PENTAX Medical EPK-3000 Imaging System with Camera Head PVK-J10, PENTAX Medical ENT Imaging System with Camera Head PVK-J10
K220465 · Pentax of America, Inc. · EOB · Nov 16, 2022 · Ear, Nose, Throat
Device Facts
Record ID
K220465
Device Name
PENTAX Medical EPK-3000 Imaging System with Camera Head PVK-J10, PENTAX Medical ENT Imaging System with Camera Head PVK-J10
Applicant
Pentax of America, Inc.
Product Code
EOB · Ear, Nose, Throat
Decision Date
Nov 16, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4760
Device Class
Class 2
Indications for Use
PENTAX Medical EPK-3000 Video Processor is intended to be used with the PENTAX Medical camera heads with PENTAX sinuscopes (8890, 8892, 8893), PENTAX Medical VNL8-J10, VNL11-J10, and VNL15-J10 endoscopes, PENTAX Medical Laryngeal Strobe, video monitors and other ancillary equipment for ENT endoscopic observation and ENT diagnosis, treatment and video observation with or without stroboscopy. The PENTAX Medical EPK-3000 Video Processor includes PENTAX i-Scan™, a digital, post-processing imaging enhancement technology. i-Scan is intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling. The PENTAX Medical Video Processor (EPK-i5010) is intended to be used with the PENTAX Medical camera heads with PENTAX sinuscopes (8890, 8891, 8892, 8893), PENTAX Medical VNL-1570STK, VNL8-J10, VNL11-110, and VNL15-J10 endoscopes, PENTAX Medical Laryngeal Strobe, video monitors and other ancillary equipment for ENT endoscopic observation and ENT diagnosis, treatment and video observation with or without stroboscopy. The PENTAX Medical EPK-i5010 Video Processor includes PENTAX i- Scan™, digital, post-processing imaging enhancement technology.
Device Story
System captures images via endoscope sensor or rigid sinuscope coupled with PVK-J10 camera head; signals processed by EPK-3000 or EPK-i5010 video processor; output displayed on monitor. Processor provides i-Scan digital post-processing for image enhancement. Used in clinical ENT settings by physicians for observation and diagnosis. Xenon lamp provides illumination via light guide. Camera head features zoom function (6.5mm-13mm). System allows adjustment of brightness, color, and contrast. Benefits include expanded compatibility with rigid scopes and enhanced visualization of mucosal structures.
Clinical Evidence
Clinical image capture study performed as part of optical/color performance testing. Representative models (PVK-J10 with Sinuscopes 8893 and 8891) compared to predicate systems. Results indicate subject device visualizes structure, vascularity, and mucosal surface as well or better than predicate.
Technological Characteristics
System includes video processor, camera head (PVK-J10), and endoscopes. Camera head uses CCD sensor. Illumination via xenon lamp. Connectivity: wired connection between camera head, processor, and monitor. Reprocessing: high-level disinfection (System 1E). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18. Software: i-Scan digital post-processing enhancement.
Indications for Use
Indicated for ENT endoscopic observation, diagnosis, treatment, and video observation (with or without stroboscopy) in patients requiring ENT procedures.
Regulatory Classification
Identification
A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.
Predicate Devices
PENTAX Medical EPK-3000 Video Imaging System (K172156, K182846)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 16, 2022
PENTAX of America, Inc. William Goeller Vice President, Ouality Assurance and Regulatory Affairs 3 Paragon Drive Montvale, New Jersey 07645-1782
Re: K220465
Trade/Device Name: PENTAX Medical EPK-3000 Video Imaging System PENTAX Medical ENT Video Imaging System Regulation Number: 21 CFR 874.4760 Regulation Name: Nasopharyngoscope (Flexible or Rigid) and Accessories Regulatory Class: Class II Product Code: EOB. PEA. EOL. OUG Dated: October 14, 2022 Received: October 17, 2022
Dear William Goeller:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Joyce C. Lin -S
for Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220465
Device Name PENTAX Medical EPK-3000 Video Imaging System
### Indications for Use (Describe)
PENTAX Medical EPK-3000 Video Processor is intended to be used with the PENTAX Medical camera heads with PENTAX sinuscopes (8890, 8892, 8893), PENTAX Medical VNL8-J10, VNL11-J10, and VNL15-J10 endoscopes, PENTAX Medical Laryngeal Strobe, video monitors and other ancillary equipment for ENT endoscopic observation and ENT diagnosis, treatment and video observation with or without stroboscopy.
