The Family of UltraPulse CO2 Surgical and Aesthetic Lasers, Delivery Devices and Accessories
K220467 · Lumenis Be, Ltd. · GEX · May 18, 2022 · General, Plastic Surgery
Device Facts
Record ID
K220467
Device Name
The Family of UltraPulse CO2 Surgical and Aesthetic Lasers, Delivery Devices and Accessories
Applicant
Lumenis Be, Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
May 18, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The UltraPulse Alpha CO2 Laser System, Delivery Devices and Accessories (members of the UltraPulse CO2 Laser Systems Family) are indicated for use in surgical or aesthetic applications requiring: ablation, vaporization, excision, or coagulation of soft tissue in medical specialties including: dermatology and plastic surgery (aesthetic), podiatry , gynecology, general and oral surgery, dental and oral surgery and genitourinary surgery as follows: Dermatology & Plastic Surgery The ablation, vaporization, excision, and coagulation of soft tissue in dermatology and plastic surgery in the performance of: - · Laser skin resurfacing - · Laser derm-abrasion - · Laser burn debridement - -Laser skin resurfacing (ablation and/or vaporization) for treatment of: - · Wrinkles, rhytids, and furrows (including fine lines and texture irregularities). - -Laser skin resurfacing (ablation and/or vaporization) of soft tissue for the reduction, removal, and/or treatment of: - Keratoses, including actinic and sebortheic keratosis, sebortheic wart and verruca sebortheica. - · Vermillionectomy of the lip - · Cutaneous horns - · Solar/actinic elastosis - · Cheilitis, including actinic cheilitis - · Lentigines, including lentigo maligna or Hutchinson's malignant freckle - · Uneven pigmentation/dyschromia - Acne scars - · Surgical scars - · Keloids including acne keloidalis nuchae - · Hemangiomas (including Buccal, port wine and pyogenic granuloma pyogenicum/granuloma telangiectaticum) - Tattoos - · Telangiectasia - · Removal of small skin tumors, including periungual (Koenen) and subungual fibromas - · Superficial pigmented lesions - · Adenosebaceous hypertrophy or sebaceous hyperplasia - Rhinophyma reduction - · Cutaneous papilloma (skin tags) - · Milia - · Debridement of eczematous or infected skin - · Basal and squamous cell carcinoma, including keratoacanthomas, Bowen's disease (Erythroplasia of Queyrat), and Bowenoid Papulosis (BP) lesions - · Nevi, including spider, epidermal and protruding - · Neurofibromas - · Laser de-epithelialization - · Tricoepitheliomas - Xanthelasma palpebrarum - · Syringoma -Laser ablation, vaporization and/or excision for complete and partial nail matrixectomy. Vaporization of: - · Benign and malignant vascular/avascular skin lesions - · Moh's Surgery - Lipectomy • Verrucae and seborrhoecae vulgares, including paronychial, periungal, and subungual warts -Laser incision and/or excision of soft tissue for the performance of upper and lower eyelid blepharoplasty. -Laser incision and/or excision of soft tissue for the creation of recipient sites for hair transplantation Podiatry Laser ablation, vaporization, and/or excision of soft tissue for the reduction, removal, and/or treatment of: - · Verrucae vulgares/plantar (warts), including paronychial, periungal and subungual warts - Porokeratoma ablation - · Ingrown nail treatment - · Neuromas/fibromas, including Morton's neuroma - Debridement of ulcers - · Other soft tissue lesions Laser ablation, vaporization, and/or excision for complete and partial matrixectomy Gynecology (GYN) Laser incision, excision, ablation and/or vaporization of soft tissue in gynecology for the treatment of: - · vulvar and vaginal intraepithelial neoplasia (VIN, VAIN) - · Condyloma acuminata, including genital, vulvar, perineal, and Bowen's disease (Erythrolasia of Queyrat) and Bowenoid papulosa (BP) lesions - · Leukoplakia (vulvar dystrophies) - · Incision and drainage (I&D) of Bartholin's and nubuthian cysts - · Herpes vaporization - · Urethral caruncle vaporization - Benign and malignant tumors - · Hemangiomas General Surgery - Debridement of traumatic wounds - Debridement of decubitus and diabetic ulcers - · Microsurgery - · Artificial joint revision - PMMA removal General and Thoracic Surgery Incision, excision and vaporization of soft tissue in general and thoracic surgery including endoscopic and open procedures. Applications include: - · Debridement of decubitus ulcers, stasis, diabetic, and other