Canary Tibial Extension (CTE) with Canary Health Implanted Reporting Processor (CHIRP) System
K220413 · Canary Medical USA, LLC · QPP · Jun 22, 2022 · Orthopedic
Device Facts
Record ID
K220413
Device Name
Canary Tibial Extension (CTE) with Canary Health Implanted Reporting Processor (CHIRP) System
Applicant
Canary Medical USA, LLC
Product Code
QPP · Orthopedic
Decision Date
Jun 22, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3600
Device Class
Class 2
Indications for Use
The Canary Tibial Extension (CTE) with Canary Health Implanted Reporting Processor (CHIRP) System is intended to provide objective kinematic data from the implanted medical device during a patient's total knee arthroplasty (TKA) postsurgical care. The kinematic data are an adjunct to other physiological parameter measurement tools applied or utilized by the physician during the course of patient monitoring and treatment post-surgery.
Device Story
CTE with CHIRP system provides objective kinematic data post-TKA; adjunct to other physiological measurement tools. System includes: CTE implant, OR base station (BS1), home base station (BS2), Canary Cloud platform, and CMGP software. Implant collects internal motion sensor data; transmits to base stations via communication interface; data uploaded to cloud. Updated version adds low-power accelerometer for activity monitoring (step counting) and motion-triggered kinematic data collection to optimize battery life. Used by patients and HCPs; data assists physician monitoring between office visits. Not intended to support clinical decision-making; no clinical benefit shown.
Clinical Evidence
Bench testing only. Non-clinical performance data included shock survival, electrical life testing, battery longevity, electronic functionality, and software verification.
Technological Characteristics
Implantable tibial extension with integrated CHIRP processor. Includes internal motion sensors and low-power accelerometer. Connectivity via communication interface to base stations and cloud. Software-based data collection and processing. Materials, geometry, and sterility identical to predicate.
Indications for Use
Indicated for patients undergoing a cemented TKA procedure who are normally indicated for at least a 58mm sized tibial stem extension.
Regulatory Classification
Identification
An implantable post-surgical kinematic measurement knee device is a device that provides objective kinematic data after total knee arthroplasty surgery. The kinematic data provided by the device are used as an adjunct to other physiological parameter measurement tools utilized during the course of patient monitoring and treatment post surgery.
Special Controls
In combination with the general controls of the FD&C Act, the implantable post-surgical kinematic measurement knee device is subject to the following special controls:
- 1. Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following tests must be conducted:
- a. Mechanical testing must evaluate the mechanical function (mechanical fatigue, static mechanical strength) and durability of the implant;
- b. Simulated use testing must evaluate the ability of the device to be sized, inserted and sufficiently secured to any compatible components;
- Testing must demonstrate the accuracy, reliability, and reproducibility of c. kinematic measurements; and
- d. Testing must demonstrate diagnostic and therapeutic ultrasound conditions for safe use.
- Testing must demonstrate that the device performs as intended under anticipated e. conditions of use demonstrating the following performance characteristics, if applicable:
- i. Magnetic pulse output testing
- ii. Magnetic and electrical field testing
- iii. Testing of the safety features built into the device.
- f. Testing must demonstrate hermeticity of any electronic component enclosures.
- 2. Performance testing must evaluate the compatibility of the device in a magnetic resonance (MR) environment.
- 3. Human factors testing must demonstrate that the intended user(s) can correctly use the device for its intended use, including for implantation and post-procedure data access.
- 4. Performance data must demonstrate the sterility of the device implant and patientcontacting components.
- 5. Performance data must validate the reprocessing instructions for the reusable components of the device.
- 6. The patient-contacting components of the device must be demonstrated to be biocompatible.
- 7. Design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
- 8. Performance testing must demonstrate the electromagnetic compatibility/interference (EMC/EMI), electrical safety, thermal safety, battery safety, and wireless performance of the device.
- 9. Software verification, validation, and hazard analysis must be performed.
- 10. The labeling must include the following:
- a. A shelf life;
- b. Physician and patient instructions for use, including images that demonstrate how to interact with the device;
- Detailed instruction of the surgical technique; C.
- d. Hardware and software requirements for interacting with the device;
- A clear description of the technological features of the device including e. identification of the device materials, compatible components, and the principles of operation;
- f. Identification of magnetic resonance (MR) compatibility status;
- Validated methods and instructions for reprocessing of any reusable components; g. and
- h. A statement regarding the limitations of the clinical significance of the kinematic data.
*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following tests must be conducted:
(i) Mechanical testing must evaluate the mechanical function (mechanical fatigue, static mechanical strength) and durability of the implant.
(ii) Simulated use testing must evaluate the ability of the device to be sized, inserted, and sufficiently secured to any compatible components.
