K220207 · Zhejiang Innuovo Rehabilitation Devices Co.,Ltd · INI · Mar 24, 2022 · Physical Medicine
Device Facts
Record ID
K220207
Device Name
Mobility Scooter, Model W3468
Applicant
Zhejiang Innuovo Rehabilitation Devices Co.,Ltd
Product Code
INI · Physical Medicine
Decision Date
Mar 24, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
Device Story
Mobility Scooter (Model W3468) is a motor-driven, three-wheeled transportation vehicle for disabled or elderly users. Device features steel frame, seat, tiller console, electromagnetic brake, and two rechargeable lead-acid batteries. Rider controls movement via throttle lever, speed control dial, and handle on tiller console. Electromagnetic brake engages automatically when not in use. Intended for indoor and restricted outdoor use on flat, paved surfaces. Provides mobility benefit to users with limited walking ability. No clinical data included; safety and performance verified via bench testing against ISO 7176 series standards.
Clinical Evidence
No clinical data. Bench testing only. Compliance with ISO 7176 series (stability, brakes, energy consumption, dimensions, fatigue, climatic, obstacle-climbing, EMC, batteries) and ISO 10993 (biocompatibility) verified design specifications.
Technological Characteristics
Steel frame; electromagnetic braking; 250W motor; 2x 12V 22Ah lead-acid batteries; 24V/2A off-board charger. Dimensions: 1075mm x 510mm x 930mm. Connectivity: USB port (power transfer only). Standards: ISO 7176 series (performance, stability, EMC, batteries), ISO 10993-5/10 (biocompatibility).
Indications for Use
Indicated for disabled or elderly persons limited to a seated position requiring indoor and outdoor mobility assistance.
Regulatory Classification
Identification
A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is a blue square with the letters FDA in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 24, 2022
Zhejiang Innuovo Rehabilitation Devices Co., Ltd % Jarvis Wu Consultant Shanghai Sungo Management Consulting Company Limited 14th Floor, 1500# Central Avenue Shanghai, Shanghai 200122 China
Re: K220207
Trade/Device Name: Mobility Scooter (Model: W3468) Regulation Number: 21 CFR 890.3800 Regulation Name: Motorized Three-Wheeled Vehicle Regulatory Class: Class II Product Code: INI Dated: January 25, 2022 Received: January 25, 2022
Dear Jarvis Wu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Heather L. Dean -S Heather Dean, PhD
Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220207
Device Name Mobility Scooter (Model: W3468)
Indications for Use (Describe)
It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary
## Document prepared date: 2022/1/21
#### Applicant: A.
Zhejiang Innuovo Rehabilitation Devices Co.,Ltd Address: No.196 Industry Road, Hengdian Movie Zone, Dongyang, Zhejjang.China Contact Person: Leo Zheng Tel: + 86 18358936043
Submission Correspondent: Primary contact: Mr. Jarvis Wu Shanghai SUNGO Management Consulting Co., Ltd. 14th Floor, Dongfang Building, 1500# Central Ave., Shanghai 200122, China Tel: +86-21-58817802 Email: haiyu.wang@sungoglobal.com Secondary contact: Mr. Raymond Luo 14th Floor, Dongfang Building, 1500# Central Ave., Shanghai 200122, China Tel: +86-21-68828050 Email: fda.sungo@@gmail.com
## B. Device:
Trade Name: Mobility Scooter Common Name: Scooter Model: W3468
Regulatory Information
Classification Name: Motorized three-wheeled vehicle
Classification: Class II.
Product code: INI
Regulation Number: 890.3800
Review Panel: Physical Medicine
### C. Predicate device:
510Knumber: K182471 Device Name: Scooter Model: KPL001 Tianjin Kepler Vehicle Industry Co. Ltd.
#### D. Indications for use of the device :
{4}------------------------------------------------
It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
## E. Device Description:
The Mobility Scooter, Models W3468, has a base with Steal frame, two front wheels, a seat, a tiller console, electromagnetic brake, front light, USB charging port (Transfer power only), 2 rechargeable Lead-acid Batteries with an off-board charger. The movement of the scooter is controlled by the rider who operates the throttle lever, speed control dial and handle on the tiller console. The device is installed with an electromagnetic brake that will engage automatically when the scooter is not in use and the brake cannot be used manually. The Scooter only can be operated on the flat road.
