PAPAYA & PAPAYA Plus are digital extraoral source X-ray system intended to produce panoramic and cephalometric(option) images of the oral and craniofacial anatomy for a precise treatment planning in adult and pediatric care. The system is used for dental & skull radiographic examination and diagnosis of teeth, jaw, oral structure, and skull by exposing an X-ray image receptor to ionizing radiation, with a digital imaging capability for taking both panoramic and cephalometric images. And This system can be equipped cust(Tomographic) option, which is capable of taking crosssectional radiographic images provide dimensional information for dental implant planning and information about location of impacted teeth.
Device Story
Digital extraoral X-ray system; acquires panoramic, cephalometric, and tomographic (CUST) images of oral/craniofacial anatomy. Input: ionizing radiation exposure to digital CMOS sensor. Transformation: system captures projection radiographs; optional CUST mode reconstructs cross-sectional images. Output: 2D/3D dental/skull images for clinical review. Used in dental clinics; operated by trained dental professionals. Software (Triana/Theia) processes images; provides 3D visualization, 2D analysis, MPR, and measurement tools. Output facilitates treatment planning (implants, impacted teeth). Benefits: precise diagnostic imaging for dental/craniofacial conditions.
Clinical Evidence
Bench testing only. Performance verified via electrical, mechanical, and environmental safety testing (IEC 60601-1, 60601-1-3, 60601-2-63). Image quality validated via MTF, DQE, and dynamic range measurements for panoramic and cephalometric sensors per FDA guidance for Solid State X-ray Imaging Devices. Software validated per IEC 62304. Cybersecurity and DICOM compliance (NEMA PS 3.1-3.20) confirmed.
Indicated for adult and pediatric patients requiring dental and skull radiographic examination and diagnosis of teeth, jaw, oral structure, and skull, including treatment planning for dental implants and localization of impacted teeth.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's name on the right. The symbol on the left is the Department of Health & Human Services logo. The right side of the logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
GENORAY Co., Ltd. % Kaitlynn Min Business Development GENORAY America Inc. 147 E. Bristol Lane ORANGE CA 92865
Re: K220423
Trade/Device Name: PAPAYA & PAPAYA Plus Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: Class II Product Code: MUH Dated: February 17, 2022 Received: February 18, 2022
Dear Kaitlynn Min:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
May 19, 2022
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Laurel Burk, Ph.D. Assistant Director Diagnostic X-ray Systems Team DHT 8B: Division of Radiological Imaging and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K220423
Device Name PAPAYA & PAPAYA Plus
#### Indications for Use (Describe)
PAPAYA & PAPAYA Plus are digital extraoral source X-ray system intended to produce panoramic and cephalometric(option) images of the oral and craniofacial anatomy for a precise treatment planning in adult and pediatric care. The system is used for dental & skull radiographic examination and diagnosis of teeth, jaw, oral structure, and skull by exposing an X-ray image receptor to ionizing radiation, with a digital imaging capability for taking both panoramic and cephalometric images. And This system can be equipped cust(Tomographic) option, which is capable of taking crosssectional radiographic images provide dimensional information for dental implant planning and information about location of impacted teeth.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
#### Exhibit 5 510(k) Summary
Date of Summary Preparation: Feb 16, 2022
## 1. Submitter and US Official Correspondent
| Submitter | : GENORAY Co., Ltd. |
|---------------|-----------------------------------------------------------------------|
| Address | : 512, 560, Dunchon-daero, Junwong-gu, Seongnam-si, Gyeonggi-Do Korea |
| Telephone No. | : +82-31-5178-5770 |
| Fax | : +82-31-5178-5699 |
Official Correspondent (U.S): Kaitlynn Min - Business Manager
| Correspondent | : GENORAY America Inc. |
|---------------|---------------------------------------------|
| Address | : 147 E. Bristol Lane, Orange, CA 92865 USA |
| Telephone No. | : +1-855-436-6729 |
| Fax | : +1-714-786-8919 |
| Email | : kaitlynn@genorayamerica.com |
## 2. Establishment Registration Number
: 3005843418
## 3. Device Information
Trade Name / Proprietary Name: PAPAYA & PAPAYA Plus Common Name: Digital X-ray Imaging System Classification Name / Product Code: Extraoral Source X-ray System / MUH Regulation Number: 21 CFR 872.1800 Device Class: Class II
## 4. Predicate Device (Equivalent Legally Marketed Device)
Trade Name / Proprietary Name: PAPAYA Plus (K141700, GENORAY Co., Ltd) Common Name: Digital X-ray Imaging System Classification Name / Product Code: Extraoral Source X-ray System / MUH Regulation Number: 21 CFR 872.1800 Device Class: Class II
# 5. Reference Device (Equivalent Legally Marketed Device)
Trade / Proprietary Name : Triana (K103182) Device Class: Class II Regulation Number : 21 CFR 892.2050 Classification Name / Product Code: System, Image Processing, Radiological / LLZ Indications for Use(IFU) : Triana is intended for use as a software package which obtains medical images from CT,
Cephalometric / Panoramic X-ray system & etc., stores those and provides 3D visualization, 2D analysis, various MPR(Multi-Planar Reconstruction) functions for further rapid and precise diagnosis.
