K220311 · Allengers Medical Systems Limited · OWB · May 3, 2022 · Radiology
Device Facts
Record ID
K220311
Device Name
Cardiovascular Angiography System
Applicant
Allengers Medical Systems Limited
Product Code
OWB · Radiology
Decision Date
May 3, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Cardiovascular Angiography System is digital system with high frequency X-Ray generator for application in cardiovascular procedures. This system used in catheterization labs uses high power x-ray pulses and digital imaging system to visualize the vascular structures of human body. Clinical applications may include (but not limited to) interventional cardiovascular procedures, Neurovascular procedures, pacemaker implantable and high end investigations. This system should be handled by persons who have been briefed in its professional handling and who have familiarized themselves with the product by means of instructions for use. Intended use also means following the user manual and observing the conditions for inspection and maintenance. Exclusion: This system is not recommended for Mammography. Contraindication: Exposure of X-Ray should be avoided during pregnancy.
Device Story
Cardiovascular Angiography System performs fluoroscopic and digital radiographic studies in catheterization labs. System inputs include X-ray pulses generated by high-frequency generators and captured by solid-state X-ray detectors (FPD). The system transforms these inputs into digital images for visualization of vascular structures. Operated by trained clinicians, the system includes a C-arm (floor, ceiling, or mobile), patient table, and ceiling-suspended displays. Output is viewed on monitors to assist in interventional cardiovascular, neurovascular, and pacemaker procedures. The device aids clinical decision-making by providing real-time imaging for diagnostic and interventional guidance, potentially improving patient outcomes through precise visualization of complex anatomical structures.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including functional testing of motion systems, imaging performance evaluation, and compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-43, and IEC 60601-2-54 standards.
Technological Characteristics
High-frequency, microprocessor-controlled X-ray generator; rotating anode X-ray tubes; amorphous silicon or CMOS solid-state X-ray detectors; C-arm configuration (floor/ceiling/mobile); DICOM 3.0 connectivity; software-based digital imaging system. Safety compliance per IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-43, and IEC 60601-2-54.
Indications for Use
Indicated for cardiovascular, neurovascular, and interventional procedures, including pacemaker implantation and high-end investigations, in adult and pediatric populations. Contraindicated for mammography and during pregnancy.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Allura Xper FD Series and Allura Xper OR Table serie (K162859)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 3, 2022
Allengers Medical Systems Limited % Sanjeev Marjara Director Technical FDA Hall Unit-2, Bhankarpur, Mubarakpur Road, Derabassi. Distt. Mohali, Punjab 140507 INDIA
Re: K220311
Trade/Device Name: Cardiovascular Angiography System Regulation Number: 21 CFR 892.1650 Regulation Name: Image-Intensified Fluoroscopic X-Ray System Regulatory Class: Class II Product Code: OWB, JAA, OXO Dated: January 27, 2022 Received: February 2, 2022
Dear Sanjeev Marjara:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurel Burk Assistant Director Diagnostic X-ray Systems Team DHT 8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
K220311
Device Name
Cardiovascular Angiography System
#### Indications for Use (Describe)
The Cardiovascular Angiography System is digital system with high frequency X-Ray generator for application in cardiovascular procedures. This system used in catheterization labs uses high power x-ray pulses and digital imaging system to visualize the vascular structures of human body.
Clinical applications may include (but not limited to) interventional cardiovascular procedures, Neurovascular procedures, pacemaker implantable and high end investigations.
This system should be handled by persons who have been briefed in its professional handling and who have familiarized themselves with the product by means of instructions for use. Intended use also means following the user manual and observing the conditions for inspection and maintenance.
Exclusion: This system is not recommended for Mammography.
Contraindication: Exposure of X-Ray should be avoided during pregnancy.
Type of Use (Select one or both, as applicable)
7 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
{3}------------------------------------------------
### 510(k) SUMMARY
This summary of 510(k) is being submitted in accordance with requirements of 21 CFR Part 807.92.
