K220280 · Arrow International, LLC (A subsidiary of Teleflex, Inc.) · DRB · Aug 30, 2022 · Cardiovascular
Device Facts
Record ID
K220280
Device Name
Arrow Stiffening Stylet
Applicant
Arrow International, LLC (A subsidiary of Teleflex, Inc.)
Product Code
DRB · Cardiovascular
Decision Date
Aug 30, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1380
Device Class
Class 2
Indications for Use
The Arrow Stiffening Stylet is intended to be placed inside a catheter to stiffen the catheter for placement.
Device Story
The Arrow Stiffening Stylet is a sterile, single-use, percutaneous peripheral vasculature catheter accessory. It consists of a PTFE-coated solid nitinol core wire with a straight tip, available in 0.015-0.017-inch diameters and 29.5-inch length. The device is provided with a flushable hub (cleared under K790408) that maintains stylet position, minimizes blood loss, and allows for flushing during catheter introduction. It is intended for transient use (up to 24 hours) and is packaged as a component within convenience kits for Arrow Peripherally Inserted Central Catheters (PICCs). The stylet is operated by a clinician during catheter placement to provide structural support, facilitating navigation and insertion. The device is removed after the catheter is successfully placed.
Clinical Evidence
No clinical data was required. Substantial equivalence is supported by bench testing, including mechanical/physical performance (corrosion resistance, fracture, flexure, tensile, torque, stiffness, coating integrity, and hub interface testing) and biocompatibility testing (cytotoxicity, sensitization, intracutaneous injection, acute systemic toxicity, pyrogenicity, and hemolysis).
Technological Characteristics
Solid nitinol core wire with PTFE coating; straight tip; 0.015-0.017 inch diameter; 29.5 inch length; includes flushable hub with side arm and slide clamp. Sterilized via Ethylene Oxide. Biocompatible per ISO 10993-1. Mechanical testing per ISO 11070.
Indications for Use
Indicated for placement within a percutaneous catheter to stiffen the device for placement.
Regulatory Classification
Identification
A catheter stylet is a wire that is run through a catheter or cannula to render it stiff.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and full name on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.
August 31, 2022
Arrow International, LLC (a subsidiary of Teleflex, Inc.) Fallon Young Regulatory Affairs Specialist 3015 Carrington Mill Blvd Morrisville, North Carolina 27560
Re: K220280
Trade/Device Name: Arrow Stiffening Stylet Regulation Number: 21 CFR 870.1380 Regulation Name: Catheter Stylet Regulatory Class: Class II Product Code: DRB Dated: July 25, 2022 Received: July 28, 2022
Dear Fallon Young:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Hetal Odobasic Assistant Director DHT2A: Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration<br>Indications for Use | Form Approved: OMB No. 0910-0120<br>Expiration Date: 06/30/2023<br>See PRA Statement below. |
|------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| 510(k) Number (if known) | K220280 |
|--------------------------|-------------------------|
| Device Name | Arrow Stiffening Stylet |
Indications for Use (*Describe*)
The Arrow Stiffening Stylet is intended to be placed inside a catheter to stiffen the catheter for placement.
Type of Use (*Select one or both, as applicable*)
| <div> <span> <svg class="bi bi-check-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> <path d="M10.97 4.97a.75.75 0 0 1 1.071 1.05l-3.992 4.99a.75.75 0 0 1-1.08.02L4.324 8.384a.75.75 0 1 1 1.06-1.06l2.094 2.093 3.473-4.425a.236.236 0 0 1 .02-.022z" fill-rule="evenodd"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> <svg class="bi bi-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review
instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of
information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it
displays a currently valid OMB number."*
FORM FDA 3881 (6/20)
Page 1 of 1 PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Arrow International. The word "ARROW" is in large, white, sans-serif font. Below it, in a smaller font, is the word "INTERNATIONAL". The background is a solid blue color.
## 510(k) SUMMARY
(as required by the Safe Medical Devices Act of 1990 and in accordance with 21 CFR 807.92)
# Arrow Stiffening Stylet
#### 1. Applicant Information:
| Name: | Arrow International LLC (subsidiary of Teleflex Incorporated) | |
|---------------------------|---------------------------------------------------------------|--|
| Address: | 3015 Carrington Mill Blvd<br>Morrisville, NC 27560 USA | |
| Contact Person: | Fallon Young | |
| Email: | fallon.young@teleflex.com | |
| Telephone Number: | (610) 984-7188 | |
| Fax Number: | (610) 478-3128 | |
| Date Prepared: | August 30, 2022 | |
| Device Name: | | |
| • Trade/Proprietary Name: | Arrow Stiffening Stylet | |
| • Classification Name: | Catheter, Stylet | |
| • CFR Number: | 21 CFR 870.1380 | |
- . Device Class: II
- DRB Product Code: .
## 3. Predicate Device(s):
2.
- . K182660: Galt Medical Corp's GaltTWS Stylet
{4}------------------------------------------------
#### 4. Description of Device:
The Arrow Stiffening Stylet is a sterile, single use, percutaneous peripheral vasculature catheter stylet intended for transient use (up to 24 hours) to be placed within a catheter to render it stiff to aid in catheter placement. The proposed device will be delivered as a sterile accessory within a convenience kit for the Arrow Peripherally Inserted Central Catheter (PICC)s. The Stylet is a PTFE coated solid nitinol core wire with straight tip. The stylet is available in 0.015-0.017-inch diameters and in 29.5-inch length with distance markings along the stylet body.