# 6. Description of the Device
The proposed devices PAPAYA & PAPAYA Plus are diagnostic imaging system which consists of multiple image acquisition modes; panorama and cephalometric. The difference beteween PAPAYA & PAPA YA Plus is only optional of the cephalometric detector. (Without cephalometric detector, we model named PAPAYA, and with cephalometric detector, we model named PAPAYA Plus.)
The proposed devices have the CUST imaging option which is used to reconstruct tomographic images from a set of pre-acquired projection radiographic images of the object.
{4}------------------------------------------------
The differences from predicate device(K141700) are change of power voltage, addition of image processing software(Theia, Triana).
First of all, the input voltage (power rating) of predicate device was 120V~, 60Hz. However due to change of the power board, proposed device is available to 100-240V~. 50/60Hz. Due to this power specification change, critical components change was verified via safety/EMC test report performed by CSA and SGS according to IEC 60601-1 (3.0 ed), IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-63.
Second, the proposed device can be used with image processing software, Triana(K103182) or Thiea. Triana (K103182) and Theia are image processing software that processes image acquired through PAPAYA & PAPAYA Plus. The two software uses not only the purpose of use, but also acquires and stores the image, and Inquires and processes the acquired image (deletion, enlargement, reduction, distance and angle measurement, drawing information in the image, 3D image reconstruction, etc.), it is the same in that it has the function to finally transmit to the DICOM Server. However, in Theia, by rearranging the configuration within the UI to improve the user's convenience, there is only a visual difference between Triana and Theia were validated throughout IEC 62304:2006/AC: 2008.
### 7. Indications for use
PAPAYA & PAPAYA Plus are a digital extraoral source X ray system intended to produce panoramic and cephalometric(option) images of the oral and craniofacial anatomy for a precise treatment planning in adult and pediatric care. The system is used for dental & skull radiographic examination and diagnosis of teeth, jaw, oral structure, and skull by exposing an X ray image receptor to ionizing radiation, with a digital imaging capability for taking both panoramic and cephalometric images. And this system can be equipped CUST (Tomographic) option, which is capable of taking cross sectional radiographic images. These images provide dimensional information for dental implant planning and information about location of impacted teeth
{5}------------------------------------------------
# 7. Substantial equivalence chart
| Criteria | Proposed device | | Predicate Device | | |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|--|
| Name | PAPAYA & PAPAYA Plus | | PAPAYA Plus | | |
| Manufacturer | GENORAY Co., Ltd. | | GENORAY Co., Ltd. | | |
| 510(k) No. | - | | K141700 | | |
| Figure | Image: PAPAYA | Image: PAPAYA Plus | Image: PAPAYA | Image: PAPAYA Plus | |
| | <PAPAYA> | <PAPAYA Plus> | <PAPAYA> | <PAPAYA Plus> | |
| Indications<br>for use | PAPAYA & PAPAYA Plus are digital extraoral source X-ray<br>system intended to produce panoramic and cephalometric(option)<br>images of the oral and craniofacial anatomy for a precise treatment<br>planning in adult and pediatric care. The system is used for dental &<br>skull radiographic examination and diagnosis of teeth, jaw, oral<br>structure, and skull by exposing an X-ray image receptor to ionizing<br>radiation, with a digital imaging capability for taking both<br>panoramic and cephalometric images. And This system can be<br>equipped cust(Tomographic) option, which is capable of taking<br>cross-sectional radiographic images. These images provide<br>dimensional information for dental implant planning and information<br>about location of impacted teeth. | | PAPAYA & PAPAYA Plus are digital extraoral source X-ray<br>system intended to produce panoramic and cephalometric(option)<br>images of the oral and craniofacial anatomy for a precise treatment<br>planning in adult and pediatric care. The system is used for dental &<br>skull radiographic examination and diagnosis of teeth, jaw, oral<br>structure, and skull by exposing an X-ray image receptor to ionizing<br>radiation, with a digital imaging capability for taking both<br>panoramic and cephalometric images. And This system can be<br>equipped cust(Tomographic) option, which is capable of taking<br>cross-sectional radiographic images. These images provide<br>dimensional information for dental implant planning and<br>information about location of impacted teeth. | | |
| Performance Specification | Panoramic | Panoramic and Cephalometric | Panoramic | Panoramic and Cephalometric | |
| Input Voltage | 100-240 V~, 50/60Hz | 100-240 V~, 50/60Hz | 120 V~, 60Hz | 120 V~, 60Hz | |
| Tube Voltage | 60-90 kV | 60-90 kV | 60-90 kV | 60-90 kV | |
| Tube Current | 4-12 mA | 4-12 mA | 4-12 mA | 4-12 mA | |
| Focal Spot Size | 0.5 mm | 0.5 mm | 0.5 mm | 0.5 mm | |
| Exposure Time | Panorama: max 17 sec. | Panorama: max 17 sec.<br>Cephalo : max 12 sec. | Panorama: max 17 sec. | Cephalo : max 12 sec. | |
