K220529 · Merits Health Products Co., Ltd. · ITI · Mar 18, 2022 · Physical Medicine
Device Facts
Record ID
K220529
Device Name
Merits P335 Powered Wheelchair
Applicant
Merits Health Products Co., Ltd.
Product Code
ITI · Physical Medicine
Decision Date
Mar 18, 2022
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 890.3860
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Merits P335 Powered Wheelchair is intended to provide indoor mobility to persons limited to a seating position that are capable of operating a powered wheelchair.
Device Story
Battery-powered, front-wheel-drive wheelchair; controlled via PG R-net 120A controller with joystick interface. Features power seating system (tilt, recline, optional elevate) actuated by 24V DC motorized linear actuators. Welded steel frame; two front drive wheels; rear pivoting casters. Joystick input signals controller to release solenoid electromechanical brakes for movement; releasing joystick triggers automatic braking. Used by patients with mobility limitations in indoor/outdoor environments. Healthcare providers/caregivers may assist with setup; patient operates via joystick. Output is physical mobility; enhances patient independence. Safety features include anti-tip wheels and gas spring suspension.
Clinical Evidence
Bench testing only. No clinical data. Device performance validated against ISO 7176 standards for stability, braking, energy consumption, dimensions, speed, fatigue strength, obstacle climbing, and electromagnetic compatibility (IEC 60601-1-2). Biocompatibility evaluated per ISO 10993-1 based on identical materials to previously cleared devices.
Technological Characteristics
Welded steel frame; 24V DC motor; 2x 12V 60Ah lead-acid batteries; PG R-net 120A controller. Front-wheel drive; solenoid electromechanical brakes. Upholstery: Vinyl fabric, PU foam (ISO 7176-16 compliant). Dimensions: 39.6"L x 24.7"W x 42"H. Max speed 6 mph; 7.5° gradient capacity. Off-board 6A charger. Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for persons limited to a seating position who are capable of operating a powered wheelchair, for indoor and outdoor mobility.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 18, 2022
Merits Health Products Co., LTD. % Dave Yungvirt CEO Third Party Review Group, LLC 25 Independence Blvd Warren, New Jersey 07059
Re: K220529
Trade/Device Name: Merits P335 Powered Wheelchair Regulation Number: 21 CFR 890.3860 Regulation Name: Powered wheelchair Regulatory Class: Class II Product Code: ITI Dated: February 20, 2022 Received: February 24, 2022
Dear Mr. Yungvirt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220529
Device Name Merits P335 Powered Wheelchair
Indications for Use (Describe)
The Merits P335 Powered Wheelchair is intended to provide indoor mobility to persons limited to a seating position that are capable of operating a powered wheelchair.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY - K220529
#### I. Submitter:
Merits Health Products Co., LTD. No.18, Jingke Rd., Nantun Dist. Taichung City 40852, Taiwan (R.O.C.)
Phone: +886-4-23594985 ext.200 Fax: +886-4-23594992
Contact Person: Martin Tseng Date Prepared: November 26, 2021
#### II. Device:
Name of Device: Merits P335 Powered Wheelchair Common or Usual Name: Powered Wheelchair Classification Name: Powered Wheelchair (21 CFR 890.3860) Regulatory Class: II Product Code: ITI
# III. Predicate Device:
Permobil F3 FWD Powered Wheelchair (#K143180) This predicate has not been subject to a design-related recall.
No reference devices were used in this submission.
### IV. Device Description:
The Merits P335 Powered Wheelchair is battery powered, front wheel motor driven and is controlled by the PG power wheelchair R-net 120 amp controller. The user interface is a joystick. P335 is powered by two 12 VDC 60ah batteries. The batteries are charged by 6A off-board charger connect with 3-pin Microphone Connector to charging socket on joystick. The approximate driving range on fully charged batteries is up to 30km (19mi). The chair frame is a welded steel construction and includes two front drive wheels with drive units (including motor, gear, and brake), batteries and rear pivoting casters. Depending on users' needs, the joystick motor control is mounted to the left or right armrest. When the user activates the joystick, the controller receives a signal to release the brakes. With the brakes released, the wheelchair is allowed to move in the direction the joystick is actuated. When the user releases the joystick, the chair slows to a stop and the brakes are automatically re-engaged. The solenoid electromechanical brakes allow the user stop by letting go of the joystick.