The Arrow Stiffening Stylet is provided with an existing, commercially available flushable hub with side arm and slide clamp cleared under K790408. The flushable hub maintains the stylet position within the catheter during catheter placement, is intended to minimize blood loss during catheter introduction, and allows the clinician to flush through the catheter to aid in catheter placement, if needed. There is no change to the flushable hub cleared under K790408 introduced by this submission.
#### ನ. Indications for Use:
The Arrow Stiffening Stylet is intended to be placed inside a catheter to stiffen the catheter for placement.
#### Substantial Equivalence 6.
# Intended Use / Indications for Use:
The intended use of the subject Arrow Stiffening Stylet is equivalent to the intended use of the predicate stylet. The subject and predicate devices are both intended to be placed within a catheter to stiffen the catheter for placement.
The indications for use of the Arrow Stiffening Stylet and the predicate stylet include the placement within a percutaneous catheter to stiffen the device for placement. The indications for use of the subject Arrow Stiffening Stylet are equivalent to the indications for use of the predicate stylet.
# Technological Characteristics:
The subject Arrow Stiffening Stylet incorporates the same fundamental technology as the predicate stylet. The subject Arrow Stiffening Stylet and the predicate stylet are both metal body stylets of similar configurations with a flushable hub.
{5}------------------------------------------------
Table 2 summarizes the substantial equivalence comparison of the subject Arrow Stiffening Stylet with the predicate stylet.
| | Proposed Device<br>Arrow Stiffening Stylet | Predicate Device – K182660<br>GaltTWS Stylet |
|--------------------------|---------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The Arrow Stiffening Stylet is intended<br>to be placed inside a catheter to stiffen<br>the catheter for placement. | The GaltTWS Stylet is a percutaneous catheter<br>stylet placed inside a catheter, lead, or cannula<br>to stiffen the device for placement. |
| Device Design Feature | | |
| Design | A solid nitinol wire | A stainless steel twisted/braided wire |
| Accessories | Existing, commercially available<br>flushable hub with side arm (K790408) | Existing, commercially available<br>flushable hub with side arm |
| Tip Type | Straight on both ends | Straight on both ends |
| Coating | PTFE | N/A |
| Core Material | Nitinol | Stainless Steel |
| Coil Material | N/A - no coil | N/A - no coil |
| Distal<br>Atraumatic Tip | Yes | Yes |
| Diameter | Diameter: 0.015-0.017 inch (0.39 mm-<br>0.44 mm) | Diameter: 0.012-0.015 inch (0.30 mm-0.39 mm) |
| Length | Length: 29.5 inch (75 cm) | Length: 30-72 inches (76 cm-182 cm) |
| Depth Control | Distance markings every 5 cm starting at<br>20 cm from distal end of stylet | No distance markings |
| Sterilization | | |
| Sterility | Ethylene Oxide | Ethylene Oxide |
| Biocompatibility | | |
| Biocompatibility | Biocompatible materials used<br>(per ISO 10993-1) | Biocompatible materials used<br>(per ISO 10993-1) |
Table 2 - Substantial Equivalence Comparison Summary
## 7. Non-Clinical Performance Data.
The technological differences between the subject Arrow Stiffening Stylet and the predicate stylet have been evaluated through verification testing. The testing conducted that supports substantial equivalence of the subject Arrow Stiffening Stylet is listed below.
Testing Included:
- Mechanical/Physical testing in accordance with ISO 11070, Guidance ●
{6}------------------------------------------------
Documents, and Arrow International Internal Requirements:
- Surface: Extraneous Matter and Defects o
- Corrosion Resistance o
- Radio-detectability O
- Guide wire Fracture O
- Guide wire Flexure O
- Guide wire Tensile O
- Retainer Leak (Flushable Hub Interface) O
- Retainer Grip (Flushable Hub Interface) O
- Coating Integrity O
- Particulate O
- Torque Test O
- Stiffness O
- Removal Force from Catheter O
- Human Factors/Usability Testing ●
- Biocompatibility testing in accordance with ISO 10993-1:
- Cytotoxicity (MEM Elution) o
- Sensitization Kligman Maximization Test (polar and non-polar o extracts)
- Intracutaneous Injection (polar and non-polar extracts) o
- Acute Systemic Toxicity (polar and non-polar extracts) O
- Materials Mediated Pyrogenicity о
- Hemolysis (direct and indirect contact) O
- The Arrow Stiffening Stylet will be packaged as a convenience kit ● component with Arrow Peripherally Inserted Central Catheter (PICC) products. There are no changes to the packaging introduced by this submission.
#### 8. Clinical Performance Data.
No human clinical data was required to support substantial equivalence.
### 9. Conclusion Regarding Substantial Equivalence
The information included in this premarket notification supports the substantial equivalence of the subject Arrow Stiffening Stylet to the
{7}------------------------------------------------
predicate stylet. The subject device has a substantially equivalent intended use, indications for use, and incorporates the same fundamental technology as the legally marketed predicate device to which it is compared.
Performance, human factors/usability, and biocompatibility testing has been conducted and has passed all acceptance criteria to verify that the proposed Arrow Stiffening Stylet meets its design, physical integrity, functional, and safety requirements and that any dimensional or material differences between it and the predicate stylet do not raise new issues of safety and effectiveness. The results of the testing included in this premarket notification support a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.