| Exposure mode | Panorama: Panoramic mode<br>TMJ mode<br>SINUS mode<br>CUST mode | Panorama: Panoramic mode<br>TMJ mode<br>SINUS mode<br>CUST mode<br>Cephalo: Cephalo mode | Panorama: Panoramic mode<br>TMJ mode<br>SINUS mode<br>CUST mode | Panorama: Panoramic mode<br>TMJ mode<br>SINUS mode<br>CUST mode<br>Cephalo: Cephalo mode | |
| Image Receptor | Panoramic sensor : CMOS | Panoramic & cephalometric<br>sensor : CMOS | Panoramic sensor : CMOS | Panoramic & cephalometric<br>sensor : CMOS | |
| Image<br>processing<br>Software | S/W | Triana(K103182) | Theia | | |
| | Intended<br>Use | Triana and Theia are intended for use as a software package<br>software which obtains medical images from CT,<br>Cephalometric / Panoramic X-ray system & etc., stores those<br>and provides 3D visualization, 2D analysis, various MPR<br>(Multi-Planar Reconstruction) functions for further rapid and<br>precise diagnosis. | | | |
| | General<br>Function | - obtaining medical images<br>- storing medical images<br>- Inquires and processes the<br>acquired image (deletion,<br>enlargement, reduction,<br>rotation, distance and<br>anglemeasurement, drawing<br>information in the image)<br>- 3D visualization<br>- 2D analysis<br>- various MPR (Multi-Planar<br>Reconstruction) | - obtaining medical images<br>- storing medical images<br>- Inquires and processes the<br>acquired image (deletion,<br>enlargement, reduction,<br>rotation, distance and<br>anglemeasurement, drawing<br>information in the image)<br>- 3D visualiziation<br>- 2D analysis<br>- various MPR (Multi-Planar<br>Reconstruction) | | |
| Result | | | The proposed devices, PAPAYA & PAPAYA Plus, are was developed from the predicate device PAPAYA Plus (K141700), and<br>modifications are changing of input power, addition of image processing software.<br><br>First of all, the input voltage (power rating) of predicate device was 120V~, 60Hz. However due to change of the power board,<br>proposed device is available to 100-240V~. 50/60Hz. Due to this power specification change, critical components change was<br>verified via safety/EMC test report performed by CSA and SGS according to IEC 60601-1 (3.0 ed), IEC 60601-1-2, IEC 60601-<br>1-3, IEC 60601-2-63. | | |
{6}------------------------------------------------
{7}------------------------------------------------
| Second, the proposed device can be used with image processing software, Triana or Theia. The difference between Triana and |
|-----------------------------------------------------------------------------------------------------------------------------------|
| Theia is only UI, and the Theia was developed for marketing purpose only. Triana and Theia were validated throughout IEC |
| 62304:2006/AC: 2008. |
| Third, Theia is a newly developed software after the release of Triana, but as can be seen from the table above, both Triana and |
| Theia can be used for PAPAYA & PAPAYA Plus, and the technical characteristics and applications, including the Intended |
| Use, are essentially the same. can be judged to be equivalent. Also, the general functions of both software are the same overall. |
| The General function is the same for both Theia and Triana. However, Theia added 3D TMJ Screen (a function to view the |
| TMJ part separately) and Airway screen (a function to indicate the airway), which are detailed functions within the 3D |
| visualization function, to improve the user's convenience in the functions of Triana. |
{8}------------------------------------------------
## 8. Safety, EMC and Performance data comparison to Predicate
Device Safety, EMC and Performance data is same to predicate device which has been established in 510(k) submission K141700 as below. All test results were satisfactory.
- Electrical, mechanical, environmental safety and performance testing to standard IEC 60601-1, IEC 60601-1-3, , IEC 60601-1-6, IEC 60601-2-63, IEC 62366 and IEC 62304 were performed.
- EMC testing was conducted in accordance with standard IEC 60601-1-2.
- Content of Premarket Submissions for Management of Cybersecurity in Medical Devices Guidance for Industry and Food and Drug Administration Staff Document Issued on: October 2, 2014
- Guidance for Industry and FDA Staff Guidance for the Content of Premarket Submissions for Software contained in Medical Devices, Document issued on: May 11, 2005 Medical Devices, Document issued on: May 11, 2005
- Guidance for Industry and Food and Drug Administration Staff Pediatric Information for X-ray Imaging Device Premarket Notifications, November 2017
- FDA Guidance "Guidance for the submissions of 510(k)'s for Solid State X-ray Imaging Devices" was performed. The tests include the MTF, DQE and the dynamic range of the panoramic sensor and the cephalometric sensor.
- PAPAYA & PAPAYA Plus also meet the provisions of NEMA PS 3.1-3.20. Digital Imaging and Communications in Medicine (DICOM) Set.
- PAPAYA & PAPAYA Plus were tested for safety and effectiveness in Clinical Evaluation Report and bench.
#### 9.Conclusion
Proposed device has the same indication for use as the predicate device. And there have been no changes that impact either the fundamental technology or the indication for use. The proposed device with modification outlined in this Premarket Notification is substantially equivalent to the currently commercially available predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.