K220220 · Zhuhai Linte Medical Instrument Co., Ltd. · DXN · Oct 3, 2022 · Cardiovascular
Device Facts
Record ID
K220220
Device Name
Electronic Sphygmomanometer (Model: LT-P30)
Applicant
Zhuhai Linte Medical Instrument Co., Ltd.
Product Code
DXN · Cardiovascular
Decision Date
Oct 3, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
The Electronic Sphygmomanometer is intended for used by a person older than twelve (12) years to measure the systolic and diastolic blood pressure and pulse rate.
Device Story
Electronic Sphygmomanometer (LT-P30) is an automated, non-invasive blood pressure monitor for OTC use. Device uses an inflatable arm cuff; internal pump; and exhaust valve. User initiates measurement via 'START/STOP' button. Device employs oscillometric method to determine systolic/diastolic blood pressure and pulse rate. Features include irregular heartbeat detection; maximum pressure safety limit (300 mmHg); and LCD display for results. Data is stored in device memory. Cuff deflates automatically post-measurement. Device is intended for home use by patients >12 years old. Output allows patients to monitor blood pressure trends, aiding in self-management of cardiovascular health.
Clinical Evidence
Clinical validation conducted per ISO 81060-2:2018. Study included 120 subjects (61 female, 59 male) aged 13 to 85. Results met criterion 1 and criterion 2 of ISO 81060-2. No adverse effects or complications reported.
Technological Characteristics
Oscillometric measurement; automatic pump/valve system. Materials: 210D raw silk cloth (arm cuff). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 80601-2-30, ISO 10993-1/-5/-10, ISO 81060-2, IEC 80369-5. Power: 4 AA batteries or USB Type-C. Display: LCD. Connectivity: None (standalone).
Indications for Use
Indicated for persons older than 12 years to measure systolic and diastolic blood pressure and pulse rate.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
A&D Medical UA-651Digital Blood Pressure Monitor (K141160)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 3, 2022
Zhuhai Linte Medical Instrument Co., Ltd. % Becky Chen Registered Engineer Feiying Drug & Medical Consulting Technical Service Group Rm 2401 Zhenye International Business Center, No. 3101-90, Qianhai Road Shenzhen, GuangDong 518052 China
Re: K220220
Trade/Device Name: Electronic Sphygmomanometer (Model: LT-P30) Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: August 30, 2022 Received: August 30, 2022
Dear Becky Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220220
Device Name
Electronic Sphygmomanometer (Models: LT-P30)
Indications for Use (Describe) Measure blood pressure (systolic and diastolic) and pulse rate.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510 (k) Summary
This "510(k) Summary" of 510(k) safety and effectiveness information is submitted in accordance with requirements of Title 21, CFR Section 807.92.
## (1) Applicant information:
| 510(k) owner's name: | Zhuhai Linte Medical Instrument Co., Ltd. |
|----------------------------|-----------------------------------------------------------------------------------------------------------------|
| Address: | 4th Floor, Building 1, No.66, Yongda Road, Hongqi Town,<br>Jinwan District, Zhuhai, Guangdong, 519090 P.R.China |
| Contact person: | Wallace Huang |
| Phone number: | +86 13823146691 |
| Email: | wallace.huang@lintemed.com |
| Date of summary prepared: | January 10, 2022 |
| Reason for the submission: | New device, there were no prior submissions for the<br>device. |
## (2) Proprietary name of the device
| Trade name/model: | Electronic Sphygmomanometer /Model: LT-P30 |
|----------------------|---------------------------------------------------|
| Common name: | Blood Pressure Monitor |
| Classification name: | System, measurement, blood-pressure, non-invasive |
| Regulation number: | 21CFR 870.1130 |
| Product code: | DXN |
| Review panel: | Cardiovascular |
| Regulation class: | Class II |
## (3) Predicate device
| Sponsor | A&D Engineering, Inc. |
|-----------------------|--------------------------------------------------|
| Device Name and Model | A&D Medical UA-651Digital Blood Pressure Monitor |
| 510(k) Number | K141160 |
| Product Code | DXN |
| Regulation Number | 21CFR 870.1130 |
| Regulation Class | Class II |
## (4) Description/ Design of device:
The Electronic Sphygmomanometer uses an inflatable cuff which is wrapped around the patient's upper arm. After the user pushes the "START/STOP" button, the cuff is inflated automatically by an internal pump. The systolic and diastolic blood pressures are determined by oscillometric method. The deflation rate is controlled by the internal
{4}------------------------------------------------
exhaust valve. There is a quick exhaust mechanism so that the cuff pressure can be completely released urgently. There is a maximum pressure safety setting at 300 mmHg. The device will not inflate the cuff higher than 300 mmHg. The device will turn on an irregular heartbeat indicator if an irregular heartbeat was detected during the measurement process. At the end of the measurement, the systolic and diastolic pressures with pulse rate are shown on the LCD and stored in the device memory. The cuff is also deflated automatically to 0 mmHg at the same time.
