# Platform® Acetabular System (K220483)

_Total Joint Orthopedics, Inc. · LZO · Sep 23, 2022 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K220483

## Device Facts

- **Applicant:** Total Joint Orthopedics, Inc.
- **Product Code:** [LZO](/productcode/LZO.md)
- **Decision Date:** Sep 23, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3353
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Klassic HD Hip System is intended for prosthetic replacement in treatment of the following: Patient conditions of non-inflammatory degenerative joint disease (NIDJD): avascular necrosis, osteoarthritis, ankylosis, protrusio acetabuli and painful hip dysplasia. Patient conditions of inflammatory joint disease (IJD): rheumatoid arthritis. Revision of a previously failed hip arthroplasty. Patients who require a total hip replacement. Those patients with failed previous surgery where pain, deformity, or dysfunction persists.

## Device Story

Platform Acetabular System is an expansion of the Klassic HD Hip System; provides acetabular shells and inserts for hip arthroplasty. Shells fabricated from Ti6Al4V; inserts from Vitamin E-stabilized UHMWPE. Components available in various sizes (44mm-64mm OD) to match patient anatomy; compatible with 28mm, 32mm, and 36mm femoral heads. Used by orthopedic surgeons in clinical settings to restore hip joint biomechanics. Provided sterile for single-use. System functions as a prosthetic interface to replace diseased or damaged hip joint structures, alleviating pain and restoring function.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing and engineering analysis, including material wear evaluation, range of motion analysis, and mechanical testing per ASTM F1820-13 (push-out, lever-out, axial torque), ASTM F2582-20 (impingement), ASTM F3090-20 (fatigue), and ISO 7206-12 (deformation). LAL testing for endotoxins was also performed.

## Technological Characteristics

Acetabular shell (Ti6Al4V per ASTM F136) and insert (Vitamin E UHMWPE). Semi-constrained design. Sizes 44mm-64mm OD. Compatible with 28/32/36mm heads. Sterile, single-use. Mechanical fixation.

## Regulatory Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

## Predicate Devices

- Klassic HD Hip System ([K100445](/device/K100445.md))
- [K141972](/device/K141972.md)
- [K173104](/device/K173104.md)
- [K180929](/device/K180929.md)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is a blue square with the letters 'FDA' in white. Next to the blue square, the words 'U.S. FOOD & DRUG' are written in blue, with the word 'ADMINISTRATION' appearing below in a smaller font size, also in blue.

September 23, 2022

Total Joint Orthopedics, Inc. % Holly Rhodes Vice President, Orthopedic Regulatory Affairs MCRA, LLC 803 7th Street NW Washington, District of Columbia 20001

Re: K220483

Trade/Device Name: Platform® Acetabular System Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: Class II Product Code: LZO, LPH, MBL, OQG Dated: July 25, 2022 Received: July 25, 2022

Dear Holly Rhodes:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-

combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

· Patient conditions of inflammatory joint disease (IJD): rheumatoid arthritis.

510(k) Number (if known)

#### K220483

Device Name Platform® Acetabular System

Indications for Use (Describe)

ankylosis, protrusio acetabuli and painful hip dysplasia.

• Revision of a previously failed hip arthroplasty. · Patients who require a total hip replacement.

· Those patients with failed previous surgery where pain, deformity, or dysfunction persists.

Type of Use (Select one or both, as applicable)

| <div> <span> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)     </span> </div> |
|--------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☐</span> Over The Counter Use (21 CFR 801 Subpart C)     </span> </div>  |

The Klassic HD Hip System is intended for prosthetic replacement in treatment of the following:

· Patient conditions of non-inflammatory degenerative joint disease (NIDJD): avascular necrosis, osteoarthritis,

