← Product Code [LYZ](/productcode/LYZ) · K220469

# Disposable Vinyl Examination Glove (K220469)

_Chifeng Huawei Medical Science&Technology Co., Ltd. · LYZ · Jun 16, 2022 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K220469

## Device Facts

- **Applicant:** Chifeng Huawei Medical Science&Technology Co., Ltd.
- **Product Code:** [LYZ](/productcode/LYZ.md)
- **Decision Date:** Jun 16, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

The DISPOSABLE VINYL EXAMINATION GLOVE is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

## Device Story

Disposable vinyl examination glove; non-sterile; ambidextrous; clear color. Used by healthcare examiners to provide a barrier against contamination between patient and examiner. Device is worn on hands during medical examinations. No electronic, software, or algorithmic components. Bench testing confirms compliance with ASTM standards for physical dimensions, tensile strength, elongation, and freedom from holes. Biocompatibility testing confirms non-cytotoxic, non-irritating, and non-sensitizing properties.

## Clinical Evidence

No clinical data. Bench testing only. Compliance verified via ASTM D5250 (dimensions/thickness), ASTM D5151 (watertightness/holes), ASTM D6124 (powder content), ASTM D412 (tensile strength/elongation), and ISO 10993-5/10 (biocompatibility).

## Technological Characteristics

Material: Vinyl. Design: Ambidextrous, powder-free, clear. Dimensions: Length ≥230mm; Thickness ≥0.08mm (finger/palm). Standards: ASTM D5250, ASTM D5151, ASTM D6124, ISO 10993-5, ISO 10993-10. Non-sterile.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Vinyl Examination Glove (Clear, Non-Colored) ([K180861](/device/K180861.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

June 16, 2022

Chifeng Huawei Medical Science&Technology Co.,Ltd. % Boyle Wang Official Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.1801,No.161 Lujiazui East Rd.,Pudong Shanghai, Shanghai 200120 China

Re: K220469

Trade/Device Name: Disposable Vinyl Examination Glove Regulation Number: 21 CFR 880.6250 Regulation Name: Non-powdered patient examination glove Regulatory Class: Class I, reserved Product Code: LYZ Dated: June 8, 2022 Received: June 16, 2022

Dear Boyle Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Bifeng Qian, M.D., Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K220469

Device Name DISPOSABLE VINYL EXAMINATION GLOVE

Indications for Use (Describe)

The DISPOSABLE VINYL EXAMINATION GLOVE is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary K220469

This summary of 510(k) is being submitted in accordance with 21 CFR 807.92.

#### 1.0 Submitter's Information

CHIFENG HUAWEI MEDICAL SCIENCE & TECHNOLOGY CO.,LTD. Name: Address: No. 2-4, Second-stage Standardized Plant, Songshan (Anqing) Industrial Park, Chifeng, China. Phone Number: +86-18248056569 Contact: Li Xiaohong Date of Preparation: Jan.7, 2022

#### Designated Submission Correspondent

Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 1801, No. 161 East Lujiazui Rd., Pudong,Shanghai 200120 ,China Tel: +86-21-50313932 Email: Info@truthful.com.cn

#### 2.0 Device Information

Trade name: DISPOSABLE VINYL EXAMINATION GLOVE Common name: Vinyl Patient Examination Glove Classification name: Non-powdered patient examination glove Model(s): S, M, L, XL

#### 3.0 Classification

Production code: LYZ Regulation number: 21CFR880.6250 Classification: Class I Panel: General Hospital

#### 4.0 Predicate Device Information

Manufacturer: ZHICHENG TRADING CO., LTD. Device: Vinyl Examination Glove (Clear, Non-Colored) 510(k) number: K180861

{4}------------------------------------------------

### 5.0 Indication for Use

The DISPOSABLE VINYL EXAMINATION GLOVE is a disposable device intended for medical purposes that is worn on the examiner's hands to prevent contamination between patient and examiner.

#### 6.0 Device Description

The subject device is powder free vinyl examination gloves. The subject device's color is clear. It can be available in four specifications: S,M,L and XL. The subject device is non-sterile.

