← Product Code [ITI](/productcode/ITI) · K220213

# Empulse R90 (K220213)

_Sunrise Medical (Us), LLC · ITI · Mar 30, 2022 · Physical Medicine · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K220213

## Device Facts

- **Applicant:** Sunrise Medical (Us), LLC
- **Product Code:** [ITI](/productcode/ITI.md)
- **Decision Date:** Mar 30, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.3860
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The Empulse R90 is intended to be used to provide power assistance to manual wheelchair users. It is designed to augment the manual propulsion provided by the user, reducing the effort required by the user to propel the wheelchair.

## Device Story

Empulse R90 is an auxiliary power module for manual wheelchairs; utilizes single electric drive wheel to provide augmented propulsion. Device mounts to rigid wheelchair camber tubes or folding wheelchair compressible axles. On-board sensors (inclinometer, accelerometer, motor current) detect user intent to move forward and terrain incline/decline. Actuator automatically engages/disengages locking latch and compresses pre-load traction spring for drive wheel contact. Removable Lithium-Ion battery pack powers system. Operated by wheelchair user; provides power assistance to reduce physical effort. Control box connects via Bluetooth LE 5.2 to drive unit for speed adjustments, motor de-energization, and actuator deployment. Benefits include reduced user fatigue and increased mobility range. Used in indoor/outdoor environments, including care facilities and private residences.

## Clinical Evidence

Bench testing only. Performance demonstrated via testing per ISO 7176 series (energy consumption, speed/acceleration, static/fatigue, obstacle climbing, EMC, and battery safety per IEC 62133-2). Biocompatibility assessed per ISO 10993-1. Software validated per IEC 62304.

## Technological Characteristics

Materials: Aluminum and steel structure, plastic housing. Sensing: Inclinometer, accelerometer, motor current. Energy: 36 Vdc Li-Ion battery, 100-160 Whr capacity. Connectivity: Bluetooth LE 5.2. Dimensions: 150mm width, 310mm depth, 377-428mm height. Mass: 6.6 kg total. Sterilization: N/A. Software: Embedded control system with smartphone app interface.

## Regulatory Identification

A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

## Predicate Devices

- Smart Drive MX2 / Max Mobility (Permobil) ([K151199](/device/K151199.md))
- JWX-2 (Navione) / Yamaha ([K140204](/device/K140204.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

March 30, 2022

Sunrise Medical (US) LLC Devin Mcelroy Senior Director Quality Assurance and Regulatory Affairs 2842 N Business Park Ave. Fresno, California 93727

Re: K220213

Trade/Device Name: Empulse R90 Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: January 21, 2022 Received: January 26, 2022

Dear Devin Mcelroy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K220213

Device Name Empulse R90

Indications for Use (Describe)

The Empulse R90 is intended to be used to provide power assistance to manual wheelchair users. It is designed to augment the manual propulsion provided by the user, reducing the effort required by the user to propel the wheelchair.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

#### DATE PREPARED

January 10, 2021

### MANUFACTURER AND 510(K) OWNER

Sunrise Medical (US) LLC 2842 N Business Park Ave., Fresno, CA 93727 USA Telephone: 1-559-294-2374 Official Contact: Mr. Devin I. McElroy, Senior Director Quality Assurance and Regulatory Affairs

#### REPRESENTATIVE / CONTACT

Devin I. McElroy, Senior Director Quality Assurance and Regulatory Affairs Sunrise Medical (US) LLC 2842 N Business Park Ave., Fresno, CA 93727 USA Telephone: 1-559-294-2374 Email: devin.mcelroy@sunmed.com Website: www.sunmed.com

## DEVICE INFORMATION

| Proprietary Name/Trade Name: | Empulse R90                                           |
|------------------------------|-------------------------------------------------------|
| Common Name:                 | Power Assist Device                                   |
| Regulation Number:           | 21 CFR 890.3860                                       |
| Class:                       | II                                                    |
| Product Code:                | ITI: Powered Wheelchair                               |
| Premarket Review:            | Neurological and Physical Medicine Devices (OHT5)     |
|                              | Neuromodulation and Physical Medicine Devices (DHT5B) |
| Review Panel:                | Physical Medicine                                     |