### 1. Contact Person and Address
| Company Name: | Allengers Medical Systems Limited |
|------------------|-----------------------------------------------------------------------------------------|
| Company Address: | FDA Hall, Unit-2, Bhankarpur, Mubarakpur Road, Derabassi, Distt<br>Mohali-140507, India |
| Telephone No: | +91 1762-282600<br>+919872980168<br>rnd@allengers.net |
| Contact Person: | Sanjeev K. Marjara |
| Date Prepared: | 27.01.2022 |
### 2. Proposed Device:
| Device (trade) name: | Cardiovascular Angiography System | | |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Model Name: | Proxima FDX, Altima FDX Adv<br>Digiscan FDX-V8043, Digiscan FDX-V8030, Digiscan FDX-V8020<br>Digiscan FDX-V6543, Digiscan FDX-V6530, Digiscan FDX-V6520<br>Digiscan FDX-V4043, Digiscan FDX-V4030, Digiscan FDX-V4020<br>Digiscan FDX-V2743, Digiscan FDX-V2730, Digiscan FDX-V2720 | | |
| Classification Name : | Image-Intensified Fluoroscopic X-Ray System | | |
| Classification Panel: | Radiology | | |
| Regulation Number : | 21 CFR 892.1650 | | |
| Device Class: | Class II | | |
| Product Code: | Primary Code : OWB<br>Subsequent Code : JAA and OXO for 'Digiscan FDX-VXXYY | | |
### 3. Predicate Device:
| Device (trade) name: | Allura Xper FD Series and Allura Xper OR Table serie | |
|-----------------------|------------------------------------------------------|--|
| 510(K) Number: | K162859 | |
| Clearance Date: | December 2, 2016 | |
| Classification Name: | Image-Intensified Fluoroscopic X-Ray System, | |
| Classification Panel: | Radiology | |
| Regulation Number: | 21 CFR 892.1650 | |
| Device Class: | Class II | |
| Product Code: | Primary Code : OWB<br>Subsequent Code : JAA | |
¹ Where XX may be 80/65/40/27 and YY may be 43/30/20
{4}------------------------------------------------
### 4. Reference devices:
| Device (trade) name: | Ziehim Vision RFD |
|-----------------------|--------------------------------------------------|
| 510(K) Number: | K203428 |
| Clearance Date: | March 17, 2021 |
| Classification Name: | Image-Intensified Fluoroscopic X-Ray System |
| Classification Panel: | Radiology |
| Regulation Number: | 21 CFR 892.1650 |
| Device Class: | Class II |
| Product Code: | Primary Code : OWB<br>Subsequent Code : JAA, OXO |
Allengers Medical Systems Ltd. Supplies Solid State X-Ray Image Detectors that have been previously cleared by FDA or tested and evaluated per quidance for submission of 510(K) for solid state X-Ray imaging devices. Table 1 provides the list of solid state detectors used with device.
| Solid State Detectors | 510(K) Numbers |
|----------------------------------------------|----------------|
| Varex Imaging corporation – Paxscan 2020 DXV | K200218 |
| Varex Imaging corporation – Paxscan 3030 DXV | K200218 |
| Varex Imaging corporation - Paxscan 4343DXV | -- |
| MX Imaging - CFP 2222 | K171755 |
| MX Imaging - CFP 3131 | K171755 |
| Thales Group – Pixium Surgical 2121S-A | K183040 |
| Thales Group - Pixium 2121S-AU | K200218 |
| Thales Group - Pixium Surgical 3030S-A | K172822 |
| Thales Group - Pixium 3030S-AU | K200218 |
| IRAY Technology - Mercu 0909F | K200218 |
| Allengers Medical Systems Ltd - FP 2121 RF | -- |
| Allengers Medical Systems Ltd - FP 2121 HR | -- |
| Allengers Medical Systems Ltd - FP 3030 HR | -- |
Table 1 List of Solid State X-Rav Image Detectors
### 5. Device description:
The Cardiovascular Angiography system is designed to perform fluoroscopic & digital radiographic studies and are used in interventional examinations. This system covers the complete range of angiographic applications, cardiac angiography, neuroangiography, general angiography, surgery and surgical angiography, multipurpose angiography, rotational angiography and digital radiographic/ fluoroscopic procedures. The following components are configured to create the Cardiovascular Angiography system:
- 1. Floor mounted C-arm (Altima FDX Adv) or Ceiling Mounted C-Arm (Proxima FDX) or Mobile C-Arm (2Digiscan FDX-VXXYY), X-ray tube assembly and Solid State X-Ray image detector
- 2. Patient Table
- 3. Ceiling suspended display(s)
<sup>2</sup> Where XX may be 80/65/40/27 and YY may be 43/30/20
{5}------------------------------------------------
- 4. Footswitch for releasing radiation
- 5. Control console for controlling the stand, patient table, collimator and imaging system.