## (5) Intended use:
The Electronic Sphygmomanometer is intended for used by a person older than twelve (12) years to measure the systolic and diastolic blood pressure and pulse rate.
## (6) Indications for use:
Measure blood pressure (systolic and diastolic) and pulse rate.
## (7) Materials
| Component name | Material of<br>Component | Body Contact<br>Category | Contact Duration |
|----------------|--------------------------|--------------------------|--------------------|
| Arm cuff | 210D raw silk cloth | Surface skin contact | Less than 24 hours |
We have directly purchased arm cuff from qualified supplier which has obtained Biocompatibility test reports. For details, please refer to "Biocompatibility Discussion".
| Item | Targeted device | Predicate device | Remark |
|-------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| Trade name | Electronic<br>Sphygmomanometer<br>(Models: LT-P30) | A&D Medical<br>UA-651Digital Blood<br>Pressure Monitor | / |
| 510 (k) number | Pending | K141160 | / |
| Regulation<br>number | 21CFR 870.1130 | 21 CFR 870.1130 | Same |
| Regulation<br>description | Noninvasive blood<br>pressure measurement<br>system | Noninvasive blood pressure<br>measurement system | Same |
| Product code | DXN | DXN | Same |
| Class | II | II | Same |
| Indications for<br>use | Measure blood pressure<br>(systolic and diastolic) and<br>pulse rate. | Measure blood pressure<br>(systolic and diastolic) and<br>pulse rate. | Same |
| Intended use | The Electronic<br>Sphygmomanometer is | The A&D Medical UA-651<br>family digital blood | Same |
| Item | Targeted device | Predicate device | Remark |
| | intended for used by a<br>person older than twelve<br>(12) years to measure the<br>systolic and diastolic<br>blood pressure and pulse<br>rate. | pressure monitor is intended<br>for used by a person older<br>than twelve (12) years to<br>measure the systolic and<br>diastolic blood pressure and<br>pulse rate. | |
| Intended patient | Person older than 12 years | Person older than 12 years | Same |
| Application sites | Upper arm | Upper arm | Same |
| Location for use | OTC | OTC | Same |
| Operation<br>principle | Oscillometric | Oscillometric | Same |
| Measurement<br>range | Pressure: 20-280mmHg<br>Pulse Rate: 30-200bpm | Pressure: 20-280mmHg<br>Pulse Rate: 30-200bpm | Same |
| Accuracy | Pressure: ±3mmHg or<br>±2% of measured value,<br>whichever is greater;<br>Pulse Rate: ±5% | Pressure: ±3mmHg or ±2%<br>of measured value,<br>whichever is greater;<br>Pulse Rate: ±5% | Same |
| Display type | LCD | LCD | Same |
| Pressurization<br>source | Automatic internal pump | Automatic internal pump | Same |
| Cuff deflation<br>method | Standard exhaust valve | Standard exhaust valve | Same |
| Cuff attachment<br>method | By plastic hose connected to monitor | By plastic hose connected to monitor | Same |
| Cuff design | D-ring cuffs | D-ring cuffs | Same |
| Cuff<br>circumference | Small cuff: 17 ~ 22 cm<br>Medium cuff: 22 ~ 32cm<br>Large cuff: 22 ~ 42 cm | Small cuff: 16 ~ 24 cm<br>Medium cuff: 23 ~ 37cm<br>Large cuff: 31 ~ 45 cm | Similar -<br>Within<br>the scope<br>of<br>predicate<br>device |
| Memory | 45×3 recording of<br>measurement data for<br>multiple person | 30 memories | Similar -<br>does not<br>affect the<br>safety and<br>effectiveness of the<br>device |
| Power source | 4 AA size batteries and | 4 AA size batteries and AC | Similar |
| Item | Targeted device | Predicate device | Remark |
| Operating<br>Environment | USB type-C cable as an<br>option | adaptor as an option | |
| | Temperature:<br>+5°C~+40°C;<br>Humidity:<br>15~80%RH | Temperature:<br>+10°C~+40°C;<br>Humidity:<br>30~85%RH | Similar-<br>does not<br>affect the<br>safety and<br>effectiveness of the<br>device |
| | | | |
| Storage<br>Environment | Temperature:<br>-40°C~+55°C;<br>Humidity:<br>$≤$ 93%RH | Temperature:<br>-20°C~+60°C;<br>Humidity:<br>10~95%RH | Similar-<br>does not<br>affect the<br>safety and<br>effectiveness of the<br>device |
| | | | |
| | | | |
| Compliance with<br>voluntary<br>standards | IEC 60601-1;<br>IEC 60601-1-2;<br>IEC 60601-1-11;<br>IEC 80601-2-30;<br>ISO 10993-1,-5,-10<br>ISO 81060-2,<br>IEC 80369-5 | IEC 60601-1;<br>IEC 60601-1-2;<br>IEC 60601-1-11;<br>IEC 80601-2-30;<br>ISO 10993-1,-5,-10<br>ISO 81060-2 | Same |
#### (8) Technological characteristics and substantial equivalence:
{5}------------------------------------------------
{6}------------------------------------------------
Zhuhai Linte Medical Instrument Co., Ltd. 510(k)s - Section 8. 510 (k) Summary
#### Conclusion:
Electronic Sphygmomanometer (Models: LT-P30) is substantial equivalent to the predicate device.