#### PREDICATE DEVICE IDENTIFICATION

The Empulse R90 is substantially equivalent to the following predicates:

| 510(k) Number | Predicate Device Name / Manufacturer      | Primary Predicate |
|---------------|-------------------------------------------|-------------------|
| K151199       | Smart Drive MX2 / Max Mobility (Permobil) | $\checkmark$      |
| K140204       | JWX-2 (Navione) / Yamaha                  |                   |

#### DEVICE DESCRIPTION

The Empulse R90 is an auxiliary power module to provide augmented power to manual wheelchair users. The device utilizes a single electric drive wheel to provide the augmented power. The device mounts to a camber tube (rigid wheelchair) or a compressible axle (folding wheelchair) to provide the assisted power. On-board sensors, including an inclinometer, accelerometer and motor current sensing to determine when the user wishes to move forward and if the system is on an incline or decline.

{4}------------------------------------------------

The removable Lithium lon battery pack is easily switched out to an alternate battery pack if additional range is required. The battery can also be removed if the user wishes to reduce the overall weight.

The actuator is used to automatically engage and disengage the locking latch as well as compress the pre-load traction spring when the user wishes to disconnect the EMPULSE R90 from the wheelchair. In the extended position, the actuator allows the pre-load spring to provide sufficient pre-load to the drive wheel to ensure sufficient traction.

# INTENDED USE

The EMPULSE R90 is designed to be used on both indoor and outdoor environments, smooth flat surfaces, inclined surfaces up to 6 degrees and lightly rough surfaces such as asphalt. It is not designed to be used on highly rough surfaces such as grass or loose soil.

#### INDICATIONS FOR USE

The Empulse R90 is intended to be used to provide power assistance to manual wheelchair users. It is designed to augment the manual propulsion provided by the user, reducing the effort required by the user to propel the wheelchair.

# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS

Sunrise Medical (US) LLC believes that the Empulse R90 Power Assist Device is substantially equivalent to the predicate devices based on the information summarized here and in Table 5.1:

The subject device has a similar design and dimensions to the device cleared in K151199 and uses similar or identical materials as the devices cleared in K140204. The subject device has the same intended use and similar technological characteristics to the devices cleared in K151199 and K140204. The subject device has the same intended use environment as the devices cleared in K151199 and K140204. The subject device uses similar software to the devices cleared in K151199 and K140204. The Empulse R90 Power Assist Device has undergone testing to ensure that any differences in technological characteristics (e.g., battery, controls, and power assistance) do not negatively affect safety and effectiveness when compared to the predicate devices.