The PENTAX Medical EPK-3000 Video Processor includes PENTAX i-Scan™, a digital, post-processing imaging enhancement technology. i-Scan is intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling.
**Type of Use (Select one or both, as applicable)**X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220465
Device Name PENTAX Medical ENT Video Imaging System
Indications for Use (Describe)
The PENTAX Medical Video Processor (EPK-i5010) is intended to be used with the PENTAX Medical camera heads with PENTAX sinuscopes (8890, 8891, 8892, 8893), PENTAX Medical VNL-1570STK, VNL8-J10, VNL11-110, and VNL15-J10 endoscopes, PENTAX Medical Laryngeal Strobe, video monitors and other ancillary equipment for ENT endoscopic observation and ENT diagnosis, treatment and video observation with or without stroboscopy.
The PENTAX Medical EPK-i5010 Video Processor includes PENTAX i- Scan™, a digital, post-processing imaging enhancement technology.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for Pentax Medical. The word "PENTAX" is in red, bold, sans-serif letters. Below that is a gray line, and below that is the word "MEDICAL" in gray, bold, sans-serif letters. The logo is simple and modern.
## 5. 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 221 CFR 807.92. All data included in this document is accurate and complete to the best of PENTAX Medical's knowledge.
#### 1 SUBMITTER
- Applicant: PENTAX Medical HOYA Corporation PENTAX Division 3 Paragon Drive Montvale, New Jersey 07645-1782
- Contact: William Goeller Vice President, Quality and Regulatory Affairs PENTAX Medical 3 Paragon Drive Montvale, New Jersey 07645-1782 Telephone: (201)571-2300 ext. 2318 FAX: (201)391-4189 Email: william.goeller@pentaxmedical.com
Date Prepared: 02/08/2022
#### SUBJECT DEVICE 2
PENTAX Medical is seeking clearance of two Imaging Systems that include a new camera head product, the PENTAX Medical Camera Head PVK-J10.
This product is used in conjunction with 3rd party non-video endoscopes (such as a rigid scope) and converts light captured by the non-video scope into video images through a processor, to be displayed on a monitor. Addition of the camera head expands utility of the video imaging systems by providing compatibility with additional ENT scopes.
PENTAX Medical Camera Head PVK-J10 is compatible with following two systems:
- PENTAX Medical EPK-3000 Video Imaging System ●
- PENTAX Medical ENT Video Imaging System ●
Compatible Endoscopes for above systems are as follows;
- Video Endoscopes : VNL-1570STK, VNL15-J10, VNL11-J10, VNL8-J10 ●
- Add-on Camera Head: PENTAX Medical camera head PVK-J10 with previously cleared Sinuscope models 8890, 8891,8892, 8893.
The regulatory classification of PENTAX Medical Camera Head PVK-J10 is identified in Table 5.1.