ulcers - · Mastectomy - Debridement of burns - · Rectal and anal hemorrhoidectomy - · Breast biopsy - · Reduction mammoplasty - Cytoreduction for metastatic disease - · Laparotomy applications - · Mediastinal and thoracic lesions and abnormalities - · Skin tag vaporization - Atheroma - · Cysts, including sebaceous cysts, pilar cysts, and mucous cysts of the lips - Pilonidal cyst removal and repair - · Abscesses - · Other soft tissue applications Dental and Oral Surgery Incision/excision and vaporization of soft tissue in dentistry and oral surgery. Applications include: - · Gingivectomy/removal of hyperplasias - · Gingivoplasty - · Incisional and excisional biopsy - · Treatment of ulcerous lesions, including aphthous ulcers - · Incision of infection when used with antibiotic therapy - · Frenectomy (frenum release) - · Excision and ablation of benign and malignant lesions - · Homeostasis - Operculectomy - · Crown lengthening - · Removal of soft tissue, cysts and tumors - · Oral cavity tumors and hemangiomas - · Abscesses - · Extraction site hemostasis - · Salivary gland pathologies - · Preprosthetic gum preparation - Leukoplakia - · Partial glossectomy - · Periodontal gum resection Genitourinary Incision/excision and vaporization of soft tissue in genitourinary procedures. Applications include: - · Benign and malignant lesions of external genitalia - · Condyloma - Phimosis - · Erythroplasia
Device Story
UltraPulse CO2 laser system; RF-modulated CO2 laser tube technology; emits 10,600 or 11,100 nm laser beams; includes red diode aiming beam. System uses computer-controlled delivery devices (microscanners, collimated/focused handpieces) to direct energy to soft tissue. Used by clinicians in OR or clinic settings for ablation, vaporization, excision, or coagulation. Output allows precise tissue removal or modification; aids in surgical procedures like skin resurfacing, tumor excision, and wound debridement. Benefits include controlled tissue interaction for various medical specialties.
Clinical Evidence
No clinical data; bench testing only. Labeling changes required no performance testing.
Technological Characteristics
RF-modulated CO2 laser tube; 10,600/11,100 nm wavelength; red diode aiming beam; computer-controlled; includes microscanners and various handpieces.
Indications for Use
Indicated for soft tissue ablation, vaporization, excision, or coagulation in dermatology, plastic surgery, podiatry, gynecology, general, thoracic, dental, oral, and genitourinary surgery. Applicable to patients requiring surgical intervention for lesions, tumors, ulcers, or aesthetic resurfacing.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Lumenis Modified Family of UltraPulse CO2 Surgical and Aesthetic Laser Systems (K203544)
Reference Devices
Lumenis Be UltraPulse SurgiTouch and Encore CO2 Lasers (K030147)
Lumenis Be UltraPulse SurgiTouch and Encore CO2 Lasers (K022060)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
May 18, 2022
Lumenis Be, Ltd. Shlomit Segman Senior Director, RA 6 Hakidma Street PO BOX 240 Yokneam, Yokneam 2069204 Israel
Re: K220467
Trade/Device Name: The Family of UltraPulse CO2 Surgical and Aesthetic Lasers, Delivery Devices and Accessories Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: February 17, 2022 Received: February 17, 2022
Dear Shlomit Segman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya, D.Eng. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K220467
#### Device Name
The UltraPulse Alpha CO2 Surgical and Aesthetic Laser, Delivery Devices and Accessories
#### Indications for Use (Describe)
The UltraPulse Alpha CO2 Laser System, Delivery Devices and Accessories (members of the UltraPulse CO2 Laser Systems Family) are indicated for use in surgical or aesthetic applications requiring:
ablation, vaporization, excision, or coagulation of soft tissue in medical specialties including: dermatology and plastic surgery (aesthetic), podiatry , gynecology, general and oral surgery, dental and oral surgery and genitourinary surgery as follows:
#### Dermatology & Plastic Surgery
The ablation, vaporization, excision, and coagulation of soft tissue in dermatology and plastic surgery in the performance of:
- · Laser skin resurfacing
- · Laser derm-abrasion
- · Laser burn debridement
- -Laser skin resurfacing (ablation and/or vaporization) for treatment of:
- · Wrinkles, rhytids, and furrows (including fine lines and texture irregularities).