{5}------------------------------------------------

| Table 5.1: Table of Technological Comparison   |                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                   |                                                                                                                |
|------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|
|                                                | Primary Predicate<br>Permobil (Max Mobility)<br>Smart Drive MX2                                                                                                                                                                                                                          | Secondary Predicate<br>Yamaha JWX-2 (Navione)                                                                                                                                                                                                                                                         | Subject Device<br>Empulse R90                                                                                                                                                                                                                                                                     | Statement of Equivalence                                                                                       |
| 1<br>Intended Use                              | K151199<br>ITI 890.3860<br>30-Jul-15                                                                                                                                                                                                                                                     | K140204<br>ITI 890.3860<br>28-Aug-14                                                                                                                                                                                                                                                                  | K220213<br>ITI 890.3860                                                                                                                                                                                                                                                                           |                                                                                                                |
| 1.1 Intended Use                               | Not provided.                                                                                                                                                                                                                                                                            | The device JWX-2 (Navione) is a Power Assist Wheelchair Conversion Kit and suitable for the manual wheelchair users who are limited in their field of activities because of their physical conditions. The device can expand their field of activities by assisting their wheelchair operating force. | The Empulse R90 is intended to be used to provide power assistance to manual wheelchair users. It is designed to augment the manual propulsion provided by the user, reducing the effort required by the user to propel the wheelchair.                                                           | Substantially equivalent to Secondary Predicate. No new issues of safety or effectiveness.                     |
| 1.2 Indications for Use                        | The SmartDrive MX2 Wheelchair Power Assist is intended to provide auxiliary power to manual wheelchairs to reduce the pushing power needed by their users, including pediatrics. It is intended to be used by users capable of operating and maneuvering a powered and manual wheelchair | The Yamaha JWX-2 is a Power Assist Wheelchair Conversion Kit and suitable for the manual wheelchair users who are limited in their field of activities because of their physical conditions. The device can expand their field of activities by assisting their wheelchair operating force.           | The Empulse R90 is intended to be used to provide power assistance to manual wheelchair users. It is designed to augment the manual propulsion provided by the user, reducing the effort required by the user to propel the wheelchair.                                                           | Substantially Equivalent to Predicates. No new issues of safety or effectiveness.                              |
|                                                | Primary Predicate<br>Permobil (Max Mobility)<br>Smart Drive MX2                                                                                                                                                                                                                          | Secondary Predicate<br>Yamaha JWX-2 (Navione)                                                                                                                                                                                                                                                         | Subject Device<br>Empulse R90                                                                                                                                                                                                                                                                     | Statement of Equivalence                                                                                       |
| 1.3 Where Used                                 | Indoors and outdoors;<br>care facilities and private<br>residences.                                                                                                                                                                                                                      | Indoors and outdoors;<br>care facilities and private<br>residences.                                                                                                                                                                                                                                   | Indoors and outdoors;<br>care facilities and private<br>residences.                                                                                                                                                                                                                               | Identical to Predicates. No<br>new issues of safety or<br>effectiveness                                        |
| 2 Safety                                       |                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                   |                                                                                                                |
| 2.1 Mechanical Safety                          | Subject to ANSI/RESNA or<br>ISO 7176 Requirements<br>for Power Wheelchairs,<br>where applicable.                                                                                                                                                                                         | Subject to ANSI/RESNA or<br>ISO 7176 Requirements<br>for Power Wheelchairs,<br>where applicable.                                                                                                                                                                                                      | Subject to ANSI/RESNA or<br>ISO 7176 Requirements<br>for Power Wheelchairs,<br>where applicable.                                                                                                                                                                                                  | Identical to Predicates. No<br>new issues of safety or<br>effectiveness.                                       |
| 2.2 Electrical Safety                          | Subject to ANSI/RESNA or<br>ISO 7176 Requirements<br>for Power Wheelchairs,<br>where applicable.                                                                                                                                                                                         | Subject to ANSI/RESNA or<br>ISO 7176 Requirements<br>for Power Wheelchairs,<br>where applicable.                                                                                                                                                                                                      | Subject to ANSI/RESNA or<br>ISO 7176 Requirements<br>for Power Wheelchairs,<br>where applicable.                                                                                                                                                                                                  | Identical to Predicates. No<br>new issues of safety or<br>effectiveness.                                       |
| 2.3 Biocompatibility                           | No indications of<br>compliance                                                                                                                                                                                                                                                          | Compliant to ISO 10993-1,<br>-5                                                                                                                                                                                                                                                                       | Compliant to ISO 10993-1,<br>-5 (See Section 15 of this<br>submission for details).                                                                                                                                                                                                               | Identical to Secondary<br>Predicate. No new issues<br>of safety or effectiveness.                              |
| 3 Overall System                               |                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                   |                                                                                                                |
| 3.1 Weight Capacity                            | 14 kg to 150 kg                                                                                                                                                                                                                                                                          | 130 kg                                                                                                                                                                                                                                                                                                | 136 kg                                                                                                                                                                                                                                                                                            | Substantially equivalent to<br>Primary Predicate. No new<br>issues of safety or<br>effectiveness.              |
