K220350 · AliveCor, Inc. · DXH · May 25, 2022 · Cardiovascular
Device Facts
Record ID
K220350
Device Name
KardiaMobile 6L
Applicant
AliveCor, Inc.
Product Code
DXH · Cardiovascular
Decision Date
May 25, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2920
Device Class
Class 2
Attributes
AI/ML
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
ECG Rhythm Analysis
KardiaAI platform
—
—
—
—
—
—
Indications for Use
KardiaMobile 6L is intended to record, store and two-channel electrocardiogram (ECG) rhythms. In single channel mode, KardiaMobile 6L can record Lead-I. In two channel mode, KardiaMobile 6L can record Lead-I and Lead-II simultaneously and derive Lead-III and unipolar limb leads aVR, aVF and aVL. KardiaMobile 6L also displays ECG rhythms and output of ECG analysis from AliveCor's KardiaAI platform including detecting the presence of normal sinus rhythm, atrial fibrillation, bradycardia, and others. KardiaMobile 6L is intended for use by healthcare professionals, patients with known or suspected heart conditions and health conscious individuals. The device has not been tested and is not intended for pediatric use.
Device Story
KardiaMobile 6L is a portable, wireless ECG event recorder using three dry electrodes. Users complete a circuit by placing fingers/hands on electrodes; device transmits ECG signals via Bluetooth to a smartphone/tablet (Mobile Computing Platform). The Kardia Core app on the mobile device displays the ECG and utilizes the KardiaAI platform to analyze rhythms. Outputs include 1- or 6-lead ECG strips and automated rhythm classifications (e.g., Normal Sinus Rhythm, Atrial Fibrillation, Bradycardia, Tachycardia). Used by patients for self-monitoring or by healthcare professionals for remote monitoring. Healthcare providers review the ECG rhythm and QT interval measurements to inform clinical decision-making. The device enables remote heart health monitoring, potentially facilitating earlier detection of cardiac conditions.
Clinical Evidence
No clinical testing was required for the software reorganization; substantial equivalence supported by bench-level software verification testing.
Technological Characteristics
Portable wireless hardware with three dry electrodes. Frequency response 0.5Hz–40Hz; 16-bit resolution; 300 samples/second. Power: 1 Lithium Manganese Dioxide coin cell (100 hours typical). Connectivity: Bluetooth to iOS/Android mobile computing platforms. Software: Kardia Core app with KardiaAI analysis platform.
Indications for Use
Indicated for healthcare professionals, patients with known or suspected heart conditions, and health-conscious individuals to record, store, and transfer 1- or 2-channel ECG rhythms. Not intended for pediatric use.
Regulatory Classification
Identification
A telephone electrocardiograph transmitter and receiver is a device used to condition an electrocardiograph signal so that it can be transmitted via a telephone line to another location. This device also includes a receiver that reconditions the received signal into its original format so that it can be displayed. The device includes devices used to transmit and receive pacemaker signals.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
May 25, 2022
AliveCor, Inc. % Prabhu Raghavan Principal Consultant Mdar, LLC. 189 N. Bernardo Ave. Suite 100 Mountain View. California 94043
Re: K220350
Trade/Device Name: KardiaMobile 6L Regulation Number: 21 CFR 870.2920 Regulation Name: Telephone Electrocardiograph Transmitter And Receiver Regulatory Class: Class II Product Code: DXH, DPS, ODA Dated: February 24, 2022 Received: February 25, 2022
Dear Prabhu Raghavan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220350
Device Name KardiaMobile 6L
#### Indications for Use (Describe)
KardiaMobile 6L is intended to record, store and two-channel electrocardiogram (ECG) rhythms. In single channel mode, KardiaMobile 6L can record Lead-I. In two channel mode, KardiaMobile 6L can record Lead-I and Lead-II simultaneously and derive Lead-III and unipolar limb leads aVR, aVF and aVL. KardiaMobile 6L also displays ECG rhythms and output of ECG analysis from AliveCor's KardiaAI platform including detecting the presence of normal sinus rhythm, atrial fibrillation, bradycardia, and others. KardiaMobile 6L is intended for use by healthcare professionals, patients with known or suspected heart conditions and health conscious individuals. The device has not been tested and is not intended for pediatric use.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
|X Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for AliveCor. The logo is in a teal color and features the company name in a sans-serif font. A registered trademark symbol is located to the right of the name.
