← Product Code [LBH](/productcode/LBH) · K220419

# Apapro Desensitizer (K220419)

_Sangi Co., Ltd. · LBH · Feb 16, 2022 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K220419

## Device Facts

- **Applicant:** Sangi Co., Ltd.
- **Product Code:** [LBH](/productcode/LBH.md)
- **Decision Date:** Feb 16, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3260
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

APAPRO Desensitizer is a fluoride-free paste that is indicated to provide relief from tooth hypersensitivity resulting from cold, heat, acids, sweets, or contact, through its action of dentin tubule occlusion. APAPRO Desensitizer is intended to be prescribed and applied by dental physicians for use in the Adult Population.

## Device Story

APAPRO Desensitizer is a fluoride-free paste containing 20% w/w synthetic hydroxyapatite; applied by dental professionals in a clinical setting. Device functions by physically depositing hydroxyapatite onto tooth surfaces to occlude exposed dentin tubules; acts as a template for further deposition of calcium and phosphate ions to promote crystal integrity. Output is the physical occlusion of tubules, which desensitizes teeth and provides relief from sensitivity triggers. Clinical benefit is the reduction of pain associated with tooth hypersensitivity.

## Clinical Evidence

Bench testing only. Non-clinical SEM micrographs compared dentin samples treated with APAPRO Desensitizer against those treated with the predicate (NovaMin). Results showed similar physical occlusion of dentin tubules via hydroxyapatite layer deposition, demonstrating comparable effectiveness.

## Technological Characteristics

Fluoride-free paste containing 20% w/w synthetic hydroxyapatite. Mode of action: physical occlusion of dentin tubules via hydroxyapatite deposition. Supplied non-sterile. Biocompatibility assessed per ISO 10993.

## Regulatory Identification

Cavity varnish is a device that consists of a compound intended to coat a prepared cavity of a tooth before insertion of restorative materials. The device is intended to prevent penetration of restorative materials, such as amalgam, into the dentinal tissue.

## Special Controls

*Classification.* Class II (special controls). The device, when it is an external cleaning solution, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.

## Predicate Devices

- OraliefTM Therapy for Sensitive Teeth ([K040858](/device/K040858.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.

February 16, 2022

Sangi Co, Ltd % Dave Yungvirt CEO Third Party Review Group, LLC 25 Independence Blvd Warren, New Jersey 07059

Re: K220419

Trade/Device Name: Apapro Desensitizer Regulation Number: 21 CFR 872.3260 Regulation Name: Cavity varnish Regulatory Class: Class II Product Code: LBH Dated: February 7, 2022 Received: February 14, 2022

Dear Dave Yungvirt:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K220419

Device Name APAPRO Desensitizer

Indications for Use (Describe)

APAPRO Desensitizer is a fluoride-free paste that is indicated to provide relief from tooth hypersensitivity resulting from cold, heat, acids, sweets, or contact, through its action of dentin tubule occlusion.

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### K220419

Image /page/3/Picture/2 description: The image is a logo for SANGI. The logo is a purple square with three horizontal white lines inside. Below the lines, the word "SANGI" is written in white, block letters.

SANGI CO., LTD., 3-11-6 TSUKIJI, CHUO-KU, TOKYO 104-8440, JAPAN TEL. +81 70-2680-2931 FAX. +81 3--3543-3651

Submitter Information:

| Name:             | Sangi Co., Ltd.                               |
|-------------------|-----------------------------------------------|
| Address:          | 3-11-6 Tsukiji, Chuo-ku, Tokyo 104-8440 Japan |
| Phone:            | +81 70-2680-2931                              |
| Facsimile:        | +81 3-3543-3651                               |
| Contact Person:   | Douglas Mercadante                            |
| Preparation Date: | 12 May 2021                                   |

- 1. Device Nomenclature:

| Trade Name:          | APAPRO Desensitizer                              |
|----------------------|--------------------------------------------------|
| Common Name:         | Tooth Desensitizing Paste                        |
| Classification Name: | Cavity Varnish (21 CFR 872.326 Product code LBH) |

- 2. Legally Marketed Predicate Device:

| Device Name:   | OraliefTM Therapy for Sensitive Teeth |
|----------------|---------------------------------------|
| 510(k) Number: | K040858                               |
| Applicant:     | NovaMin Technology, Inc.              |

- 3. Device Description:
APAPRO Desensitizer is a fluoride-free tooth desensitizing paste that utilizes hydroxyapatite, the main component of teeth, as its active ingredient to relieve tooth sensitivity by physically occluding exposed dentin tubules. The hydroxyapatite deposited into dentin tubules by APAPRO Desensitizer then acts as a template for the further deposition of large amounts of calcium and phosphate ions, promoting crystal integrity and growth.

{4}------------------------------------------------

Sangi Co., Ltd. Traditional 510(k) APAPRO Desensitizer

- 4. Indications for Use
APAPRO Desensitizer is a fluoride-free paste that is indicated to provide relief from tooth hypersensitivity resulting from cold, heat, acids, sweets, or contact, through its action of dentin tubule occlusion.

- 5. Intended Use
APAPRO Desensitizer is a fluoride-free paste that is indicated to provide relief from tooth hypersensitivity resulting from cold, heat, acids, sweets, or contact, through its action of dentin tubule occlusion. APAPRO Desensitizer is intended to be prescribed and applied by dental physicians for use in the Adult Population.