{5}------------------------------------------------
| Table 5.1: Regulatory Classification of PENTAX Medical Imaging Systems with PVK-J10 |
|-------------------------------------------------------------------------------------|
| Camera Head |
| Device Names | PENTAX Medical EPK-3000 Video Imaging System |
|----------------------|-------------------------------------------------------|
| Common Name | Naso-Pharyngo-Laryngoscope |
| Classification Name | Nasopharyngoscope (Flexible or Rigid) and Accessories |
| Regulation No. | 874.4760 |
| Device Class | II |
| Product Code | EOB, PEA, EQL, OUG |
| Classification Panel | Ear Nose & Throat |
| Device Names | PENTAX Medical ENT Video Imaging System |
|----------------------|-------------------------------------------------------|
| Common Name | Naso-Pharyngo-Laryngoscope |
| Classification Name | Nasopharyngoscope (Flexible or Rigid) and Accessories |
| Regulation No. | 874.4760 |
| Device Class | II |
| Product Code | EOB, PEA, EQL, OUG |
| Classification Panel | Ear Nose & Throat |
#### 3 PREDICATE DEVICE
The Predicate Devices for each Subject Devices are identified in Table 5.2. And Figure 5.1 shows system difference between Predicate Devices and Subject Devices Table 5.2: Subject Device and Predicate Device relation
| Subject Device | Predicate Device | Predicate Device 510(k)<br>Clearance Number |
|----------------------------------------------------|----------------------------------------------------|---------------------------------------------|
| PENTAX Medical<br>EPK-3000 Video Imaging<br>System | PENTAX Medical<br>EPK-3000 Video Imaging<br>System | K172156, K182846 |
| PENTAX Medical<br>ENT Video Imaging System | PENTAX Medical<br>ENT Video Imaging System | K183691 |
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the logo for Pentax Medical. The word "PENTAX" is in red, bold, sans-serif letters. Below it is a gray line, and below that is the word "MEDICAL" in gray, bold, sans-serif letters. The logo is simple and modern.
Image /page/6/Figure/3 description: The image shows a comparison between a subject device and a predicate device. The subject device includes a PENTAX rigid scope with the label (Schoelly: K142249), an adaptor AP-RV3, a PVK-J10, a PENTAX light guide cable labeled (510(k) exempt), a connector 8879 (K944656), an EPK-3000/EPK-i5010 video processor light source, a monitor, and a 9310HD/9372HD video capture module. The predicate device includes a VNL11-J10, an EPK-3000/EPK-i5010 video processor light source, a monitor, and a 9310HD/9372HD video capture module.
Figure 5.1. System configuration difference Predicate Devices and Subject Devices
#### DEVICE DESCRIPTION 4
PENTAX Medical EPK-3000 and ENT Video Imaging Systems are used for ENT endoscopic observation and nasopharyngo-laryngoscopic (ENT) diagnosis, treatment, and video observation. The System functions by receiving image signals from the image sensor in an endoscope, which are processed within a video processor and then output to a monitor. Brightness, color balance, and other properties of the displayed images can be adjusted using the buttons on the system's control panel. The light from a xenon lamp at the distal end of the endoscope illuminates the body cavities of the patient through the endoscope connected to the video processor.
The new component to be added to the subject system is the PENTAX Medical Camera Head, model PVK-J10. The PVK-J10 is used in conjunction with a non-video endoscope, video processor, and light guide. The non-video endoscope is connected to the video processor via the camera head and the light guide. The light supplied from the video processor is transmitted to the distal end of the endoscope via the light guide and connector in order to illuminate the patient's body cavity. Light captured by the endoscope are converted into the video signals by the camera head, and the signals are transferred to the video processor to display the video images on the monitor.
The PVK-J10 camera head connects to rigid, non-video sinuscopes (K142249). It takes images formed on the ocular lens by CCD sensor inside the control head, and transmits the image signal to the video processor. Image signal from CCD sensor is adjusted in terms of color, contrast, etc.,
{7}------------------------------------------------
Image /page/7/Picture/2 description: The image shows the logo for Pentax Medical. The word "PENTAX" is in red, bold, sans-serif letters. Below the word "PENTAX" is a gray line. Below the gray line is the word "MEDICAL" in gray, bold, sans-serif letters.
inside the video processor, and displayed on video equipment such as a monitor.
The camera head has a zoom function with focal lengths of 6.5 mm for the wide end and 13mm for the telephoto end.
The video processors to be connected to the PVK-J10 camera head and light guides are PENTAX Medical Video Processors EPK-i5010 and EPK-3000, both of which have been cleared.
