Noodle is intended to reduce the pain of hypodermic injections into the upper extremities (leg), buttocks, abdomen, and deltoid areas in patients 18 to 70 years of age. The Noodle device may be used in the home by patients who self-administer injections or in a medical setting by professional healthcare providers. Body parts intended to be conditioned by the Noodle device are sites that are commonly the target of injections such as the upper and lower extremities (arms and legs), buttocks, abdomen, and deltoid and is not intended to be used on smaller body parts such as the hands, feet, head (including ears, nose and around the eyes), the neck, and the abdominal area in pregnant women.
Device Story
Noodle is a battery-operated handheld device combining TENS stimulation and mechanical tapping to temporarily anesthetize injection sites. Input: 9V battery power; Output: electrical stimulation via hydrogels and mechanical tapping via a rotating polyurethane sponge stick. The sponge is impregnated with 70% isopropanol for simultaneous site disinfection. Used in home or clinical settings by patients or healthcare providers. Operation: device runs for a fixed 25-second cycle before automatic shut-off. Healthcare providers/patients use the device to precondition skin prior to injection; the resulting temporary numbness reduces perceived injection pain. Benefits include non-chemical pain management for routine injections.
Clinical Evidence
Multi-site, sham-controlled, double-blind trial with 18 patients. Subjects received two needle sticks, preconditioned with either Noodle or a sham device. Primary endpoints: peak and average VAS pain scores. Results: 77% of subjects reported >=3 point reduction in VAS (approx. 30% reduction) with Noodle. Average VAS was 35% lower with Noodle compared to sham. No adverse events reported.
Indicated for pain reduction during hypodermic injections in patients 18-70 years old. Applicable sites: upper/lower extremities, buttocks, abdomen, deltoid. Contraindicated for hands, feet, head, neck, and abdomen in pregnant women.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 9, 2022
Pourang Bral, DDS, LLC President 155 Albion Street Passaic. New Jersey 07055
Re: K220573
Trade/Device Name: Noodle Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: NUH Dated: February 15, 2022 Received: February 28, 2022
Dear Dr. Bral:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Pamela Scott Assistant Director DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220573
Device Name Noodle
Indications for Use (Describe)
Noodle is intended to reduce the pain of hypodermic injections into the upper extremities (leg), buttocks, abdomen, and deltoid areas in patients 18 to 70 years of age.
The Noodle device may be used in the home by patients who self-administer injections or in a medical setting by professional healthcare providers.
Body parts intended to be conditioned by the Noodle device are sites that are commonly the target of injections such as the upper and lower extremities (arms and legs), buttocks, abdomen, and deltoid and is not intended to be used on smaller body parts such as the hands, feet, head (including ears, nose and around the eyes), the neck, and the abdominal area in pregnant women.
Type of Use (Select one or both, as applicable)
| Residential Use (Per 24 CFR 982.1 Subpart E) | <span style="text-decoration: overline;">☒</span> |
|-----------------------------------------------|---------------------------------------------------|
| Homeownership Option (Per 24 CFR 982.625-641) | ☐ |
| | Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
Type of submission: Traditional 510(k)
Date: July 9, 2022 Submitter: Pourang Bral, DDS, LLC Address 155 Albion Street, Passaic, NJ 07055 Phone: 201 264-4862 Contact: Pourang Bral Email: Pourangb@gmail.com
### Identification of the Subject Device
| Device name | Noodle |
|-----------------------------|------------------------------------------------------------------------|
| Common or usual name | Electrical and mechanical stimulator for injection site pain reduction |
| Classification product code | NUH |
| Device classification | II |
| Regulation number | 882.5890 |
| Regulation description | Transcutaneous electrical nerve stimulator for pain relief. |
| Review panel | Neurology |
This is an application for a 510K approval of a device called Noodle that uses electrical and physical stimulation on the skin to temporarily anesthetize an injection to reduce the pain of the injection without the use of chemicals.
Noodle is a battery-operated handheld device that includes a TENS unit and a motor operated by a 9V commercial battery. The classification product code for Noodle is a Class II Device: § 882.5890 - Transcutaneous electrical nerve stimulator for pain relief. Once it is activated, it operates for 25 seconds and shuts off automatically.
### ldentification of the predicate device
Predicate Device K130802
| Device name | OTC Electrical Stimulator LT3060 |
|-----------------------------|--------------------------------------------|
| Common or usual name | Transcutaneous Electrical Nerve Stimulator |
| Classification product code | NUH, NGX |
| Device classification | II |
{4}------------------------------------------------
| Regulation number | 21 CFR 882.5890 |
|------------------------|------------------------------------------------------------|
| Regulation description | Transcutaneous Electrical Nerve Stimulator for Pain Relief |
| Review panel | Neurology |
The predicate device is the OTC Electrical Stimulator LT3060, with the stated indication for use of Temporary relief of pain associated with sore and aching muscles in the shoulder, waist, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities. It has a 510(K) number of K130802, a Product Code NUH, and a Classification Name of Stimulator, Nerve, Transcutaneous, over the counter.