## F. Non-Clinical TestConclusion
Non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- ISO 10993-5: 2009 Biological Evaluation of Medical Devices -- Part 5: Tests For In Vitro > Cytotoxicity
- ISO 10993-10: 2010 Biological Evaluation of Medical Devices Part 10: Tests For Irritation And A Skin Sensitization
- ISO 7176-1: 2014, Wheelchairs Part 1: Determination of static stability >
- A ISO 7176-2:2017, Wheelchairs - Part 2: Determination of dynamic stability of Powered Wheelchairs
- ISO 7176-3: 2012, Wheelchairs Part 3: Determination of effectiveness of brakes A
- > ISO 7176-4, Third edition 2008-10-01, Wheelchairs - Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range
- ISO 7176-5, Second edition 2008-06-01, Wheelchairs Part 5: Determination of overall dimensions, A mass and manoeuvring space
- ISO 7176-6: 2018. Wheelchairs Part 6: Determination of maximum speed, acceleration and A deceleration of Powered Wheelchairs
- ISO 7176-7. Wheelchairs Part 7: Measurement of seating and wheel dimensions >
- ISO 7176-8:2014. Wheelchairs Part 8: Requirements and test methods for static, impact and > fatigue strengths
- ISO 7176-9:2009, Wheelchairs Part 9: Climatic tests for Powered Wheelchairs A
- ISO 7176-10:2008. Wheelchairs Part 10: Determination of obstacle-climbing ability of electrically A powered wheelchairs
- ISO 7176-11:2012 Wheelchairs Part 11: Test dummies. A
- A ISO 7176-13, First edition 1989-08-01, Wheelchairs - Part 13: Determination of coefficient of friction of test surfaces
- A ISO 7176-14:2008, Wheelchairs - Part 14: Power and control systems for electrically powered wheelchairs and scooters - Requirements and test methods
- A ISO 7176-15:1996, Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling
- > ISO 7176-16:2012 Wheelchairs - Part 16: Resistance to ignition of postural support devices.
- A ISO 7176-21:2009 Wheelchairs - Part 21: Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and scooters, and battery chargers
- > ISO 7176-25:2013 Wheelchairs - Part 25: Batteries and chargers for powered wheelchairs
# G. Clinical Test Conclusion
{5}------------------------------------------------
No clinical study is included in this submission.
| Elements of<br>Comparison | Subject Device(s) | Predicate Device (K182471) | Remark |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|
| Manufacturer | Zhejiang Innuovo Rehabilitation<br>Devices Co.,Ltd | Tianjin Kepler Vehicle Industry Co.<br>Ltd. | -- |
| Common or<br>Usual name | Scooter | Scooter | Same |
| Model(s) | W3468 | KPL001 | -- |
| Indications<br>for use | It is a motor driven, indoor and<br>outdoor transportation vehicle with the<br>intended use to provide mobility to a<br>disabled or elderly person limited to a<br>seated position. | It is a motor driven, indoor and<br>outdoor transportation vehicle with the<br>intended use to provide mobility to a<br>disabled or elderly person limited to a<br>seated position. | S.E. |
| Overall<br>dimension | 1075 mm x 510 mm x 930 mm | 1030mm/40.6" x 630mm/24.8" x<br>920mm/36.2" | Similar |
| Frame<br>Material | Steel | Steel | S.E. |