| 3.2 Expected Product Life                      | 5 years                                                                                                                                                                                                                                                                                  | Not provided                                                                                                                                                                                                                                                                                          | 5 years                                                                                                                                                                                                                                                                                           | Identical to Primary<br>Predicate. No new issues<br>of safety or effectiveness.                                |
| 4 Mechanical Enclosure and Mounting            |                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                   |                                                                                                                |
| 4.1 Structural Materials                       | Aluminum and Steel                                                                                                                                                                                                                                                                       | Aluminum and steel                                                                                                                                                                                                                                                                                    | Aluminum and steel                                                                                                                                                                                                                                                                                | Identical to Predicates. No<br>new issues of safety or<br>effectiveness.                                       |
| 4.2 Width                                      | 141 mm                                                                                                                                                                                                                                                                                   | 91mm (both left and right<br>wheels)                                                                                                                                                                                                                                                                  | 150mm                                                                                                                                                                                                                                                                                             | Substantially equivalent to<br>Primary Predicate. No new<br>issues of safety or<br>effectiveness. Slightly     |
|                                                | Primary Predicate<br>Permobil (Max Mobility)<br>Smart Drive MX2                                                                                                                                                                                                                          | Secondary Predicate<br>Yamaha JWX-2 (Navione)                                                                                                                                                                                                                                                         | Subject Device<br>Empulse R90                                                                                                                                                                                                                                                                     | Statement of Equivalence                                                                                       |
| 4.3 Depth                                      | 389mm                                                                                                                                                                                                                                                                                    | Same as height                                                                                                                                                                                                                                                                                        | 310mm                                                                                                                                                                                                                                                                                             | wider width does not<br>interfere with safe<br>operation of manual<br>wheelchairs.                             |
| 4.4 Height                                     | 242mm                                                                                                                                                                                                                                                                                    | 609 mm (24")<br>559 mm (22")                                                                                                                                                                                                                                                                          | 377mm – 428mm                                                                                                                                                                                                                                                                                     | Substantially equivalent to<br>Primary Predicate. No new<br>issues of safety or<br>effectiveness.              |
| 4.5 Overall Mass                               | 5.7 kg                                                                                                                                                                                                                                                                                   | 17 kg (Ni-MH)<br>17.7 kg (Li-Ion)                                                                                                                                                                                                                                                                     | 5.5 kg + 1.1 kg (battery)                                                                                                                                                                                                                                                                         | Substantially equivalent to<br>Primary Predicate. No new<br>issues of safety or<br>effectiveness.              |
| 4.6 Manual Wheelchair<br>Mount                 | Rigid: Clamp mount to<br>the camber tube. Drive<br>assembly pivotally mount<br>onto clamp.<br>Folding: Same as rigid but<br>with the addition of a<br>cross member that<br>extends horizontally<br>between the two axle<br>plates, thus simulating a<br>camber tube.                     | Quick Release axle mount.<br>One wheel per side.                                                                                                                                                                                                                                                      | Rigid: Receiver fixed<br>mounted to the camber<br>tube. Drive assembly free<br>to move vertically relative<br>to mount.<br>Folding: Same as rigid but<br>with the addition of a<br>cross member that<br>extends horizontally<br>between the two axle<br>plates, thus simulating a<br>camber tube. | Substantially equivalent to<br>Primary Predicate. No new<br>issues of safety or<br>effectiveness.              |
| 4.7 Housing Material                           | Plastic                                                                                                                                                                                                                                                                                  | Aluminum                                                                                                                                                                                                                                                                                              | Plastic                                                                                                                                                                                                                                                                                           | Identical to Primary<br>Predicate. No new issues<br>of safety or effectiveness.                                |
| 5 Performance                                  |                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                   |                                                                                                                |
|                                                | Primary Predicate<br>Permobil (Max Mobility)<br>Smart Drive MX2                                                                                                                                                                                                                          | Secondary Predicate<br>Yamaha JWX-2 (Navione)                                                                                                                                                                                                                                                         | Subject Device<br>Empulse R90                                                                                                                                                                                                                                                                     | Statement of Equivalence                                                                                       |
