Easi-Vue™ embolic microspheres System is intended for embolization of arteriovenous malformations and hypervasular tumors.
Device Story
Easi-Vue embolic microspheres System consists of radiopaque, non-compressible, non-resorbable glass microspheres (tantalum-barium-boron-sodium-silicon oxide glass) provided with an administration kit and refill syringe. Used by physicians in clinical settings for targeted vascular embolization. Microspheres are delivered via catheter into target arteries to mechanically occlude blood flow to hypervascular tumors or arteriovenous malformations. Unlike compressible polymer-based predicates, these glass spheres are non-deformable. Radiopacity allows for visualization during delivery without requiring additional contrast media. Benefits include controlled, targeted vessel occlusion and potential reduction in blood flow to target tissues.
Clinical Evidence
No human clinical data provided. Evidence consists of bench testing (simulated use, MR compatibility, performance verification) and a GLP animal study. Animal study confirmed ease of delivery and effective arterial occlusion with no clinically significant pathological or systemic abnormalities. Biocompatibility testing performed per ISO 10993.
Technological Characteristics
Materials: Tantalum-barium-boron-sodium-silicon oxide glass microspheres in saline. Sizes: 50, 100, 150 microns. Principle: Mechanical occlusion via catheter delivery. Radiopaque, non-compressible, non-resorbable. MR Safe. Sterilization: Gamma radiation (SAL 10^-6). Packaging: Tyvek pouch. Compatible with catheters (min ID 0.021", length 110-150 cm).
Indications for Use
Indicated for embolization of arteriovenous malformations and hypervascular tumors in patients requiring vascular occlusion.
Regulatory Classification
Identification
A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 7, 2022
ABK Biomedical Inc Brandi Woods Director, Regulatory Affairs 155 Chain Lake Drive Unit 32 Halifax, NS B3S 1B3 Canada
Re: K220567
Trade/Device Name: Easi-Vue embolic microspheres System Regulation Number: 21 CFR 870.3300 Regulation Name: Vascular Embolization Device Regulatory Class: Class II Product Code: KRD Dated: August 18, 2022 Received: August 19, 2022
Dear Brandi Woods:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm = identifies = combination = product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical
{1}------------------------------------------------
device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-DICE website assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Misti Malone Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220567
Device Name Easi-Vue™ embolic microspheres System
Indications for Use (Describe)
Easi-Vue™ embolic microspheres System is intended for embolization of arteriovenous malformations and hypervasular tumors.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K220567 PAGE 1 OF 4
#### 1 510(k) Summary
The following information is provided as required by 21 CFR § 807.92 and the Safe Medical Devices Act 1990.
#### 1.1 General Information
| Company: | ABK Biomedical<br>155 Chain Lake Drive<br>Unit 32<br>Halifax Nova Scotia B3S 1B3<br>Canada |
|----------|--------------------------------------------------------------------------------------------|
| Contact: | Brandi Woods<br>Director, Regulatory Affairs<br>919-619-6417<br>b.woods@abkbiomedical.com |
6 September 2022 Date of Summary:
#### 1.2 Device Name and Classification
| Proprietary Name: | Easi-Vue™ embolic microspheres System |
|----------------------|----------------------------------------------|
| Common Name: | Device, Vascular, For Promoting Embolization |
| Classification Name: | Vascular Embolization Device |
| Regulatory Class: | 2 |
| Regulation: | 870.3300 |
| Product Codes: | KRD |
#### 1.3 Predicate Device
| Proprietary Name: | Embosphere® Microspheres |
|-------------------|-------------------------------------------------------------------|
| Common Name: | Device, Vascular, For Promoting Embolization |
| 510(k) Number: | K021397 |
| Regulatory Class: | 2 |
| Regulation: | 870.3300 |
| Product Codes: | KRD |
| | Predicate device has not been subject to a design-related recall. |
#### 1.4 Indication for Use
Easi-Vue™ embolic microspheres System is intended for embolization of arteriovenous malformations and hypervascular tumors.
{4}------------------------------------------------
Easi-Vue™ embolic microspheres System
#### 1.5 Device Description
Easi-Vue embolic microspheres are biocompatible, radiopaque, non-compressible, non-resorbable glass microspheres. The Easi-Vue embolic microspheres are included with the Easi-Vue embolic microspheres System which is comprised of Easi-Vue embolic microspheres Administration Kit and the Easi-Vue embolic microspheres Refill Syringe. The device utilizes the Administration Kit for controlled and targeted delivery of imageable radiopaque microspheres. The Easi-Vue™ embolic microspheres System is offered in three sizes: 50 um, and 150 um to occlude various size arteries for the purpose of blocking blood flow to a target tissue (as summarized in Table 1).