| Frame style | Foldable seat, removable battery pack,<br>disassemble for transport | Foldable seat, removable battery pack,<br>disassemble for transport | S.E. |
| Front wheel<br>size | 225 x 65 mm | 190 x 54 mm | Analysis |
| Front Wheels<br>Quantity | 2 | 2 | S.E. |
| Rear wheel<br>size | 225 x 65 mm | 215 x 70 mm | Analysis |
| Rear Wheels<br>Quantity | 2 | 2 | S.E. |
| Ground<br>clearance | 85mm(3.3") | 45 mm(1.8") | Analysis |
| Max<br>Loading( on<br>level ground) | 160kg | 120 kg/265 lbs approx. | Analysis |
| Turn Radius | 1800mm / 70" | 1300 mm(51.2") | Analysis |
| Motor output | 250W | 24 V 180W | Analysis |
| Drive System | Rear Wheel Drive | Rear Wheel Drive | S.E. |
| Brakes | Electromagnetic brake | Electromagnetic brake | S.E. |
| Battery | Lead Acid 12V x 2 (22Ah) | Lead-acid 24V/12AH | S.E. |
| Charger | 24V/2A | 24V/2A | S.E. |
| Max Speed | 6.4 km/h | 6.6 km/h | S.E. |
| Slope Grade<br>Ability | 9 degree | 0-12 degree | Analysis |
| Travel<br>Distance | 20km / 14 mile (22Ah) | 15 km/9.32 Miles | Analysis |
| Arm Rests<br>(Distance<br>between<br>armrests) | 45 cm | No available data | N/A |
| Controller | PG45A | PG45A | S.E. |
| Time to<br>brake | 0.7-1s | < 1 s | S.E. |
| Brake<br>Distance-<br>Normal<br>operation<br>(Horizontal-<br>Forwa rd-<br>Max speed) | $≤$ 1.5m | 1m | S.E. |
| Battery<br>weight | 14kg (31lb) | 10.2 kg/22.5 lbs | Analysis |
| Base weight<br>(not<br>including<br>battery) | 58kg | 50 kg/110 lbs | Analysis |
| Operating<br>surface &<br>environment | Indoor use and restricted outdoor use<br>on pavements or paved footpaths only. | Indoor use and restricted outdoor use<br>on pavements or paved footpaths only. | S.E. |
| Remote<br>control | None | None | S.E. |
| Item | Proposed Device | Predicate Devices | Results |
| Biocompatibility | All user directly contacting<br>materials are compliance with<br>ISO10993-5 and<br>ISO10993-10 requirements. | All user directly contacting<br>materials are compliance with<br>ISO10993-5 and<br>ISO10993-10 requirements. | S.E. |
| EMC | ISO7176-21 | ISO7176-21 | S.E. |
| Performance | ISO7176 series | ISO7176 series | S.E. |
| Label and labeling | Conforms to FDA Regulatory | Conforms to FDA Regulatory | S.E. |
# H. Comparison with predicate Device Table 1 General Comparison
{6}------------------------------------------------
# I. Difference analysis
The design and technological characteristics of the Electric Scooter is similar to the predicates chosen. There are minor differences between the devices including Front & Rear wheel size, Ground clearance, Max Loading(on level ground), Turn Radius, Motor output, Slope Grade Ability, Travel Distance, Battery weight and Base weight (not including battery). All of the parameter with difference have been tested according to ISO7176 series standards and the test records support its safety and effectiveness. There is no deleterious effect on safety and effectiveness due to the minor differences do not influence the intended use of the device. Therefore, the proposed Scooter is substantially
{7}------------------------------------------------
equivalent (SE) to The Scooter (K182471).