- -Laser skin resurfacing (ablation and/or vaporization) of soft tissue for the reduction, removal, and/or treatment of:
- Keratoses, including actinic and sebortheic keratosis, sebortheic wart and verruca sebortheica.
- · Vermillionectomy of the lip
- · Cutaneous horns
- · Solar/actinic elastosis
- · Cheilitis, including actinic cheilitis
- · Lentigines, including lentigo maligna or Hutchinson's malignant freckle
- · Uneven pigmentation/dyschromia
- Acne scars
- · Surgical scars
- · Keloids including acne keloidalis nuchae
- · Hemangiomas (including Buccal, port wine and pyogenic granuloma pyogenicum/granuloma telangiectaticum)
- Tattoos
- · Telangiectasia
- · Removal of small skin tumors, including periungual (Koenen) and subungual fibromas
- · Superficial pigmented lesions
- · Adenosebaceous hypertrophy or sebaceous hyperplasia
- Rhinophyma reduction
- · Cutaneous papilloma (skin tags)
- · Milia
- · Debridement of eczematous or infected skin
- · Basal and squamous cell carcinoma, including keratoacanthomas, Bowen's disease (Erythroplasia of Queyrat), and Bowenoid Papulosis (BP) lesions
- · Nevi, including spider, epidermal and protruding
- · Neurofibromas
- · Laser de-epithelialization
- · Tricoepitheliomas
- Xanthelasma palpebrarum
{3}------------------------------------------------
#### · Syringoma
-Laser ablation, vaporization and/or excision for complete and partial nail matrixectomy. Vaporization of:
- · Benign and malignant vascular/avascular skin lesions
- · Moh's Surgery
- Lipectomy
• Verrucae and seborrhoecae vulgares, including paronychial, periungal, and subungual warts
-Laser incision and/or excision of soft tissue for the performance of upper and lower eyelid blepharoplasty.
-Laser incision and/or excision of soft tissue for the creation of recipient sites for hair transplantation
### Podiatry
Laser ablation, vaporization, and/or excision of soft tissue for the reduction, removal, and/or treatment of:
- · Verrucae vulgares/plantar (warts), including paronychial, periungal and subungual warts
- Porokeratoma ablation
- · Ingrown nail treatment
- · Neuromas/fibromas, including Morton's neuroma
- Debridement of ulcers
- · Other soft tissue lesions
Laser ablation, vaporization, and/or excision for complete and partial matrixectomy
### Gynecology (GYN)
Laser incision, excision, ablation and/or vaporization of soft tissue in gynecology for the treatment of:
- · vulvar and vaginal intraepithelial neoplasia (VIN, VAIN)
- · Condyloma acuminata, including genital, vulvar, perineal, and Bowen's disease (Erythrolasia of Queyrat) and Bowenoid papulosa (BP) lesions
- · Leukoplakia (vulvar dystrophies)
- · Incision and drainage (I&D) of Bartholin's and nubuthian cysts
- · Herpes vaporization
- · Urethral caruncle vaporization
- Benign and malignant tumors
- · Hemangiomas
### General Surgery
- Debridement of traumatic wounds
- Debridement of decubitus and diabetic ulcers
- · Microsurgery
- · Artificial joint revision
- PMMA removal
### General and Thoracic Surgery
Incision, excision and vaporization of soft tissue in general and thoracic surgery including endoscopic and open procedures. Applications include:
- · Debridement of decubitus ulcers, stasis, diabetic, and other ulcers
- · Mastectomy
- Debridement of burns
- · Rectal and anal hemorrhoidectomy
- · Breast biopsy
- · Reduction mammoplasty
- Cytoreduction for metastatic disease
- · Laparotomy applications
- · Mediastinal and thoracic lesions and abnormalities
- · Skin tag vaporization
- Atheroma
- · Cysts, including sebaceous cysts, pilar cysts, and mucous cysts of the lips
- Pilonidal cyst removal and repair
{4}------------------------------------------------
