<br>lesions and axillary<br>lymph nodes for<br>histologic examination.<br>The NeoNavia biopsy<br>system is to be used<br>only by healthcare<br>professionals in<br>hospitals or healthcare<br>facilities. | The Mammotome elite®<br>Biopsy System is<br>indicated to obtain<br>tissue samples from the<br>breast or axillary lymph<br>nodes for diagnostic<br>analysis of breast<br>abnormalities.<br>• The Mammotome<br>elite® Biopsy System is<br>intended to provide<br>breast tissue for<br>histologic examination<br>with partial or complete<br>removal of the imaged<br>abnormality.<br>• The Mammotome<br>elite® Biopsy System is<br>intended to provide<br>breast tissue for<br>histologic examination<br>with partial removal of<br>a palpable abnormality.<br>The extent of a<br>histologic abnormality<br>cannot always be<br>readily determined<br>from the palpation or<br>imaged appearance.<br>Therefore, the extent of<br>removal of the palpated<br>or imaged evidence of<br>an abnormality does<br>not predict the extent<br>of removal of a<br>histologic abnormality,<br>e.g., malignancy. When<br>the sampled<br>abnormality is not<br>histologically benign, it | NeoNavia is not<br>intended for complete<br>removal of the imaged<br>abnormality whereas<br>Mammotone is<br>designed to provide<br>breast tissue for<br>histologic examination<br>with partial or complete<br>removal of the imaged<br>abnormality. This<br>difference does not<br>affect NeoNavia's<br>intended use or<br>indications for use for<br>obtaining tissue<br>samples from both<br>breast lesions and<br>axillary lymph nodes for<br>diagnostic analysis of<br>breast abnormalities. |
| Characteristics | NeoNavia® Biopsy<br>System (subject device) | Mammotome elite®<br>Biopsy System<br>(Primary predicate) | Differences and<br>Comments |
| | | is essential that the<br>tissue margins be<br>examined for<br>completeness of<br>removal using standard<br>surgical procedures.<br>In instances when a<br>patient presents with a<br>palpable abnormality<br>that has been classified<br>as benign through<br>clinical and/or<br>radiological criteria<br>(e.g., fibroadenoma,<br>fibrocystic lesion), the<br>Mammotome elite®<br>Biopsy System may also<br>be used to partially<br>remove such palpable<br>lesions. Whenever<br>breast tissue is<br>removed, histological<br>evaluation of the tissue<br>is the standard of care.<br>When the sampled<br>abnormality is not<br>histologically benign, it<br>is essential that the<br>tissue margins be<br>examined for<br>completeness of<br>removal using standard<br>surgical procedures. | |
| Intended users | Healthcare<br>professionals<br>(The NeoNavia system is<br>to be used only by<br>healthcare professionals<br>trained in subcutaneous<br>biopsy procedures | Healthcare<br>professionals<br>(Minimally invasive<br>procedures should be<br>performed only by<br>persons having<br>adequate training and | Same |
| Characteristics | NeoNavia® Biopsy<br>System (subject device) | Mammotome elite®<br>Biopsy System<br>(Primary predicate) | Differences and<br>Comments |
| | guided by ultrasound<br>imaging equipment.) | familiarity with<br>minimally invasive<br>techniques.) | |
| Image guidance<br>modality | Ultrasound | Ultrasound | Same |
| Patient contacting<br>materials | Stainless steel | Stainless steel | Same |
| Handheld procedure? | Yes | Yes | Same |
| Method of needle<br>insertion and<br>advancement | Manual positioning<br>Optional short<br>longitudinal movement<br>to aid advancing the<br>needle millimeter by<br>millimeter. | Manual positioning | NeoNavia may be used<br>manually and with short<br>longitudinal movement<br>that provides greater<br>control over manual<br>positioning. |
| Vacuum | Vacuum Assisted<br>System | Vacuum Assisted<br>System | Same |
| Single use disposable vs<br>Reusable device | Single use, disposable<br>biopsy probe; reusable<br>driver and base unit | Single use, disposable<br>biopsy probe; reusable<br>driver | Same |
{5}------------------------------------------------
{6}------------------------------------------------
{7}------------------------------------------------
