K220493 · Gcs Co. Limited · QVJ · Oct 27, 2023 · General, Plastic Surgery
Device Facts
Record ID
K220493
Device Name
Plaxpot Multi Plasma
Applicant
Gcs Co. Limited
Product Code
QVJ · General, Plastic Surgery
Decision Date
Oct 27, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
PLAXPOT™ MULTI PLASMA is intended for the removal and destruction of skin lesions and coagulation of tissue.
Device Story
Handheld electrosurgical device; utilizes high-frequency current to ionize air, creating plasma arc between electrode tip and skin; enables non-contact tissue sublimation and coagulation; powered by internal rechargeable lithium-ion battery; operated by clinician in clinical setting; device features OLED display and mode selection (0.7W, 1W, 2W); output is monopolar; provides precise tissue removal without heat transfer to surrounding areas; benefits patient through targeted lesion destruction.
Clinical Evidence
Bench testing only. Testing included basic safety and essential performance (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and thermal effects testing per FDA guidance for electrosurgical devices. All tests passed.
Technological Characteristics
Monopolar electrosurgical device; 70 kHz frequency; 3 power modes (0.7W, 1W, 2W); internal 7.4 VDC 1000mAh Li-ion battery; OLED display; non-sterile; reusable; complies with IEC 60601-1 and IEC 60601-1-2 standards.
Indications for Use
Indicated for the removal and destruction of skin lesions and coagulation of tissue in patients requiring electrosurgical treatment.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 27, 2023
GCS Co. Limited % Su Kyung Park Director Compliance Insight 497 Circle Freeway Dr. Unit 230 West Chester, Ohio 45246
Re: K220493
Trade/Device Name: Plaxpot Multi Plasma Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: QVJ Dated: [NOTE: Use date of most recent supplement] Received: September 29, 2023
Dear Su Kyung Park:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Digitally signed by Mark Mark Trumbore -S Trumbore -S Date: 2023.10.27
> 16:19:00 -04'00' Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220493
Device Name PLAXPOT™ MULTI PLASMA
Indications for Use (Describe) PLAXPOT™ MULTI PLASMA is intended for the removal and destruction of skin lesions and coagulation of tissue.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <div> <span>☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## PLAXPOT™ MULTI PLASMA
# 510(k) Summary
#### General Information 5.1
18 February 2022 Preparation Date:
26 October 2023 Revision Date:
## Submitter/Holder
Su Kyung, Park
Director
GCS Co., Ltd. #1008, 555, Dunchon-daero, Jungwon-gu, Seongnam-si, Gyeonggi-do, Korea Tel: 82-2-555-1536 Fax : 82-2-3142-7670 Email: kara21c(@gcsgroup.co.kr
Website: http://www.bestgcs.com/
## Primary Contact
Cynthia Ipach President/CEO
Email: cipach@compliance-insight.com
Office +1 513-860-3512 ext. 306 Compliance Insight 497 Circle Freeway Dr Suite 230. Cincinnati, OH 45246
#### 5.2 Regulatory Information
| Subject Device Trade Name | PLAXPOT TM MULTI PLASMA |
|---------------------------|-------------------------------------------------------------------|
| Classification Name | Low Power Electrosurgical Devices For Skin Lesion<br>Destruction |
| Device Classification | II |
| Regulation Description | Electrosurgical cutting and coagulation device and<br>accessories |
| FDA Product Code | QVJ |
| CFR References | 21 CFR 878.4400 |
| Review Panel | General & Plastic Surgery |
{4}------------------------------------------------
#### 5.3 Identification of Predicate Devices
The primary predicate device for PLAXPOT™ MULTI PLASMA is the SubNovii Advanced Plasma Technology (K201738) and the reference device is Plasma IQ (K192813).
#### 5.4 Subject Device Description
PLAXPOT™ MULTI PLASMA is a device that consists of a handpiece which is the main body and a charger (charging cradle, and power adapter). The high frequency current is exposed to the air and the air is ionized. The electrode tip can perform treatment without direct contact with the tissue.
After the power is applied through the internal battery, the mode is set and the Shot button is pressed, a high voltage alternating current pulse is released according to the set value, and the air is ionized to form a plasma arc.
This device uses plasma energy. This causes excess skink and sublimate (direct passage from the solid to the gaseous state of matter) without affecting the deeper skin layers. The difference in potential distance between device's handpiece and the skin represents the area where the voltaic arc will take place. In this gap, the gases in the air are ionized, thus Plasma is generated. The tissues touched by plasma "sublime" without heat transfer to other unwanted tissues.
#### ર્સ્ટ Indications for Use
PLAXPOT™ MULTI PLASMA is intended for the removal and destruction of skin lesions and coagulation of tissue.
#### Substantial Equivalence Discussion 5.6
In Table 1 on the following page please find a characteristics comparison summary. The review of the indications for use and comparison characteristics provided in Table 1 demonstrate that PLAXPOT™ MULTI PLASMA is substantially equivalent to the primary predicate device, the SubNovii Advanced Plasma Technology K201738.