- · Abscesses
- · Other soft tissue applications
Dental and Oral Surgery
Incision/excision and vaporization of soft tissue in dentistry and oral surgery. Applications include:
- · Gingivectomy/removal of hyperplasias
- · Gingivoplasty
- · Incisional and excisional biopsy
- · Treatment of ulcerous lesions, including aphthous ulcers
- · Incision of infection when used with antibiotic therapy
- · Frenectomy (frenum release)
- · Excision and ablation of benign and malignant lesions
- · Homeostasis
- Operculectomy
- · Crown lengthening
- · Removal of soft tissue, cysts and tumors
- · Oral cavity tumors and hemangiomas
- · Abscesses
- · Extraction site hemostasis
- · Salivary gland pathologies
- · Preprosthetic gum preparation
- Leukoplakia
- · Partial glossectomy
- · Periodontal gum resection
#### Genitourinary
Incision/excision and vaporization of soft tissue in genitourinary procedures. Applications include:
- · Benign and malignant lesions of external genitalia
- · Condyloma
- Phimosis
- · Erythroplasia
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known)
K220467
#### Device Name
The UltraPulse SurgiTouch and UltraPulse Encore CO2 Surgical and Aesthetic Laser, Devices and Accessories
#### Indications for Use (Describe)
The UltraPulse SurgiTouch and UltraPulse Encore CO2 Laser System, Delivery Devices and Accessories (members of the UltraPulse CO2 Laser Systems Family) are indicated for use in surgical or aesthetic applications requiring: ablation, vaporization, excision, or coagulation of soft tissue in medical specialties including: aesthetic (dermatology and plastic surgery), podiatry, otolaryngology (including laparoscopy), neurosurgery, orthopedics (soft tissue), arthroscopy (knee), general and thoracic surgery (including open and endoscopic), dental and oral surgery and genitourinary surgery as follows:
#### Dermatology & Plastic Surgery
The ablation, vaporization, excision, and coagulation of soft tissue in dermatology and plastic surgery in the performance of:
- · Laser skin resurfacing
- · Laser derm-abrasion
- · Laser burn debridement
Laser skin resurfacing (ablation and/or vaporization) for treatment of:
- · Wrinkles, rhytids, and furrows (including fine lines and texture irregularities).
Laser skin resurfacing (ablation and/or vaporization) of soft tissue for the reduction, removal, and/or treatment of:
- · Keratoses, including actinic and sebortheic keratosis, sebortheic wart and veruca sebortheica.
- · Vermillionectomy of the lip
- · Cutaneous horns
- Solar/actinic elastosis
- · Cheilitis, including actinic cheilitis
- · Lentigines, including lentigo maligna or Hutchinson's malignant freckle
- · Uneven pigmentation/dyschromia
- Acne scars
- · Surgical scars
- · Keloids including acne keloidalis nuchae
- · Hemangiomas (including Buccal, port wine and pyogenic granuloma pyogenicum/granuloma
- telangiectaticum)
- · Tattoos
- · Telangiectasia
- · Removal of small skin tumors, including periungual (Koenen) and subungual fibromas
- · Superficial pigmented lesions
- · Adenosebaceous hypertrophy or sebaceous hyperplasia
- · Rhinophyma reduction
- · Cutaneous papilloma (skin tags)
- · Milia
- · Debridement of eczematous or infected skin
- · Basal and squamous cell carcinoma, including keratoacanthomas, Bowen's disease (Erythroplasia of Queyrat), and Bowenoid Papulosis (BP) lesions
- · Nevi, including spider, epidermal and protruding
- · Neurofibromas
- · Laser de-epithelialization
- · Tricoepitheliomas
{6}------------------------------------------------
- Xanthelasma palpebrarum
- · Syringoma
Laser ablation, vaporization and/or excision for complete and partial nail matrixectomy.