| 5.1 Max Speed (fwd)                            | 5.5 mph                                                                                                                                                                                                                                                                                  | Not Specified                                                                                                                                                                                                                                                                                         | 5.5 mph                                                                                                                                                                                                                                                                                           | Identical to Primary<br>Predicate. No new issues<br>of safety or effectiveness.                                |
| 5.2 Max Speed (rev)                            | N/A                                                                                                                                                                                                                                                                                      | Not Specified                                                                                                                                                                                                                                                                                         | N/A                                                                                                                                                                                                                                                                                               | Identical to Predicates. No<br>new issues of safety or<br>effectiveness.                                       |
| 5.3 Max Safe Slope                             | As specified by manual<br>wheelchair manufacturer                                                                                                                                                                                                                                        | 6°                                                                                                                                                                                                                                                                                                    | 6°                                                                                                                                                                                                                                                                                                | Identical to Secondary<br>Predicate. No new issues<br>of safety or effectiveness.                              |
| 5.4 Maximum Range on<br>Full Charge            | 19.8 km                                                                                                                                                                                                                                                                                  | 20 km (NiMH)<br>40 km (Li-ion)                                                                                                                                                                                                                                                                        | 19 ± 1.5 km                                                                                                                                                                                                                                                                                       | Substantially equivalent to<br>Primary Predicate. No new<br>issues of safety or<br>effectiveness.              |
| 5.5 Bluetooth                                  | Bluetooth LE                                                                                                                                                                                                                                                                             | Not Available                                                                                                                                                                                                                                                                                         | Bluetooth LE 5.2                                                                                                                                                                                                                                                                                  | Identical to Primary<br>Predicate. No new issues<br>of safety or effectiveness.                                |
| 6 Power System                                 |                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                   |                                                                                                                |
| 6.1 Motor Power                                | 250 W                                                                                                                                                                                                                                                                                    | 2x 110 W                                                                                                                                                                                                                                                                                              | 250 W                                                                                                                                                                                                                                                                                             | Identical to Primary<br>Predicate. No new issues<br>of safety or effectiveness.                                |
| 6.2 Battery Type                               | Li-lon                                                                                                                                                                                                                                                                                   | Ni-MH or Li-Ion                                                                                                                                                                                                                                                                                       | Li-Ion                                                                                                                                                                                                                                                                                            | Identical to Primary<br>Predicate. No new issues<br>of safety or effectiveness.                                |
| 6.3 Battery Capacity                           | 122 Whr                                                                                                                                                                                                                                                                                  | 160 Whr (Ni-MH)<br>283 Whr (Li-ion)                                                                                                                                                                                                                                                                   | 100 Whr – 160 Whr                                                                                                                                                                                                                                                                                 | Substantially equivalent to<br>Predicates. No new issues<br>of safety or effectiveness.                        |
| 6.4 Battery Charger Input                      | 100-240 Vac, 50/60Hz                                                                                                                                                                                                                                                                     | 100-240 Vac, 50/60Hz                                                                                                                                                                                                                                                                                  | 100-240 Vac, 50/60Hz                                                                                                                                                                                                                                                                              | Identical to Predicates. No<br>new issues of safety or<br>effectiveness.                                       |
| 6.5 Battery Charger Output                     | 2.0 A                                                                                                                                                                                                                                                                                    | 2.6 A                                                                                                                                                                                                                                                                                                 | 2.0 A                                                                                                                                                                                                                                                                                             | Identical to Primary<br>Predicate. No new issues                                                               |