# Table 2 Safety comparison
# Table 3 Safety comparison
| Item | Proposed Device | Predicate Devices | Results |
|-------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| ISO7176-1 | The Static stability has been<br>determined after the testing according<br>to the ISO 7176-1, and test results<br>meet its design specification. | The Static stability has been<br>determined after the testing according<br>to the ISO 7176-1, and test results meet<br>its design specification. | S.E. |
| ISO7176-2 | The dynamic stability has been<br>determined after the testing according<br>to the ISO 7176-2, and test results<br>meet its design specification. | The dynamic stability has been<br>determined after the testing according<br>to the ISO 7176-2, and test results meet<br>its design specification. | S.E. |
| ISO7176-3 | The effectiveness of brakes has been<br>determined after the testing according<br>to the ISO 7176-3, and test results<br>meet its design specification. | The effectiveness of brakes has been<br>determined after the testing according<br>to the ISO 7176-3, and test results meet<br>its design specification. | S.E. |
| ISO7176-4 | The theoretical distance range has<br>been determined after the testing<br>according to the ISO 7176-4, and test<br>results meet its design specification. | The theoretical distance range has<br>been determined after the testing<br>according to the ISO 7176-4, and test<br>results meet its design specification. | S.E. |
| ISO7176-5 | The dimensions, mass has been<br>determined after the testing according<br>to the ISO 7176-5, | The dimensions, mass has been<br>determined after the testing according<br>to the ISO 7176-5, | S.E. |
| ISO7176-6 | The dimensions, mass has been<br>determined after the testing according<br>to the ISO 7176-5, | The dimensions, mass has been<br>determined after the testing according<br>to the ISO 7176-5, | S.E. |
| ISO7176-7 | The seating and wheel dimensions<br>has been determined after the testing<br>according to the<br>ISO 7176-7, | The seating and wheel dimensions<br>has been determined after the testing<br>according to the<br>ISO 7176-7, | S.E. |
| ISO7176-8 | All test results meet the<br>requirements in Clause 4 of ISO<br>7176-8 | All test results meet the<br>requirements in Clause 4 of ISO<br>7176-8 | S.E. |
| | | S.E. | |
| ISO7176-9 | The test results shown that the device<br>under tests could continue to<br>function according to manufacturer's<br>specification after being subjected to<br>each of the tests specified in Clause 8<br>of ISO 7176-9 | The test results shown that the device<br>under tests could continue to function<br>according to manufacturer's<br>specification after being subjected to<br>each of the tests specified in Clause 8<br>of ISO 7176-9 | S.E. |
| ISO7176-10 | The obstacle-climbing ability of<br>device has been determined after the<br>testing according to the ISO 7176-<br>10, | The obstacle-climbing ability of<br>device has been determined after the<br>testing according to the ISO 7176-<br>10, | S.E. |
| ISO7176-11 | The test dummies used in the testing<br>of ISO 7176 series are meet the<br>requirements of ISO 7176-11 | The test dummies used in the testing<br>of ISO 7176 series are meet the<br>requirements of ISO 7176-11 | S.E. |
| ISO7176-13 | The coefficient of friction of test<br>surfaces has<br>been determined, which could be used<br>in other 7176 series tests involved | The coefficient of friction of test<br>surfaces has<br>been determined, which could be used<br>in other 7176 series tests involved | S.E. |
| ISO7176-14 | All test results meet the<br>requirements in Clause 7, 8, 9, 10,<br>11, 12, 13, 14, 15, 17 of ISO 7176-<br>14 | All test results meet the<br>requirements in Clause 7, 8, 9, 10,<br>11, 12, 13, 14, 15, 17 of ISO 7176-<br>14 | S.E. |
| ISO7176-15 | The test results shown that<br>information disclosure,<br>documentation and labelling of<br>device meet the requirements of<br>ISO 7176-15 | The test results shown that<br>information disclosure,<br>documentation and labelling of<br>device meet the requirements of ISO<br>7176-15 | S.E. |
| ISO7176-16 | The performance of resistance to<br>ignition meet the requirements of<br>ISO 7176-16 | The performance of resistance to<br>ignition meet the requirements of<br>ISO 7176-16 | S.E. |
| ISO 7176-21 | The EMC performance results meet<br>the requirements of ISO 7176-21 | The EMC performance results meet the<br>requirements of ISO 7176-21 | S.E. |
| ISO7176-25 | The performance of batteries and<br>charger of device meet the<br>Requirements in Clause 5 and 6 of<br>ISO 7176-25 | The performance of batteries and<br>charger of device meet the<br>Requirements in Clause 5 and 6 of ISO<br>7176-25 | S.E. |
{8}------------------------------------------------
Zhejiang Innuovo Rehabilitation Devices Co.,Ltd No.196 Industry Road, Hengdian Movie Zone, Dongyang, Zhejiang,China
### J. Conclusion
The conclusion drawn from the nonclinical tests demonstrates that the subject device in 510(K) submission, the Mobility scooter, model W3468, is as safe, as effective, and performs as well as the legally marketed predicate device cleared under K182471.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.