K220217 · NormaTec Industries, LP · IRP · Feb 25, 2022 · Physical Medicine
Device Facts
Record ID
K220217
Device Name
Normatec 3
Applicant
NormaTec Industries, LP
Product Code
IRP · Physical Medicine
Decision Date
Feb 25, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5650
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The Normatec 3 is an air pressure massager intended to temporarily relieve minor muscle aches and to temporarily increase circulation to the treated areas.
Device Story
Normatec 3 is a powered inflatable tube massager for temporary muscle ache relief and circulation increase. Device uses external power supply or internal lithium-ion battery to drive a compressor and valve system; sequentially inflates/deflates appliance cells to simulate manual kneading/stroking. Operated by users in clinics, hospitals, athlete training, or home environments via on-device buttons or Bluetooth-connected mobile app (Android/iOS). App serves as alternative user interface for setting treatment time, intensity (pressure), and attachment selection. Output is mechanical pressure applied to limbs/hips. Healthcare providers or patients use device to manage muscle recovery; benefits include temporary pain relief and improved circulation.
Clinical Evidence
Bench testing only. No clinical data provided. Performance, functionality, and reliability were established through non-clinical verification and validation testing, demonstrating the device is as safe and effective as the predicate.
Technological Characteristics
Powered inflatable tube massager; 15 VDC power supply (IEC 60601-1) or internal lithium-ion battery (IEC 62133-2). Materials: 200 denier nylon with polyurethane laminate/extrusion; molded ABS (94V0) housing. Sequential gradient, peristaltic, and pulsing inflation/deflation. Pressure range 0-110 mmHg. Bluetooth connectivity for app-based control. Complies with AAMI ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-1-6, and ANSI C63.27-2017.
Indications for Use
Indicated for adults in good health for temporary relief of minor muscle aches/pains and temporary increase in circulation to treated areas (leg, arm, hip).
Regulatory Classification
Identification
A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 25, 2022
NormaTec Industries, LP % Prithul Bom Responsible Third Party Official Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114
Re: K220217
Trade/Device Name: Normatec 3 Regulation Number: 21 CFR 890.5650 Regulation Name: Powered inflatable tube massager Regulatory Class: Class II Product Code: IRP Dated: January 25, 2022 Received: January 26, 2022
Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K220217
Device Name Normatec 3
Indications for Use (Describe)
The Normatec 3 is an air pressure massager intended to temporarily relieve minor muscle aches and to temporarily increase circulation to the treated areas.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## NormaTec Industries, LP
| Date of Preparation: | February 22, 2022 |
|---------------------------------------------------------|---------------------------------------------------------|
| 480 Pleasant Street<br>Suite A200<br>Watertown MA 02472 | Tel - 800.355.0960<br>Fax - 866.279.2579 |
| Official Contact: | Steve Henderson, Quality Systems and Regulatory Manager |
| Proprietary or Trade<br>Name: | Normatec 3 |
| Subject Device: | Normatec 3 |
| Common/Usual Name: | Massager, Powered Inflatable Tube |
| Classification Name | Powered inflatable tube massager |
| Regulation Number: | 21 CFR 890.5650 |
| Product Code: | IRP |
| Regulation Medical<br>Speciality: | Physical Medicine |
| Predicate Device: | K183169 – NormaTec Pulse 2.0 and Pulse Pro 2.0 |
| Classification Name | Powered inflatable tube massager |
| Common/Usual Name | Massager, Powered Inflatable Tube |
| Regulation Number: | 21 CFR 890.5650 |
| Product Code | IRP |
| Regulation Medical<br>Speciality: | Physical Medicine |
{4}------------------------------------------------
### Device Description
The Normatec 3 is a powered inflatable tube massager (Product Code "IRP"). It is intended to temporarily relieve minor muscle aches and/or pains, and to temporarily increase circulation to the treated areas. It simulates manual kneading and stroking of tissues by the use of an inflatable pressure cuff. The devices are to be used by people who are in good health. The devices are powered from an external IEC 60601-1 compliant power supply and can also be powered by an internal IEC 62133-2 compliant lithium-ion battery.