PENTAX Medical sinuscopes 8890, 8891,8892, 8893, manufactured by Schoelly Fiberoptic GmbH and cleared under K142249. The light guides 8899HT used to connect PVK-J10 with these endoscopes are also manufactured by Schoelly Fiberoptic GmbH. PENTAX Medical sell Schoelly Sinuscopes and light guide as private brand, Model Name combination are shown in Table 5.3
| Sinuscopes | | Light Guides | | K#<br>Schoelly |
|--------------------------------|-------------------------|--------------------------------|-------------------------|----------------|
| PENTAX Medical<br>Model Number | Schoelly<br>Item number | PENTAX Medical<br>Model Number | Schoelly<br>Item number | |
| 8893 | 31.0801a | 8899HT | 05.0090I.ht | K142249 |
| 8890 | 31.0804a | | | |
| 8892 | 31.0810a | 8898HT | 05.0088I.ht | |
| 8891 | 31.0811a | | | |
Table 5.3:PENTAX Private Brand Name and Schoelly Item Number Combination
#### 5 INTENDED USE AND INDICATIONS FOR USE
### PENTAX Medical EPK-3000 Video Imaging System
PENTAX Medical EPK-3000 Video Processor is intended to be used with the PENTAX Medical camera heads with PENTAX sinuscopes (8890, 8891, 8892, 8893), PENTAX Medical VNL8-J10, VNL11-J10, and VNL15-J10 endoscopes, PENTAX Medical Laryngeal Strobe, video monitors and other ancillary equipment for ENT endoscopic observation and ENT diagnosis, treatment and video observation with or without stroboscopy.
The PENTAX Medical EPK-3000 Video Processor includes PENTAX i-Scan™, a
digital, post-processing imaging enhancement technology. i-Scan is intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling.
### PENTAX Medical ENT Video Imaging System
The PENTAX Medical Video Processor (EPK-i5010) is intended to be used with the PENTAX Medical camera heads with PENTAX sinuscopes (8890, 8891, 8892, 8893),
{8}------------------------------------------------
Image /page/8/Picture/2 description: The image contains the logo for Pentax Medical. The word "PENTAX" is in red, bold, sans-serif font. Below "PENTAX" is a gray line. Below the gray line is the word "MEDICAL" in gray, bold, sans-serif font.
PENTAX Medical VNL-1570STK, VNL8-J10, VNL11-J10, and VNL15-J10 endoscopes, PENTAX Medical Laryngeal Strobe, video monitors and other ancillary equipment for ENT endoscopic observation and ENT diagnosis, treatment and video observation with or without stroboscopy.
The PENTAX Medical EPK-i5010 Video Processor includes PENTAX i- Scan™, digital, post-processing imaging enhancement technology.
#### COMPARISON OF TECHNOLOGY CHARACTERISTICS WITH THE 6 PREDICATE
The subject Imaging systems that include the PENTAX Medical PVK-J10 Camera Head are functionally equivalent to the predicate devices, PENTAX Medical EPK-3000 Video Imaging (K172156, K182846) and PENTAX Medical ENT Video Imaging System (K183691) . The main difference between the subject and the predicate are minor technological changes that add a compatible Camera Head with sinuscope to the predicate systems.
The changes in the subject device have been evaluated through performance testing and raise no issue of safety and effectiveness of the device as these differences have no effect on the performance, function or general intended use of the devices.
The optical components of the PENTAX Medical Camera Head have the same fundamental technology ( Software, Type of Image Sensor, CCD Pixel ) as the optical components of the predicate device. Both the PENTAX Medical Camera Head and the predicate device are intended for illuminating and viewing the inside of the human body. Subject devices that add PVK-J10 with sinuscope to the predicate system configurations have the same visual capabilities as the predicate systems. This has been successfully substantiated by a series of bench tests and the clinical image capture study.
Both subject and the predicate scopes are reprocessed by the user.
#### 7 NON-CLINICAL PERFORMANCE DATA
The PENTAX Medical Imaging Systems with PVK-J10 Camera Head have been successfully tested for their function, performance and safety as per FDA recognized consensus standards. The following performance data are provided in support of the substantial equivalence determination.
Operational and Reprocessing Instructions for Use are provided for the PVK-J10.
