BD Vacutainer® Push Button Blood Collection Set, BD Vacutainer® Push Button Blood Collection Set with Pre-Attached Holder
K220212 · Becton, Dickinson and Company · JKA · Mar 24, 2023 · Clinical Chemistry
Device Facts
Record ID
K220212
Device Name
BD Vacutainer® Push Button Blood Collection Set, BD Vacutainer® Push Button Blood Collection Set with Pre-Attached Holder
Applicant
Becton, Dickinson and Company
Product Code
JKA · Clinical Chemistry
Decision Date
Mar 24, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1675
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K220212 · Mar 24, 2023
BD Vacutainer® Push Button Blood Collection Set, BD Vacutainer® Push Button Blood Collection Set with Pre-Attached Holder
Becton, Dickinson and Company
Published clinical literature (Ibarra & Villanueva, 2019); Published clinical literature (Merrill, 2021)
The literature is used to support the clinical performance and safety of the device in patients with difficult vein access (DVA), demonstrating that the device maintains specimen quality and is preferred by clinicians in routine clinical settings.
Merrill, 2021 (Assessing Phlebotomy device preference and specimen quality in an oncology outpatient clinic); Observational clinical study
89 oncology outpatients ≥18 years of age; Sample Size: 89
Not applicable for this study
Specimen quality (LDH, K, and plasma free hemoglobin levels)
Ibarra AF and Villanueva, 2019 (Evaluation of phlebotomy-related anxiety, pain and safety in a Mexican general hospital); Clinical evaluation in a general hospital setting
General hospital patients; Number of Sites: 1 (Mexican general hospital)
Conventional (straight) blood collection needle
Patient and phlebotomist satisfaction
Indications for Use
The BD Vacutainer® Push Button Blood Collection Set is a sterile, multi-sample, single-use fixed winged blood collection set intended for use in the general population by healthcare professionals experienced with venipuncture to obtain blood specimens from patients, including those patients with difficult vein access who may have small, fragile, and or non-palpable veins, into evacuated blood collection tubes and/or bottles. When used without the male Luer adapter, the device allows the clinician to obtain a blood specimen from the female Luer connector with a syringe, if necessary. The device can be used by healthcare professionals with infusion experience for short-term, single infusions with consideration given to patient size and appropriateness for the solution being infused. The device is not to be left in place and is to remain under the direct supervision of a clinician. The recommended use of the device is to activate the needle safety feature prior to removal from the venipuncture site. The retraction of the intravenous (IV) end of the needle aids in the prevention of accidental needlestick injury.
Device Story
Sterile, single-use, winged blood collection set; used for venipuncture to obtain blood specimens or short-term fluid infusion. Input: venous blood via stainless steel cannula. Operation: user depresses button on hub to trigger internal spring mechanism; needle retracts and locks inside device barrel; prevents accidental needlestick. Output: blood collected into evacuated tubes/bottles or syringe. Used in clinical settings by healthcare professionals. Benefits: improved access for patients with difficult veins (DVA); enhanced safety via automated needle retraction.
Clinical Evidence
Clinical literature review provided. Study 1 (Ibarra & Villanueva, 2019): Phlebotomist preference and satisfaction compared to conventional needles. Study 2 (Merrill, 2021): 89 oncology outpatients (≥18 years) with DVA; compared specimen quality (LDH, K, plasma free hemoglobin) using winged sets. Results: successful blood collection in all patients; analyte levels within normal reference ranges, indicating no compromise to sample quality.
Technological Characteristics
Winged blood collection set; stainless steel 304 cannula; 302 stainless steel spring; polycarbonate hub; polyolefin wings; PVC tubing. Sterilized via gamma irradiation. Compliant with ISO 10993 (biocompatibility), ISO 594-1/2 (luer fittings), and EN ISO 23908 (sharps protection).