The user interface on the Normatec 3 is a series of buttons with one small display the treatment time. The user interface provides for:
- Starting and stopping the massage treatment; .
- Adjusting time and intensity (pressure) of the treatment; .
- Selection of Attachment in use (Leg, Arm, Hip); .
- . Selection of Zone Boost which provides an increase of 10 mm Hg to the Zone selected
In addition to the user interface on the devices the proposed devices have Bluetooth capability that allows the use of a NormaTec app to control the device. The Bluetooth app allows the user to use a compatible Android or iOS phone to select and set device parameters listed above for convenience.
Intended User
ОТС
Patient Population
Adults
#### Indications for Use
The NormaTec Normatec 3 is an air pressure massager intended to temporarily relieve minor muscle aches and/or pains, and to temporarily increase circulation to the treated areas.
#### Intended Use Environment:
Clinics, hospital, athlete training, and home environments.
{5}------------------------------------------------
#### 510(k) Summary Page 3 of 9
### Table 1 – Table of Device Comparisons and Differences
| | Predicate Device | Proposed New Device | Comment | Characteristic | Pulse 2.0 / Pulse<br>Pro 2.0 | Normatec 3 |
|---------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|------------|
| Model Name<br>510(k) Number | NormaTec Pulse 2.0 and Pulse Pro 2.0<br>510(k) K183169 | Normatec 3<br>510(k) K220217 | N/A | Power On - The system recalls and displays the last used settings for: | | |
| Manufacturer | NormaTec Industries, LP | Same | N/A | Treatment time | X | X |
| Prescriptive | No, OTC | Same | N/A | Treatment mode | X | N/A |
| Indications for use | The NormaTec Pulse 2.0 and Pulse<br>Pro 2.0 are air pressure massagers<br>intended to temporarily relieve minor<br>muscle aches and/or pains, and to<br>temporarily increase circulation to the<br>treated areas. | The NormaTec Normatec 3 is an air<br>pressure massager intended to<br>temporarily relieve minor muscle aches<br>and/or pains, and to temporarily<br>increase circulation to the treated<br>areas. | N/A | Therapy mode<br>Rehab mode - Recalls zone focus<br>Custom - recalls all settings<br>Recovery Flush | Pulse 2.0 – N/A<br>Pulse Pro 2.0 – X | N/A |
| Intended Use<br>Environment | Clinics, hospital, athlete training, and<br>home environments | Same | N/A | Rehab<br>Preset NormaTec PULSE time / PULSE pressure values<br>Leg appliance - Foot/ankle, calf, knee, lower quad, upper quad<br>Arm appliance - Hand/wrist, forearm, elbow, bicep, shoulder<br>Hip appliance - Quadriceps, hip | Pulse 2.0 – N/A<br>Pulse Pro 2.0 – X | N/A |
| Power Source(s) | 15 VDC via an IEC 60601-1<br>compliant power supply (100- 240 VAC<br>input)<br>Integrated rechargeable battery | Same | N/A | Appliance Type | Pulse 2.0 – N/A<br>Pulse Pro 2.0 – X | X |
| Software / Firmware<br>Micro-processor<br>Control | Microprocessor | Same | N/A | Pressure Level | X | X |
| Technology | Compressor and valve system that<br>sequentially inflates cells of appliance.<br>Bluetooth communication ability. | Same | N/A | Number of zones | X | X |
| Compliance with<br>Voluntary standards | AAMI ANSI ES60601-1, IEC 60601-1-<br>2, IEC 60601-1-11 and ANSI C63.27-<br>2017. | Same and includes IEC 60601-1-6<br>compliance | The additional testing of 60601-1-6<br>compliance does not raise any new<br>device safety concerns | Rest time | X | x |
| | Predicate Device | Proposed New Device | Comment | Time | | |
| Device Pressure<br>Range | 0-110 mmHg | Same | N/A | Add time<br>Before treatment<br>During treatment | X | X |
