K220554 · Taiwan Main Orthopaedic Biotechnology Co., Ltd. · SBF · Dec 16, 2022 · Neurology
Device Facts
Record ID
K220554
Device Name
Caduceus S
Applicant
Taiwan Main Orthopaedic Biotechnology Co., Ltd.
Product Code
SBF · Neurology
Decision Date
Dec 16, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
Caduceus S is intended as an intraoperative guidance system to enable minimally invasive surgery. Intraoperatively registered surgical Instruments are tracked by a passive marker sensor system to virtual computer image space on a patient's preoperative or intraoperative 2D or 3D image data. Caduceus S enables image-guide navigation of medical image data, which can either be acquired preoperatively CT or intraoperatively C-arm by an appropriate image acquisition system. Caduceus S offers pedicle screw implant size planning and navigation on rigid bone structures with intraoperatively registered surgical Instruments. Caduceus S is indicated for L5~T6 spine surgery where reference to a rigid anatomical structure can be identified relative to the acquired patient imagery (CT or C-arm). The headset (Surglasses) of the Caduceus S system is an optional heads up display that projects the 2D stereotaxic screens of the system's display.
Device Story
Caduceus S is an optical surgical navigation system for spine surgery; tracks surgical instruments and patient anatomy using passive marker sensors; registers instruments to preoperative CT or intraoperative C-arm images. System components include navigation cart, touch screen, and optional 'Surglasses' head-mounted display (HMD) for augmented reality visualization. Operates in the OR; used by surgeons to plan and navigate pedicle screw placement. Inputs include patient image data and real-time optical tracking signals; transforms inputs via software to display instrument position and trajectory relative to anatomy on HMD or monitor. Provides visual guidance to assist in accurate screw placement; potentially improves surgical precision and reduces invasiveness.
Clinical Evidence
No clinical data. Bench testing only. Performance validated via phantom and cadaver studies (L5-T6 spine). Accuracy metrics: CT mode mean positional error 1.91mm (99% UBL 2.07mm), angular error 1.59° (99% UBL 1.71°); C-arm mode mean positional error 1.80mm (99% UBL 2.06mm), angular error 1.65° (99% UBL 1.80°). All results statistically significantly lower than 3mm/3° thresholds.
Technological Characteristics
Optical tracking system using two infrared cameras (1.3-2m range). 6-DOF tracking via passive reflective spheres. Components: navigation cart, touch screen, HMD (Surglasses), disposable/reusable tools. Connectivity: DICOM import via CD/DVD/USB. Power: Li-ion battery for HMD. Software: Spine Navigation Software. Sterilization: EO for surgical instruments. Standards: ISO 14971, ISO 11135, ISO 10993, IEC 62304, IEC 60601-1.
Indications for Use
Indicated for L5~T6 spine surgery in patients where reference to a rigid anatomical structure can be identified relative to acquired CT or C-arm imagery for minimally invasive surgery.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
December 16, 2022
Image /page/0/Picture/1 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Taiwan Main Orthopaedic Biotechnology Co., Ltd. Min-Liang Wang, CEO 1F., No. 46, Keya Rd., Daya Dist. Taichung City, 42881 Tw
### Re: K220554
Trade/Device Name: Caduceus S Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: November 30, 2022 Received: November 30, 2022
### Dear Min-Liang Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Jesse Muir -S
For; Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K220554
Device Name Caduceus S
### Indications for Use (Describe)
Caduceus S is intended as an intraoperative gudance system to enable minimally invasive surgery. Intraoperatively registered surgical Instruments are tracked by a passive marker sensor system to virtual computer image space on a patient's preoperative or intraoperative 2D or 3D image data.
Caduceus S enables image-guide navigation of medical image data, which can either be acquired preoperatively CT or intraoperatively C-arm by an appropriate image acquisition system.
Caduceus S offers pedicle screw implant size planning and navigation on rigid bone structures with intraoperatively registered surgical Instruments.
Caduceus S is indicated for L5~T6 spine surgery where reference to a rigid anatomical structure can be identified relative to the acquired patient imagery (CT or C-arm).