#### (10) Performance Data:
The performance data includes "Non-Clinical Data" and "Clinical Data", brief description of which are shown as below:
## Non-Clinical Data
The following performance data have been conducted to verify that the Electronic Sphygmomanometer (Models: LT-P30) meets all design specifications which supports the conclusion that it's Substantially Equivalent (SE) to the predicate device. The testing results demonstrate that the targeted device complies with the following standards:
#### Biocompatibility Testing:
The biocompatibility evaluation for the body-contacting component (arm cuff) of this
{7}------------------------------------------------
device was conducted in accordance with the "Use of International Standard ISO 10993-1, Biological Evaluation of Medical Device - Part 1: Evaluation and Testing Within a Risk Management Process", as recongnized by FDA. The arm cuff has performed and passed the Biocompatibility test. So we have reason to believe that the arm cuff is safe for the users. The arm cuff complies with the following standards:
- ISO 10993-5, Biological Evaluation Of Medical Devices -- Part 5: Tests For A InVitro Cytotoxicity
- A ISO 10993-10, Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization.
## Electrical and EMC Safety:
The electrical safety and EMC safety testing was performed to, and passed, the following standards:
- ANSI AAMI ES60601-1, Medical electrical equipment -- Part 1: General A requirements for basic safety and essential performance
- IEC 60601-1-11, Medical electrical equipment -- Part 1-11: General requirements A for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- > IEC 60601-1-2, Medical electrical equipment -- Part 1-2: General requirements for basic safety and essential performance -- Collateral standard: Electromagnetic disturbances - Requirements and tests
## Performance:
The performance testing was performed to, and passed, the following standards:
- IEC 80601-2-30, Medical Electrical Equipment -- Part 2-30: Particular A Requirements For The Basic Safety And Essential Performance Of Automated Non-Invasive Sphygmomanometers
- > IEC 80369-5. Small-bore connectors for liquids and gases in healthcare applications - Part 5: Connectors for limb cuff inflation applications
## Software:
We have also conducted Software verification and validation test according to the requirements of the FDA "Guidance for Pre Market Submissions and for Software Contained in Medical Devices".
## Clinical Data
The clinical testing has been conducted per ISO 81060-2: 2018 Non-Invasive Sphygmomanometers -- Part 2: Clinical Validation of Automated Measurement Type on the Electronic Sphygmomanometer.
{8}------------------------------------------------
There are 120 subjected involved in the study, of which 61 female and 59 male. All the subjects are aged from 13 to 85. And the size distribution meets the requirements described in clause 5.1.4 of the standard.
The subject devices use the oscillometric method to measure blood pressure, the accuracy of determination depends on the sensor, race and ethnicity differences do not have an effect on the measuring result in the specified measuring scope.
During the testing, there were no adverse effects and complications occurred, and the results both meet criterion 1 and criterion 2 of ISO 81060-2.
#### Summary:
Based on the above non-clinical data and clinical data as documented in this application, the Electronic Sphygmomanometer was found to have a safety and effectiveness profile that is similar to the predicate device.
## (11) Conclusion
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the comparison of intended use, design, materials and performance, the Electronic Sphygmomanometer is to be concluded substantial equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.