The following in Table 2 are the specific components for various configurations of the system. A complete system will consist of a selection of one of the devices in each category.
| Component | Manufacture | Model | |
|-----------------|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| X-Ray Generator | Allengers | Proxima FDX<br>Altima FDX Adv | XGEN-100CV |
| | | Proxima FDX<br>Altima FDX Adv | |
| X-Ray Generator | Allengers | Digiscan FDX-V8043<br>Digiscan FDX-V8030<br>Digiscan FDX-V8020 | XGEN-80CV |
| | | | |
| | | | |
| X-Ray Generator | Allengers | Digiscan FDX-V6543<br>Digiscan FDX-V6530<br>Digiscan FDX-V6520 | XGEN-65CV |
| | | | |
| | | | |
| X-Ray Generator | Allengers | Digiscan FDX-V4043<br>Digiscan FDX-V4030<br>Digiscan FDX-V4020 | XGEN-40CV |
| | | | |
| | | | |
| X-Ray Generator | Allengers | Digiscan FDX-V2743<br>Digiscan FDX-V2730<br>Digiscan FDX-V2720 | XGEN-27CV |
| | | | |
| | | | |
| X-Ray Tube | Siemens<br>Healthcare<br>GmbH | Proxima FDX<br>Altima FDX Adv | Megalix CAT<br>125/35/80 125 GW<br>Megalix CAT<br>125/15/40/80122 GW |
| | | | |
| | | | |
| X-Ray Tube | Canon<br>Electron Tubes<br>and devices | Proxima FDX<br>Altima FDX Adv | E79039X<br>E79030X<br>E79016X |
| | | | |
| | | | |
| X-Ray Tube | Varex Imaging | Proxima FDX<br>Altima FDX Adv<br>Digiscan FDX-V8043<br>Digiscan FDX-V8030<br>Digiscan FDX-V8020<br>Digiscan FDX-V6543<br>Digiscan FDX-V6530<br>Digiscan FDX-V6520<br>Digiscan FDX-V4043<br>Digiscan FDX-V4030<br>Digiscan FDX-V4020 | G1082, G-1582BI,<br>G2090, G1092 |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
{6}------------------------------------------------
| X-Ray Tube | Varex Imaging | Digiscan FDX-V2743<br>Digiscan FDX-V2730<br>Digiscan FDX-V2720 | A-145, RAD-99,<br>RAD-99B, RAD-15 |
|------------------------------------------|--------------------|-----------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|
| Solid State X-<br>Ray Image<br>Detectors | Varex Imaging | Proxima FDX<br>Altima FDX Adv<br>Digiscan FDX-V8020<br>Digiscan FDX-V6520<br>Digiscan FDX-V4020<br>Digiscan FDX-V2720 | Paxscan 2020 DXV |
| Solid State X-<br>Ray Image<br>Detectors | Varex Imaging | Proxima FDX<br>Altima FDX Adv<br>Digiscan FDX-V8030<br>Digiscan FDX-V6530<br>Digiscan FDX-V4030<br>Digiscan FDX-V2730 | Paxscan 3030 DXV |
| Solid State X-<br>Ray Image<br>Detectors | Varex Imaging | Proxima FDX<br>Altima FDX Adv<br>Digiscan FDX-V8043<br>Digiscan FDX-V6543<br>Digiscan FDX-V4043<br>Digiscan FDX-V2743 | Paxscan 4343DXV |
| Solid State X-<br>Ray Image<br>Detectors | IRAY<br>Technology | Proxima FDX<br>Altima FDX Adv<br>Digiscan FDX-V8020<br>Digiscan FDX-V6520<br>Digiscan FDX-V4020<br>Digiscan FDX-V2720 | Mercu 0909F |
| Solid State X-<br>Ray Image<br>Detectors | Thales Group | Proxima FDX<br>Altima FDX Adv<br>Digiscan FDX-V8020<br>Digiscan FDX-V6520<br>Digiscan FDX-V4020<br>Digiscan FDX-V2720 | Pixium 2121S-AU<br>Pixium Surgical<br>2121S-A |
| Solid State X-<br>Ray Image<br>Detectors | Thales Group | Proxima FDX<br>Altima FDX Adv<br>Digiscan FDX-V8030<br>Digiscan FDX-V6530<br>Digiscan FDX-V4030<br>Digiscan FDX-V2730 | Pixium 3030S-AU<br>Pixium Surgical<br>3030S-A |
| Solid State X-<br>Ray Image<br>Detectors | MX Imaging | Proxima FDX<br>Altima FDX Adv<br>Digiscan FDX-V8020 | CFP 2222 |
{7}------------------------------------------------
