← Product Code [FTL](/productcode/FTL) · K220586

# Symbotex Composite Mesh, ProGrip Self-Gripping Polyester Mesh, Parietex Hydrophilic 2D 3D Anatomical Mesh, Versatex Monofilament Mesh (K220586)

_Sofradim Production · FTL · Jun 22, 2022 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K220586

## Device Facts

- **Applicant:** Sofradim Production
- **Product Code:** [FTL](/productcode/FTL.md)
- **Decision Date:** Jun 22, 2022
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.3300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Symbotex™ Composite Mesh is intended for the reinforcement of abdominal wall soft tissue where a weakness exists. ProGrip™ Self-Gripping Polyester Mesh is intended for use in reinforcement of abdominal wall soft tissues where a weakness exists. Parietex™ Hydrophilic 2D 3D Anatomical Mesh is intended for reinforcement of abdominal wall soft tissue where a weakness exists. Versatex™ Monofilament Mesh is intended for reinforcement of abdominal wall soft tissue where a weakness exists.

## Device Story

Surgical meshes (Symbotex, ProGrip, Parietex, Versatex) used for abdominal wall reinforcement during hernia repair; implanted by surgeons in clinical settings. Symbotex: 3D monofilament polyester with absorbable porcine collagen/glycerol film. ProGrip: knitted monofilament polyester with resorbable polylactic acid pins for self-fixation. Parietex: 2D/3D multifilament polyester mesh. Versatex: macroporous 3D monofilament polyester. Devices provide structural support to abdominal wall tissue; facilitate tension-free repair. Update to Instructions for Use (IFU) to comply with EU 2017/745; includes added information on trocar compatibility, fixation, follow-up, and MRI compatibility. No changes to device design, materials, or manufacturing.

## Clinical Evidence

Bench testing only. In vitro testing performed to evaluate trocar compatibility of the subject devices; results met all established acceptance criteria. No clinical data provided as device design and materials remain unchanged from predicate devices.

## Technological Characteristics

Polyester-based surgical meshes. Symbotex: 3D monofilament polyester with porcine collagen/glycerol film. ProGrip: knitted monofilament polyester with resorbable polylactic acid pins. Parietex: 2D/3D multifilament polyester. Versatex: macroporous 3D monofilament polyester. Features include dyed monofilament markers (D&C Green No. 6) for orientation. Devices are permanent implants.

## Regulatory Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

## Predicate Devices

- Symbotex™ Composite Mesh ([K142908](/device/K142908.md))
- ProGrip™ Self-Gripping Polyester Mesh ([K142900](/device/K142900.md))
- Parietex™ Hydrophilic 2D 3D Anatomical Mesh ([K173796](/device/K173796.md))
- Versatex™ Monofilament Mesh ([K150091](/device/K150091.md))

## Reference Devices

- Symbotex™ Composite Mesh ([K131969](/device/K131969.md))
- ProGrip™ Self-Gripping Polyester Mesh ([K081050](/device/K081050.md))
- Parietex™ Hydrophilic 2D 3D Anatomical Mesh ([K982532](/device/K982532.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

June 22, 2022

Sofradim Production % Wing Ng Regulatory Affairs Director Covidien LLC. 15 Hampshire Street Mansfield, Massachusetts 02048

Re: K220586

Trade/Device Name: Symbotex Composite Mesh, ProGrip Self-Gripping Polyester Mesh, Parietex Hydrophilic 2D 3D Anatomical Mesh, Versatex Monofilament Mesh Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical Mesh Regulatory Class: Class II Product Code: FTL Dated: May 5, 2022 Received: May 23, 2022

Dear Wing Ng:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Deborah Fellhauer, RN, BSN Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K220586

Device Name

Symbotex™ Composite Mesh

Indications for Use (Describe)

Symbotex™ Composite Mesh is intended for the reinforcement of abdominal wall soft tissue where a weakness exists, in procedures involving primary abdominal wall and incisional hernia surgeries.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K220586

Device Name

ProGrip™ Self-Gripping Polyester Mesh

Indications for Use (Describe)

ProGrip™ Self-Gripping Polyester Mesh is intended for the reinforcement of abdominal wall soft tissues where a weakness exists in procedure involving inguinal and incisional hernias repair.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known) K220586

Device Name

ParietexTM Hydrophilic 2D 3D Anatomical Mesh

Indications for Use (Describe)