Indications for Use
Indicated for general population, including patients with difficult vein access (DVA) (small, fragile, or non-palpable veins), for venipuncture to obtain blood specimens into evacuated tubes/bottles or via syringe. Also indicated for short-term, single infusions (up to 2 hours) by healthcare professionals. Not for indwelling use.
Regulatory Classification
Identification
A blood specimen collection device is a device intended for medical purposes to collect and to handle blood specimens and to separate serum from nonserum (cellular) components prior to further testing. This generic type device may include blood collection tubes, vials, systems, serum separators, blood collection trays, or vacuum sample tubes.
Predicate Devices
BD Vacutainer® Push Button Blood Collection Set (K030573)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 24, 2023
Becton, Dickinson and Company Kelly Hilliger Staff Regulatory Affairs Specialist 1 Becton Drive Franklin Lakes, New Jersey 07417
### Re: K220212
Trade/Device Name: BD Vacutainer® Push Button Blood Collection Set, BD Vacutainer® Push Button Blood Collection Set with Pre-Attached Holder Regulation Number: 21 CFR 862.1675 Regulation Name: Blood Specimen Collection Device Regulatory Class: Class II Product Code: JKA, FPA Dated: February 24, 2023 Received: February 24, 2023
Dear Kelly Hilliger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
David Walloschek
David Wolloscheck, Ph.D. Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K220212
#### Device Name
BD Vacutainer® Push Button Blood Collection Set, BD Vacutainer® Push Button Blood Collection Set with Pre-Attached Holder
#### Indications for Use (Describe)
The BD Vacutainer® Push Button Blood Collection Set is a sterile, multi-sample, single-use fixed winged blood collection set intended for use in the general population by healthcare professionals experienced with venipuncture to obtain blood specimens from patients, including those patients with difficult vein access who may have small, fragile, and or non-palpable veins, into evacuated blood collection tubes and/or bottles. When used without the male Luer adapter, the device allows the clinician to obtain a blood specimen from the female Luer connector with a syringe, if necessary. The device can be used by healthcare professionals with infusion experience for short-term, single infusions with consideration given to patient size and appropriateness for the solution being infused. The device is not to be left in place and is to remain under the direct supervision of a clinician.
The recommended use of the device is to activate the needle safety feature prior to removal from the venipuncture site. The retraction of the intravenous (IV) end of the needle aids in the prevention of accidental needlestick injury.
| Type of Use (Select one or both, as applicable) |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="padding-right: 20px;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K220212 510(K) SUMMARY
### BD Vacutainer® Push Button Blood Collection Set
#### Summary Preparation Date:
March 24, 2023
#### Submitted by:
Becton, Dickinson and Company 1 Becton Drive Franklin Lakes, NJ 07417-1885
Phone: (201) 847-6800
### Contact:
Kelly Hilliger Staff Regulatory Affairs Specialist email: Kelly.hilliger@bd.com
Phone: (551) 554-1675
### Proprietary Names:
BD Vacutainer® Push Button Blood Collection Set, BD Vacutainer® Push Button Blood Collection Set with Pre-Attached Holder
#### Common or Usual Names:
Blood Collection Tubes, Vials, Systems, Serum Separators
### Regulatory Information
Classification Name: Blood Collection Set Classification Regulation: 21 CFR §862.6175 Regulatory Class: Class II Primary Product Code: JKA Secondary Product Code: FPA Classification Panel: Clinical Chemistry
#### Predicate Device:
BD Vacutainer® Push Button Blood Collection Set (K030573)
{4}------------------------------------------------
#### Device Description
The BD Vacutainer® Push Button Blood Collection Set is a sterile, multiple-sample, single-use device used for venipuncture to obtain blood specimens from patients and may be used for the short-term intravenous administration of fluids. The BD Vacutainer® Push Button Blood Collection Set is a winged blood collection set with flexible tubing, a female luer connector and an optional male luer adapter. The BD Vacutainer® Push Button Blood Collection Set also contains a needle protector. The wingset is designed with a safety mechanism to protect against needlesticks. When the button is depressed by the user, the needle fully retracts and is enclosed and locked within the barrel of the device. The retraction of the intravenous (IV) end of the needle aids in the prevention of accidental needlestick injury.