|                                                | Primary Predicate<br>Permobil (Max Mobility)<br>Smart Drive MX2                                                                                                                                                                                                                          | Secondary Predicate<br>Yamaha JWX-2 (Navione)                                                                                                                                                                                                                                                         | Subject Device<br>Empulse R90                                                                                                                                                                                                                                                                     | Statement of Equivalence                                                                                       |
| 6.6 Operating Voltage                          | 36 Vdc                                                                                                                                                                                                                                                                                   | 24 Vdc                                                                                                                                                                                                                                                                                                | 36 Vdc                                                                                                                                                                                                                                                                                            | of safety or effectiveness.<br>Identical to Primary<br>Predicate. No new issues<br>of safety or effectiveness. |
| 7 Control System                               |                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                   |                                                                                                                |
| 7.1 Control Type                               | Constant power on. Stop<br>and start of power assist<br>is controlled by wrist<br>strap sensor with<br>Bluetooth communication<br>to the drive unit.                                                                                                                                     | Torque sensing handrim<br>applies proportional<br>power based on torque<br>applied to the handrim.<br>Power is pulsed on for<br>specified duration. Gain of<br>applied power is<br>adjustable.                                                                                                        | The device will sense<br>acceleration (torque) and<br>apply a power pulse to the<br>drive motor for a specified<br>duration. Gain of applied<br>power is adjustable.                                                                                                                              | Substantially equivalent to<br>Predicates. No new issues<br>of safety or effectiveness.                        |
| 7.2 Operating Voltage                          | 36 Vdc                                                                                                                                                                                                                                                                                   | 24 Vdc                                                                                                                                                                                                                                                                                                | 36 Vdc                                                                                                                                                                                                                                                                                            | Identical to Primary<br>Predicate. No new issues<br>of safety or effectiveness.                                |
| 7.3 Moisture Resistance                        | ISO 7176-9                                                                                                                                                                                                                                                                               | ISO 7176-9                                                                                                                                                                                                                                                                                            | ISO 7176-9                                                                                                                                                                                                                                                                                        | Identical to Predicates. No<br>new issues of safety or<br>effectiveness                                        |
| 7.4 Operating<br>Temperature                   | -25°C to +50°C                                                                                                                                                                                                                                                                           | -25°C to +50°C                                                                                                                                                                                                                                                                                        | -25°C to +50°C                                                                                                                                                                                                                                                                                    | Identical to Predicates. No<br>new issues of safety or<br>effectiveness                                        |
| 7.5 Storage Temperature                        | -40°C to +65°C                                                                                                                                                                                                                                                                           | -40°C to +65°C                                                                                                                                                                                                                                                                                        | -40°C to +65°C                                                                                                                                                                                                                                                                                    | Identical to Predicates. No<br>new issues of safety or<br>effectiveness                                        |
| 8 EMC                                          |                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                   |                                                                                                                |
| 8.1 EMC - Immunity                             | Compliant to ANSI/RESNA<br>WC2-21 or ISO 7176-21                                                                                                                                                                                                                                         | Compliant to ANSI/RESNA<br>WC2-21 or ISO 7176-21                                                                                                                                                                                                                                                      | Compliant to ANSI/RESNA<br>WC2-21 or ISO 7176-21                                                                                                                                                                                                                                                  | Identical to Predicates. No<br>new issues of safety or<br>effectiveness.                                       |
| 8.2 EMC - Radiated<br>Emissions                | Compliant to ANSI/RESNA<br>WC2-21 or ISO 7176-21                                                                                                                                                                                                                                         | Compliant to ANSI/RESNA<br>WC2-21 or ISO 7176-21                                                                                                                                                                                                                                                      | Compliant to ANSI/RESNA<br>WC2-21 or ISO 7176-21                                                                                                                                                                                                                                                  | Identical to Predicates. No<br>new issues of safety or<br>effectiveness                                        |
|                                                | Primary Predicate<br>Permobil (Max Mobility)<br>Smart Drive MX2                                                                                                                                                                                                                          | Secondary Predicate<br>Yamaha JWX-2 (Navione)                                                                                                                                                                                                                                                         | Subject Device<br>Empulse R90                                                                                                                                                                                                                                                                     | Statement of Equivalence                                                                                       |