### Indications for use of the subject device
Noodle is intended to reduce injection-site pain of hypodermic injections into the upper extremities (arms), lower extremities (leg), buttocks, abdomen, and deltoid areas in patients 18 to 70 years of age.
The Noodle device may be used in the home by patients who self-administer injections or in a medical setting by professional healthcare providers.
Body parts intended to be conditioned by the Noodle device are sites that are commonly the target of injections such as the upper and lower extremities (arms and legs), buttocks, abdomen, and deltoid and is not intended to be used on smaller body parts such as the hands, feet, head (including ears, nose and around the eyes), the neck, and the abdominal area in pregnant women.
### Description of the Device
Noodle consists of a TENS unit and a small motor that causes a polyurethane sponge stick to rotate and tap the skin. The combination of the TENS application and tapping on the skin for 25 seconds numbs the skin temporarily for a subsequent injection. The sponge stick is impregnated in 70% isopropanol and disinfects the injection site at the same time it taps it.
The device body has 5 pieces, the enclosure, the bottom, the battery door, and the receptacle. Noodle further includes a PCB that is controlled by a firmware. The PCB supplies TENS stimulation that is transferred to the skin by a pair of hydrogels that come in electrical contact with the skin around the injection site.
### Statement of Conformity
List of FDA recognized voluntary consensus standards cited in this submission.
| IEC 60601-1 |
|----------------|
| IEC 60601-1-2 |
| IEC 60601-2-10 |
{5}------------------------------------------------
IEC 60601-1-11
## Discussion of non-clinical tests
- ANSI AAMI ES60601-1: 2005/(R)2012 And A1:2012, C1:2009((R)2012 And A2:2010/(R)2012 (Consolidated i. Text) Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance (IEC 60601-1:2005, MOD)
- ii. ANSI AAMI IEC 60601-1-2:2014, Medical Electrical Equipment -- Part 1-2: General Requirements for Basic Safety and Essential Performance -- Collateral Standard: Disturbances -- Requirements and Test
- iii. IEC 60601-2-10 Edition 2.1 2016-04, Medical Electrical Equipment Part 2-10: Particular Requirements for The Basic Safety and Essential Performance of Nerve and Muscle Stimulators
- iv. IEC 60601-1-11 Edition 2.0 2015-01. Medical Electrical Equipment Part 1-11: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment (General II (ES/EMC))
Noodle has passed all these non-clinical tests.
## Discussion of clinical tests
The clinical performance data supporting the safety and effectiveness of the Noodle device for reducing the pain of hypodermic injections into the extremities, buttocks, abdomen, and deltoid (as described in the Indications for Use statement) was based on a multi-site sham-controlled double-blind causal-comparative trial using the Noodle device in 18 patients at 3 sites (one doctor's office and two pharmacies). The sham device was identical to the Noodle but did not deliver stimulation.
The 18 patients were divided into two groups: 9 who received treatment with the Noodle device first and 9 getting the sham first. Each subject gave themselves two needle sticks. Prior to each needle stick, they preconditioned the needle stick site either by the Noodle or a sham. The VAS pain scale was used to assess the level of pain. The sponsor assessed the improvement in acute pain from baseline before the needlestick to after, in comparison to the sham using two metrics: the peak pain score (peak VAS) and average pain score (average VAS).
77% of subjects (14 of 18) reported a reduction in pain scores of at least 3 or greater on the VAS scale, about 30% pain reduction, when the injection site was preconditioned with Noodle. No adverse events were reported in the 18 subjects. Average VAS was lower with Noodle than with the sham device by 35%. The difference seen in the clinical results was considered to provide clinical benefit to patients given that a sham-controlled, double-blind trial and the study results did not indicate any significant adverse effects or safety risks for this device. Hence, the results were considered to provide clinical benefit in reducing the pain of hypodermic injections into the extremities, buttocks, abdomen, and deltoid.
### Substantial Equivalence Determination
Both Noodle and the predicate device use TENS stimulation to achieve their desired effects.
{6}------------------------------------------------
A side-by-side comparison of the technological characteristics of Noodle and the cited predicate device is included in the table below.