The BD Vacutainer® Push Button Blood Collection Set consists of:
- Stainless steel cannula (Intravenous end and non-patient end of cannula) .
- Stainless steel spring ●
- Hub, front and rear barrel .
- Wings (color coded according to needle gauge)
- Tubing
- Female luer connector and an optional male luer adapter
- Intravenous (IV) needle protector (covers the needle before use)
- . Pre-attached holder (connected by male luer adapter in some models)
The intravenous needle of the blood collection set is bonded to one end of the hub. The other end of the hub is bonded to the blood collection set tubing, the end of which is then bonded to a female luer connector with an optional attached male luer adapter. Some models come with a pre-attached holder connected to the male luer adapter for user convenience. A spring is loaded onto the front barrel of the hub. Once the sample is collected, and without removing the needle from the patient vein, the user depresses the raised push button on top of the hub which activates the spring mechanism to automatically retract the needle from the device. In this retracted (locked) position, the IV point of the needle is fully contained within the body of the device. This will prevent the needle from coming out of the front barrel once it has retracted as well as preventing accidental overriding of the safety feature.
The blood collection set with a male luer adapter contains threads for attachment to a Vacutainer® Brand Needle Holder, and a non-patient cannula for blood collection into evacuated blood collection tubes and/or blood culture bottles. For models without the pre-attached holder, the user is instructed to attach a holder before using the device. The pre-attached holder models come with the BD Vacutainer® Brand Holder pre-attached for user convenience. The non-patient end cannula of the luer adapter has a sleeve that recovers over the cannula to stop blood flow in between collection of multiple tubes.
{5}------------------------------------------------
When used without the male adapter, the device allows the clinician to obtain a blood specimen from the female hub with a syringe, if necessary. The device can be used for short-term, single infusions (up to 2 hours) with consideration given to patient size and appropriateness for the solution being infused.
#### Indications for Use
| Subject Device | Predicate Device |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| BD Vacutainer® Push Button Blood Collection<br>Set<br>K220212 | BD Vacutainer® Push Button Blood Collection<br>Set<br>K030573 |
| The BD Vacutainer® Push Button Blood Collection<br>Set is a sterile, multi-sample, single-use fixed<br>winged blood collection set intended for use in the<br>general population by healthcare professionals<br>experienced with venipuncture to obtain blood<br>specimens from patients, including those patients<br>with difficult vein access who may have small,<br>fragile, and/or non-palpable veins, into<br>evacuated blood collection tubes and/or blood<br>culture bottles. When used without the male<br>Luer adapter, the device allows the clinician to<br>obtain a blood specimen from the female Luer<br>connector with a syringe, if necessary. The<br>device can be used by healthcare professionals<br>with infusion experience for short-term, single<br>infusions with consideration given to patient size<br>and appropriateness for the solution being infused.<br>The device is not to be left in place and is to<br>remain under the direct supervision of a<br>clinician. | The BD Vacutainer® Push Button Blood Collection<br>Set is a sterile, multiple-sample, single-use winged<br>blood collection set intended for venipuncture to<br>obtain blood specimens from patients.<br><br>The BD Vacutainer® Push Button Blood<br>Collection Set is also indicated for the<br>intravenous administration of fluids as indicated<br>in 21 CFR 880.5440. It may be used for any<br>patient population with consideration given to<br>patient size and appropriateness for the solution<br>being infused and duration of therapy. |
| The recommended use of the device is to activate<br>the needle safety feature prior to removal from the<br>venipuncture site. The retraction of the intravenous<br>(IV) end of the needle aids in the prevention of<br>accidental needlestick injury. | The recommended use of the device is to activate<br>the needle safety feature prior to removal from the<br>venipuncture site. The retraction of the intravenous<br>(IV) end of the needle aids in the prevention of<br>accidental needlestick injury. |
The proposed indications for use are similar to the predicate indications for use, except for the addition of healthy patients and patients with vascular anatomy and/or co-morbidities which may cause difficulty in locating a vein. Healthy patients with these conditions often exhibit small, fragile, and/or non-palpable veins, commonly referred to as DVA. The addition of DVA in the revised indications for use statement does not significantly change the intended use as butterfly needles have historically been used on patients with small veins and similar statements were included for previous device iterations and were inadvertently removed in subsequent submissions. This is supported by the Clinical Laboratory Standards Institute (CLSI) guideline GP41', which
Clinical Laboratory Standards Institute (CLSI) Guidelines. Collection of Diagnostic Venous Blood Specimens. 7th ed. CLSI stardard GP4. Wayne, PA: Clinical and Laboratory Standards Institute; 2017.