ParietexTM hydrophilic 2-dimensional mesh, Parietex-1M hydrophilic 3-dimensional mesh and ParietexTM 1ydrophilic anatomical mesh are intended for reinforcement of abdominal wall soft tissue where a weakness exists, in procedures involving inguinal and incisional hernia repairs.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20)

{5}------------------------------------------------

### Indications for Use

510(k) Number (if known) K220586

Device Name

VersatexIM Monofilament Mesh

Indications for Use (Describe)

VersatexTM Monofilament Mesh is intended for reinforcement of abdominal wall soft tissue where a weakness exists, in procedures involving abdominal wall hernia repairs.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (6/20)

{6}------------------------------------------------

K220586, Page 1 of 6

# 510(k) Summary

| Date Prepared:          | June 22, 2022                                                                                                                                                                       |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:              | Sofradim Production (subsidiary of Covidien llc)<br>116, avenue du Formans<br>01600 Trevoux, France<br>Telephone: +33 (0)4 74 08 90 00<br>Fax: +33 (0) 4 74 08 90 02                |
| Contact:                | Paula Paz<br>Regulatory Operations Specialist<br>116, avenue du Formans<br>01600 Trevoux, France<br>Phone: +33 (0)4 74 08 90 00<br>Email: paula.paz@medtronic.com                   |
| Name of device:         | Symbotex™ Composite Mesh<br>ProGrip™ Self-Gripping Polyester Mesh<br>Parietex™ Hydrophilic 2D 3D Anatomical Mesh<br>Versatex™ Monofilament Mesh                                     |
| Trade/Proprietary name: | Surgical Mesh                                                                                                                                                                       |
| Common name:            | Mesh, Surgical, Absorbable, Abdominal Hernia<br>Panel number and product code: 79 FTL                                                                                               |
| Classification name:    | Regulation number: 21 CFR 878.3300                                                                                                                                                  |
| Predicate Device:       |                                                                                                                                                                                     |
| Trade/Proprietary name: | Symbotex™ Composite Mesh<br>ProGrip™ Self-Gripping Polyester Mesh<br>Parietex™ Hydrophilic 2D 3D Anatomical Mesh<br>Versatex™ Monofilament Mesh                                     |
| Common name:            | Surgical Mesh                                                                                                                                                                       |
| Classification name:    | Mesh, Surgical, Absorbable, Abdominal Hernia<br>Panel number and product code: 79 FTL<br>Regulation number: 21 CFR 878.3300                                                         |
| 510(k) Number:          | Symbotex™ Composite Mesh: K142908<br>ProGrip™ Self-Gripping Polyester Mesh: K142900<br>Parietex™ Hydrophilic 2D 3D Anatomical Mesh: K173796<br>Versatex™ Monofilament Mesh: K150091 |
| Manufacturer:           | Sofradim Production (subsidiary of Covidien llc)<br>116, avenue du Formans<br>01600 Trevoux, France                                                                                 |

{7}------------------------------------------------

Reason for 510(k) The purpose of this 510(k) is to notify the Agency of the update Submission: of Instructions for Use for the predicate devices Symbotex™ Composite Mesh (K142908), ProGrip™ Self-Gripping Polyester Mesh (K142900), Parietex™ Hydrophilic 2D 3D Anatomical Mesh (K173796) and Versatex™ Monofilament Mesh (K150091). These changes are made to comply with the requirements of the new European Requlation (EU) 2017/745. and, as part of continuous improvement, for improved readability and clarity and alignment with the state-of-the-art for surgical mesh implants. Device Description The purpose of this Special 510(k) is to notify the Agency the changes made on Instructions for Use. There is no change performed on the devices itself, and the device description is identical to those provided for predicate devices Symbotex™ Composite Mesh (K142908), ProGrip™ Self-Gripping Polyester Mesh (K142900), Parietex™ Hydrophilic 2D 3D Anatomical Mesh (K173796) and Versatex™ Monofilament Mesh (K150091). Symbotex™ Composite Mesh Symbotex™ Composite Mesh is made out of a threedimensional monofilament polyester textile which is covered with an absorbable, continuous and hydrophilic film on one of its sides. This film is made up of collagen from porcine origin and glycerol. The collagen film is essentially degraded in less than 1 month. For "Flat sheet" (SYM reorder codes): A dyed monofilament polyester (D&C Green No. 6) marking is positioned on the center of the textile, on the opposite side of the film, and helps center and orient the mesh. For "Flat Sheet with sutures" (SYM-F reorder codes): Nonabsorbable pre-placed sutures are tied to the three-dimensional mesh. A dyed monofilament polyester (D&C Green No. 6) marking is positioned on the center of the textile, on the opposite side of the film, and helps center and orient the mesh. For "With flap" (SYM-OS reorder codes): A dyed (D&C Green No. 6) bi-dimensional monofilament polvester textile flap is attached to the threedimensional reinforcement and helps place and fix the mesh.