- 6. Technological Characteristics:
Both APAPRO Desensitizer and the selected predicate device Oralief™ Therapy for Sensitive Teeth are intended to be used to treat the same disorder, namely for tooth hypersensitivity, by the same mode of action, which is the deposition of a calcium hydroxyapatite layer onto the tooth surface resulting in the physical occlusion of dentin tubules. The two devices differ in that Oralief The Therapy for Sensitive Teeth uses calcium sodium phosphosilicate (NovaMin®) as its active ingredient, which when introduced to the physiological fluid of the mouth, reacts to form a hydroxycarbonate apatite layer on the teeth, whereas APAPRO Desensitizer introduces synthetic hydroxyapatite directly onto the tooth surface. APAPRO Desensitizer is to be prescribed and applied by a dental professional. Oralief Therapy for Sensitive Teeth is also a home-use device to be prescribed by dental professionals.

- 7. Safety and Performance Data:
The ability of APAPRO Desensitizer to provide relief from tooth hypersensitivity was shown by non-clinical tests demonstrating that it physically occludes exposed dentin tubules. The SEM micrographs of dentin samples with tubules occluded by APAPRO Desensitizer were compared to SEM Micrographs of NovaMintreated dentin samples. The SEM Micrographs of APAPRO Desensitizer and NovaMin-treated samples showed similar occlusion of the dentin tubules, which was achieved in both cases by the deposition of a hydroxyapatite layer onto the surface of the tooth. Thus APAPRO Desensitizer was shown to be as effective as the NovaMin® found in the predicate device in occluding dentin tubules.

| Trait                     | APAPRO Desensitizer<br>New Device | Oralief™ Therapy for<br>Sensitive Teeth<br>Predicate Device | Equivalence<br>Determination |
|---------------------------|-----------------------------------|-------------------------------------------------------------|------------------------------|
| 510(k) Number             | TBD                               | K040858                                                     | N/A                          |
| Classification<br>Number  | 872.3260                          | 872.3260                                                    | Equivalent                   |
| Product Code              | LBH                               | LBH                                                         | Equivalent                   |
| Product<br>Classification | Class II                          | Class II                                                    | Equivalent                   |

- 8. Device Comparison Table:

{5}------------------------------------------------

## Sangi Co., Ltd. Traditional 510(k) APAPRO Desensitizer

| Prescription/<br>Over-the-<br>counter use     | Prescription                                                                                                                                                                    | Prescription                                                                                                                                                                                                                    | Equivalent  |
|-----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Intended Use                                  | For dental hypersensitivity relief                                                                                                                                              | For dental hypersensitivity relief                                                                                                                                                                                              | Equivalent  |
| Indications for<br>Use                        | Provides rapid and continual relief<br>from tooth hypersensitivity due to<br>cold, heat, acids, sweets, or<br>contact through its action of the<br>occlusion of dentin tubules. | Provides rapid and continual<br>relief from tooth hypersensitivity<br>due to cold, heat, acids, sweets,<br>or contact through its action of<br>the occlusion of dentin tubules.                                                 | Equivalent  |
| Active<br>Ingredient                          | 20% w/w Calcium Phosphate<br>(Hydroxyapatite)                                                                                                                                   | 7.5% w/w Calcium Sodium<br>Phosphosilicate (NovaMin®)                                                                                                                                                                           | Equivalent* |
| Device Action                                 | APAPRO Desensitizer directly<br>occludes dentin tubules by<br>depositing a crystalline calcium<br>phosphate layer on the tooth<br>surface (i.e. hydroxyapatite).                | The NovaMin® particles react<br>with the user's saliva to release<br>calcium and phosphate ions,<br>which precipitate onto the tooth<br>surface as a crystalline calcium<br>phosphate layer (i.e.<br>hydroxycarbonate apatite). | Equivalent  |
| Material<br>Produced to<br>Occlude<br>Tubules | Hydroxyapatite                                                                                                                                                                  | Hydroxycarbonate Apatite                                                                                                                                                                                                        | Equivalent* |
| Performance                                   | The hydroxyapatite layer occludes<br>exposed dentin tubules, which<br>results in desensitization.                                                                               | The hydroxycarbonate apatite<br>layer occludes exposed dentin<br>tubules, which results in<br>desensitization.                                                                                                                  | Equivalent  |
| Biocompatibility                              | Biocompatible, non-antigenic                                                                                                                                                    | Biocompatible, non-antigenic                                                                                                                                                                                                    | Equivalent  |
| Sterility                                     | Supplied non-sterile                                                                                                                                                            | Supplied non-sterile                                                                                                                                                                                                            | Equivalent  |

*Details are provided in Section 12

- 9. Biocompatibility
Biocompatibility tests in accordance with ISO 10993 have been performed on APAPRO Desensitizer. A biological risk assessment, captured in the enclosed Biological Evaluation Report, has also been performed and included in this Traditional 510(k) application package. The biocompatibility assessment indicated that APAPRO Desensitizer has no hazardous effects to the consumer when used as directed in the Adult Population.

- 10. Conclusions:

{6}------------------------------------------------

Sangi Co., Ltd. Traditional 510(k) APAPRO Desensitizer

APAPRO Desensitizer is considered to be substantially equivalent to the legally marketed predicate device, Oralief™ Therapy for Sensitive Teeth. Both products are indicated for the relief of tooth hypersensitivity through the occlusion of dentin tubules. Both products achieve this by producing a calcium hydroxyapatite layer on the tooth surface. Additionally, both products have been shown to have comparable effectiveness with regards to tubule occlusion. No additional safety concerns have been observed when comparing APAPRO Desensitizer to Oralief™ Therapy for Sensitive Teeth.

---

**Source:** [https://fda-staging.innolitics.com/device/K220419](https://fda-staging.innolitics.com/device/K220419)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).