{6}------------------------------------------------
states that winged blood collection sets are recommended for accessing smaller, fragile, or difficult-to-locate veins. Additionally, per the World Health Organization (WHO) Guidelines', winged blood collection sets can provide easier access and movement and better precision when drawing blood from patients with small or difficult veins. BD believes that the addition of specific reference to patients with DVA does not raise new questions of safety/effectiveness.
Furthermore, the use of a syringe was part of the instructions for use cleared under K030573, but was unintentionally omitted from the indications for use. The inclusion of the use of a syringe has been added back to the indications for use. Based on this, BD has concluded the use of a syringe with this device is not a new intended use. BD believes this does not raise new questions of safety/effectiveness.
#### Substantial Equivalence3
The subject and predicate device are substantially equivalent as described in Table 1.
| Characteristic | Subject Device<br>BD Vacutainer® Push Button<br>Blood Collection Set | Predicate Device<br>BD Vacutainer® Push Button<br>Blood Collection Set<br>K030573 | Comparison |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Population | General use including patients<br>with difficult vein access<br>(DVA), who may have small,<br>fragile, and/or non-palpable<br>veins | General use | Different<br>Comment # 1<br>Added specific<br>reference to patients<br>with DVA which is a<br>subset of 'General<br>Use' – no effect on<br>device safety or<br>effectiveness |
| Needle Diameter<br>OD | 21G, 23G and 25G | 21G, 23G and 25G | Same |
| Needle Diameter<br>ID | Thin wall | Thin wall | Same |
| Needle Point<br>Geometry | 3 Bevel | 3 Bevel | Same |
| Needle Length | $\frac{3}{4}$ inch | $\frac{3}{4}$ inch | Same |
| Wing | Polyolefin | Polyolefin | Same material |
| Hub | Polycarbonate | Polycarbonate | Same material |
| Button Ink | UV Curable Ink | UV Curable Ink | Same material |
| Table 1: | Substantial Equivalence Comparison |
|----------|------------------------------------|
|----------|------------------------------------|
WHO guidelines on drawing blood: best practices in phlebotomy. World Health Organization; Geneva, Switzerland, 2010.
<sup>3</sup> The term "substantial equivalence" as used in this 510(k) notification of substantial equivalence as found in the Federal Food, Drug and Cosmetic Act, as amplied under 21 CFR 807, Subpart E under which a device can be marketed without pre-market approval or reclassification of substantial equivalency under this notification is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statements related to, or in support of substantial equivalence herein shall be construed as an admission against interest under the US Patent Laws or the courts.