| 8.3 EMC - ESD                                  | Compliant to ANSI/RESNA<br>WC2-21 or ISO 7176-21                                                                                                                                                                                                                                         | Compliant to ANSI/RESNA<br>WC2-21 or ISO 7176-21                                                                                                                                                                                                                                                      | Compliant to ANSI/RESNA<br>WC2-21 or ISO 7176-21                                                                                                                                                                                                                                                  | Identical to Predicates. No<br>new issues of safety or<br>effectiveness.                                       |
| 8.4 Power Frequency<br>Magnetic Field Immunity | Compliant to ANSI/RESNA<br>WC2-21 or ISO 7176-21                                                                                                                                                                                                                                         | Compliant to ANSI/RESNA<br>WC2-21 or ISO 7176-21                                                                                                                                                                                                                                                      | Compliant to ANSI/RESNA<br>WC2-21 or ISO 7176-21                                                                                                                                                                                                                                                  | Identical to Predicates. No<br>new issues of safety or<br>effectiveness.                                       |
| 9 User Input Controls                          |                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                   |                                                                                                                |
| 9.1 On/Off Button                              | Located on the drive unit                                                                                                                                                                                                                                                                | Located on the drive unit                                                                                                                                                                                                                                                                             | Located on the drive unit                                                                                                                                                                                                                                                                         | Identical to Predicates. No<br>new issues of safety or<br>effectiveness.                                       |
| 9.3 Speed Up/Speed Down                        | Available by tapping on<br>the wrist control.<br>Maximum speed<br>adjustable through App.                                                                                                                                                                                                | Maximum speed<br>adjustable on the drive<br>unit.                                                                                                                                                                                                                                                     | Speed adjustable by<br>applying an<br>acceleration/retardation<br>torque to the manual<br>wheelchair handrims.                                                                                                                                                                                    | Substantially equivalent to<br>Predicates. No new issues<br>of safety or effectiveness.                        |
| 9.4 Battery state of charge<br>indication      | LED Indicator                                                                                                                                                                                                                                                                            | LED Indicator                                                                                                                                                                                                                                                                                         | LED Indicator                                                                                                                                                                                                                                                                                     | Identical to Predicates. No<br>new issues of safety or<br>effectiveness.                                       |
| 9.5 Programming Tool                           | Smart phone app                                                                                                                                                                                                                                                                          | Not programmable                                                                                                                                                                                                                                                                                      | Smart phone app                                                                                                                                                                                                                                                                                   | Identical to Primary<br>Predicate. No new issues<br>of safety or effectiveness.                                |
| 9.6 Bluetooth functionality                    | Parameter setting<br>Over the air firmware<br>updates<br>Distance, pushes and time<br>tracker<br>Route Tracking                                                                                                                                                                          | Not available                                                                                                                                                                                                                                                                                         | The Control Box connects<br>with the primary drive<br>unit through a wireless<br>Bluetooth connection. The<br>Control Box provides<br>secondary speed increase,<br>speed decrease and motor<br>de-energization through<br>the control box. The                                                    | Substantially equivalent<br>to the Primary<br>Predicate. No new<br>issues of safety or<br>effectiveness.       |
| Primary Predicate                              | Secondary Predicate                                                                                                                                                                                                                                                                      | Subject Device                                                                                                                                                                                                                                                                                        | Statement of Equivalence                                                                                                                                                                                                                                                                          |                                                                                                                |
| Permobil (Max Mobility)<br>Smart Drive MX2     | Yamaha JWX-2 (Navione)                                                                                                                                                                                                                                                                   | Empulse R90                                                                                                                                                                                                                                                                                           | Control Box also provides<br>a means to toggle<br>between lifting and<br>deploying the drive wheel<br>actuator.                                                                                                                                                                                   |                                                                                                                |

Table 5.1: Table of Technological Comparison

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

# SUMMARY OF NON-CLINICAL TESTING

The following tests were performed to demonstrate substantial equivalence based on current industry and FDA recognized standards.

- Energy consumption (per ISO 7176-4)
- Maximum speed, acceleration, and deceleration (per ISO 7176-6)
- . Static, impact, and fatigue (per ISO 7176-8)
- Climatic test (per ISO 7176-9)
- Obstacle climbing (per ISO 7176-10)
- Power and control systems for power wheelchairs (per ISO 7176-14)
- Documentation and labeling (per ISO 7176-15)
- EMC testing (per ISO 7176-21)
- . Vocabulary (per ISO 7176-26)
- Alkaline or other non-acid electrolyte batteries safety testing (IEC 62133-2)
- Evaluation and testing within a risk management process (per ISO 10993-1)
- Software life cycle process (per IEC 62304) ●

## CONCLUSION

Based on the testing performed, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate devices. The similar indications for use, technological characteristics, and performance characteristics for the proposed Empulse R90 Power Assist Device is assessed to be substantially equivalent to the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K220213](https://fda-staging.innolitics.com/device/K220213)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).