The device is equipped with a Power Seating System consists of a power tilt unit, a power recline module and an optional power elevating seat module. The tilt, reclining and elevating systems are actuated by 24V DC motorized linear actuator. The tilt system includes one motorized linear actuator caused
{4}------------------------------------------------
the seat frame to shift forward. This enhances stability since the center of gravity is kept substantially in place while the user is tilting. The recline system include one motorized linear actuator change the position of the backrest with respect to the seat pan. The optional elevating seat module use one motorized linear actuator which allows the user to elevate the seat pan for easy accessing things stored in higher places.
The upholstery of the device complies with ISO 7176-16:2012: Wheelchairs -- Part 16: Resistance to ignition of postural support devices.
The device can be operated on dry, level surfaces composed of concrete, blacktop, or asphalt under normal driving conditions.
The Merits P335 is substantially equivalent to Permobil F3 FWD Powered Wheelchair (#K143180). Both products are battery power, motorized vehicles designed for use as personal power mobility aids. Performance characteristics and drive mechanisms are similar and all have the same intended function and use which is to provide indoor and outdoor mobility to persons limited to a seating position that are capable of operating a powered wheelchair. Additional, they are all constructed from the same basic materials, have the same basic operational principles and all use DC batteries as their source of power.
Although there are some minor differences between P335 and its predicate device. But they raise no new issues of safety or effectiveness. Performance data demonstrate that Merits P335 is safe. The non-clinical testing and the predicate comparisons demonstrate that any differences in their technological characteristics do no raise any new questions of safety or effectiveness.
# V. Indications for Use:
The Merits P335 Powered Wheelchair is to provide indoor and outdoor mobility to persons limited to a seating position that are capable of operating a powered wheelchair.
The Indications for Use for the Merits P335 is identical to the predicate device.
## VI. Comparison of Technological Characteristics with The Predicate Device:
The Merits P335 Powered Wheelchair is substantially equivalent to the Permobil F3 Powered Wheelchair (K143180). Both products are battery power, motorized vehicles designed for use as personal power mobility aids. Performance characteristics and drive mechanisms are similar and all have the same intended function and use. The detailed comparison table is as follow:
| Description | Merits P335 Power Wheelchair<br>K220529 | Permobil F3 Powered Wheelchair<br>K143180 |
|--------------|-------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Merits Health Products Co., LTD. | Permobil AB |
| Model | P335 | F3 |
| Intended Use | To provide indoor and outdoor<br>mobility to persons limited to a<br>seating position that are capable<br>of operating a powered<br>wheelchair. | To provide indoor and outdoor<br>mobility to persons limited to a<br>seating position that are capable<br>of operating a powered<br>wheelchair. |
{5}------------------------------------------------
| Framework | | Steel welded frame | Steel welded frame |
|-----------------------|-----------|---------------------------------------------------------------|---------------------------------------------------------------|
| Overall<br>Dimensions | Length | 39.6" (±0.6") | 40" |
| | Width | 24.7" (±0.6") | 24" |
| | Height | 42" (±0.6") | 38" |
| Seat<br>Dimensions | Width | 14~22" (±0.6") | 17~23" |
| | Depth | 18" (±0.6") | 18" |
| | Height | 17.5~19.5" (±0.6") From Ground | 17.5~19.5" From Ground |
| Weight | Total | 386.1lb | 386 lb |
| | Base | 147 lb | 169 lb |
| | Seat | 122 lb | 140 lb |