The headset (Surglasses) of the Caduceus S system is an optional heads up display that projects the 2D stereotaxic screens of the system's display.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Traditional 510(k) Summary
# K220554 510(k) SUMMARY
| 5.1 | Type of Submission: | Traditional |
|-----|---------------------|--------------------------------------------------------------------------|
| 5.2 | Date of Summary: | February 25, 2022 |
| 5.3 | Submitter: | Taiwan Main Orthopaedic Biotechnology Co., Ltd. |
| | Address: | 1F., No. 46, Keya Rd., Daya Dist., Taichung City 428, Taiwan<br>(R.O.C.) |
| | Phone: | +886-4-3707-3159 |
| | Fax: | +886-4-2565-3330 |
| | Contact: | Jacky Fan (Regulatory Affair) |
#### 5.4 Identification of the Device:
| Proprietary/Trade name: | Caduceus S |
|------------------------------|------------------------|
| Classification Product Code: | OLO |
| Regulation Number: | 882.4560 |
| Regulation Description: | Stereotaxic instrument |
| Review Panel: | Orthopedic |
| Device Class: | II |
| Basis for the Submission: | New Device |
#### ર્ રંડ Identificationof the Predicate Device:
| Predicate Device Name: | Spine & Trauma Navigation |
|------------------------------|---------------------------|
| Submitter: | Brainlab AG |
| Classification Product Code: | OLO |
| Regulation number: | 882.4560 |
| Device Class: | II |
| 510(k) Number: | K183605 |
#### 5.6 Identificationof the Reference Device:
| Reference Device Name: | xvision Spine system (XVS) |
|------------------------------|----------------------------|
| Submitter: | Augmedics Ltd. |
| Classification Product Code: | OLO |
{4}------------------------------------------------
| Regulation number: | 882.4560 |
|--------------------|----------|
| Device Class: | II |
| 510(k) Number: | K190929 |
### 5.7 Indications for Use
Caduceus S is intended as an intraoperative guidance system to enable minimally invasive surgery. Intraoperatively registered surgical Instruments are tracked by a passive marker sensor system to virtual computer image space on a patient's preoperative or intraoperative 2D or 3D image data.
Caduceus S enables image-guide navigation of medical image data, which can either be acquired preoperatively CT or intraoperatively C-arm by an appropriate image acquisition system.
Caduceus S offers pedicle screw implant size planning and navigation on rigid bone structures with intraoperatively registered surgical Instruments.
Caduceus S is indicated for L5~T6 spine surgery where reference to a rigid anatomical structure can be identified relative to the acquired patient imagery (CT or C-arm).
The headset (Surglasses) of the Caduceus S system is an optional heads up display that projects the 2D stereotaxic screens of the system's display.
### 5.8 Device Description
Caduceus S surgical navigation system is an image guidance system, which is composed of Navi Tracker, Surglasses, other hardware (Navigation Cart with an arm, Touch Screen, Router, Control System, Connection System), disposable tools (Disposable Passive Sphere, Straight Guide Pin, Navi Clamp Kit, Instrument Adapter Type A&B, Calibration Plate, Registration Kit, Correction Tool), reusable tools (Calibration Board, Instrument Holder), and Spine Navigation Software. Caduceus S is an optical tracking and guiding system for spine surgery. It can track the marks on the surgical instruments and patient's anatomical structure with marks, and register with the preoperative or intraoperative images of the patients. During the surgery, it can be displayed on the Touch Screen and the head mounted display Surglasses to provide the navigation for the surgical instruments.
{5}------------------------------------------------
### Traditional 510(k) Summary
# 5.9 Substantial Equivalence Determination
Equivalence, same and difference among the subject, predicate and reference devices are cited as below.