{7}------------------------------------------------
| Characteristic | Subject Device | Predicate Device | Comparison |
|--------------------------------------------|----------------------------------------------------|---------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | BD Vacutainer® Push Button<br>Blood Collection Set | BD Vacutainer® Push Button<br>Blood Collection Set<br>K030573 | |
| Front Barrel | Polypropylene | Polypropylene | Same material |
| Rear Barrel | Acrylic | Acrylic | Same material |
| Rear Barrel Lubricant | Silicone | Silicone | Same material |
| Spring | Stainless Steel 302 | Stainless Steel 302 | Same material |
| IV Protector<br>(Cannula<br>Protector) | Polyethylene | Polyethylene | Same material |
| IV Cannula/NP<br>Cannula | Stainless Steel 304 | Stainless Steel 304 | Same material |
| Tubing | Polyvinyl Chloride | Polyvinyl Chloride | Same material |
| Cannula<br>Lubricant | Silicone | Silicone | Same material |
| Cannula Adhesive | UV cured adhesive | UV cured adhesive | Same material |
| Viscous Fluid | Not included in manufacture of<br>device | Included in manufacture of<br>device | Material is not part<br>of subject device.<br>The difference does<br>not raise new<br>questions on the<br>safety and<br>effectiveness of the<br>subject device. |
| Hub-Tubing<br>Adhesive | UV cured adhesive | UV cured adhesive | Same material |
| Female Luer<br>Adapter<br>(Connectors) | Acrylonitrile Butadiene Styrene | Acrylonitrile Butadiene Styrene | Same material |
| Luer Adapter Hub | Polypropylene | Polypropylene | Same material |
| NP Sleeve | Synthetic Isoprene Rubber | Synthetic Isoprene Rubber | Same material |
| Luer Adhesive | Heat Curing Epoxy | Heat Curing Epoxy | Same material |
| Luer Cannula<br>Lubricant | Medical Grade Silicone | Medical Grade Silicone | Same material |
| Luer Cap | Polypropylene | Polypropylene | Same material |
| Top Web | Paper | Paper | Same material |
| Blister | Polyethylene terephthalate -<br>glycol modified | Polyethylene terephthalate -<br>glycol modified | Same material |
| Characteristic | Subject Device | Predicate Device | Comparison |
| | BD Vacutainer® Push Button<br>Blood Collection Set | BD Vacutainer® Push Button<br>Blood Collection Set<br>K030573 | |
| Pre-attached<br>holder | Polypropylene | N/A | Different<br>Comment #2<br>Model numbers were<br>added to include<br>devices assembled<br>with the legally<br>marketed holder pre-<br>attached for user<br>convenience. This<br>non-significant<br>change does not<br>affect safety or<br>effectiveness. |
| Materials | Compliant with ISO 10993<br>series | Compliant with ISO 10993<br>series | Same |
| Non-pyrogenic | Yes | Yes | Same |
| Non-toxic | Yes | Yes | Same |
| Sterile | Yes | Yes | Same |
| Sterility<br>Assurance Level<br>(SAL) 10-6 | Yes | Yes | Same |
| Sterilization<br>Method | Gamma | Gamma | Same |
| Shelf Life | 2 years | 2 years | Same |
| Models | 367323 | 367323 | Model numbers were<br>added to include<br>devices assembled<br>with the legally<br>marketed holder pre-<br>attached for user<br>convenience. This<br>change does not<br>raised new questions<br>of safety or<br>effectiveness. |
| | 367324 | 367324 | |
| | 367326 | 367326 | |
| | 367335 | 367335 | |
| | 367336 | 367336 | |
| | 367338 | 367338 | |
| | 367341 | 367341 | |
| | 367342 | 367342 | |
| | 367344 | 367344 | |
| | 367352 | | |
| | 367354 | | |
| | 367355 | | |
| | 368656 | | |
| | 368657 | | |
| | 368658 | | |
| | 368659 | | |
{8}------------------------------------------------
### Substantial Equivalence Discussion:
## Comment #1
The subject device indicates that the intended population is for general use, including a subset of patients who may have difficult veins, whereas the predicate device population is general use. Both devices are used on all patient populations regardless of their vein status. Adding a reference to patients with difficult veins has no effect on clinical safety or effectiveness. This is supported by
{9}------------------------------------------------
the Clinical Laboratory Standards Institute (CLSI) guideline GP41', which states that winged blood collection sets are recommended for accessing smaller, fragile, or difficult-to-locate veins. Additionally, per the World Health Organization (WHO) Guidelines2, winged blood collection sets can provide easier access and movement and better precision when drawing blood from patients with small or difficult veins.
