← Product Code [NUH](/productcode/NUH) · K220573

# Noodle (K220573)

_Pourang Bral, Dds, LLC · NUH · Jul 9, 2022 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K220573

## Device Facts

- **Applicant:** Pourang Bral, Dds, LLC
- **Product Code:** [NUH](/productcode/NUH.md)
- **Decision Date:** Jul 9, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5890
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

Noodle is intended to reduce the pain of hypodermic injections into the upper extremities (leg), buttocks, abdomen, and deltoid areas in patients 18 to 70 years of age. The Noodle device may be used in the home by patients who self-administer injections or in a medical setting by professional healthcare providers. Body parts intended to be conditioned by the Noodle device are sites that are commonly the target of injections such as the upper and lower extremities (arms and legs), buttocks, abdomen, and deltoid and is not intended to be used on smaller body parts such as the hands, feet, head (including ears, nose and around the eyes), the neck, and the abdominal area in pregnant women.

## Device Story

Noodle is a battery-operated handheld device combining TENS stimulation and mechanical tapping to temporarily anesthetize injection sites. Input: 9V battery power; Output: electrical stimulation via hydrogels and mechanical tapping via a rotating polyurethane sponge stick. The sponge is impregnated with 70% isopropanol for simultaneous site disinfection. Used in home or clinical settings by patients or healthcare providers. Operation: device runs for a fixed 25-second cycle before automatic shut-off. Healthcare providers/patients use the device to precondition skin prior to injection; the resulting temporary numbness reduces perceived injection pain. Benefits include non-chemical pain management for routine injections.

## Clinical Evidence

Multi-site, sham-controlled, double-blind trial with 18 patients. Subjects received two needle sticks, preconditioned with either Noodle or a sham device. Primary endpoints: peak and average VAS pain scores. Results: 77% of subjects reported >=3 point reduction in VAS (approx. 30% reduction) with Noodle. Average VAS was 35% lower with Noodle compared to sham. No adverse events reported.

## Technological Characteristics

Handheld TENS stimulator; 9V battery-powered. Materials: Polypropylene enclosure, polyurethane sponge. Sensing/Actuation: TENS via hydrogel electrodes; mechanical tapping via motor-driven sponge. Dimensions: 120x45x47mm. Weight: 119g. Connectivity: None. Sterilization: N/A (disinfection via 70% isopropanol). Software: Firmware-controlled, fixed 25-second cycle, automatic shut-off. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10, IEC 60601-1-11.

## Regulatory Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

## Predicate Devices

- OTC Electrical Stimulator LT3060 ([K130802](/device/K130802.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 9, 2022

Pourang Bral, DDS, LLC President 155 Albion Street Passaic. New Jersey 07055

Re: K220573

Trade/Device Name: Noodle Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: NUH Dated: February 15, 2022 Received: February 28, 2022

Dear Dr. Bral:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Pamela Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K220573

Device Name Noodle

Indications for Use (Describe)

Noodle is intended to reduce the pain of hypodermic injections into the upper extremities (leg), buttocks, abdomen, and deltoid areas in patients 18 to 70 years of age.

The Noodle device may be used in the home by patients who self-administer injections or in a medical setting by professional healthcare providers.

Body parts intended to be conditioned by the Noodle device are sites that are commonly the target of injections such as the upper and lower extremities (arms and legs), buttocks, abdomen, and deltoid and is not intended to be used on smaller body parts such as the hands, feet, head (including ears, nose and around the eyes), the neck, and the abdominal area in pregnant women.

Type of Use (Select one or both, as applicable)

| Residential Use (Per 24 CFR 982.1 Subpart E)  | <span style="text-decoration: overline;">☒</span> |
|-----------------------------------------------|---------------------------------------------------|
| Homeownership Option (Per 24 CFR 982.625-641) | ☐                                                 |

| | Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

Type of submission: Traditional 510(k)

Date: July 9, 2022 Submitter: Pourang Bral, DDS, LLC Address 155 Albion Street, Passaic, NJ 07055 Phone: 201 264-4862 Contact: Pourang Bral Email: Pourangb@gmail.com

### Identification of the Subject Device

| Device name                 | Noodle                                                                 |
|-----------------------------|------------------------------------------------------------------------|
| Common or usual name        | Electrical and mechanical stimulator for injection site pain reduction |
| Classification product code | NUH                                                                    |
| Device classification       | II                                                                     |
| Regulation number           | 882.5890                                                               |
| Regulation description      | Transcutaneous electrical nerve stimulator for pain relief.            |
| Review panel                | Neurology                                                              |

This is an application for a 510K approval of a device called Noodle that uses electrical and physical stimulation on the skin to temporarily anesthetize an injection to reduce the pain of the injection without the use of chemicals.