## 510(k) Summary
Prepared in accordance with the requirements of 21 CFR 807.92
#### Submitter Information [807.92(a)(1)]
| Submitter/Applicant | AliveCor, Inc.<br>189 N. Bernardo Ave. Suite 100<br>Mountain View, CA 94043<br>Phone: 650-396-8553<br>Fax: 650-282-7932 |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary Contact Person | Prabhu Raghavan<br>Regulatory Consultant for AliveCor<br>Principal Consultant, MDQR, LLC<br>189 N. Bernardo Ave. Suite 100<br>Mountain View, CA 94043<br>Phone: 408-316-5707<br>Fax: 650-282-7932<br>Email: prabhu@mdqr.solutions |
| Submitter and<br>Secondary Contact Person | Susan Noriega<br>VP of Regulatory Affairs, Clinical Affairs and Quality<br>189 N. Bernardo Ave. Suite 100<br>Mountain View, CA 94043<br>Phone: 650-793-1966<br>Fax: 650-282-7932<br>Email: snoriega630@alivecor.com |
| Date Prepared | February 02, 2022 |
| Device Information [807.92(a)(2)] | |
| Trade Name | KardiaMobile 6L |
| Common Name | Transmitters and Receivers, Electrocardiograph, Telephone |
| Classification | 21 CFR§870.2920 |
| Device Class | Class II |
| Product Code | DXH |
| Subsequent Product Code | DPS, ODA |
#### Predicate Information [807.92(a)(3)]
AliveCor, Inc., K210753, KardiaMobile 6L Predicate(s)
## Device Description [807.92(a)(4)]
KardiaMobile 6L is a trans-telephonic (transmission by smartphone) electrocardiogram (ECG) event recorder that records, stores, transfers, and analyzes single-channel or two channel ECG rhythm recordings. KardiaMobile 6L provides output of one or six ECG leads, including Lead I, Lead II, Lead III, aVL, aVR and aVF. The device utilizes the computing power of iOS-based or
{4}------------------------------------------------
510(k) Summary for KardiaMobile 6L AliveCor, Inc.
Android-based devices (referred to as "Mobile Computing Platforms" (MCP) within this 510(k)) to record and analyze ECG signals. KardiaMobile 6L consists of KardiaMobile 6L Hardware (portable small wireless hardware with electrodes) and the Kardia Core app. which is installed on an MCP (i.e., iOS or Android devices). KardiaMobile 6L Hardware uses Bluetooth to transmit the ECG signal from the electrodes to the Kardia Core app on the MCP, which then displays the recorded ECG on the MCP's screen. The device is intended to be used by patients with known or suspected heart conditions and health conscious individuals as well as by healthcare professionals (HCPs) who want to remotely monitor their patient's heart health. The device is available for Over-the-Counter (OTC) as well prescription use. The Kardia Core app provides the complete ECG recording and analysis workflow, from acquisition of the signal from the KardiaMobile 6L hardware, to the display of the ECG and analysis results, to printing of the ECG rhythm strip. The app utilizes KardiaAI (K181823, K201985) to provide ECG analysis, which includes the determinations of Normal Sinus Rhythm, Atrial Fibrillation, Bradycardia, Tachycardia, or Unclassified to OTC users and additionally, Sinus Rhythm determinations (Sinus Rhythm with Wide ORS, Sinus Rhythm with Supraventricular Ectopy, and Sinus Rhythm with Premature Ventricular Contractions) for prescription-use only users.
The Kardia Core app may be distributed as is or be incorporated into other apps ("variants"), by AliveCor or by third-parties, that can provide additional functionality such patient/user onboarding, storage and review of past data, as well as providing secondary MDDS displays of other medical devices. The main software app variant, called the Kardia App, is the default software marketed with KardiaMobile 6L. Kardia App includes workflows for OTC user account creation, integration with AliveCor's KardiaPro MDDS portal, secondary display of Bluetoothenabled blood pressure monitors, and other non-regulated or enforcement discretion features. Kardia App also provides workflows for patients to forward the recorded ECG to their HCP, who can review the ECG rhythm and measure the QT interval. An alternate software variant, called the KardiaStation App, incorporates the Kardia Core app to be used exclusively within hospitals and clinics by adding patient administration workflows. Finally, a third variant called Kardia Rx App incorporates the Kardia Core app to provide a simple workflow to transfer the data for remote patient monitoring applications.
#### Indications for use [807.92(a)(5)]
KardiaMobile 6L is intended to record, store and transfer one- and two-channel electrocardiogram (ECG) rhythms. In single channel mode, KardiaMobile 6L can record Lead-I. In two channel mode, KardiaMobile 6L can record Lead-I simultaneously and derive Lead-III and unipolar limb leads aVR, aVF and aVL. KardiaMobile 6L also displays ECG rhythms and output of ECG analysis from AliveCor's KardiaAI platform including detecting the presence of normal sinus rhythm, atrial fibrillation, bradycardia, and others. KardiaMobile 6L is intended for use by healthcare professionals, patients with known or suspected heart conditions and health conscious individuals. The device has not been tested and is not intended for pediatric use.