## Comment #2
The subject device is available in models with and without a pre-attached holder. The predicate device is only available in models without the pre-attached holder and requires the user to add a holder in order to use the device for blood collection. Adding models with the pre-attached holder to the subject device is a user convenience and does not affect clinical safety or effectiveness.
## Performance Testing - Non-Clinical Bench Summary
Non-Clinical Bench, Biocompatibility, and Sterilization testing were conducted on the subject device to validate that the device performs as intended over the course of the product shelf life and was substantially equivalent (SE) to the predicate device. Results of testing demonstrate acceptable performance for the subject device and complies to applicable parts of the following standards:
#### Performance Testing I.
- . EN ISO 23908:2013 Sharps injury protection - Requirements and test methods - Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling
- EN ISO 9626:2016 Stainless steel needle tubing for the manufacture of medical devices -● Requirements and test methods
- ISO 594-1-1986 Conical fittings with 6 % (Luer) taper for syringes, needles and certain ● other medical equipment - Part 1: General requirements
- ISO 594-2-1998 Conical fittings with 6 % (Luer) taper for syringes, needles and certain ● other medical equipment -Part 2: Lock fittings
- . EN ISO 14971:2019 Medical Devices - Application of risk management to medical devices
#### II. Sterilization, Package Integrity, Shipping and Shelf Life
The devices are sterilized via gamma irradiation. The sterilization and package integrity, shipping and shelf life validations to support the shelf life claim was conducted according to the following standards:
- EN ISO 11137-1:2015/A2:2019 Sterilization of health care products Radiation Part 1: . Requirements for development, validation, and routine control of a sterilization process for medical devices
{10}------------------------------------------------
- EN ISO 11137-2:2015 Sterilization of health care products Radiation Part 2: ● Establishing the sterilization dose
- EN ISO 11737-1:2018 Sterilization of health care products Microbiological methods -● Part 1: Determination of a population of microorganisms on products
- EN ISO 11737-2:2019 Sterilization of medical devices Microbiological methods Part . 2: Tests of sterility performed in the definition, validation and maintenance of a sterilization process
- ISO 11607-1:2019 Packaging for terminally sterilized medical devices Part 1: . Requirements for materials, sterile barrier systems and packaging systems
- . ISO 11607-2:2019 Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes
- ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
- . ASTM D4169-16 Standard Practice for Performance Testing of Shipping Containers and Svstems
- ASTM F2096-11 Standard Test Method for Detecting Gross Leaks in Packaging by ● Internal Pressurization (Bubble Test)
- ASTM F88 Standard Test Method for Seal Strength of Flexible Barrier Materials .
#### III. Biocompatibility
The BD Vacutainer® Push Button Blood Collection Set is classified as an externally communicating device that comes into direct contact with circulating blood for a limited (≤ 24 hours) duration. Biocompatibility was evaluated on the final, finished device with the following endpoints:
- Cytotoxicity (ISO 10993-5:2009) ●
- Sensitization (ISO 10993-10:2010)
- Irritation (ISO 10993-10:2010)
- Acute Systemic Toxicity (ISO 10993-11:2017) ●
- Pyrogenicity (ISO 10993-11:2017, USP<151>) ●
- Hemocompatibility: Coagulation, Platelet Activation, Complement Activation and Hemolysis (ISO 10993-4:2017, ISO 10993-7:2017)
- Particulates (USP<788>) ●
{11}------------------------------------------------
#### Non-clinical Performance Summary
The non-clinical performance tests below were conducted to verify that the proposed devices met all design specifications and performance standards and are each Substantially Equivalent (SE) to the predicate device.