Noodle is a battery-operated handheld device that includes a TENS unit and a motor operated by a 9V commercial battery. The classification product code for Noodle is a Class II Device: § 882.5890 - Transcutaneous electrical nerve stimulator for pain relief. Once it is activated, it operates for 25 seconds and shuts off automatically.

### ldentification of the predicate device

Predicate Device K130802

| Device name                 | OTC Electrical Stimulator LT3060           |
|-----------------------------|--------------------------------------------|
| Common or usual name        | Transcutaneous Electrical Nerve Stimulator |
| Classification product code | NUH, NGX                                   |
| Device classification       | II                                         |

{4}------------------------------------------------

| Regulation number      | 21 CFR 882.5890                                            |
|------------------------|------------------------------------------------------------|
| Regulation description | Transcutaneous Electrical Nerve Stimulator for Pain Relief |
| Review panel           | Neurology                                                  |

The predicate device is the OTC Electrical Stimulator LT3060, with the stated indication for use of Temporary relief of pain associated with sore and aching muscles in the shoulder, waist, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities. It has a 510(K) number of K130802, a Product Code NUH, and a Classification Name of Stimulator, Nerve, Transcutaneous, over the counter.

### Indications for use of the subject device

Noodle is intended to reduce injection-site pain of hypodermic injections into the upper extremities (arms), lower extremities (leg), buttocks, abdomen, and deltoid areas in patients 18 to 70 years of age.

The Noodle device may be used in the home by patients who self-administer injections or in a medical setting by professional healthcare providers.

Body parts intended to be conditioned by the Noodle device are sites that are commonly the target of injections such as the upper and lower extremities (arms and legs), buttocks, abdomen, and deltoid and is not intended to be used on smaller body parts such as the hands, feet, head (including ears, nose and around the eyes), the neck, and the abdominal area in pregnant women.

### Description of the Device

Noodle consists of a TENS unit and a small motor that causes a polyurethane sponge stick to rotate and tap the skin. The combination of the TENS application and tapping on the skin for 25 seconds numbs the skin temporarily for a subsequent injection. The sponge stick is impregnated in 70% isopropanol and disinfects the injection site at the same time it taps it.

The device body has 5 pieces, the enclosure, the bottom, the battery door, and the receptacle. Noodle further includes a PCB that is controlled by a firmware. The PCB supplies TENS stimulation that is transferred to the skin by a pair of hydrogels that come in electrical contact with the skin around the injection site.

### Statement of Conformity

List of FDA recognized voluntary consensus standards cited in this submission.

| IEC 60601-1    |
|----------------|
| IEC 60601-1-2  |
| IEC 60601-2-10 |

{5}------------------------------------------------

IEC 60601-1-11

## Discussion of non-clinical tests

- ANSI AAMI ES60601-1: 2005/(R)2012 And A1:2012, C1:2009((R)2012 And A2:2010/(R)2012 (Consolidated i. Text) Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance (IEC 60601-1:2005, MOD)
- ii. ANSI AAMI IEC 60601-1-2:2014, Medical Electrical Equipment -- Part 1-2: General Requirements for Basic Safety and Essential Performance -- Collateral Standard: Disturbances -- Requirements and Test
- iii. IEC 60601-2-10 Edition 2.1 2016-04, Medical Electrical Equipment Part 2-10: Particular Requirements for The Basic Safety and Essential Performance of Nerve and Muscle Stimulators
- iv. IEC 60601-1-11 Edition 2.0 2015-01. Medical Electrical Equipment Part 1-11: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment (General II (ES/EMC))

Noodle has passed all these non-clinical tests.

## Discussion of clinical tests

The clinical performance data supporting the safety and effectiveness of the Noodle device for reducing the pain of hypodermic injections into the extremities, buttocks, abdomen, and deltoid (as described in the Indications for Use statement) was based on a multi-site sham-controlled double-blind causal-comparative trial using the Noodle device in 18 patients at 3 sites (one doctor's office and two pharmacies). The sham device was identical to the Noodle but did not deliver stimulation.