| Pressure Levels | Level 1: 40 mm Hg max<br>Level 2: 50 - 60 mm Hg max<br>Level 3: 60 - 70 mm Hg max<br>Level 4: 70 - 80 mm Hg max<br>Level 5: 80 mm Hg max<br>Level 6: 80 - 90 mm Hg max<br>Level 7: 100 mm Hg max | Same | N/A | Decrease time<br>Before treatment<br>During treatment | X | x |
| ZoneBoost | Provides an increase of 10 mm Hg to<br>the Zone selected | Same | N/A | Continuous mode | X | N/A |
| Treatment Time | Stays on until the user turns it off or<br>can be set up to turn off in a range of<br>10 minutes to continuous | Stays on until the user turns it off or<br>can be set up to turn off in a range of<br>15 minutes to 60 minutes | Changed setup times for treatment of<br>new device to a max of 60 minutes<br>instead of continuous to match user<br>typical sessions with device. | Counter | X | X |
| Inflation/Deflation<br>Cycle Type | Sequential Gradient, Peristaltic and<br>Pulsing | Same | N/A | Count down - 1 second increments | X | X |
| Appliance Contact<br>Surface Material | 200 denier nylon with a polyurethane<br>laminate/extrusion | Same | N/A | Count up (continuous mode) - 1 second increments | X | N/A |
| Number of Inflatable<br>appliance segments | 5 or less | Same | N/A | Compliance - Trip meter | Pulse 2.0 – N/A<br>Pulse Pro 2.0 – X | X |
| Weight | 3.6 pounds (incl. battery) | Same | N/A | Chronometer Odometer - cannot be reset by User | X | X |
| Dimensions<br>(W x H x D) | 4.4" x 3.8" x 8.1" | 4.17 x 3.66 x 8.25 | N/A | Pressure | | |
| Housing Materials<br>and Constructions | Molded ABS enclosure (94V0) | Same | N/A | | | |
| Patient contact | Non-conductive appliances | Same | N/A | | | |
{6}------------------------------------------------
## K220217
# 510(k) Summary
Page 4 of 9
{7}------------------------------------------------
## Differences Between Models
{8}------------------------------------------------
K220217
# 510(k) Summary
Page 6 of 9
| Characteristic | Pulse 2.0 / Pulse<br>Pro 2.0 | Normatec 3 |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|------------|
| Increase level | | |
| Before treatment | × | × |
| During treatment | | |
| Decrease level | | |
| Before treatment | × | × |
| During treatment | | |
| Characteristic | Pulse 2.0 / Pulse<br>Pro 2.0 | Normatec 3 |
| Treatment Mode – Can ONLY be changed before treatment begins | | |
| NormaTec Pulse | X | X |
| Sequential | X | N/A |
| Rest Time | | |
| View or change rest time – before treatment | X | N/A |
| # of Zones | | |
| Change number of zones – before treatment | X | N/A |
| Treatment | | |
| Starting | X | X |
| Pause | X | X |
| Un-pause | X | X |
| End - If timer reaches 0:0000, treatment will continue until Rest period<br>is reached. | X | X |
| Remote operation via Bluetooth app | X | X |
| Battery Charging | | |
| Displays battery level | X | X |
| Displays when battery is charging | X | X |
| Power off | | |
| No treatment in process | X | X |
| Treatment is process | X | X |
| Appliance type | | |
| Select appliance – boot, arm, hip<br>NOTE: If hip appliance is selected, treatment mode defaults to 2-zone<br>treatment | X | X |
| Therapy Mode | | |
| Recovery Flush<br>Preset NormaTec PULSE time / PULSE pressure values<br>Appliance type applicable - boot, arm, hip | Pulse 2.0 – N/A<br>Pulse Pro 2.0 – X | N/A |
| Rehab<br>Preset NormaTec PULSE time / PULSE pressure values<br>Leg appliance - Foot/ankle, calf, knee, lower quad, upper quad<br>Arm appliance - Hand/wrist, forearm, elbow, bicep, shoulder<br>Hip appliance - Quadriceps, hip | Pulse 2.0 – N/A<br>Pulse Pro 2.0 – X | N/A |
| Characteristic | Pulse 2.0 / Pulse<br>Pro 2.0 | Normatec 3 |
| Custom Settings<br>Allows user to program NormaTec PULSE pressure and time values for<br>each zone<br>Pressure range – 30-110 mmHg, user can select in 10 mmHg<br>increments<br>Time range - 15 seconds to 4 minutes, user can select in 15 second<br>increments | Pulse 2.0 – N/A<br>Pulse Pro 2.0 - Χ | N/A |
