K220560 · 3M · OMP · Jul 14, 2022 · General, Plastic Surgery
Device Facts
Record ID
K220560
Device Name
Dermatac Drape
Applicant
3M
Product Code
OMP · General, Plastic Surgery
Decision Date
Jul 14, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Dermatac™ Drape is an accessory to the following V.A.C.® Therapy Negative Pressure Wound Therapy Systems: 3M™ ActiV.A.C.™, 3M™ V.A.C.® Simplicity, and 3M™ V.A.C.Via™ Negative Pressure Wound Therapy Systems, which are integrated wound management systems for use in acute, extended and home care settings. The V.A.C.ULTA™ and V.A.C.RX4™ Negative Pressure Wound Therapy Systems are integrated wound management systems for use in acute care settings and other professional healthcare environments where product use is conducted by or under the supervision of a qualified healthcare professional. When used on open wounds, these systems are intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention, by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation, and by removing exudate and infectious material. Open wound types include: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure or venous insufficiency) flaps and grafts. When used on closed surgical incisions, they are intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudates via the application of negative pressure wound therapy. The Dermatac™ Drape is also an accessory to the V.A.C. VERAFLO™ Therapy (Instillation) provided by the V.A.C.ULTA™ Therapy Unit. V.A.C. VERAFLO™ Therapy is indicated for patients who would benefit from vacuum assisted drainage and controlled delivery of topical wound treatment solutions and suspensions over the wound bed. V.A.C. VERAFLO™ Therapy is indicated for patients with chronic, sub-acute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure or venous insufficiency) flaps and grafts.
Device Story
Dermatac™ Drape is a sterile, single-use, semi-occlusive wound dressing accessory for V.A.C.® and V.A.C. VERAFLO™ Negative Pressure Wound Therapy (NPWT) systems. Composed of polyurethane film with acrylic adhesive and a perforated silicone layer, it secures to periwound skin to create a sealed environment. The drape enables the NPWT system to apply negative pressure to the wound bed, promoting healing by removing exudate, reducing edema, and facilitating granulation tissue formation. In VERAFLO™ configurations, it also supports controlled delivery of topical solutions. Used in acute, extended, and home care settings, the drape is applied by healthcare professionals. By maintaining a sealed, moist wound environment, the device supports wound bed preparation and closure. It does not contain active electronic components but functions as a critical interface for the therapy unit's suction and fluid management.
Clinical Evidence
Bench testing only. No clinical data was required. Performance testing verified dressing functionality post-ETO sterilization, including dressing extensibility, moisture vapor transmission rate, peel adhesion force, and release liner testing. Biocompatibility testing was conducted per ISO 10993 standards, including cytotoxicity, sensitization, irritation, acute systemic toxicity, material-mediated pyrogenicity, subacute systemic toxicity, genotoxicity, and implantation, all yielding favorable results.
Technological Characteristics
Semi-occlusive drape; polyurethane film with acrylic adhesive and perforated silicone layer. Single-use, sterile. Functions as a sealed interface for NPWT systems. Biocompatibility per ISO 10993. Sterilization via ETO.
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
July 14, 2022
3M Melanie Avila Regulatory Affairs Manager 6203 Farinon Drive San Antonio, Texas 78249
Re: K220560
Trade/Device Name: Dermatac™ Drape Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP Dated: February 28, 2022 Received: February 28, 2022
Dear Melanie Avila:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Julie Morabito, Ph.D. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220560
Device Name Dermatac™ Drape
#### Indications for Use (Describe)
The Dermatac™ Drape is an accessory to the following V.A.C.® Therapy Negative Pressure Wound Therapy Systems:
3MTM ActiV.A.C.TM, 3M™ V.A.C.® Simplicity, and 3MTM V.A.C.Via™ Negative Pressure Wound Therapy .
Systems, which are integrated wound management systems for use in acute, extended and home care settings. . The V.A.C.ULTA™ and V.A.C.RX4™ Negative Pressure Wound Therapy Systems are integrated wound management systems for use in acute care settings and other professional healthcare environments where product use is conducted by or under the supervision of a qualified healthcare professional.
When used on open wounds, these systems are intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention, by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation, and by removing exudate and infectious material. Open wound types include: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure or venous insufficiency) flaps and grafts.
When used on closed surgical incisions, they are intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudates via the application of negative pressure wound therapy.