### Table 2: Comparison of NeoNavia VacuPulse probe to predicate
| Characteristics | NeoNavia VacuPulse<br>configurations | Mammotome elite®<br>Biopsy System | Differences and<br>Comments |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
| Marketed needle size | 10 Gauge | 10 & 13 Gauge | Mammotome has an<br>additional smaller size<br>needle. |
| Method of tissue<br>sampling | Vacuum assisted side-cut<br>needle with rotating<br>inner cutting cannula<br>advanced forward to<br>sever tissue sample. | Vacuum assisted side-cut<br>needle with rotating<br>inner cutting cannula<br>advanced forward to<br>sever tissue sample. | Same |
| Method of sample<br>collection | Vacuum transport to<br>sample collection basket. | Vacuum transport to<br>sample collection basket. | Same |
{8}------------------------------------------------
| Characteristics | NeoNavia VacuPulse<br>configurations | Mammotome elite®<br>Biopsy System | Differences and<br>Comments |
|-----------------|--------------------------------------|-----------------------------------|-----------------------------|
| | Multiple samples possible. | Multiple samples possible. | |
#### Table 3: Comparison of NeoNavia Flexipulse probe to the predicates
| Characteristics | NeoNavia FlexiPulse<br>configurations | Mammotome elite®<br>Biopsy System | Differences and<br>Comments |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Marketed needle size | 14 Gauge | 10 & 13 Gauge | Difference in needle sizes.<br>No significant impact to<br>substantial equivalence<br>per testing performed<br>with FlexiPulse needle as<br>compared to<br>Mammotome 10G<br>needle. Ref. NeoNavia<br>Biopsy System Needle<br>Size Rational report |
| Method of tissue<br>sampling | Front loaded vacuum<br>assisted needle with<br>rotational severing<br>capability, advanced<br>forward millimeter by<br>millimeter using short<br>longitudinal movement. | Vacuum assisted side-cut<br>needle with rotating<br>inner cutting cannula<br>advancing forward to<br>sever tissue sample. | NeoNavia needle is front<br>loaded compared to side<br>cut. Safety and<br>performance of<br>needles/probes validated<br>by animal testing. |
| Method of sample<br>collection | Single sample inside front<br>loaded needle. | Vacuum transport to<br>sample collection basket.<br>Multiple samples<br>possible. | FlexiPulse collects a single<br>sample whereas<br>Mammotome collects<br>multiple samples. |
# Performance testing:
# Sterilization, Shelf Life and Packaging
The VacuPulse and FlexiPulse probes of the NeoNavia Biopsy System are provided sterilization validation, sterile packaging validation and shelf life testing have been performed for the sterile disposable probes.
Biocompatibility Testing
{9}------------------------------------------------
Biocompatibility testing results for cytotoxicity, irritation, sensitization, acute systemic toxicity and materials mediated pyrogenicity indicate the NeoNavia VacuPulse probe is safe for its indicated use.
# Software and Security
The Level of Concern associated with the software of the NeoNavia Biopsy System is considered to be Moderate. Testing was completed in accordance with the guidance document, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (Final, 2005).
# Electromagnetic Compatibility and Electrical Safety
Electromagnetic compatibility and electrical safety testing was performed on the NeoNavia Biopsy System.
# Bench Testing
Various bench testing was performed for verification of the NeoNavia Biopsy System. Testing included comparison performance testing, usability validation and performance verification testing.
# Animal Testing
Animal testing was conducted to compare the NeoNavia VacuPulse, NeoNavia Flexipulse and Mammotome biopsy devices. Biopsy samples from sheep and pigs were evaluated for biopsy sample quality (gross qualitative assessment and via histopathology), acute bleeding rate, and hematoma.
# Conclusion
Based on the information provided the NeoNavia Biopsy System is substantially equivalent in terms of indications for use, technology and performance to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.