(iii) Testing must demonstrate the accuracy, reliability, and reproducibility of kinematic measurements.
(iv) Testing must demonstrate diagnostic and therapeutic ultrasound conditions for safe use.
(v) Testing must demonstrate that the device performs as intended under anticipated conditions of use demonstrating the following performance characteristics, if applicable:
(A) Magnetic pulse output testing;
(B) Magnetic and electrical field testing; and
(C) Testing of the safety features built into the device.
(vi) Testing must demonstrate hermeticity of any electronic component enclosures.
(2) Performance testing must evaluate the compatibility of the device in a magnetic resonance (MR) environment.
(3) Human factors testing must demonstrate that the intended user(s) can correctly use the device for its intended use, including for implantation and post-procedure data access.
(4) Performance data must demonstrate the sterility of the device implant and patient-contacting components.
(5) Performance data must validate the reprocessing instructions for the reusable components of the device.
(6) The patient-contacting components of the device must be demonstrated to be biocompatible.
(7) Design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(8) Performance testing must demonstrate the electromagnetic compatibility/interference, (EMC/EMI), electrical safety, thermal safety, battery safety, and wireless performance of the device.
(9) Software verification, validation, and hazard analysis must be performed.
(10) The labeling must include the following:
(i) A shelf life;
(ii) Physician and patient instructions for use, including images that demonstrate how to interact with the device;
(iii) Detailed instruction of the surgical technique;
(iv) Hardware and software requirements for interacting with the device;
(v) A clear description of the technological features of the device including identification of the device materials, compatible components, and the principles of operation;
(vi) Identification of magnetic resonance (MR) compatibility status;
(vii) Validated methods and instructions for reprocessing of any reusable components; and
(viii) A statement regarding the limitations of the clinical significance of the kinematic data.
Predicate Devices
Canary Tibial Extension (CTE) with Canary Health Implanted Reporting Processor (CHIRP) System (DEN200064)
Reference Devices
Zimmer Biomet Persona Personalized Knee System
Submission Summary (Full Text)
{0}------------------------------------------------
June 22, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
Canary Medical USA LLC Nora York Vice President, Regulatory Affairs and Quality Affairs 2710 Loker Ave. West, Suite 350 Carlsbad, California 92010
#### Re: K220413
Trade/Device Name: Canary Tibial Extension (CTE) with Canary Health Implanted Reporting Processor (CHIRP) System Regulation Number: 21 CFR 888.3600 Regulation Name: Implantable Post-Surgical Kinematic Measurement Knee Device Regulatory Class: Class II Product Code: QPP Dated: Mav 26, 2022 Received: May 27, 2022
Dear Nora York:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ting Song, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K220413
Device Name
Canary Tibial Extension (CTE) with Canary Health Implanted Reporting Processor (CHIRP) System
Indications for Use (Describe)
The Canary Tibial Extension (CTE) with Canary Health Implanted Reporting Processor (CHIRP) System is intended to provide objective kinematic data from the implanted medical device during a patient's total knee arthroplasty (TKA) postsurgical care. The kinematic data are an adjunct to other physiological parameter measurement tools applied or utilized by the physician during the course of patient monitoring and treatment post-surgery.
The device is indicated for use in patients undergoing a cemented TKA procedure that are normally indicated for at least a 58mm sized tibial stem extension.
The objective kinematic data generated by the CTE with CHIRP System are not intended to support clinical decisionmaking and have not been shown to provide any clinical benefit.
The CTE with CHIRP System is compatible with Zimmer Personalized Knee System.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Traditional 510(k) Submission
Canary Tibial Extension (CTE) With Canary Health Implanted Reporting Processor (Chirp)
Image /page/3/Picture/3 description: The image shows the logo for Canary. The logo features a yellow bird above the word "canary" in black, lowercase letters. The bird is stylized with simple shapes, and the word "canary" is in a bold, sans-serif font. There is a trademark symbol to the right of the word.
