K220426 · Quanta System Spa · GEX · Mar 16, 2022 · General, Plastic Surgery
Device Facts
Record ID
K220426
Device Name
Fiber Dust PRO
Applicant
Quanta System Spa
Product Code
GEX · General, Plastic Surgery
Decision Date
Mar 16, 2022
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Fiber Dust PRO is intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy, gastroenterological surgery and gynecological surgery.
Device Story
Fiber Dust PRO is a Thulium laser surgical system (1940 nm, up to 60W) for soft tissue surgery and lithotripsy. It delivers laser energy via quartz optical fibers (up to 1000 um) to target tissues or calculi. The system comprises a laser source, power electronics, optical delivery, control electronics, and a cooling system. Operated by a physician in clinical settings (OR/clinic) via a footswitch, the device allows selection of settings for incision, ablation, or fragmentation. The output is laser radiation; the physician uses this to perform surgical procedures or stone fragmentation. The device benefits patients by providing a minimally invasive tool for urological, gastroenterological, and gynecological conditions. The PRO model is a mobile console version of the desktop predicate.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1 (general safety) and IEC 60601-1-2 (electromagnetic compatibility) confirmed. Software verification and validation performed per FDA guidance. Biocompatibility remains unchanged from predicate.
Technological Characteristics
Thulium laser source; 1.94 μm wavelength; 60W max average power; CW/pulsed emission; 532nm aiming beam (<5 mW). Quartz optical fiber delivery (up to 1000 um). Mobile console form factor. Standards: IEC 60601-1, IEC 60601-1-2. Software-controlled functions.
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 16, 2022
Quanta System Spa Dario Bandiera RA Manager Via Acquedotto, 109 Samarate, 21017 Italy
Re: K220426
Trade/Device Name: Fiber Dust PRO Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: February 7, 2022 Received: February 14, 2022
Dear Dario Bandiera:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K220426
Device Name Fiber Dust PRO
### Indications for Use (Describe)
Fiber Dust PRO is intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy, gastroenterological surgery and gynecological surgery.
Urology
- · Ablation of Benign Prostatic Hyperplasia (Hypertrophy) [BPH]
- · Laser Resection of the Prostrate (LRP)
- · Laser Enucleation of the Prostate (LEP)
- · Laser Ablation of the Prostate (LAP)
- · Transurethral Incision of the Prostate (TUIP)
- Condylomas
- · Urethral strictures
- Lesions of external genitalia
- · Bladder neck incisions (BNI)
- · Ablation and resection of bladder tumors, urethral tumors, and ureteral tumors
- · Endoscopic fragmentation of urethral, ureteral, bladder, and renal calculi
- · Treatment of distal impacted fragments remaining in the ureters following lithotripsy
Lithotripsy and Percutaneous Urinary Lithotripsy Indications
- · Endoscopic fragmentation of ureteral, bladder and renal calculi including cystine, calcium oxalate,
- monohydrate and calcium oxalate dehydrate stones
- Endoscopic fragmentation of renal calculi
- · Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed
### Gastroenterology
Open and endoscopic gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and haemostasis) including:
- · Appendectomy
- · Angiodysplasia
- · Polyps
- · Colorectal cancer
- · Biopsy
- · Telangiectasias
- Gall Bladder calculi
- · Telangiectasias of the Osler-Weber-Renu disease
- Biliary/Bile duct calculi
- Vascular Malformation
- Ulcers
- · Gastritis
- · Gastric ulcers
- · Esophagitis
- · Duodenal ulcers
- · Esophageal ulcers
{3}------------------------------------------------
- Non Bleeding Ulcers
- · Varices
- Pancreatitis
- Colitis
- Haemorrhoids
- Mallory-Weiss tear
- ·Cholecystectomy
- · Gastric Erosions
- ·Benign and Malignant Neoplasm
### Gynecology
Open and laparoscopic gynecological surgery (incision, excision, ablation, vaporization, coagulation and hemostasis) of soft tissue.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
K220426
### 510(K) SUMMARY
| Applicant /<br>Manufacturer<br>Name and Address: | Quanta System S.p.A.<br>Via Acquedotto, 109<br>Samarate (VA) Italy,<br>21017 |
|--------------------------------------------------|------------------------------------------------------------------------------------------|
| 510(k) Contact Person: | Dario Bandiera<br>Regulatory Affairs Manager<br>Quanta System S.p.A. |
| | Email: dario.bandiera@quantasystem.com<br>Phone: +39-0331-376797<br>Fax: +39-0331-367815 |
| Date Prepared: | 7thFebruary 2022 |
| Trade Name: | Fiber Dust PRO |
| Device: | Powered Laser Surgical Instrument |
| Classification: | Class II |
| Classification Name: | Laser surgical instrument for use in general and plastic<br>surgery and in dermatology. |
| Regulation Number: | 21 CFR 878.4810 |
| Product Code: | GEX |
| Basis for submission: | Addition of a new model |
| Predicate Device | FiberDust - Quanta System S.p.A. (K210142) |
### Description of the device:
Fiber Dust PRO is a laser system that emits a wavelength of 1940 nm with a laser output power up to 60W. Laser radiation is delivered to the patient via a quartz optical fiber having a diameter up to 1000 um. The main subsystems of the device are the laser, the power electronics, the optical delivery system, the control electronics, and the cooling system. Software controls the device functions and allows the user select device settings. Laser emission is triggered by a footswitch.