{8}------------------------------------------------

### ProGrip™ self-aripping polvester mesh

The mesh and the overlapping flaps of the pre-cut versions are made of knitted monofilament polyester and have polylactic acid monofilament resorbable pins on one of the sides. These pins facilitate placing, positioning and fixation of the overlapping flap and the mesh to the surrounding tissue. A colored yarn marker on the medial edge of the pre-cut mesh helps orientation.

The monofilament polylactic acid pins are bioresorbable and contribute to the fixation of the mesh to surrounding tissue during at least 8 weeks. The polylactic acid pins degrade and resorb in vivo by hydrolysis and are metabolized by the body into CO2 and H2O.

### Parietex™ Hydrophilic 2D 3D Anatomical Mesh

Two or three-dimensional multifilament polyester wall reinforcements.

The two-dimensional mesh, with rectangular pores, is available in two different textures: a standard version (TEC references) and a rigid version (TECR references). The textures and transparencies of this mesh make them particularly suitable for the treatment of parietal affections, in particular inguinal hernias, through laparoscopic approach.

The three-dimensional mesh (TET references) has hexagonal pores. The flexibility, porosity and low density of this mesh make it particularly suitable for the treatment of parietal affections, in particular incisional and inquinal hernias, through open approach.

The Parietex™ hydrophilic 2-dimensional mesh, hydrophilic 3dimensional mesh and hydrophilic anatomical mesh has been adapted for various techniques of abdominal repair. The rectangular mesh is designed for the repair of inguinal and incisional hernias in a pre-peritoneal or pre-muscular approach. The pre-cut and slit mesh is suitable for the repair of inguinal hernias via anterior approach using the tension free technique. The folding mesh (in two-dimensional textile) is designed for the repair of direct or indirect inguinal hernias through a laparoscopic approach (trans-abdominal or pre-peritoneal or totally extra-peritoneal). Some codes have a slit for the passage of the cord.

The anatomical mesh is mainly designed for the repair of inquinal hernias via laparoscopic or posterior open procedures and is available for the left and/or right side.

{9}------------------------------------------------

K220586, Page 4 of 6

|                                                                  | Versatex™ Monofilament Mesh<br>Versatex™ Monofilament Mesh is made of macroporous three-<br>dimensional monofilament polyester textile. Largest<br>sizes include a green dyed monofilament polyester (D&C Green<br>No. 6) marking that is positioned in the center of the textile to<br>help center and orient the mesh. |
|------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use                                                     | Symbotex™ Composite Mesh is intended for the reinforcement<br>of abdominal wall soft tissue where a weakness exists.                                                                                                                                                                                                     |
|                                                                  | ProGrip™ Self-Gripping Polyester Mesh is intended for use in<br>reinforcement of abdominal wall soft tissues where a weakness<br>exists.                                                                                                                                                                                 |
|                                                                  | Parietex™ Hydrophilic 2D 3D Anatomical Mesh is intended for<br>reinforcement of abdominal wall soft tissue where a weakness<br>exists.                                                                                                                                                                                   |
|                                                                  | Versatex™ Monofilament Mesh is intended for reinforcement of<br>abdominal wall soft tissue where a weakness exists.                                                                                                                                                                                                      |
| Indications for use                                              | Symbotex™ Composite Mesh is intended for the reinforcement<br>of abdominal wall soft tissue where a weakness exists, in<br>procedures involving primary abdominal wall and incisional<br>hernia surgeries.                                                                                                               |
|                                                                  | ProGrip™ Self-Gripping Polyester Mesh is intended for use in<br>reinforcement of abdominal wall soft tissues where a weakness<br>exists in procedure involving inguinal and incisional hernias<br>repair.                                                                                                                |
|                                                                  | Parietex™ Hydrophilic 2D 3D Anatomical Mesh is intended for<br>reinforcement of abdominal wall soft tissue where a weakness<br>exists, in procedures involving inguinal and incisional hernia<br>repairs.                                                                                                                |
|                                                                  | Versatex™ Monofilament Mesh is intended for reinforcement of<br>abdominal wall soft tissue where a weakness exists, in<br>procedures involving abdominal wall hernia repairs.                                                                                                                                            |
| Summary comparing<br>the technological<br>characteristics of the | In order to claim compliance with new European Medical Device<br>Regulation 2017/745 (MDR), and, as part of continuous<br>improvement, for improved readability and clarity and alignment                                                                                                                                |