The 18 patients were divided into two groups: 9 who received treatment with the Noodle device first and 9 getting the sham first. Each subject gave themselves two needle sticks. Prior to each needle stick, they preconditioned the needle stick site either by the Noodle or a sham. The VAS pain scale was used to assess the level of pain. The sponsor assessed the improvement in acute pain from baseline before the needlestick to after, in comparison to the sham using two metrics: the peak pain score (peak VAS) and average pain score (average VAS).

77% of subjects (14 of 18) reported a reduction in pain scores of at least 3 or greater on the VAS scale, about 30% pain reduction, when the injection site was preconditioned with Noodle. No adverse events were reported in the 18 subjects. Average VAS was lower with Noodle than with the sham device by 35%. The difference seen in the clinical results was considered to provide clinical benefit to patients given that a sham-controlled, double-blind trial and the study results did not indicate any significant adverse effects or safety risks for this device. Hence, the results were considered to provide clinical benefit in reducing the pain of hypodermic injections into the extremities, buttocks, abdomen, and deltoid.

### Substantial Equivalence Determination

Both Noodle and the predicate device use TENS stimulation to achieve their desired effects.

{6}------------------------------------------------

A side-by-side comparison of the technological characteristics of Noodle and the cited predicate device is included in the table below.

|                                       | New Device              | Predicate Device                         | Substantial Equivalence<br>Determination                  |
|---------------------------------------|-------------------------|------------------------------------------|-----------------------------------------------------------|
| 1. 510(k) Number                      |                         | K130802                                  |                                                           |
| 2. Device Name,<br>Model              | Noodle                  | OTC Electrical Stimulator<br>LT3060      |                                                           |
| 3. Manufacturer                       | Painless World          | Shenzhen Dongdixin<br>Technology Co. Ltd |                                                           |
| 4. Power Source(s)<br>Battery         | (9V - user replaceable) | 9V Battery                               | Same                                                      |
| - Method of Line Current<br>Isolation | N/A                     | N/A                                      | Same                                                      |
| - Patient Leakage Current             | N/A                     |                                          |                                                           |
| - Normal condition                    | N/A                     | 0.61uA                                   | This parameter is not<br>applicable to the new<br>device. |
| - Single fault condition              | N/A                     | 0.68uA                                   | This parameter is not<br>applicable to the new<br>device. |
| 5. Number of Output<br>Modes          | 1                       | 1                                        | Same                                                      |

| 6. Number of Output<br>Channels | 1   | Alternating | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device. |
|---------------------------------|-----|-------------|----------------------------------------------------------------------------------------------|
| Synchronous or<br>Alternating?  | N/A | Alternating | This parameter is not<br>applicable to the new<br>device.                                    |

{7}------------------------------------------------

| - Method of Channel<br>Isolation                | N/A               | By electrical circuit and<br>software | This parameter is not<br>applicable to the new<br>device.                                    |
|-------------------------------------------------|-------------------|---------------------------------------|----------------------------------------------------------------------------------------------|
| 7. Regulated Current or<br>Regulated Voltage?   | Regulated Voltage | Regulated Current                     | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device. |
| 8.Software/Firmware/<br>Microprocessor Control? | Yes               | Yes                                   | Same                                                                                         |
| 9. Automatic Overload<br>Trip?                  | Yes               | Yes                                   | Same                                                                                         |
| 10. Automatic<br>No-Load Trip?                  | No                | Yes                                   | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device. |
| 11. Automatic Shut<br>Off?                      | Yes               | Yes                                   | Same                                                                                         |
| 12. Patient Override<br>Control?                | No                | Yes                                   | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device. |
| 13. Indicator Display:                          |                   |                                       |                                                                                              |
| - On/Off Status?                                | Yes               | Yes                                   | Same                                                                                         |
| - Low Battery?                                  | No                | Yes                                   | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device. |