| | Footplate | 40.1lb | |
| | Battery | 38.5lb*2pcs | 38.5lb*2pcs |
| Unit Configuration | | Front Wheel Drive | Front Wheel Drive |
| Weight Capacity | | 300 lb | 300 lb |
| Gradient | | 7.5° | 6° |
| Obstacle Climbing | | 2.4"(60mm) | 2.4"(60mm) |
| Turning Radius | | 26.6" | 26.6" |
| Ground Clearance | | 3" | 3" |
| Drive Wheel Tires | | O.D.=355.6mm (14") Foam Filled<br>Tire | O.D.=355.6mm (14") Foam Filled<br>Tire |
| | | I.D.=203.2mm (8")<br>Width=76.2mm(3") | I.D.=203.2mm (8")<br>Width=76.2mm(3") |
| Caster Tires | | 225x70 mm (9”) Foam Filled Tire | 210x65 mm (8") Foam Filled Tire |
| Anti-Tip Wheel | | 100x24 mm (4") Non-Removable | 100x24 mm (4") Non-Removable |
| Suspension | | Front/ Rear Gas Spring | Front/ Rear Gas Spring |
| Max. Speed | | 6 mph | 6 mph |
| Range up to | | 19miles | 20 miles |
| Brake System | | Electronic braking by drive motor.<br>Magnetic parking brakes | Electronic braking by drive motor.<br>Magnetic parking brakes |
| Motor | | 2 Pole, 24Vdc Motor | 2 Pole, 24Vdc Motor |
| Controller | | PG R-net 120A | PG R-net 120A |
| Battery | | 2*12V 60Ah Lead Acid Battery | 2*12V 60Ah Lead Acid Battery |
| | | Type: Off-board charger | Type: Off-board charger |
| Charger | | Input: 100-240Vac, 50/60Hz, 2.5A | Input: 100-240Vac, 50/60Hz, 3.8A |
| | | Output: 6A /24Vdc | Output: 8A /24Vdc |
| Footplate | | Yes | Yes |
| Tilt | | Yes | Yes |
| Recline | | Yes | Yes |
| Elevate | | Optional | Optional |
| Backrest Cover | | Vinyl Fabric | Vinyl Fabric |
| Backrest Upholstery | | PU Foam | PU Foam |
| Cushion cover | | Vinyl Fabric | Vinyl Fabric |
| Cushion Upholstery | | PU Foam | PU Foam |
| Armrest Type | | PU Foam | PU Foam |
There are some differences between P335 and its predicate device:
- a. Dimension: The dimensions of the Merits P335 is equivalent in size to the Premobil F3 (K143180). Although there are some differences, the device passed ISO 7176-1 (Determination of Static Stability) and ISO 7176-2 (Determination of Dynamic Stability of electric wheelchairs). So there is no deleterious affection of safety and effectiveness about the difference on Dimension with predicated device.
- b. Gradient: The Merits P335 has a larger Gradient than the Permobil F3 (K143180). The device has passed ISO 7176-2 (Determination of Dynamic
{6}------------------------------------------------
Stability of electric wheelchairs) & ISO 7176-3 (Determination of Efficiency of Brake of Electric Wheelchair) test. So there is no deleterious affection of safety and effectiveness about the difference on Gradient with predicated device.
- c. Rear Caster size: The Merits P335 is equipped with 2 larger 9" Casters. which give the same or better curb climbing ability than the Permobil F3 (#K143180) 8" Castors while moving backward. There is no deleterious affection of safety and effectiveness about the difference on Rear Castor size with predicated device.
- d. Weight: Although the total weight of P335 is slightly heavier than Permobil F3 Powered Wheelchair (K143180). The device has passed ISO 7176-8 (Requirements and test methods for static, impact and fatigue strengths) test. So there is no deleterious affection of safety and effectiveness about the difference on Weight with predicated device.
- e. Charger: The Merits P335 uses a 6A off-board charger which is smaller than the 8A off-board charger used on Permobil F3 (K143180). This will slightly increase the charging time needed for the same 60Ah batteries. But the charger has passed the ISO 7176-25 (Batteries and charger for powered wheelchairs) test. So there is no deleterious affection of safety and effectiveness about the difference on charger with predicated device.
The minor differences between P335 and its predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the P335 is safe.
The non-clinical testing and the predicate comparisons demonstrate that any differences in their technological characteristics do no raise any new questions of safety or effectiveness.