The Dermatac™ Drape is also an accessory to the V.A.C. VERAFLO™ Therapy (Instillation) provided by the V.A.C.ULTATM Therapy Unit.
V.A.C. VERAFLO™ Therapy is indicated for patients who would benefit from vacuum assisted drainage and controlled delivery of topical wound treatment solutions and suspensions over the wound bed.
V.A.C. VERAFLO™ Therapy is indicated for patients with chronic, sub-acute and dehisced wounds, partial-thickness burns. ulcers (such as diabetic, pressure or venous insufficiency) flaps and grafts.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
3M Health Care Business Group 6203 Farinon San Antonio. TX 78249
Contact Person: Melanie Avila Email: mavila9@mmm.com Phone: 210-275-5038 Date Prepared: 14 February 2022
Name of Subiect Device: Dermatac™ Drape Predicate Device: V.A.C. DERMATAC™ Drape (K200390) Common or Usual Name: Dressing component of Negative Pressure Wound Therapy (NPWT) System Classification Name: Neqative Pressure Wound Therapy Powered Suction Pump (and components) Regulatory Number: 21 CFR 878.4780 Regulatory Class: Class II Product Code: OMP
## Device Description
The Dermatac™ Drape is a semi-occlusive wound drape that is used as an accessory to the V.A.C.® Therapy and V.A.C. VERAFLO™ Therapy Systems. The Dermatac™ Drape is a single-use, sterile covering that provides a sealed environment for delivery of these therapies. It also allows for a moist wound environment.
The drape consists of a polyurethane film with acrylic adhesive with a perforated silicone layer. The perforations in the silicone laver expose the acrylic adhesive coated on the polyurethane film. The acrylic adhesive secures the drape to the periwound skin, thus creating a sealed wound environment.
The therapy systems with which it is used are comprised of the following:
- Software controlled therapy unit that provides negative pressure and, in the case of ● V.A.C. VERAFLO™ Therapy, a pump for controlled delivery of topical wound solutions.
- Disposable canister which collects wound exudate and, in the case of V.A.C. ● VERAFLO™ Therapy, instilled solutions
- Foam dressing for placement into the wound ●
- Semi-occlusive drape that covers the dressing ●
- Tubing set that connects the dressing to the therapy unit. ●
#### Intended Use / Indications for Use
The Dermatac™ Drape is an accessory to the following V.A.C.® Therapy Negative Pressure Wound Therapy Systems:
{5}------------------------------------------------
3M™ ActiV.A.C.™, 3M™ V.A.C.® Simplicity, and 3M™ V.A.C.Via™ Negative Pressure Wound . Therapy Systems, which are integrated wound management systems for use in acute, extended and home care settings.
The V.A.C.ULTA™ and V.A.C.RX4™ Negative Pressure Wound Therapy Systems are . integrated wound management systems for use in acute care settings and other professional healthcare environments where product use is conducted by or under the supervision of a qualified healthcare professional.
When used on open wounds, these systems are intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention, by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and perfusion, and by removing exudate and infectious material. Open wound types include: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure or venous insufficiency) flaps and grafts.
When used on closed surgical incisions, they are intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudates via the application of negative pressure wound therapy.
The Dermatac™ Drape is also an accessory to the V.A.C. VERAFLO™ Therapy (Instillation) provided by the V.A.C.ULTA™ Therapy Unit.
V.A.C. VERAFLO™ Therapy is indicated for patients who would benefit from vacuum assisted drainage and controlled delivery of topical wound treatment solutions and suspensions over the wound bed.
V.A.C. VERAFLO™ Therapy is indicated for patients with chronic, acute, traumatic, sub-acute and dehisced wounds, partial-thickness burns, ulcers and venous insufficiency), flaps and grafts.
#### Summary of Technological Characteristics
Comparison of subject and predicate device:
- Minor changes to the material formulation .
- . Minor changes to device packaging.
- . Additional manufacturing location
- . Minor changes to labeling
The indications for use, technological characteristics and principles of operation have not changed.
A table comparing the key features of the subject and predicate devices is provided below.