# 7.0 Technological Characteristic Comparison Table

| Item                                  | Subject Device<br>(K220469)                                                                                                                                                                                     | Predicated Device<br>(K180861)                                                                                                                                                                                               | Remark                |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|
| Product Code                          | LYZ                                                                                                                                                                                                             | LYZ                                                                                                                                                                                                                          | Same                  |
| Regulation No.                        | 21CFR880.6250                                                                                                                                                                                                   | 21CFR880.6250                                                                                                                                                                                                                | Same                  |
| Class                                 | I                                                                                                                                                                                                               | I                                                                                                                                                                                                                            | Same                  |
| Intended Use /<br>Indications for Use | The DISPOSABLE<br>VINYL EXAMINATION<br>GLOVE is a disposable<br>device intended for<br>medical purposes that is<br>worn on the examiner's<br>hands to prevent<br>contamination between<br>patient and examiner. | The Vinyl Examination<br>Glove (Clear,<br>Non-Colored) is a<br>disposable device<br>intended for medical<br>purposes that is worn on<br>the examiner's hands to<br>prevent contamination<br>between patient and<br>examiner. | Same                  |
| Material                              | Vinyl                                                                                                                                                                                                           | Vinyl                                                                                                                                                                                                                        | Same                  |
| Powdered or<br>Powered free           | Powdered free                                                                                                                                                                                                   | Powdered free                                                                                                                                                                                                                | Same                  |
| Design Feature                        | Ambidextrous                                                                                                                                                                                                    | Ambidextrous                                                                                                                                                                                                                 | Same                  |
| Colorant                              | Clear                                                                                                                                                                                                           | Clear                                                                                                                                                                                                                        | Same                  |
| Labeling<br>Information               | Single-use indication,<br>powder free, device<br>color, device name,<br>glove size and quantity,<br>Non-Sterile                                                                                                 | Single-use indication,<br>powder free, device<br>color, device name,<br>glove size and quantity,<br>Non-Sterile                                                                                                              | Same                  |
| Dimensions(mm)                        | Length:<br>S/M/L/XL: ≥230;<br>Width:<br>S: 85±5;                                                                                                                                                                | Length:<br>S≥230;<br>M≥235;<br>L≥245;                                                                                                                                                                                        | Similar<br>Analysis 1 |

#### Table1-General Comparison

{5}------------------------------------------------

|                        | M: 95±5;<br>L: 105±5;<br>XL: 115±5 | XL≥245;<br>Width:<br>S: 85±5;<br>M: 95±5;<br>L: 105±5;<br>XL: 115±5                                                  |                                                                                                                      |                    |
|------------------------|------------------------------------|----------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|--------------------|
| Thickness(mm)          | Finger: ≥0.08;<br>Palm: ≥0.08      | Finger: ≥0.05;<br>Palm: ≥0.08                                                                                        | Similar Analysis 1                                                                                                   |                    |
|                        |                                    |                                                                                                                      |                                                                                                                      |                    |
| Physical<br>Properties | Before<br>Aging                    | Tensile Strength<br>11MPa, min<br>Ultimate Elongation<br>300% min                                                    | Tensile Strength<br>15MPa, min<br>Ultimate Elongation<br>380% min                                                    | Similar Analysis 2 |
|                        |                                    |                                                                                                                      |                                                                                                                      |                    |
|                        | After<br>Aging                     | Tensile Strength<br>11MPa, min<br>Ultimate Elongation<br>300%min                                                     | Tensile Strength<br>15MPa, min<br>Ultimate Elongation<br>380%min                                                     | Similar Analysis 2 |
|                        |                                    |                                                                                                                      |                                                                                                                      |                    |
| Freedom from<br>Holes  |                                    | Be free from holes when<br>tested in accordance with<br>ASTMD5151<br>AQL=2.5                                         | Be free from holes when<br>tested in accordance with<br>ASTMD5151<br>AQL=1.5                                         | Similar Analysis 3 |
| Powder Content         |                                    | Meet the requirements of<br>ASTM D6124                                                                               | Meet the requirements of<br>ASTM D6124                                                                               | Same               |
| Biocompatibility       |                                    | ISO 10993-10;<br>Under the conditions of<br>the study, not an irritant<br>or a sensitizer                            | Comply with<br>ISO10993-10                                                                                           | Same               |
|                        |                                    | Under conditions of the<br>study, did not show<br>potential toxicity to L-929<br>cells. Complies with ISO<br>10993-5 | Under conditions of the<br>study, did not show<br>potential toxicity to L-929<br>cells. Complies with ISO<br>10993-5 | Same               |

Analysis 1: The physical dimensions are different with that of the predicate, but they all meet the requirements of ASTM D5250, so the differences do not raise any new safety or performance questions.