| Display Settings Adjustment | | |
| Brightness controls | Pulse 2.0 – N/A<br>Pulse Pro 2.0 - Χ | N/A |
{9}------------------------------------------------
#### 510(k) Summary Page 7 of 9
{10}------------------------------------------------
K220217
#### 510(k) Summary Page 8 of 9
### Determination of Substantial Equivalence
The Normatec 3 is substantially equivalent to currently marketed and cleared devices (K183169) because:
#### Indications for Use
The Normatec 3 indications for use are identical to the predicate, NormaTec Pulse 2.0 and Pulse Pro 2.0, 510(k) K183169.
#### Prescriptive
The Normatec 3 and the predicate (K183169) are OTC devices.
#### Design, Technology, and Principle of Operation
The Normatec 3 has equivalent design and features when compared to the predicate and have similar technology to the predicate, NormaTec Pulse 2.0 and Pulse Pro 2.0 510(k) K183169.
The proposed device has Bluetooth capability that allows the use of a NormaTec app to control the device. In addition to the user interface on the devices, the proposed device has Bluetooth capability that allows the use of a NormaTec app to control the device. The Bluetooth app allows the user to use a compatible Android or iOS phone to select and set device parameters listed above for convenience.
#### Performance and Specifications
The Normatec 3 has equivalent specifications of performance when compared to the predicate, NormaTec Pulse 2.0 and Pulse Pro 2.0 510(k) K183169.
#### Compliance with Standards
The predicate devices are compliant with AAM ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-11 and ANSI C63.27-2017.
The Normatec 3 complies with AAM ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-2, IEC 60601-1-11 and ANSI C63.27-2017.
#### Materials
The patient-contacting materials of the Normatec 3 are the inflatable appliances that are identical to the predicate, NormaTec Pulse 2.0 and Pulse Pro 2.0 510(k) K183169.
{11}------------------------------------------------
#### Intended Use Environment
Clinics, hospital, athlete training, and home environments, which are identical to the predicate.
#### Features
The Normatec 3 has equivalent features when compared to the predicate, NormaTec Pulse 2.0 and Pulse Pro 2.0 510(k) K183169.
### Conclusion
The Normatec 3 is substantially equivalent to the predicate, NormaTec Pulse 2.0 and Pulse Pro 2.0 510(k) K183169 in:
- Patient population, .
- Environment of use, .
- . Technology characteristics,
- Materials, ●
- Specifications / performance, and .
- Compliance with international standards. .
The inclusion of the Bluetooth capability and associated app does not affect the equivalency of the proposed devices and the predicate. The app functions only as an alternative user interface. No additional features are provided or unlocked by the app. Nor is the therapy provided by the devices affected by using the app.
Other minor differences as detailed in the substantial equivalence table do not raise questions of safety and effectiveness.
#### Indications for Use
The Normatec 3 Indications for Use and Contraindications are identical to the predicate NormaTec Pulse 2.0 and Pulse Pro 2.0 510(k) K183169.
Verification and Validation activities required that establish the performance, functionality, and reliability characteristics of the Normatec 3 with respect to the predicate were performed. Testing performed demonstrated that the proposed device meets defined requirements and performance claims.
The summary includes the conclusions drawn from the nonclinical tests (discussed above) that demonstrate that the device is as safe, as effective, and performs as well as the predicate device according to 807.92(b)(3).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.