{7}------------------------------------------------
# VII. Performance Data
The following Non-clinical Tests have been Performed in support of the substantial equivalence determination:
| STD No. | Title of Standard | FDA Rec. No. |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| ISO 7176-1:2014 | Determination of Static Stability | 16-195 |
| ISO 7176-2:2017 | Determination of Dynamic Stability of electric<br>wheelchairs | 16-202 |
| ISO 7176-3:2012 | Determination of effectiveness of brakes | 16-192 |
| ISO 7176-4:2008 | Energy consumption of electric wheelchairs and<br>scooters for determination of theoretical distance<br>range | 16-162 |
| ISO 7176-5:2008 | Determination of overall dimensions, mass and<br>maneuvering space | 16-163 |
| ISO 7176-6:2018 | Determination of maximum speed, acceleration and<br>deceleration of electric wheelchairs | 16-204 |
| ISO 7176-7:1998 | Method of Measurement of Seating and Wheel<br>Dimensions | 16-196 |
| ISO 7176-8:2014 | Requirements and test methods for static, impact and<br>fatigue strengths | 16-197 |
| ISO 7176-9:2009 | Climatic tests for wheelchairs | 16-167 |
| ISO 7176-10:2008 | Determination of obstacle-climbing ability of<br>electrically power wheelchairs | 16-164 |
| ISO 7176-11:2012 | Test dummies | 16-190 |
| ISO 7176-13:1989 | Determination of coefficient of friction of test surfaces | 16-25 |
| ISO 7176-14:2008 | Power and control systems for electrically powered<br>wheelchairs and scooter- Requirements and test<br>methods. | 16-165 |
| ISO 7176-15:1996 | Requirements for Information Disclosure,<br>Documentation and Labeling | 16-27 |
| ISO 7176-16:2012 | Resistance to ignition of postural support devices | 16-191 |
| ISO 7176-21:2009 | Requirements and test methods for electromagnetic<br>compatibility of electrically powered wheelchairs and<br>scooters, and battery chargers. | 16-166 |
| ISO 7176-25:2013 | Batteries and charger for powered wheelchairs | 16-194 |
| ISO 10993-1:2018 | Biological evaluation of medical devices - Part 1:<br>Evaluation and testing within a risk management<br>process | 2-258 |
| IEC60601-1-2:2020 | Medical electrical equipment - Part 1-2: General<br>requirements for basic safety and essential<br>performance - Collateral Standard: Electromagnetic<br>disturbances - Requirements and tests | 19-36 |
The results of the testing confirm that the device meets specifications and is substantially equivalent to the predicate device.
{8}------------------------------------------------
# Biocompatibility Evaluation
A Biocompatibility Evaluation has been performed according to the ISO 10993-1:2018. As the skin contact components listed below are identical to those used on our premarket devices in formulation, processing sterilization and geometry, and no other chemical have been added (e.q. plasticizers, fillers, color additives, cleaning agents, mold release agents). There is no need for any more Biocompatibility tests.
| | Merits<br>P335 | Merits<br>P322 | Merits<br>Model R Series | Same or<br>Different |
|-------------------------------------|----------------------------------|----------------------------------|-------------------------------------|----------------------|
| K Number | K220529 | K122357 | K113577 | N/A |
| Product Code | ITI | ITI | ITI | Same |
| Seat Cushion:<br>Supplier: | Vinyl Fabric<br>ROODINSEAT Corp. | Vinyl Fabric<br>ROODINSEAT Corp. | Vinyl Fabric<br>ROODINSEAT<br>Corp. | Same |
| Backrest Cushion:<br>Supplier: | Vinyl Fabric<br>ROODINSEAT Corp. | Vinyl Fabric<br>ROODINSEAT Corp. | Vinyl Fabric<br>ROODINSEAT<br>Corp. | Same |
| Armrest Pad<br>Material: Supplier: | PU Foam<br>WELL YI CO., LTD. | PU Foam<br>WELL YI CO., LTD. | PU Foam<br>WELL YI CO., LTD. | Same |
| Footrest Pad<br>Material: Supplier: | PU Foam<br>WELL YI CO., LTD. | PU Foam<br>WELL YI CO., LTD. | PU Foam<br>WELL YI CO., LTD. | Same |
## VIII. Conclusions
According to comparison table, the differences on function of elevate, recline and dimension of width do not deleteriously affect the safety and effectiveness of the device.
So based on the design, performance specifications and testing and intended use, the Merits P335 Powered Wheelchair is substantially equivalent to Permobil F3 FWD Powered Wheelchair (#K143180).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.