{8}------------------------------------------------

| 14. Timer Range                                                                                 | 25 seconds                                                       | 1-60 minutes                                                     | New device is several<br>times safer than the<br>predicate because of the<br>short usage time. |
|-------------------------------------------------------------------------------------------------|------------------------------------------------------------------|------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| 15. Compliance with<br>Voluntary Standards? (If<br>yes, specify)                                | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-10<br>IEC 60601-1-11 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-10<br>ISO 10993-5/10 | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device.   |
| 16. Compliance* with 21<br>CFR 898? Yes /<br>No                                                 | Yes                                                              | Yes                                                              | Same                                                                                           |
| 17. Weight                                                                                      | 119 grams, including a<br>9V battery                             | 128 grams<br>(including<br>batteries)                            | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device.   |
| 18. Dimensions (in.)<br>[W x H x D]                                                             | 120 x 45 x 47                                                    | 117 x 60 x 34 mm                                                 | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device.   |
| 19. Housing Materials<br>and Construction                                                       | Polypropylene                                                    | ABS                                                              | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device.   |
| Waveform (e.g.,<br>pulsed monophasic,<br>biphasic)                                              | Monophasic                                                       | Biphasic                                                         | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device.   |
| Shape (e.g.,<br>rectangular, spike,<br>rectified sinusoidal)                                    | Rectangular                                                      | Square                                                           | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device.   |
| Maximum Output<br>Voltage@ 500Ω                                                                 | 28.6V                                                            | 96+-20%<br>(48+-20% (Vp))                                        | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device.   |
| Maximum Output<br>Voltage @ 2kΩ                                                                 | 94.3V                                                            | 228+-20%<br>(114+-20% (Vp))                                      | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device.   |
| Maximum Output<br>Voltage @ 10kΩ                                                                | 125.7V                                                           | 230+-20%<br>(115+-20% (Vp))                                      | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device.   |
| Maximum Output<br>Current @ 500Ω                                                                | 58mA                                                             | 96+-20%                                                          | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device.   |
| Maximum Output<br>Current @ 2kΩ                                                                 | 50mA                                                             | 57+-20%                                                          | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device.   |
| Maximum Output<br>Current @ 10kΩ                                                                | 14mA                                                             | 11.5 +-20%                                                       | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device.   |
| Pulse Width                                                                                     | 9us - 113us                                                      | 50 - 300uS                                                       | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device.   |
| Frequency                                                                                       | 150Hz                                                            | 1-150Hz                                                          | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device.   |
| Net Charge (mC per<br>pulse) @ 500Ω                                                             | 7uC                                                              | 0uC@500Ω                                                         | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device.   |
| Maximum Phase<br>Charge, (mC) @<br>500Ω                                                         | 7uC                                                              | 0.0288@500Ω                                                      | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device.   |
| Maximum Current<br>Density, (mA/cm²) @<br>500Ω                                                  | 4.23mA/cm2                                                       | 1.15@500Ω                                                        | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device.   |
| Maximum Power<br>Density, (mW/cm²) @<br>500Ω                                                    | 120mW/cm2                                                        | Unavailable                                                      |                                                                                                |
| Max. Average Power<br>Density, mW/cm²<br>(using smaller<br>electrode conductive<br>surface area | 2.03                                                             | 0.373@500Ω                                                       |                                                                                                |

{9}------------------------------------------------

{10}------------------------------------------------

The electronic characteristics listed in the table above were designed to those of the predicate. They were then measured and confirmed in the lab. The power source in Noodle and predicate are the same. In lines 5 and 6, the output modes and output channels in the predicate are more numerous. The predicate offers more output modes, and the user opts for one of them. Noodle offers only one output mode and one output channel that are offered by the predicate.

In line 14, the time range for Noodle is significantly less than the predicate.

Similarly in lines 15 and 16, the predicate offers more choices in waveform and in pulse width range than Noodle. Noodle is preset to offer only one of those options offered by the predicate. Line 22 shows what the Noodle enclosure is made of. This material meets all applicable industrial standards. Conclusion

{11}------------------------------------------------

All nonclinical tests recommended by the FDA were performed and found Noodle to meet all the requirements of the stated standards. In addition, Noodle was tested clinically and was shown to effectively and safely reduce pain of hypodermic injections into the upper extremities (arms), lower extremities (leg), buttocks, abdomen, and deltoid areas in patients 18 to 70 years of age.

---

**Source:** [https://fda-staging.innolitics.com/device/K220573](https://fda-staging.innolitics.com/device/K220573)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com