| | New Device | Predicate Device | Substantial Equivalence<br>Determination |
|---------------------------------------|-------------------------|------------------------------------------|-----------------------------------------------------------|
| 1. 510(k) Number | | K130802 | |
| 2. Device Name,<br>Model | Noodle | OTC Electrical Stimulator<br>LT3060 | |
| 3. Manufacturer | Painless World | Shenzhen Dongdixin<br>Technology Co. Ltd | |
| 4. Power Source(s)<br>Battery | (9V - user replaceable) | 9V Battery | Same |
| - Method of Line Current<br>Isolation | N/A | N/A | Same |
| - Patient Leakage Current | N/A | | |
| - Normal condition | N/A | 0.61uA | This parameter is not<br>applicable to the new<br>device. |
| - Single fault condition | N/A | 0.68uA | This parameter is not<br>applicable to the new<br>device. |
| 5. Number of Output<br>Modes | 1 | 1 | Same |
| 6. Number of Output<br>Channels | 1 | Alternating | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device. |
|---------------------------------|-----|-------------|----------------------------------------------------------------------------------------------|
| Synchronous or<br>Alternating? | N/A | Alternating | This parameter is not<br>applicable to the new<br>device. |
{7}------------------------------------------------
| - Method of Channel<br>Isolation | N/A | By electrical circuit and<br>software | This parameter is not<br>applicable to the new<br>device. |
|-------------------------------------------------|-------------------|---------------------------------------|----------------------------------------------------------------------------------------------|
| 7. Regulated Current or<br>Regulated Voltage? | Regulated Voltage | Regulated Current | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device. |
| 8.Software/Firmware/<br>Microprocessor Control? | Yes | Yes | Same |
| 9. Automatic Overload<br>Trip? | Yes | Yes | Same |
| 10. Automatic<br>No-Load Trip? | No | Yes | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device. |
| 11. Automatic Shut<br>Off? | Yes | Yes | Same |
| 12. Patient Override<br>Control? | No | Yes | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device. |
| 13. Indicator Display: | | | |
| - On/Off Status? | Yes | Yes | Same |
| - Low Battery? | No | Yes | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device. |
{8}------------------------------------------------
| 14. Timer Range | 25 seconds | 1-60 minutes | New device is several<br>times safer than the<br>predicate because of the<br>short usage time. |
|-------------------------------------------------------------------------------------------------|------------------------------------------------------------------|------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| 15. Compliance with<br>Voluntary Standards? (If<br>yes, specify) | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-10<br>IEC 60601-1-11 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-10<br>ISO 10993-5/10 | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device. |
| 16. Compliance* with 21<br>CFR 898? Yes /<br>No | Yes | Yes | Same |
| 17. Weight | 119 grams, including a<br>9V battery | 128 grams<br>(including<br>batteries) | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device. |
| 18. Dimensions (in.)<br>[W x H x D] | 120 x 45 x 47 | 117 x 60 x 34 mm | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device. |
| 19. Housing Materials<br>and Construction | Polypropylene | ABS | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device. |
| Waveform (e.g.,<br>pulsed monophasic,<br>biphasic) | Monophasic | Biphasic | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device. |
| Shape (e.g.,<br>rectangular, spike,<br>rectified sinusoidal) | Rectangular | Square | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device. |
| Maximum Output<br>Voltage@ 500Ω | 28.6V | 96+-20%<br>(48+-20% (Vp)) | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device. |
| Maximum Output<br>Voltage @ 2kΩ | 94.3V | 228+-20%<br>(114+-20% (Vp)) | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device. |
| Maximum Output<br>Voltage @ 10kΩ | 125.7V | 230+-20%<br>(115+-20% (Vp)) | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device. |
| Maximum Output<br>Current @ 500Ω | 58mA | 96+-20% | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device. |
| Maximum Output<br>Current @ 2kΩ | 50mA | 57+-20% | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device. |
| Maximum Output<br>Current @ 10kΩ | 14mA | 11.5 +-20% | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device. |
| Pulse Width | 9us - 113us | 50 - 300uS | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device. |
| Frequency | 150Hz | 1-150Hz | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device. |
| Net Charge (mC per<br>pulse) @ 500Ω | 7uC | 0uC@500Ω | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device. |
| Maximum Phase<br>Charge, (mC) @<br>500Ω | 7uC | 0.0288@500Ω | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device. |
| Maximum Current<br>Density, (mA/cm²) @<br>500Ω | 4.23mA/cm2 | 1.15@500Ω | Different but does not<br>adversely impact safety<br>and effectiveness of the<br>new device. |
| Maximum Power<br>Density, (mW/cm²) @<br>500Ω | 120mW/cm2 | Unavailable | |
| Max. Average Power<br>Density, mW/cm²<br>(using smaller<br>electrode conductive<br>surface area | 2.03 | 0.373@500Ω | |
{9}------------------------------------------------
{10}------------------------------------------------
The electronic characteristics listed in the table above were designed to those of the predicate. They were then measured and confirmed in the lab. The power source in Noodle and predicate are the same. In lines 5 and 6, the output modes and output channels in the predicate are more numerous. The predicate offers more output modes, and the user opts for one of them. Noodle offers only one output mode and one output channel that are offered by the predicate.
In line 14, the time range for Noodle is significantly less than the predicate.
Similarly in lines 15 and 16, the predicate offers more choices in waveform and in pulse width range than Noodle. Noodle is preset to offer only one of those options offered by the predicate. Line 22 shows what the Noodle enclosure is made of. This material meets all applicable industrial standards. Conclusion
{11}------------------------------------------------
All nonclinical tests recommended by the FDA were performed and found Noodle to meet all the requirements of the stated standards. In addition, Noodle was tested clinically and was shown to effectively and safely reduce pain of hypodermic injections into the upper extremities (arms), lower extremities (leg), buttocks, abdomen, and deltoid areas in patients 18 to 70 years of age.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.