Summary of the technological characteristics of the device compared to the predicate device [21 CFR 807.92(a)(6)]
{6}------------------------------------------------
| Characteristic | Subject Device: | Predicate Device: Dermatac Drape, K200390 |
|----------------------------------------------|------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indicated Wound Types | Identical | Chronic Acute Traumatic Subacute Dehisced wounds Partial-thickness burns Ulcers (such as diabetic, pressure or venous insufficiency) Flaps and grafts Surgically closed incisions (For V.A.C. Therapy only) |
| V.A.C. Negative Pressure Wound Therapy Units | 3M™ ActiV.A.C.™ 3M™ V.A.C.® Simplicity 3M™ V.A.C.Via™ V.A.C.ULTA™ V.A.C.RX4™ | ACTIV.A.C.™ V.A.C. SIMPLICITY™ V.A.C.VIA™ V.A.C. FREEDOM™ V.A.C.ULTA™, INFOV.A.C™ V.A.C.RX4™ |
| Use environment/Care Setting of dressing kit | Identical | Acute, extended and home care settings |
## Performance Data
Summary of non-clinical tests conducted for determination of substantial equivalence:
- . V.A.C.™ Negative Pressure Maintenance System Test demonstrates the Dermatac™ Drape maintains negative pressure within specifications and manages fluid exudate without unexpected alarms.
- . Package Integrity testing to ensure the sterile barrier integrity is maintained throughout its labeled shelf life.
- Product performance testing of dressing components after ETO sterilization to verify . the product functions as intended throughout its labeled shelf life.
- o Dressing Extensibility
- Moisture Vapor Transmission Rate o
- Peel Adhesion Force o
- Wet Peel Adhesion Testing o
- o Release Liner Testing
- Summary of biocompatibility Testing .
{7}------------------------------------------------
| Endpoint | Study Type<br>(Test System) | Guidelines |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------|
| Cytotoxicity | MEM Elution<br>(L-929 Mouse Fibroblast Cells) | USFDA GLP 21CFR58<br>ISO 10993-5 (2009)<br>ISO 10993-12 (2007) |
| Sensitization | Guinea Pig Maximization Sensitization<br>(Hartley guinea pigs) | USFDA GLP 21CFR58<br>ISO 10993-10 (2010)<br>ISO 10993-12 (2012) |
| Irritation | Intracutaneous Reactivity<br>(New Zealand White Rabbit) | USFDA GLP 21CFR58<br>ISO 10993-10 (2010)<br>ISO 10993-12 (2012) |
| Acute Systemic<br>Toxicity | Acute Systemic Injection<br>(Swiss Mice) | USFDA GLP 21CFR58<br>ISO 10993-11 (2006)<br>ISO 10993-12 (2012) |
| Material-Mediated<br>Pyrogenicity | Rabbit Pyrogenicity Test<br>(New Zealand White Rabbits) | USFDA GLP 21CFR58<br>ISO 10993-11 (2006)<br>ISO 10993-12 (2012) |
| Subacute Systemic<br>Toxicity | 32-day Repeated Dose Subacute Toxicity<br>(Sprague Dawley rats) | US FDA GLP 21CFR58<br>ISO 10993-11 (2006)<br>ISO 10993-12 (2012) |
| Genotoxicity | Bacterial Mutagenicity - Ames Assay<br>(Salmonella typhimurium [TA97a, TA98, TA100<br>and TA1535] and Escherichia coli [WP2-uvrA]) | USFDA GLP 21CFR58<br>ISO 10993-3 (2014)<br>ISO 10993-12 (2012) |
| Implantation | In Vitro Mouse Lymphoma with Extended<br>Treatment<br>(L5178Y cells) | USFDA GLP 21CFR58<br>ISO 10993-6 (2016) |
The product had favorable biocompatibility test data for all relevant endpoints.
In all instances, the Dermatac™ Drape functioned as intended and all test results observed were as expected.
Clinical and Pre-clinical testing were not necessary to demonstrate equivalence. In addition, human factors enqineering testing was not required since the subject device has the same user interface and use environment as the predicate.
# Conclusions
The subject device is as safe and effective as the predicate device. The subject device's fundamental technology and principles of operation are unchanged compared to the predicate device. The subject device's Intended Use remains the same from the predicate device as cleared under K200390.
The minor differences between the subject device and its predicate do not significantly affect the safety or effectiveness of the device, nor did they represent a change in intended use. The performance data
{8}------------------------------------------------
demonstrates that the Dermatac™ Drape is as safe and effective as the predicate. Thus, the Dermatac™ Drape is substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.