Analysis 2: The tensile strength and ultimate elongation are different with that of the predicate, but they all meet the requirements of ASTM D5250, so the differences do not raise any new safety or performance questions.

Analysis 3: Freedom from holes of subject device is similar with that of the predicate, but they all meet the requirements of ASTMD5151, so the differences do not raise any new safety or performance questions.

{6}------------------------------------------------

#### 8.0 Summary of Non-clinical Testing

Non-clinical tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

ISO 10993-5 Third edition 2009-06-01, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity

ISO 10993-10:2010 Biological Evaluation of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.

ASTM D6124-06 (Reapproved 2017), Standard Test Method for Residual Powder on Medical Gloves

ASTM D5151-19, Standard Test Method for Detection of Holes in Medical Gloves.

ASTM D5250-19, Standard Specification for Poly (vinyl chloride) Gloves for Medical Application.

| Test<br>Method | Purpose                                                   | Acceptance Criteria                                                                                                                               | Results                                                                                                                                                                                   |                |
|----------------|-----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|
| ASTM<br>D5250  | Physical<br>Dimensions<br>Test                            | Length(mm):<br>S/M/L/XL: ≥230;<br>Width(mm):<br>S: 85±5;<br>M: 95±5;<br>L: 105±5;<br>XL: 115±5<br>Thickness (mm):<br>Finger: ≥0.08<br>Palm: ≥0.08 | Length(mm):<br>> 230/Pass;<br>Width(mm):<br>S: 85-89 /Pass<br>M: 95-97/ Pass<br>L: 104-107/ Pass<br>XL:116-118/ Pass<br>Thickness (mm):<br>Finger: 0.11-0.17/Pass<br>Palm: 0.10-0.13/Pass |                |
| ASTM<br>D5151  | Watertightness<br>Test<br>for<br>Detection<br>of<br>Holes | Meet the requirements of ASTM D5151<br>AQL 2.5                                                                                                    | 0/125/Pass                                                                                                                                                                                |                |
| ASTM<br>D6124  | Powder<br>Content                                         | Meet the requirements of ASTM D6124<br>< 2.0mg                                                                                                    | 0.06-0.09mg/Pass;                                                                                                                                                                         |                |
| ASTM<br>D412   | Physical<br>properties                                    | Before Aging<br>Tensile Strength                                                                                                                  | ≥11MPa                                                                                                                                                                                    | 11-19MPa/Pass; |
|                |                                                           | Before Aging<br>Ultimate                                                                                                                          | ≥300%                                                                                                                                                                                     | 320-369%/Pass; |

Table 2 - Summary of non-clinical performance testing

{7}------------------------------------------------

|                 |               | Elongation            |                        |        |                                                                            |
|-----------------|---------------|-----------------------|------------------------|--------|----------------------------------------------------------------------------|
|                 |               | After<br>Aging        | Tensile<br>Strength    | ≥11MPa | 11-18MPa/Pass;                                                             |
|                 |               |                       | Ultimate<br>Elongation | ≥300%  | 310-352%/Pass;                                                             |
| ISO<br>10993-5  | Cytotoxicity  | In Vitro Cytotoxicity |                        |        | Under conditions of the<br>study, device extract is<br>not cytotoxic./Pass |
| ISO<br>10993-10 | Irritation    | Non-irritating        |                        |        | Under the conditions of<br>the study, not an<br>irritant/ Pass             |
| ISO<br>10993-10 | Sensitization | Non-sensitizing       |                        |        | Under conditions of the<br>study, not a sensitizer./<br>Pass               |

#### 9.0 _Summary of Clinical Testing

Clinical testing is not needed for this device.

#### 10.0 Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the subject device DISPOSABLE VINYL EXAMINATION GLOVE is as safe, as effective, and performs as well as or better than the legally marketed predicated device K180861.

---

**Source:** [https://fda-staging.innolitics.com/device/K220469](https://fda-staging.innolitics.com/device/K220469)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).