- IV Protector Pull Test ●
- IV Cannula Removal Test ●
- Flow Rate ●
- Torque to Rotate Wing ●
- Wing Removal Test
- Front Barrel/Rear Barrel Separation Test
- Tubing to Female Luer Adapter Pull Test ●
- . Tubing to Hub Pull Test
- Product (safety feature) Pre-Activation
- Retraction and Lock-out
- Force to Override Lock Mechanism
- Luer Cap Removal Torque
- Luer Separation Test
- NP Cannula Pull Test
- Male Luer Adapter Spin Out from Holder Test
- Torque to Unseat
- NP Sleeve Function ●
- Tube Push Off
- IV Cannula Penetration Test
- Submerged Leak Test (Device) ●
- Air Leakage Test
- Package Integrity Test
- Luer Compatibility as per ISO 594-1 and ISO 594-2 ●
- Sharps Injury Protection Testing as per ISO 23908 ●
- Stainless Steel Needle Tubing testing as per ISO 9626 ●
### Performance Testing - Animal Summary
Not applicable
### Performance Testing - Clinical Summary
The real-world clinical data of the study submitted as part of this 510(k) indicate that the BD Vacutainer® Push Button Blood Collection Set was preferred by phlebotomists when compared to conventional blood collection needles as demonstrated by increased patient and phlebotomist satisfaction with the subject device as compared to a conventional (straight) blood collection needle (Ibarra AF and Villanueva, 2019)4. Additionally, winged blood collection sets, such as the subject device, were shown to improve venipuncture collection for patients with DVA. The BD
<sup>4</sup> Ibarra AF and Villanueva, S. Evaluation of phlebotomy-related anxiety, pain and safety in a Mexican general hospital using winged blood collection sets. BJSTR 2019; 13(5):10219-10221. DOI: 10.26717/BJSTR.2019.13.002455. ISSN: 2574-1241.
{12}------------------------------------------------
Vacutainer® Push Button Blood Collection Set allows healthcare professionals to draw blood from patients with small, fragile, and/or non-palpable veins (DVA) without compromising sample quality as supported by clinical literature (Merrill, 2021)5. In this study, venipuncture blood draws were collected from a total of 89 oncology outpatients ≥18 years of age using the BD Vacutainer® Push Button Blood Collection Set or devices similar to the subject device. Chemotherapy impacts the vasculature causing veins to become much smaller, fragile, and difficult to anchor for venous access (Lynn, 2011). Specimen quality (LDH, K and plasma free hemoglobin levels) were measured from specimens collected with each winged blood collection set. The results of the study indicate that the devices provided the ability to draw blood from oncology patients, who may have small, fragile and non-palpable veins (DVA), as each device achieved successful blood collections from each patient, and the sample quality was not compromised as LDH, K and plasma free hemoglobin levels were within the normal and acceptable analyte reference ranges (Merrill, 2021) 5.
#### Conclusion
In summary, the BD Vacutainer® Push Button Blood Collection Set has the same intended use as its predicate device. In addition, the BD Vacutainer® Push Button Blood Collection Set has the same technological characteristics and principles of operation as its predicate. The clinical literature sufficiently supports the determination that patients with DVA (small, fragile and nonpalpable veins) is a subset of the current patient population for the BD Vacutainer® Push Button Blood Collection Set. The risk profile of patients with small, fragile, and/or non-palpable veins is not substantially different from that of the general patient population and is supported by the clinical literature evidence presented under this submission. Furthermore, the changes made to the device since K030573, do not raise different questions of safety or effectiveness because they do not alter the device's diagnostic/therapeutic purpose, blood collection functionality, mechanism of action, or method of use. Performance testing confirms that the subject device performs as intended and is as safe and effective as the predicate device. Based on information provided in this submission the proposed device is substantially equivalent to the predicate device.
<sup>3</sup> Merrill, V.D., Ward, M.D., Diaz-McNair, J., Picket, E.A., Duh, S.H., Christenson, R.H., Assessing Phlebotomy device preference and specimen quality in an oncology outpatient clinic. J Appl Lab Med 2021; ifab109, https://doi.org/10.1093/jalm/ffab109.
<sup>6</sup> Lynn K. (2011). Challenges of the oncology draw. MLO: medical laboratory observer, 43(1), 22.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.