← Product Code [LZA](/productcode/LZA) · K220215

# Powder Free Nitrile Examination Glove, Non-sterile, Tested for Use with Chemotherapy Drugs, and the Opioid Fentanyl citrate, simulated Gastric Acid, and Fentanyl in Gastric Acid (Blue) (K220215)

_Ksg Medicare Sdn. Bhd. · LZA · Mar 26, 2022 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K220215

## Device Facts

- **Applicant:** Ksg Medicare Sdn. Bhd.
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Mar 26, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

A Nitrile powder free examination glove is a disposable device intended for medical purposes, worn on the examiner's hand or finger to prevent contamination between examiner and patient. This specialty glove has also been tested for use with Chemotherapy Drugs and the Opioid Fentanyl citrate, simulated Gastric Acid and Fentanyl in Gastric Acid.

## Device Story

Disposable nitrile examination glove; worn by healthcare providers to prevent cross-contamination between examiner and patient. Provides barrier protection against biological contaminants, specific chemotherapy drugs, and opioid exposure (Fentanyl citrate). Tested for breakthrough times against various hazardous agents; results guide selection based on specific drug exposure risks. Not for use with Carmustine or ThioTepa due to low breakthrough times. Used in clinical settings; provides patient and provider safety by minimizing contact with hazardous substances.

## Clinical Evidence

Bench testing only. Permeation testing performed per ASTM standards to determine breakthrough detection times for 21 chemotherapy drugs, Fentanyl citrate, and simulated gastric acid. Results demonstrate no breakthrough up to 240 minutes for most agents, with specific exceptions for Carmustine (23.6 min) and ThioTepa (57.8 min).

## Technological Characteristics

Material: Nitrile. Form factor: Powder-free, non-sterile examination glove. Barrier protection: Tested for chemical permeation resistance against chemotherapy agents and opioids. Connectivity: None. Energy source: None. Sterilization: Not specified.

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 26, 2022

KSG MEDICARE Sdn. Bhd. % Wava Truscott President Truscott MedSci Associates, LLC 180 Burkemeade Ct. Roswell, Georgia 30075

Re: K220215

Trade/Device Name: Powder Free Nitrile Examination Glove, Non-sterile, Tested for Use with Chemotherapy Drugs, and the Opioid Fentanyl citrate, simulated Gastric Acid, and Fentanyl in Gastric Acid (Blue) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I, reserved Product Code: LZA, LZC, OPJ, ODO Dated: January 23, 2022 Received: January 26, 2022

Dear Wava Truscott:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Clarence W. Murray, III, PhD Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

## 510(k) Number (if known) K220215

## Device Name

Powder Free Nitrile Examination Glove, Non-Sterile, Tested for Use with Chemotherapy Drugs, and the Opioid Fentanyl citrate, simulated Gastric Acid, and Fentanyl in Gastric Acid (Blue)

## Indications for Use (Describe)

Intended for use: A Nitrile powder free examination glove is a disposable device intended for medical purposes, worn on the examiner's hand or finger to prevent contamination between examiner and patient. This specialty glove has also been tested for use with Chemotherapy Drugs and the Opioid Fentanyl citrate, simulated Gastric Acid and Fentanyl in Gastric Acid.

| Chemotherapy Drug          | Concentration<br>Tested | Minimum<br>Breakthrough Detection Time |
|----------------------------|-------------------------|----------------------------------------|
| Carboplatin                | 10.0mg/ml               | No breakthrough up to 240 minutes      |
| Carmustine (BCNU)          | 3.3mg/ml                | 23.6 minutes                           |
| Cisplatin                  | 10.0mg/ml               | No breakthrough up to 240 minutes      |
| Cytarabine HCL             | 100.0mg/ml              | No breakthrough up to 240 minutes      |
| Cyclophosphamide (Cytoxan) | 20.0mg/ml               | No breakthrough up to 240 minutes      |
| Dacarbazine (DTIC)         | 10.0mg/ml               | No breakthrough up to 240 minutes      |
| Docetaxel                  | 10.0mg/ml               | No breakthrough up to 240 minutes      |
| Doxorubicin Hydrochloride  | 2.0mg/ml                | No breakthrough up to 240 minutes      |
| Etoposide (Toposar)        | 20.0mg/ml               | No breakthrough up to 240 minutes      |
| Fluorouracil               | 50.0mg/ml               | No breakthrough up to 240 minutes      |
| Gemcitabine                | 38.0mg/ml               | No breakthrough up to 240 minutes      |
| Ifosfamide                 | 50.0mg/ml               | No breakthrough up to 240 minutes      |
| Irinotecan                 | 20.0mg/ml               | No breakthrough up to 240 minutes      |
| Mechlorethamine            | 1.0mg/ml                | No breakthrough up to 240 minutes      |
| Melphalan                  | 5.0mg/ml                | No breakthrough up to 240 minutes      |
| Methotrexate               | 25.0mg/ml               | No breakthrough up to 240 minutes      |
| Mitomycin C                | 0.5mg/ml                | No breakthrough up to 240 minutes      |
| Mitoxantrone               | 2.0mg/ml                | No breakthrough up to 240 minutes      |
| Paclitaxel (Taxol)         | 6.0mg/ml                | No breakthrough up to 240 minutes      |
| ThioTepa                   | 10.0mg/ml               | 57.8 minutes                           |
| Vincristine Sulfate        | 1.0mg/ml                | No breakthrough up to 240 minutes      |

Important: Carmustine and ThioTepa have extremely low minimal breakthrough times of 23.6 minutes and 57.8 minutes respectively. Warning: Do not use with Carmustine or ThioTepa

| Opioid Drug                   | Concentration<br>Tested | Minimum Breakthrough Detection<br>Time |
|-------------------------------|-------------------------|----------------------------------------|
| Fentanyl citrate (injectable) | 100mg/2ml               | No breakthrough up to 240 minutes      |
| Gastric Acid (simulated)      | 0.2%NaCl in 0.7 HCL     | No breakthrough up to 240 minutes      |
| Fentanyl in Gastric Acid      | 50/50 mix               | No breakthrough up to 240 minutes      |

Please Note: Gloves used for protection against possible Chemotherapy Drug exposure should be selected specifically for the type of drugs used. Users should review the drug labeling or material data sheets for the drug being used to determine an adequate level of protection.

{3}------------------------------------------------

| Type of Use (Select one or both, as applicable)                                               | <table style="border:none;"><tr><td><span style="border: 1px solid #000000;"></span>Prescription Use (Part 21 CFR 801 Subpart D)</td><td><span style="border: 1px solid #000000;">☒</span>Over-The-Counter Use (21 CFR 801 Subpart C)</td></tr></table> | <span style="border: 1px solid #000000;"></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="border: 1px solid #000000;">☒</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| <span style="border: 1px solid #000000;"></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="border: 1px solid #000000;">☒</span> Over-The-Counter Use (21 CFR 801 Subpart C)                                                                                                                                                           |                                                                                               |                                                                                               |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services  

 Food and Drug Administration  

 Office of Chief Information Officer  

 Paperwork Reduction Act (PRA) Staff  

[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)

*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*

| FORM FDA 3881 (6/20) | Page 1 of 1 | PSC Publishing Services (301) 443-6740 EF |
|----------------------|-------------|-------------------------------------------|
|----------------------|-------------|-------------------------------------------|

-

---

**Source:** [https://fda-staging.innolitics.com/device/K220215](https://fda-staging.innolitics.com/device/K220215)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