The Easi-Vue™ embolic microsphere System is sterile, single use, and available only for prescription use. The device is compatible with a catheter with minimum inner diameter of 0.021'' and a length between 110-150 cm (as appropriate for the diameter of the intended treatment vessel).
| | Easi-Vue™ embolic microspheres<br>Administration Kit | Easi-Vue™ embolic microspheres<br>Refill Syringe |
|-------------|------------------------------------------------------|--------------------------------------------------|
| 50 µm size | EVA50 | EVR50 |
| 100 µm size | EVA100 | EVR100 |
| 150 µm size | EVA150 | EVR150 |
## Table 1: Product Availability
#### 1.6 Technological Comparison
Easi-Vue™ Embolic Microspheres System is substantially equivalent, for the purpose of this 510(k), to Embosphere® Microspheres (K021397), the predicate device. The subject device and predicate devices are similar in intended use, design, and principle of operation. This is based upon the comparison of the operational characteristics, product technical characteristics, performance and safety characteristics, sterility, and product handling. Easi-Vue™ embolic microspheres are made from radiopaque glass, which is a different material compared to the predicate, acrylic polymer sphere impregnated with porcine gelatin. The differences in material composition has shown no new questions of safety based upon bench, biocompatibility and animal testing.
Unlike the predicate Easi-Vue™ embolic microspheres are non-compressible, meaning they are non-deformable during delivery. In animals, Easi-Vue™ embolic microspheres has demonstrated equivalent effectiveness at occluding the target vessel, with no migration observed.
While differences in the technology characteristics exist between the Easi-Vue™ embolic microspheres System and the predicate device, these differences, which are detailed in Table 2, do not raise different questions of safety or effectiveness.
{5}------------------------------------------------
## Easi-Vue™ embolic microspheres System Table 2: Comparison to Predicate
| | Easi-Vue™ Embolic<br>Microspheres System | Embosphere® Microspheres | Comparison |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| Company | ABK Biomedical, Inc | Biosphere Medical, S.A. | |
| | Subject Device | Predicate Device | |
| 510(k) Number | K220567 | K021397 | |
| Product Code | KRD | KRD (also NAJ, NOY, & HCG) | Same |
| Intended use | Vascular Embolization Device | Vascular Embolization Device | Same |
| Indications for<br>Use | Easi-Vue™ Embolic<br>Microspheres System is intended<br>for use in embolization of<br>hypervascular tumors and<br>arteriovenous malformations | Embosphere Microspheres are indicated<br>for use in embolization of arteriovenous<br>malformations, hypervascular tumors,<br>and symptomatic uterine fibroids | Equivalent<br>Subject device<br>has more<br>restrictive<br>indication. |
| Mechanism of<br>Action | Mechanical occlusion | Mechanical occlusion | Same |
| Principle of<br>Operation | The microspheres are<br>administered into the patient's<br>artery via a catheter under<br>radiographic imaging | The microspheres are administered with<br>contrast medium into the patient's artery<br>via a catheter under radiographic imaging | Equivalent<br>Subject device<br>does not require<br>contrast medium<br>for imaging. |
| Material<br>Composition of<br>spheres | tantalum-barium-boron-sodium-<br>silicon oxide glass<br>Saline | Acrylic polymer sphere impregnated with<br>Porcine gelatin<br>Saline | No new safety<br>concern |
| Size Range | 50 ±20 microns<br>100 ±35 microns<br>150 ±50 microns | 40- 120 microns<br>100-300 microns<br>300-500 microns<br>500-700 microns<br>700-900 microns<br>900- 1200 microns | Equivalent |
| Physical<br>Characteristics | Biocompatible, radiopaque, non-<br>compressible, non-resorbable | Biocompatible, hydrophilic,<br>compressible, non-resorbable | No new safety<br>concern |
| Performance | Controlled, targeted embolization<br>at the desired level of vessel<br>occlusion | Controlled, targeted embolization at the<br>desired level of vessel occlusion | Same |
| Pyrogenicity | Non-pyrogenic | Non-pyrogenic | Same |
| MR Safe | MR Safe | MR Safe | Same |
| Sterility<br>(method) | Sterile (gamma) to SAL 10-6 | Sterile (steam) to SAL 10-6 | Equivalent |
| Quantity of<br>microspheres<br>per package | Each 1 mL syringe contains<br>approximately 1.25 g of<br>microspheres in saline. | Each 20 mL syringe or 8 mL glass vial<br>contains approximately 1.0 mL or 2.0 mL<br>of microspheres in saline. | Equivalent |
| Packaging | Tyvek pouch | A peel-away tray | Equivalent |
{6}------------------------------------------------
K220567
#### 1.7 Non-Clinical Performance Summary
Bench testing, biocompatibility, and non-clinical studies were conducted to support substantial equivalence of Easi-Vue embolic microspheres System to the predicated device. Easi-Vue embolic microspheres System was demonstrated to be biocompatible according to ISO 10993. Bench testing, including simulated use, MR compatibility and performance verification confirmed that Easi-Vue embolic microspheres System met all the acceptance criteria and performed as intended. A GLP animal study confirmed the ease of delivery, effectiveness of arterial occlusion and there were no clinically significant pathological abnormalities or systemic abnormalities. No new safety or effectiveness concerns were raised during testing.
#### 1.8 Conclusion
Based on the intended use, technological characteristics, performance characteristics and data included in this application, the Easi-Vue embolic microspheres System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.