Vaporization or coagulation of:
- · Benign and malignant vascular/avascular skin lesions
- · Moh's Surgery
- Lipectomy
- · Verrucae and seborrhoecae vulgares, including paronychial, periungal, and subungual warts
Laser incision and/or excision of soft tissue for the performance of upper and lower eyelid blepharoplasty.
Laser incision and/or excision of soft tissue for the creation of recipient sites for hair transplantation Podiatry
Laser ablation, vaporization, and/or excision of soft tissue for the reduction, removal, and/or treatment of:
- · Verrucae vulgares/plantar (warts), including paronychial, periungal and subungual warts
- Porokeratoma ablation
- · Ingrown nail treatment
- · Neuromas/fibromas, including Morton's neuroma
- · Debridement of ulcers
- · Other soft tissue lesions
Laser ablation, vaporization, and/or excision for complete and partial matrixectomy
Otolaryngology (ENT)
Laser incision, excision, ablation and/or vaporization of soft tissue in otolaryngology for the treatment of:
- · Choanal atresia
- · Leukoplakia, including oral, larynx, uvula, palatal, and upper lateral pharyngeal tissue
- · Nasal obstruction
- · Adult and juvenile papillomatosis polyps
- · Polypectomy of nose and nasal passages
- · Lymphangioma removal
- · Removal of vocal cord/fold nodules, polyps and cysts
- · Removal of recurrent papillomas in the oral cavity, larynx, pharynx and trachea, including the uvula, palatal, upper lateral pharyngeal tissue, tongue and vocal cords.
- · Laser/tumor surgery in the larynx, pharynx, nasal, ear and oral structures and tissue
- · Zenker's Diverticulum/pharyngoesophageal diverticulum (endoscopic laser-assisted esophagodiverticulostomy (ELAED))
- · Stenosis, including subglottic stenosis
- · Tonsillectomy (including tonsillar cryptolysis and neoplasma) and tonsil ablation/tonsillotomy
- · Pulmonary bronchial and tracheal lesion removal
- · Benign and malignant nodules, tumors and fibromas (larynx, trachea, tracheobronchial/endobronchial)
- · Benign and malignant lesions and fibromas (nose and nasal passages)
- · Benign and malignant tumors and fibromas (oral)
- · Stapedotomy/Stapedectomy
- · Acoustic neuroma in the ear
- · Superficial lesions of the ear, including chondrodermatitis nondularis chronica helices/Winkler's disease
- · Telangiectasia/hemangioma of larynx, pharynx and trachea (includes uvula, palatal, or upper lateral pharyngeal tissue)
- · Cordectomy, cordotomy (for the treatment of vocal fold paralysis/vocal fold motion impairment), and cordal lesions of larynx, pharynx and trachea
- · Myringotomy/tympanostomy (tympanic membrane fenestration)
- · Uvulopalatoplasty (LAUP, laser UPPP)
- · Turbinectomy and turbinate reduction/ablation
- · Septal spur ablation/reduction and septoplasty
- · Partial glossectomy
- · Tumor resection of oral, subfacial and neck tissues
- · Rhinophyma
- Verrucae vulgares (warts)
- Gingivoplasty/gingivectomy
{7}------------------------------------------------
### Gynecology (GYN)
Laser incision, excision, ablation and/or vaporization of soft tissue in gynecology for the treatment of:
- · Conization of the cervix, including cervical intraepithelial neoplasia (CIN), and vulvar and vaginal intraepithelial neoplasia (VIN, VAIN)
- · Condyloma acuminata, including cervical, genital, vulvar, perineal, and Bowen's disease (Erythrolasia of Queyrat) and Bowenoid papulosa (BP) lesions
- · Leukoplakia (vulvar dystrophies)
- · Incision and drainage (I&D) of Bartholin's and nubuthian cysts
- · Herpes vaporization
- · Urethral caruncle vaporization
- Cervical dysplasia
