The Avanos* CORTRAK*2 Enteral Access System (EAS) assists qualified clinicians to guide NG/NI feeding tubes into the stomach or small bowel.
Device Story
Device tracks path of 8Fr+ feeding tube tip during placement. Distal end of transmitting stylet contains coil winding acting as transmitter; signal detected by external receiver unit. Monitor unit processes, records, and presents raw data as visual screen tracing. Used in clinical environments by trained clinicians. Output provides real-time visual feedback to assist clinicians in guiding tube placement into stomach or small bowel. Benefits patient by facilitating accurate enteral feeding tube positioning.
Clinical Evidence
No clinical data. Non-clinical verification of labeling changes conducted via risk management process per ISO 14971:2019.
Technological Characteristics
Electrical device; non-patient contacting; reusable; non-sterile. Uses electromagnetic coil-based tube tracking technology. No changes in materials or design from predicate.
Indications for Use
Indicated for qualified clinicians to guide placement of Avanos Medical CORTRAK*2 feeding tubes (8FR or greater) into the stomach or small bowel of patients requiring enteral feeding, in conjunction with institution protocols.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 22, 2022
Avanos Medical, Inc. % Angela Cushman Senior Vice President, QA/RA (Interim) 5405 Windward Parkway Alpharetta, GA 30004
# Re: K220588
Trade/Device Name: Avanos* CORTRAK* 2 Enteral Access System (EAS) Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal Tube and Accessories Regulatory Class: Class II Product Code: KNT Dated: February 28, 2022 Received: March 1, 2022
# Dear Angela Cushman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220588
Device Name Avanos* CORTRAK* 2 Enteral Access System (EAS)
Indications for Use (Describe)
The Avanos* CORTRAK*2 Enteral Access System (EAS) utilizes tube tracking technology to assist, in conjunction with institution protocols, qualified clinicians in guiding placement of Avanos Medical CORTRAK *2 feeding tubes of 8FR or greater into the stomach or small bowel of patients requiring enteral feeding.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "AVANOS" in a bold, sans-serif font. The letters are evenly spaced and appear to be a dark gray or black color against a white background. The overall impression is clean and modern.
K220588 Page 1 of 3 5405 Windward Parkway Alpharetta, GA 30004
T 1-844-4AVANOS www.avanos.com
### 510(k) SUMMARY - K220558
### Avanos* CORTRAK* 2 Enteral Access System (EAS)
#### 1. SUBMITTER
Avanos Medical, Inc. 5405 Windward Parkway Alpharetta, Georqia 30004, U.S.A.
Correspondent Name: Angela Cushman Senior VP, RAQA (Interim) 706-344-8678 angela.cushman@avanos.com
Date Prepared: April 7, 2022
#### 2. Device
Name of Device: Common or Usual Name: Classification Name: Regulation: Regulatory Class: FDA Product Code:
Avanos* CORTRAK* 2 Enteral Access System (EAS) Enteral Access Device Gastrointestinal tube and accessories 21 CFR § 876.5980 = KNT
#### 3. Predicate Device
CORTRAK* 2 Equilateral EAS (K191340)
#### 4. Description of Device
Avanos* CORTRAK* 2 Enteral Access System (EAS) device is designed to track the path of an 8 Fr or greater Avanos Medical feeding tube tip during the patient placement procedure. A coil winding at the distal end of the transmitting stylet acts as a transmitter, and its signal is detected by the externally positioned receiver unit. The received signals are input to the attached Monitor unit. The resulting raw data is processed, recorded, and presented to the operator in a meaningful and intuitive screen tracing. The Avanos* CORTRAK*2 EAS device is an electrical device that does not contact the patient, is not sterilized, and is reusable. Like the predicate device, it is intended to be used in a clinical environment by qualified trained clinicians.
{4}------------------------------------------------
# 5. Intended Use
The Avanos* CORTRAK*2 Enteral Access System (EAS) assists qualified clinicians to guide NG/NI feeding tubes into the stomach or small bowel of patients requiring enteral feeding.
### 6. Indications for Use
The Avanos* CORTRAK*2 Enteral Access System (EAS) utilizes tube tracking technology to assist, in conjunction with institution protocols, qualified clinicians in guiding placement of Avanos Medical CORTRAK*2 feeding tubes of 8FR or greater into the stomach or small bowel of patients requiring enteral feeding.