| Solid State X-<br>Ray Image<br>Detectors | MX Imaging | Digiscan FDX-V6520<br>Digiscan FDX-V4020<br>Digiscan FDX-V2720 | |
|------------------------------------------|------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|
| | | Digiscan FDX-V8030<br>Digiscan FDX-V6530<br>Digiscan FDX-V4030<br>Digiscan FDX-V2730 | CFP 3131 |
| Solid State X-<br>Ray Image<br>Detectors | Allengers | Proxima FDX<br>Altima FDX Adv<br>Digiscan FDX-V8020<br>Digiscan FDX-V6520<br>Digiscan FDX-V4020<br>Digiscan FDX-V2720 | FP 2121 RF<br>FP 2121 HR |
| | | | |
| Solid State X-<br>Ray Image<br>Detectors | Allengers | Proxima FDX<br>Altima FDX Adv<br>Digiscan FDX-V8030<br>Digiscan FDX-V6530<br>Digiscan FDX-V4030<br>Digiscan FDX-V2730 | FP 3030 HR |
| | | | |
| Patient Table | Allengers | Proxima FDX<br>Altima FDX Adv | C Tab F FDX<br>C Tab FXL FDX<br>C Tab FCV FDX |
| Patient Table | Allengers | Digiscan FDX-V8043<br>Digiscan FDX-V8030<br>Digiscan FDX-V8020<br>Digiscan FDX-V6543<br>Digiscan FDX-V6530<br>Digiscan FDX-V6520<br>Digiscan FDX-V4043<br>Digiscan FDX-V4030<br>Digiscan FDX-V4020<br>Digiscan FDX-V2743<br>Digiscan FDX-V2730 | C Tab M FDX<br>Vascu Tab FDX |
{8}------------------------------------------------
#### 6. Indications for Use:
The Cardiovascular Angiography System is digital system with high frequency X-Ray generator for application in cardiovascular procedures. This system used in catheterization labs uses high power x-ray pulses and digital imaging system to visualize the vascular structures of human body.
Clinical applications may include (but not limited to) interventional cardiovascular procedures, Neurovascular procedures, pacemaker implantable and high end investigations.
This system should be handled by persons who have been briefed in its professional handling and who have familiarized themselves with the product by means of instructions for use. Intended use also means following the user manual and observing the conditions for inspection and maintenance.
Exclusion: This system is not recommended for Mammography.
Contraindications: Exposure of X-Ray should be avoided during pregnancy.
#### 7. Technological Characteristics Comparison to Predicate & Reference Devices:
The Cardiovascular Angiography Systems having set of components similar to the Allura Xper FD Series and Allura Xper OR Table series (K162859) and Ziehm Vision RFD (K203428) as compared in Table 3 found below in this Section. This table below shows that the systems are either similar, or the same, as the predicate & reference device.
#### 8. Software Feature
Synergy FP FDX & Synergy FDX-CR imaging software is a Digital Imaging System (DIS) provides useful functions to manage X-Ray images obtained from Cardiovascular Anqiography System.
The software feature set and functions is essentially the same as the device, with the system complying with DICOM 3.0 specifications .Refer to section 9 Image processing and storage of the table 3 for a list of top level functions
#### 9. Substantial Equivalence:
The Cardiovascular Angiography system are substantially equivalent to the commercially available Allura Xper FD series and Allura Xper OR Table series (K162859) & Ziehm Vision RFD (K203428). Functional and specification differences are identifying in the following table.