{10}------------------------------------------------

| subject and predicate<br>device: | with the state-of-the-art for surgical mesh implants,<br>modifications of the predicate devices Symbotex™ Composite<br>Mesh (K142908), ProGrip™ Self-Gripping Polyester Mesh<br>(K142900), Parietex™ Hydrophilic 2D 3D Anatomical Mesh<br>(K173796) and Versatex™ Monofilament Mesh (K150091).<br>were made. These modifications have no impact on the<br>substantially equivalence between the subject devices and the<br>predicate devices in terms of indications and design for the<br>following technological characteristics:<br>• Indications<br>• Labelling (labels)<br>• Design<br>• Raw materials<br>• Packaging<br>• Biocompatibility<br>• Stability<br>• Sterilization |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Performance data:                | The purpose of this Special 510(k) is to notify the Agency of the<br>update of Instructions of Use for the currently marketed devices<br>Symbotex™ Composite Mesh (K131969 and K142908),<br>ProGrip™ Self-Gripping Polyester Mesh (K081050 and<br>K142900), Parietex™ Hydrophilic 2D 3D Anatomical Mesh<br>(K982532 and K173796) and Versatex™ Monofilament Mesh<br>(K150091). These changes are made to comply with the<br>requirements of the new European Regulation (EU) 2017/745<br>and as part of continuous improvement initiatives and IFUs<br>portfolio homogenization. The changes made to the Instructions<br>for Use are summarized below:                             |
|                                  | • Indications for use are reworded for Versatex™<br>Monofilament Mesh: minor wording changes and indication narrowed down as part of continuous improvement initiatives and IFUs portfolio homogenization. The indications of Symbotex™ Composite Mesh, ProGrip™ Self-Gripping Polyester Mesh and Parietex™ Hydrophilic 2D 3D Anatomical Mesh were not modified.<br>• For the four products, contraindications were reworded as part of continuous improvement initiatives and IFUs portfolio homogenization. They do not lead to new underlying risks.<br>• For the four products, possible complications were added or deleted as part of continuous improvement initiatives     |

{11}------------------------------------------------

|             | and IFUs portfolio homogenization. They do not lead to             new underlying risks.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|-------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|             | <ul><li>For the four products, additional information on trocar                     compatibility and fixation means compatibility were added                     as requested by the European Medical Devices                     Regulation (Regulation (EU) 2017/745).                 </li><li>For the four products, minor wording improvements have                     been performed in the storage section.                 </li><li>For the four products, a "Follow up" and "Magnetic                     Resonance Imaging (MRI) compatibility" section was                     added as requested by the European Medical Devices                     Regulation (Regulation (EU) 2017/745).                 </li></ul> |
|             | A Patient Implant Card (PIC) which is a small portable card is             supplied with each permanent implantable product.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|             | The following performance data is provided in support of the             substantial equivalence determination:             In vitro (bench) tests have been performed to evaluate the             trocar compatibility of the subject devices. The results             demonstrate that the subject devices successfully met the             established acceptance criteria.                                                                                                                                                                                                                                                                                                                                                       |
|             | The update of labeling for the currently marketed devices does             not affect the safety and effectiveness of the products.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Conclusion: | The subject devices are substantially equivalent to the predicate             devices Symbotex™ Composite Mesh (K142908), ProGrip™             Self-Gripping Polyester Mesh (K142900), Parietex™             Hydrophilic 2D 3D Anatomical Mesh (K173796) and Versatex™             Monofilament Mesh (K150091).                                                                                                                                                                                                                                                                                                                                                                                                                      |

---

**Source:** [https://fda-staging.innolitics.com/device/K220586](https://fda-staging.innolitics.com/device/K220586)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).