# 7. Technological Characteristics
The Avanos* CORTRAK* 2 Enteral Access System is substantially equivalent to the predicate device CORTRAK* 2 Equilateral EAS (K191340).
The only change to the subject device involves labeling updates. Labeling updates include a change to the indications for use to clarify the inclusion of a secondary confirmatory method for tube placement per institution protocols, and enhanced warnings in a consolidated organized section.
There were no other changes in the design, materials, performance, and technological characteristics from the predicate device.
The difference in the updated labeling has no impact on the intended use, technological principles, safety, or effectiveness of the subject device when compared to the predicate device.
# 8. Substantial Equivalence Discussion
#### Intended Use Comparison
| Predicate Indication<br>for Use<br>(K191340) | Predicate Intended<br>Use<br>(K191340) | Subject Indication<br>for Use (K220558) | Subject Intended Use<br>(K220558) |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|
| CORTRAK*2<br>Equilateral EAS<br>CORTRAK* 2<br>Equilateral Enteral<br>Access System is an<br>electrical device<br>designed to aid<br>qualified operators in<br>the placement of<br>Avanos NG feeding<br>tubes of 8 FR or<br>greater into the<br>stomach or small<br>bowel of patients | CORTRAK*2<br>Equilateral EAS<br>Assists qualified<br>operators to guide<br>NG/NI feeding tubes<br>into stomach or small<br>bowel. | Avanos* CORTRAK*2<br>EAS<br>The Avanos*<br>CORTRAK*2 Enteral<br>Access System (EAS)<br>utilizes tube tracking<br>technology to assist, in<br>conjunction with<br>institution protocols,<br>qualified clinicians in<br>guiding placement of<br>Avanos Medical<br>CORTRAK*2 feeding<br>tubes of 8FR or | Avanos* CORTRAK*2<br>EAS<br>Assists qualified<br>clinicians to guide<br>NG/NI feeding tubes<br>into stomach or small<br>bowel. |
{5}------------------------------------------------
| Predicate Indication<br>for Use<br>(K191340) | Predicate Intended<br>Use<br>(K191340) | Subject Indication<br>for Use (K220558) | Subject Intended Use<br>(K220558) |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|--------------------------------------------------------------------------------------------|-----------------------------------|
| CORTRAK*2<br>Equilateral EAS | CORTRAK*2<br>Equilateral EAS | Avanos* CORTRAK*2<br>EAS | Avanos* CORTRAK*2<br>EAS |
| requiring enteral<br>feeding. CORTRAK* 2<br>Equilateral Enteral<br>Access System can be<br>used to confirm<br>placement of feeding<br>tubes prior to<br>commencing the<br>delivery of enteral<br>nutrition. | | greater into the<br>stomach or small<br>bowel of patients<br>requiring enteral<br>feeding. | |
# Conclusion:
The change in the Indications for Use does not change the Intended Use of assisting qualified operators to quide NG/NI feeding tubes into the stomach or small bowel.
The Indications for Use change does not raise different or new questions of safety and effectiveness.
# Technologic Characteristics Comparison
The labeling changes have no effect on the technologic characteristics. The labeling difference has no impact on the intended use, technological principles, safety, or effectiveness of the subject device when compared to the predicate device. The device design, material, and operating principle remain the same as the predicate device CORTRAK* 2 Equilateral EAS (K191340).
# 9. Summary of Non-Clinical Testing
Non-clinical verification of the labeling change was conducted through the risk management process according to ISO 14971:2019. The risk management file was updated to eliminate risks associated with the use of Avanos* CORTRAK*2 EAS as a confirmatory method for the placement of CORTRAK*2 NG/NI feeding tubes. The risk profile of the device system did not have a significant impact due to the labeling updates.
There were no other non-clinical tests performed for the labeling updates for the subject device.
#### 10. Conclusion
The difference between the predicate CORTRAK* 2 Equilateral Enteral Access Device (K191340) and subject Avanos* CORTRAK*2 Enteral Access System (EAS) (K220588) do not raise any new or different questions of safety or effectiveness. The subject Avanos* CORTRAK*2 Enteral Access System (EAS) is substantially equivalent to the predicate CORTRAK*Equilateral Enteral Access Device cleared under K191430 with respect to the intended use, technology, operating principle, material composition, and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.