- Benign and malignant tumors
- · Hemangiomas
- GYN Laparoscopy
Vaporization, incision, excision, ablation or photocoagulation of soft tissue in endoscopic surgery, including gynecological laparoscopy, for the treatment of:
- · Endometrial lesions, including ablation of endometriosis
- · Excision/lysis adhesions
- · Salpingostomy
- · Oophorectomy/ovariectomy
- · Fimbrioplasty
- · Metroplasty
- · Microsurgery (tubal)
- · Uterine myomas and fibroids
- · Ovarian fibromas and follicle cysts
- Uterosacral ligament ablation
- · Hysterectomy
Neurosurgerv
Laser incision, excision, ablation and/or vaporization of soft tissue in neurosurgery for the treatment of: Cranial
- · Posterior fossa tumors
- · Peripheral neurectomy
- · Benign and malignant turnors and cysts, for example, gliomas, meningiomas (including basal turnors), acoustic neuromas, lipomas, and large tumors
- · Arteriovenous malformation
- · Pituitary gland tumors (transphenoidal approach)
Spinal cord
- · Incision/excision and vaporization of benign and malignant tumors and cysts
- · Intra and extradural lesions
- · Laminectomy/laminotomy/microdiscectomy
Orthopedic
Incision/excision and vaporization of soft tissue in orthopedic surgery, including the following applications:
- · Arthroscopy
- · Menisectomy
- Chondromalacia
- · Chondroplasty
- · Ligament release (lateral and other)
- · Excision of plica
- · Partial synovectomy
### General
- · Debridement of traumatic wounds
- · Debridement of decubitus and diabetic ulcers
- · Microsurgery
{8}------------------------------------------------
- PMMA removal
General and Thoracic Surgery
Incision, excision and vaporization of soft tissue in general and thoracic surgery including endoscopic and open procedures. Applications include:
- · Debridement of decubitus ulcers, stasis, diabetic, and other ulcers
- · Mastectomy
- · Debridement of burns
- · Rectal and anal hemorrhoidectomy
- · Breast biopsy
- · Reduction mammoplasty
- · Cytoreduction for metastatic disease
- · Laparotomy and laparoscopic applications
- · Mediastinal and thoracic lesions and abnormalities
- · Skin tag vaporization
- Atheroma
- · Cysts, including sebaceous cysts, pilar cysts, and mucous cysts of the lips
- · Pilonidal cyst removal and repair
- · Abscesses
- · Other soft tissue applications
### Dental and Oral Surgery
Incision/excision and vaporization of soft tissue in dentistry and oral surgery. Applications include:
- · Gingivectomy/removal of hyperplasias
- · Gingivoplasty
- · Incisional and excisional biopsy
- · Treatment of ulcerous lesions, including aphthous ulcers
- · Incision of infection when used with antibiotic therapy
- · Frenectomy (frenum release)
- · Excision and ablation of benign and malignant lesions
- · Homeostasis
- · Operculectomy
- · Crown lengthening
- Removal of soft tissue, cysts and tumors
- · Oral cavity tumors and hemangiomas
- · Abscesses
- · Extraction site hemostasis
- · Salivary gland pathologies
- · Preprosthetic gum preparation
- Leukoplakia
- · Partial glossectomy
- · Periodontal gum resection
### Genitourinary
Incision/excision and vaporization of soft tissue in genitourinary procedures. Applications include:
- · Benign and malignant lesions of external genitalia
- · Condyloma
- Phimosis
- · Erythroplasia
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
<sup>·</sup> Artificial joint revision
{9}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Lumenis. The logo consists of a blue circle with a white "L" inside, followed by the word "Lumenis" in blue. Below the word "Lumenis" is the tagline "Energy to Healthcare" in a smaller font.