#### i. Reprocessing Validation
As result of the assessment, simulated use testing, cleaning, high level disinfecting and rinsing (after cleaning and after HLD) validation studies of PVK-J10 was conducted and
{9}------------------------------------------------
Image /page/9/Picture/2 description: The image is a logo for Pentax Medical. The word "PENTAX" is in red, and the word "MEDICAL" is in gray. There is a gray line separating the two words.
confirmed the effectiveness of reprocessing procedures in accordance with FDA's 2015 Final Guidance, Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling ("FDA's 2015 Reprocessing Guidance"). Acceptance criteria were established in accordance with AAMI TIR 30:2011 for amount of residual soil accumulation and extraction efficiency.
All acceptance criteria were satisfied.
- ii. Sterilization and Shelf Life
PENTAX Medical coordinated with STERIS Corporation to validate the use of System 1E liquid chemical sterilization for the sterilization of the PVK-J10. The device is not provided sterile, therefore, shelf-life is not applicable.
- iii. Biocompatibility
PVK-J10 is the camera head used in combination with scope and non-contact with patients, therefore biocompatibility evaluation is not applicable.
- iv. Software and Cybersecurity
Software verification and validation including cybersecurity assessments were conducted according to IEC 62304: 2006 and FDA Guidances for Industory and Staff "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.", "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" and "Postmarket Management of Cybersecurity in Medical Devices."
Electrical Safety and EMC V.
> The acceptable level of electrical safety (ES) and electromagnetic compatibility (EMC) for the PENTAX Medical Camera Head PVK-J10 was confirmed by the following standards: IEC 60601-1-2:2014; IEC 60601-1:2005+CORR 1:2006+CORR 2:2007+A1:2012; and IEC 60601-2-18:2009.
- vi. System Performance
The system performance of the subject devices demonstrated the equivalence to the predicate devices. Test results also demonstrated six years of the service life for the PVK-J10 and AP-RV3 (PENTAX Medical Endoscope Mount Adapter for Rigid Ocular Endoscopes)
- vii. Optical Performance
As a part of Design Verification and Validation, optical properties of imaging and illumination performances were measured for the PVK-J10 with representative Sinuscopes and light guides in conjunction with the EPK-3000 and EPK-i5010 Video Processors. All results show that the optical characteristics of the subject device is equivalent to those of the predicate device.
{10}------------------------------------------------
Image /page/10/Picture/2 description: The image contains the logo for Pentax Medical. The word "PENTAX" is in red, bold, sans-serif font. Below the word "PENTAX" is a gray line. Below the gray line is the word "MEDICAL" in gray, bold, sans-serif font.
#### viii. Photobiological Safety of lamps and lamp systems
PENTAX Medical evaluated the photobiological safety of lamps and lamp systems for the PVK-J10 with the representative Sinuscope and light guide in conjunction with the EPK-3000 and EPK-i5010 Video Processors. The results show that all hazards related to the photobiological safety are acceptable for the subject device.
- ix. Animal Image Capture Study N/A
- x. Clinical Image Capture Study
Clinical image capture study was performed as a part of optical and color performance testing. In this study, the Imaging systems with PVK-J10 and Sinuscope 8893 and PVK-J10 with Sinuscope 8891 were used as the representative model. The results indicate that the subject device is able to visualize structure, vascularity and mucosal surface as well or better than the predicate device.
#### 8 Substantial Equivalence Discussion
After analyzing the intended use, indications for use, technological characteristics (including fundamental operating principle, energy source, scientific technology, functional characteristics, design features, performance characteristics, and constituent materials), labeling, and sterilization method, PENTAX Medical concludes that the subject Imaging Systems with PVK-J10 camera head are as safe and effective as the predicate devices. There are no differences in indications for use and intended use between the subject and predicate devices The technological differences in terms of design features, performance characteristics and constituent materials are not substantive. Thus, the subject and predicate devices are substantially equivalent.
#### 9 CONCLUSION
Accordingly, PENTAX Medical believes the PENTAX Medical Imaging systems with PVK-110 Camera Head are substantially equivalent to the identified predicates. PENTAX Medical EPK-3000 video imaging system (K172156 and K182846) and PENTAX Medical ENT Video Imaging System ( K183691).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.