| Item | Subject Device | Predicate Device | Reference Device | Substantial Equivalence<br>Determination |
|---------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary Name | Caduceus S | Spine & Trauma Navigation | xvision Spine system (XVS) | |
| 510(k) No. | | K183605 | K190929 | |
| Indications for<br>Use | Caduceus S is intended as an<br>intraoperative guidance system to<br>enable minimally invasive surgery.<br>Intraoperatively registered surgical<br>Instruments are tracked by a passive<br>marker sensor system to virtual<br>computer image space on a patient's<br>preoperative or intraoperative 2D or 3D<br>image data.<br>Caduceus S enables image-guide<br>navigation of medical image data,<br>which can either be acquired<br>preoperatively CT or intraoperatively<br>C-arm by an appropriate image<br>acquisition system.<br>Caduceus S offers pedicle screw<br>implant size planning and navigation<br>on rigid bone structures with | Spine & Trauma 3D is intended as an<br>intraoperative image-guided<br>localization system to enable<br>minimally invasive surgery. It links a<br>freehand probe, tracked by a passive<br>marker sensor system to virtual<br>computer image space on a patient's<br>preoperative or Intraoperative 2D or<br>3D image data.<br>Spine & Trauma 3D enables computer-<br>assisted navigation of medical image<br>data, which can either be acquired<br>preoperatively or inter-operatively by<br>an appropriate image acquisition<br>system.<br>The software offers screw implant size<br>planning and navigation on rigid<br>bone structures with precalibrated and | The xvision Spine System, with<br>xvision Spine System Software, is<br>intended as an aid for precisely<br>locating anatomical structures in<br>either open or percutaneous spine<br>procedures. Their use is indicated<br>for any medical condition in which<br>the use of stereotactic surgery may<br>be appropriate, and where<br>reference to a rigid anatomical<br>structure, such as the spine, can be<br>identified relative to CT imagery<br>of the anatomy. This can include the<br>spinal implant procedures, such as<br>Posterior Pedicle Screw Placement<br>in the thoracic and sacro-lumbar<br>region.<br>The Headset of the xvision Spine | Equivalent<br>Both the subject and<br>predicate devices are an<br>intraoperative image-guided<br>localization system to enable<br>minimally invasive surgery<br>and the subject device is<br>focus on spine surgery.<br>Both the subject and<br>reference devices display the<br>image by a headset and a<br>screen.<br>Although there are some<br>technical differences among<br>these devices, the subject<br>device is tested and |
| | intraoperatively<br>registered surgical | additional | individually-calibrated System displays 2D stereotaxic | validated so that the |
| | Instruments. | surgical tools. The system is indicated | screens and a virtual anatomy | difference does not raise |
| | Caduceus S is indicated for L5~T6 for any medical condition in which the | | screen. The stereotaxic screen is | new issues of SE. |
| | spine surgery where reference to a rigid | use of stereotactic surgery may be | indicated for correlating the | |
| | anatomical structure can be identified | appropriate and where a reference to a | tracked instrument location to the | |
| | relative to the acquired patient imagery | rigid anatomical structure, such as the | registered patient imagery. The | |
| | (CT or C-arm). | skull, the pelvis, a long bone or | virtual screen is indicated for | |
| | The headset (Surglasses) of the | vertebra can be identified relative to | displaying the virtual instrument | |
| | Caduceus S system is an optional heads | the acquired image (CT, MR, 2D | location in relation to the virtual | |
| | up display that projects the 2D | fluoroscopic image or 3D fluoroscopic | anatomy to assist in percutaneous | |
| | stereotaxic screens of the system's | image reconstruction) and/or an image | visualization and trajectory | |
| | display. | data based model of the anatomy. | planning. | |
| | | | The virtual display should not be | |
| | | | relied upon solely for absolute | |
| | | | positional information and should | |
| | | | always be used in conjunction with | |
| | | | the displayed stereotaxic | |
| | | | information. | |
| Type of Use | Prescription Use | Prescription Use | Prescription Use | Same |
| Use Environment | Operating Room | Operating Room | Operating Room | Same |
| | • Platform including cart, computer,<br>monitor and tracking cameras | • Platform including cart, computer,<br>monitor and tracking cameras | • Headset with near eye<br>see-through display | Equivalent |
| Main Component | • Headset with near eye see-through | • Software application | • tracking camera | subject device is<br>composed of headset as the |
| | display<br>Navi tracker<br>Software application<br>Reflective markers - Spheres<br>Instrument universal adaptors | Reflective markers - Spheres<br>Accessories (Instrument adaptors,<br>referencing system) | Software application<br>Reflective markers - Flat<br>Instrument universal adaptors<br>Reference point | reference device does, and<br>the rest components are<br>similar with that of the<br>predicate device.<br>Although the main<br>component is slightly<br>different among these<br>devices, the subject device is |
| | Patient Preparation<br>System Set-up | Patient Preparation<br>System Set-up…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.