# 510(K) SUMMARY
| Sponsor: | Canary Medical USA LLC<br>2710 Loker Ave. West, Suite 350<br>Carlsbad, CA 92010 |
|-------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment<br>Registration Number: | 3015802419 |
| Contact Person: | Nora C.R. York<br>Vice President, Regulatory Affairs and Quality Affairs<br>Mobile: 760-497-6115<br>Email: nyork@canarymedical.com |
| Date of Summary<br>Preparation: | March 23, 2022 |
| Subject Device: | Canary Tibial Extension (CTE) with Canary Health Implanted Reporting<br>Processor (CHIRP) System (K220413) |
| Regulation Number: | 21 CFR 888.3600 |
| Product Code: | QPP |
| Device Class: | II |
| Predicate Device: | Canary Tibial Extension (CTE) with Canary Health Implanted Reporting<br>Processor (CHIRP) System (DEN200064) |
| Purpose of Submission: | The purpose of this submission is to add a dedicated low-power<br>accelerometer for activity monitoring (step counting) and detection of<br>significant motion to trigger kinematic data collection. The design<br>enhancements are intended to increase the kinematic data sampling day<br>collection capability using the same battery. The design enhancement<br>does not change the intended use / indications for use or fundamental<br>scientific technology of the device. The addition of the accelerometer<br>necessitates other accommodations to the printed circuit board assembly<br>(PCBA) and associated embedded software. |
| Device Description: | The Canary Tibial Extension (CTE) with Canary Health Implanted<br>Reporting Processor (CHIRP) System is comprised of the following<br>subsystems:<br>Canary Tibial Extension (CTE) implant, Operating Room (OR) Base Station System ("BS1"), Home Base Station System ("BS2"), Canary Cloud Data Management Platform ("Cloud") and Canary Medical Gait Parameters (CMGP software module). |
| Indications for Use: | The CTE and CHIRP System is intended to provide objective kinematic<br>data on patient's total knee arthroplasty (TKA) function. The kinematic<br>data produced by the CTE implant is intended as an adjunct to other<br>physiological measurement tools post TKA surgical care while providing<br>additional tibial stability afforded by traditional tibial extensions of<br>similar length. The implanted CTE collects data from internal motion<br>sensors, and when queried by a BS1 or BS2 over a communication<br>interface, transmits the motion data to the Base Station System. The<br>Base Station System, in turn, uploads the data to the Canary Cloud Data<br>Management Platform. The User is defined as the Patient with the CTE<br>and CHIRP System and their designated Health Care Professional (HCP)<br>with access to the Patient's CTE data.<br><br>The CTE is designed for use with the Zimmer Biomet Persona Personalized<br>Knee System tibial baseplate, to provide additional stability and collect<br>kinematic data to assist the physician in monitoring patient activity<br>following total knee arthroplasty (TKA) in between office visits.<br><br>The Canary Tibial Extension (CTE) with Canary Health Implanted<br>Reporting Processor (CHIRP) System is intended to provide objective<br>kinematic data from the implanted medical device during a patient's<br>total knee arthroplasty (TKA) post-surgical care. The kinematic data are<br>an adjunct to other physiological parameter measurement tools applied<br>or utilized by the physician during the course of patient monitoring and<br>treatment post-surgery.<br><br>The device is indicated for use in patients undergoing a cemented TKA<br>procedure that are normally indicated for at least a 58mm sized tibial<br>stem extension.<br><br>The objective kinematic data generated by the CTE with CHIRP System<br>are not intended to support clinical decision-making and have not been<br>shown to provide any clinical benefit.<br><br>The CTE with CHIRP System is compatible with Zimmer Persona®<br>Personalized Knee System. |
| Summary of<br>Technological<br>Characteristics and<br>Comparison: | The rationale for substantial equivalence is based on comparative<br>assessment of the following characteristics:<br>• Indications for use: Same as predicate device.<br>• Materials: Same as the predicate device.<br>• Geometry/Configuration/Size (of Tibial Extension): Same as the<br>predicate device. |
| | Sterility: Same as the predicate device. Packaging: Same as the predicate device. |
| Summary of<br>Performance Data: | The following non-clinical tests were performed to support the<br>modifications to the subject device: Shock Survival Electrical Life Test Battery Longevity Electronic Functionality Software Verification |
| Substantial Equivalence<br>Conclusion: | The subject device has the same intended use and indications for use as<br>the predicate device. The subject device has the same technological<br>characteristics to the predicate, and the performance data and analyses<br>demonstrate that: any differences do not raise new questions of safety and<br>effectiveness; and the proposed device is at least as safe and effective as the legally<br>marketed predicate device. |
{4}------------------------------------------------
K220413 Page 2 of 3
## Traditional 510(k) Submission
Canary Tibial Extension (CTE) With Canary Health Implanted Reporting Processor (Chirp)
Image /page/4/Picture/3 description: The image shows the logo for Canary. The logo features a yellow bird perched above the word "canary" in a bold, black font. The bird is stylized with a simple, streamlined design, and the word "canary" is followed by the trademark symbol.
{5}------------------------------------------------
## K220413 Page 3 of 3
## Traditional 510(k) Submission
Canary Tibial Extension (CTE) With
Canary Health Implanted Reporting Processor (Chirp)
Image /page/5/Picture/4 description: The image shows the logo for Canary. The logo features a yellow bird silhouette above the word "canary" in a bold, black sans-serif font. The bird is facing left with its wings slightly raised, and the word "canary" is positioned directly below it. There is a trademark symbol to the right of the word.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.