{9}------------------------------------------------
| Feature | Cardiovascular<br>Angiography System<br>(Subject Device) | Allura Xper FD Series<br>and Allura Xper OR Table<br>series<br>(Predicate Device) | Ziehm Vision RFD<br>(Reference Device) | Justification<br>for<br>differences | |
|-------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|
| 1. 510(k) | This submission | K162859 | K203428 | | |
| 2. Product Code | | | | | |
| Product<br>Classification<br>Code | OWB, JAA &<br>OXO for 'Digiscan FDX-<br>VXXYY' | OWB and JAA | OWB , JAA & OXO | Same | |
| 3. Product Classification | | | | | |
| Classification | 21 CFR 892.1650 | 21 CFR 892.1650 | 21 CFR 892.1650 | Same | |
| 4. Indication for Use | | | | | |
| Indications for<br>Use | The Cardiovascular<br>Angiography system is<br>digital system with high<br>frequency X-Ray generator<br>for application in<br>cardiovascular procedures.<br>This system used in<br>catheterization labs uses<br>high power x-ray pulses and<br>digital imaging system to<br>visualize the vascular<br>structures of human body.<br><br>Clinical applications may<br>include (but not limited to)<br>interventional cardiovascular<br>procedures, Neurovascular<br>procedures, pacemaker<br>implantable and high end<br>investigations.<br><br>This system should be<br>handled by persons who<br>have been briefed in its<br>professional handling and<br>who have familiarized<br>themselves with the product<br>by means of instructions for<br>use. Intended use also<br>means following the user<br>manual and observing the<br>conditions for inspection and<br>maintenance.<br><br>Exclusion: This system is<br>not recommended for<br>Mammography.<br><br>Contraindications:<br>Exposure of X-Ray should<br>be avoided during<br>pregnancy. | The Allura Xper series and<br>the Allura Xper OR Table<br>series (within the limits of<br>the used OR table) are<br>intended for use<br>on human patients to<br>perform:<br>• Vascular, cardiovascular<br>and neurovascular imaging<br>applications, including<br>diagnostic, interventional<br>and minimally<br>invasive procedures. This<br>includes, e.g., peripheral,<br>cerebral, thoracic and<br>abdominal angiography, as<br>well as PTAs, stent<br>placements, embolisations<br>and thrombolysis.<br>• Cardiac imaging<br>applications including<br>diagnostics, interventional<br>and minimally invasive<br>procedures (such as<br>PTCA, stent placing,<br>atherectomies),<br>pacemaker implantations,<br>and electrophysiology<br>(EP). • Non-vascular<br>interventions such as<br>drainages, biopsies and<br>vertebroplasties<br>procedures.<br>Additionally:<br>• The Allura Xper and<br>Allura Xper OR Table<br>series is compatible with a<br>hybrid Operating Room.<br>• Allura Xper FD10 is<br>compatible with specified<br>magnetic navigation<br>systems.<br>• Combined with a | The Ziehm Vision RFD<br>is intended for use in<br>providing medical<br>imaging for adult and<br>pediatric populations,<br>using pulsed and<br>continuous fluoroscopic<br>digital imaging, as well<br>as digital subtraction<br>and cine image capture<br>during diagnostic<br>interventional and<br>surgical procedures<br>where intra-operative<br>imaging and<br>visualization of<br>complex anatomical<br>structures of both lower<br>and higher contrast<br>density are required.<br>Such procedures may<br>include but are not<br>limited to those of<br>interventional<br>cardiology, heart<br>surgery, hybrid<br>procedures,<br>interventional<br>radiology,<br>interventional<br>angiography,<br>electrophysiology,<br>pediatrics, endoscopic,<br>urological,<br>gastroenterology,<br>orthopedic,<br>maxillofacial surgery,<br>neurology,<br>neurosurgery, critical<br>care, emergency room<br>procedures, and those<br>procedures visualizing<br>structures of the | Essentially<br>the same<br><br>Note: There<br>are no<br>differences<br>between<br>the subject<br>device and<br>the predicate<br>with respect<br>to indication<br>and intended<br>use. | |
| Feature | Cardiovascular<br>Angiography System<br>(Subject Device) | Allura Xper FD Series<br>and Allura Xper OR Table<br>series<br>(Predicate Device) | Ziehm Vision RFD<br>(Reference Device) | Justification<br>for<br>differences | |
| | | qualified, compatible OR<br>table, the Allura Xper OR<br>Table series can be used<br>for imaging in the Hybrid<br>OR within the applications<br>domains Neuro, Vascular,<br>Non Vascular and Cardiac.<br>The OR table can also be<br>used standalone for<br>surgical use in the OR. | cervical, thoracic, and<br>lumber regions of the<br>spine, and joint<br>fractures of the upper<br>and lower extremities,<br>and where digital<br>image data is required<br>for computer aided<br>surgery procedures<br>and whenever the<br>clinician benefits from<br>the high degree of<br>geometric imaging<br>accuracy, and where<br>such fluoroscopic, cine<br>and DSA imaging is<br>required. The<br>visualization of such<br>anatomical structures<br>assists the clinician in<br>the clinical outcome.<br>This device does not<br>support direct<br>radiographic film<br>exposures and is not<br>intended for use in<br>performing<br>mammography. The<br>system is not intended<br>for use near MRI<br>systems. | | |
| 5. X-Ray Generator | | | | | |
| Type | Microprocessor-controlled,<br>high frequency converter<br>generator | Microprocessor-controlled,<br>high frequency converter<br>generator | Monoblock and High<br>Frequency Generator | Same as<br>Predicate<br>device | |
| Power Rating | XGEN-100CV | 100KW | 100 KW | 30 KW…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.