# 5. 510(K) SUMMARY
UltraPulse Family of CO2 Laser System, Delivery Devices and Accessories
| Applicant Name: | Lumenis Be Ltd.<br>6 Hakidma Street PO Box 240<br>Yokneam Industrial Park,<br>Yokneam 2069204, Israel<br>Tel: +972-4-9599000<br>Fax: +972-4-9599046 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Shlomit Segman, Lumenis Be Ltd.<br>6 Hakidma Street PO Box 240,<br>Yokneam Industrial Park,<br>Yokneam 2069204, Israel<br>Tel: +972-4-9599230<br>Fax: +972-4-9599198<br>Email:Shlomit.Segman@lumenis.com |
| Date Prepared: | February 17th, 2022 |
| Trade Name: | UltraPulse Family of CO2 Surgical and Aesthetic Lasers, Delivery Devices<br>and Accessories |
| Classification Name: | Powered laser surgical instrument |
| Product Code: | GEX |
| Device Class: | Class II |
| Regulation Number: | 21 CFR 878.4810 |
| Panel: | General & Plastic Surgery |
| Predicate Devices: | Primary: K203544 Lumenis Modified Family of UltraPulse CO2 Surgical<br>and Aesthetic Laser Systems<br>Reference: K030147 and K022060 Lumenis Be UltraPulse SurgiTouch<br>and Encore CO2 Lasers, Delivery Devices and Accessories. |
### Intended Use/ Indications for Use:
The indications for use for the subject devices are as attached (see FDA Indications for Use Forms 3881.)
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the Lumenis logo. The logo consists of a blue circle with a white "L" inside of it, followed by the word "Lumenis" in blue. Below the word "Lumenis" is the phrase "Energy to Healthcare" in a smaller font.
### Device Description
The Lumenis Be Family of UltraPulse CO2 Surgical and Aesthetic Laser Systems, Delivery Devices and Accessories (subject devices) consist of the following devices:
#### Laser Systems
- o UltraPulse Alpha (also written as UltraPulse α)
- UltraPulse Encore o
- UltraPulse SurgiTouch o
### Delivery Devices and Accessories
- o DeepFX Microscanner and disposable tips
- UltraScan CPG Microscanner o
- o TrueSpot 2.0 mm Collimated Handpiece
- o 0.2 mm and 1.0 mm Focused Incisional Handpieces (also called Standard Handpieces)
Each UltraPulse laser system is an advanced computer-controlled device with RF - modulated CO2 laser tube technology that emits laser beams at a wavelength of 10,600 or 11,100 nm. Each has an additional diode laser in the red visible spectrum emitting along the same optical path as the CO2 to serve as an aiming beam.
Attached to each laser system are the delivery devices and accessories that direct the laser beams to the intended treatment site in the format as selected by the user.
### Substantial Equivalence Discussion
In comparison to the predicate Lumenis Family of UltraPulse CO₂ Surgical and Aesthetic Laser Systems, Delivery Devices and Accessories (cleared under K203544), the subject Lumenis Be Family of UltraPulse CO2 Surgical and Aesthetic Laser Systems, Delivery Devices and Accessories share the identical technological characteristics, operating principles, and manufacturing process.
Only labeling changes are to be implemented with this 510(k). The key labeling change is to reflect the new name for the manufacturer, which is Lumenis Be. In addition, there will be labeling changes to reflect the UltraPulse a ("Alpha") unit that supports only a subset of the indications cleared for its predecessor, predicate UltraPulse Model, as well as additional clarifications and incorporating the individual instruction booklets for the accessories into the laser manual.
### Summary of Testing
For the labeling changes described in the 510(k), no testing was required.
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the Lumenis logo. The logo consists of a blue circle with a white "L" inside, followed by the word "Lumenis" in blue. Below the word "Lumenis" is the phrase "Energy to Healthcare" in a smaller font.
# Conclusion
None of the identified labeling changes raises a new safety or effectiveness questions. The Lumenis Be Family of UltraPulse CO2 Surgical and Aesthetic Laser Systems, Devices and Accessories (subject devices) is substantially equivalent to the Lumenis Family of UltraPulse CO2 Surgical and Aesthetic Laser Systems, Delivery Devices and Accessories (predicate devices).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.