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

| Manufacturer:      | Total Joint Orthopedics, Inc.<br>1567 E. Stratford Avenue<br>Salt Lake City, UT 84106<br>Phone: 801.486.6070<br>Fax: 801.486.6117                                                                                                          |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:           | Mr. Chris Weaber<br>Director of Research and Development                                                                                                                                                                                   |
| Prepared By:       | MCRA, LLC<br>803 7th Street NW<br>Washington, DC 20001<br>Phone: 202.552.5800<br>Fax: 202.552.5798                                                                                                                                         |
| Date Prepared:     | September 21, 2022                                                                                                                                                                                                                         |
| Device Trade Name: | Platform® Acetabular System                                                                                                                                                                                                                |
| Common Name:       | Acetabular Shell, Acetabular Inserts                                                                                                                                                                                                       |
| Classifications:   | 21 CFR 888.3353 - Hip joint metal/ceramic/polymer semi-<br>constrained cemented or nonporous uncemented prosthesis<br>21 CFR 888.3358 - Hip joint metal/polymer/metal semi-<br>constrained porous-coated uncemented prosthesis<br>Class II |
| Product Codes:     | LZO, LPH, MBL, OOG                                                                                                                                                                                                                         |

### Indications for Use:

The Klassic HD® Hip System is intended for prosthetic replacement without bone cement in treatment of the following:

- Patient conditions of non-inflammatory degenerative joint disease (NIDJD): avascular . necrosis, osteoarthritis, ankylosis, protrusio acetabuli and painful hip dysplasia.
- Patient conditions of inflammatory joint disease (IJD): rheumatoid arthritis.
- Those patients with failed previous surgery where pain, deformity, or dysfunction persists. ●
- Revision of a previously failed hip arthroplasty. ●
- Patients who require a total hip replacement. ●

{4}------------------------------------------------

### Device Description:

The Klassic HD Hip System employs prostheses designed to help surgeons restore hip joint biomechanics. The purpose of this 510(k) is to add the Platform Acetabular System consisting of the Platform Acetabular Shell with Ti-Coat and Platform Acetabular Inserts with E-Link (Neutral and Hooded) to the Klassic HD Hip System. The Platform Acetabular Shells are fabricated from Ti6Al4V per ASTM F136 and the Platform Acetabular Inserts with E-Link are made from Vitamin E. UHMWPE material. The subject acetabular components are available in various sizes ranging in outer diameters of 44mm to 64mm to match patient anatomy and are compatible with 28mm, 32mm, and 36mm heads. The subject components are provided sterile, for single use, by prescription only.

### Predicate Devices:

The modified Klassic HD Hip System is substantially equivalent to the predicate Klassic HD® Hip System (Primary predicate: K100445; Additional predicates: K141972, K173104, K180929) with respect to indications, design, material and function. The information summarized in this 510(k) demonstrates that the subject device is substantially equivalent to the identified predicate devices.

# Comparison of Technological Characteristics:

The subject devices feature the same material (Ti6A14V per ASTM F136 and Vitamin E UHMWPE), same use as acetabular shells and liners, same articular interface design, same compatibility, and same sterilization as predicate shells and liners (K100445).

# Discussion of Non-Clinical Testing/ Performance Data:

The subject device underwent push out testing, lever out testing, axial torque disassembly testing, impingement testing, acetabular shell fatigue testing, and deformation testing. In addition, engineering analysis was performed to evaluate material wear and range of motion for the subject devices. Additionally, the Klassic HD® Hip System is in compliance with LAL testing requirements for orthopedic implants.

Non-clinical testing and engineering analysis conducted to demonstrate substantial equivalence was as follows:

- Material wear (via engineering analysis) of acetabular shells and liners ●
- Range of motion (via engineering analysis)
- Push out testing per ASTM F1820-13 ●
- Lever out testing per ASTM F1820-13 ●
- Axial Torque Disassembly Testing per ASTM F1820-13
- Impingement Testing per ASTM F2582-20
- Acetabular Shell Fatigue Testing per ASTM F3090-20
- Deformation Testing per ISO 7206-12 .

### Conclusion

Testing and engineering analyses demonstrated that the subject acetabular components perform as safe and effective compared to the predicate components and are substantially equivalent to the predicate.

---

**Source:** [https://fda-staging.innolitics.com/device/K220483](https://fda-staging.innolitics.com/device/K220483)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).