#### Substantial Equivalence
The KardiaMobile 6L subject device has the same intended use, physical characteristics, and technological characteristics as the K210753, KardiaMobile 6L predicate device. The difference between the two devices is the reorganization of the software only and the addition of an
{5}------------------------------------------------
510(k) Summary for KardiaMobile 6L AliveCor, Inc.
Application Program Interface for other apps to access Kardia Core workflows. Performance testing, utilizing test methods previously used in K183319 and K210753, were employed to ensure that the device meets its requirements and that the differences between the subject and predicate device do not raise different questions of safety or effectiveness.
## Performance Data [807.92(b)]
All necessary testing was conducted on KardiaMobile 6L to support a determination of substantial equivalence to the predicate device.
## Nonclinical Testing Summarv [807.92(b)(1)]
Since both the subject and the predicate have the same hardware, verification activities focused on the changes related to the software reorganization. The subject device successfully passed the software verification test methods previously used in K183319 and K210753.
#### Clinical Testing Summary [807.92(b)(2)]
No clinical testing was required for the minor changes made to the software.
## Conclusions [807.92(b)(3)]
The KardiaMobile 6L subject device has the same intended use as the predicate device, and the software changes to support the reorganization do not raise different questions of safety or effectiveness. The software changes between the subject device and the predicate device have been verified to ensure that the subject device is as safe and effective for its intended use as the predicate device. Therefore, the KardiaMobile 6L subject device is substantially equivalent to the predicate device.
| Feature | AliveCor KardiaMobile 6L<br>(Subject Device) | AliveCor KardiaMobile 6L<br>(K210753, Predicate Device) |
|------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | DXH, DPS, QDA | No difference |
| Mechanism of Action | User completes circuit with skin<br>contact and hardware transmits ECG<br>signal to MCP to convert and display<br>ECG waveform | No difference |
| Where used (intended use) | Mobile/active users at rest (ambulatory) | No difference |
| Use cases for HCPs | Rhythm analysis<br>QT Interval Measurement | No difference |
| Anatomical sites | Left hand fingers to right hand fingers<br>Left hand fingers to right hand fingers<br>and to left leg/knee | No difference |
| Software application<br>functions | Medical device functions<br>● Acquire ECG from<br>○ KardiaMobile 6L Hardware | Medical device functions<br>● No difference<br>○ |
| Feature | AliveCor KardiaMobile 6L<br>(Subject Device) | AliveCor KardiaMobile 6L<br>(K210753, Predicate Device) |
| | Record, transfer, and display an ECG, ECG analysis of KardiaAI (K181823 or K201985), as applicable Create a PDF rhythm strip of the recorded ECG API for integrating with other mobile applications | Provides additional nonmedical device or enforcement discretion functions such as workflows for user account creation, and onboarding, connecting to Bluetooth-enabled blood pressure monitors to provide a secondary display, share previously recorded ECG and blood pressure data with healthcare professionals, etc. |
| ECG Analysis<br>Determinations using<br>K181823 KardiaAI.<br>In OTC and Rx Only type of<br>use | Atrial Fibrillation<br>Normal Sinus Rhythm<br>Tachycardia<br>Bradycardia<br>Unclassified<br>Unreadable | No difference |
| ECG Analysis<br>Determinations using<br>K201985 KardiaAI.<br>In Rx Only type of use | Atrial Fibrillation<br>Sinus Rhythm<br>Sinus Rhythm with Wide QRS,<br>Sinus Rhythm with Supraventricular<br>Ectopy<br>Sinus Rhythm with Premature<br>Ventricular Contractions<br>Normal Sinus Rhythm<br>Tachycardia<br>Bradycardia<br>Unclassified<br>Unreadable | No difference |
| Data Acquisition: | | |
| Frequency Response | 0.5Hz – 40Hz | No difference |
| No. of ECG electrodes | Three (3) dry electrodes | No difference |
| No. of ECG channels | Single-channel and two-channel | No difference |
| Resolution | 16 bits | No difference |
| Sample Rate | 300 Samples/Second | No difference |
| Power Supply: | | |
| Battery | 1 Lithium Manganese Dioxide Coin<br>Cells | No difference |
| Battery Life (typical) | 100 hours operational | No difference |
## Comparison of Technological Characteristics with the Predicate Device [807.92(a)(6)]
{6}------------------------------------------------
# Comparison of Technological Characteristics with the Predicate Device [807.92(a)(6)]
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.