{5}------------------------------------------------
### Device comparison
The proposed modified version is a mobile console configuration and the predicate device is a desktop configuration.
Technical specifications are the same or similar
| | Proposed device | Predicate device | Comparison to<br>predicate device |
|-------------------|----------------------|----------------------|----------------------------------------------------------------------------------|
| model name | Fiber Dust PRO | Fiber Dust | - |
| manufacturer | Quanta System S.p.A. | Quanta System S.p.A. | - |
| 510(k) | - | K210142 | - |
| Laser Source | Thulium laser | Thulium laser | Same technology |
| Wavelength (nm) | 1.94 μm | 1.94 μm | Same |
| Emission | CW/pulsed | CW/pulsed | Same |
| Pulse duration | Up to 15ms | Up to 15ms | Same |
| Frequency | Up to 2500 Hz | Up to 2500 Hz | Same |
| Max average power | 60 W | 60 W | Same |
| Configuration | Mobile | Desktop | Conformance with<br>IEC 60601-1 and IEC<br>60601-1-2<br>performance<br>standards |
| Delivery system | Optical fibers | Optical fibers | Same |
| Aiming beam | 532nm laser < 5 mW | 532nm laser < 5 mW | Same |
### Intended use
The intended use and the indications for use of the modified device and the unmodified device are exactly the same, as follows.
Fiber Dust PRO is intended for incision, excision, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy, gastroenterological surgery and gynecological surgery.
{6}------------------------------------------------
# K220426
## Urology
- · Ablation of Benign Prostatic Hyperplasia (Hypertrophy) [BPH]
- · Laser Resection of the Prostrate (LRP)
- · Laser Enucleation of the Prostate (LEP)
- · Laser Ablation of the Prostate (LAP)
- · Transurethral Incision of the Prostate (TUIP)
- Condylomas
- Urethral strictures
- Lesions of external genitalia
- · Bladder neck incisions (BNI)
- Ablation and resection of bladder tumors, urethral tumors, and ureteral tumors
- · Endoscopic fragmentation of urethral, ureteral, bladder, and renal calculi
- Treatment of distal impacted fragments remaining in the ureters following lithotripsy
## Lithotripsy and Percutaneous Urinary Lithotripsy Indications
· Endoscopic fragmentation of ureteral, bladder and renal calculi including cystine, calcium oxalate, monohydrate and calcium oxalate dehydrate stones
- Endoscopic fragmentation of renal calculi
- Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed
## Gastroenterology
Open and endoscopic gastroenterology surgery (incision, resection, ablation, vaporization, coagulation and haemostasis) including:
- Appendectomy
- · Angiodysplasia
- Polyps
- Colorectal cancer
- · Biopsy
- · Telangiectasias
- Gall Bladder calculi
- · Telangiectasias of the Osler-Weber-Renu disease
- Biliary/Bile duct calculi
- Vascular Malformation
- · Ulcers
- Gastritis
- Gastric ulcers
- Esophagitis
- Duodenal ulcers
- Esophageal ulcers
- Non Bleeding Ulcers
- Varices
- Pancreatitis
- · Colitis
{7}------------------------------------------------
- Haemorrhoids
- Mallory-Weiss tear
- •Cholecystectomy
- Gastric Erosions
- •Benign and Malignant Neoplasm
### Gynecology
Open and laparoscopic gynecological surgery (incision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue
### Performance data:
Based on the nature of the modified device was subjected to performance testing in accordance with the following recognized consensus standards:
- i IEC 60601-1:2005 MOD Medical Electrical equipment Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2 Edition 4.0 2014-02 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests
The modified device passed all the required testing and is in compliance will all applicable sections of the above mentioned performance standards.
### Software Verification and Validation Testing
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
### Biocompatibility:
There are no new patient-contacting materials used in the K210142 device compared to those used in the Fiber Dust PRO device that adversely affect biocompatibility.
### Substantial Equivalence:
The modifications made to produce the Fiber Dust PRO do not raise new types of questions regarding the safety and effectiveness of the device for the proposed indications for use, and the performance testing supports that the device can be used safely and effectively for the proposed indications for use. The device is considered to be substantially equivalent to the predicate device K210142.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.