{5}------------------------------------------------
| Device<br>Characteristics | Proposed<br>Device | Primary<br>Predicate | Reference<br>Device |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Product Name | PLAXPOTTM<br>MULTI PLASMA | SubNovii<br>Advanced<br>Plasma<br>Technology | Plasma IQ |
| 510(k) | K220493 | K201738 | K192813 |
| Manufacturer | GCS Co. Ltd. | Cartessa<br>Aesthetics | Neauvia North<br>America |
| FDA Product<br>Code | QVJ | GEI | GEI |
| CFR Reference | 878.4400 | 878.4400 | 878.4400 |
| Device Class | II | II | II |
| Prescription or<br>OTC | Prescription | Prescription | Prescription |
| Implanted<br>Device | No | No | No |
| Indications for<br>Use Statement | PLAXPOT™<br>MULTI PLASMA<br>is intended for the<br>removal and<br>destruction of skin<br>lesions and<br>coagulation of<br>tissue. | The SubNovii is<br>intended for the<br>removal and<br>destruction of<br>skin lesions and<br>coagulation of<br>tissue. | Plasma IQ is<br>used in the<br>removal and<br>destruction of<br>skin lesions and<br>coagulation of<br>tissue. |
| Mode of<br>Operation | Plasma high<br>frequency current<br>ionizing the air<br>creating a plasma<br>stream. | Plasma<br>radiofrequency<br>energy ionizes<br>the air creating a<br>plasma stream. | Plasma<br>radiofrequency<br>energy ionizes<br>the air creating<br>a plasma<br>stream. |
| Output Mode | Continuous | Continuous | Continuous |
| Energy Type | High Frequency | High Frequency | High Frequency |
| Power Level | 3 modes at 0.7W,<br>1W, 2W | Max 5W | Max 5W |
| Reusable or<br>Single Use | Reusable | Reusable | Reusable |
| Output | Monopolar | Monopolar | Monopolar |
| Max Power<br>Output | 3W | 5W | 5W |
| Power Supply | Internal<br>Rechargeable<br>Lithium Ion<br>Battery 7.4 VDC<br>1000mAh | Internal<br>Rechargeable<br>Lithium Ion<br>Battery 11.1<br>VDC | Internal<br>Rechargeable<br>Lithium Ion<br>Battery 11.1<br>VDC |
| Frequency | 70 kHz | 40kHz | 40kHz |
| System<br>Components | A handpiece that<br>incorporates the<br>electrosurgical<br>generator unit,<br>OLED display,<br>software, and<br>docking station. | A handpiece that<br>incorporates the<br>electrosurgical<br>generator unit,<br>docking station<br>and an active<br>electrode. | A handpiece<br>that<br>incorporates the<br>electrosurgical<br>generator unit,<br>docking station<br>and an active<br>electrode. |
| Electrical<br>Safety<br>Standards | IEC 60601-1<br>IEC 60601-1-2 | IEC 60601-1<br>IEC 60601-1-2 | IEC 60601-1<br>IEC 60601-1-2 |
| Software | Yes | No | No |
| Software<br>Compliance | IEC 62304 | N/A | N/A |
# Table 1. Summary Comparison of Characteristics
{6}------------------------------------------------
# PLAXPOT™ MULTI PLASMA
#### 5.7 Sterilization and Shelf Life
The PLAXPOT™ MULTI PLASMA is sold non-sterile.
{7}------------------------------------------------
Shelf-life is not applicable because of low likelihood of time-dependent product degradation.
#### Performance Testing - Bench 5.8
Performance bench tests of PLAXPOT™ MULTI PLASMA have been performed, see Table 2. The results from the performance bench testing demonstrate that PLAXPOT™ MULTI PLASMA has met the functional requirements and is substantially equivalent to the predicate device. Performance bench testing is provided in detail.
| Study | Attachment | Test Method | Results |
|-------------------------------------------|------------|--------------------------------------------------------------------------------------------------|---------|
| Basic Safety and<br>Essential Performance | 18-3 | EN 60601-<br>1:2006+A1:2013 or EN<br>60601-<br>1:2006+A12:2014 and<br>test report<br>IEC60601_1K | Pass |
| Manufacturer<br>Technical File | 18-5 | N/A | Pass |
| EMC Test | 18-6 | IEC 60601-1-2:2014 | Pass |
| EMC Test | 18-7 | IEC 60601-1-2:2014 | Pass |
| Thermal Effects Test | NA | FDA Submissions for<br>Electrosurgical Devices<br>for General Surgery | Pass |
### Table 2. Performance Testing Summary
#### 5.9 Conclusion
The subject device PLAXPOT™ MULTI PLASMA has the same technology, principle of operation, indications for use and technical specifications as the primary predicate device Subnovii and the reference device Plasma IQ. PLAXPOT™ MULTI PLASMA shares a substantially equivalent design, indications for use and technology (i.e. features, materials, and principles of operation) as the predicate device and no new elements pertaining to safety or effectiveness have been identified. Performance test results demonstrate that the subject device can perform the same intended use as safely as the predicate and reference devices. Therefore, the subject device is substantially equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.