K220199 · Techfit Digital Surgery, Inc. · HRS · Mar 21, 2022 · Orthopedic
Device Facts
Record ID
K220199
Device Name
AFFINITY Proximal Tibia System
Applicant
Techfit Digital Surgery, Inc.
Product Code
HRS · Orthopedic
Decision Date
Mar 21, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
AFFINITY Proximal Tibia System is intended to treat fractures, nonunions of the proximal tibia including simple, comminuted, lateral wedge, depression, medial wedge, bicondylar combination of lateral wedge and depression, periprosthetic, and fractures with associated shaft fractures. - Simple metaphyseal fractures (Classification AO 41-A2) - Multifragmentary metaphyseal fractures (Classification AO 41-A3) - Simple bicondylar fractures (Classification AO41-C1, 41-C2) - Multifragmentary bicondylar fractures (Classification AO 41-C3) - Simple joint, simple metaphyseal fractures (Classification AO 41-C1) - Diaphisary fractures (Classification AO 42A and 42B)
Device Story
AFFINITY Proximal Tibia System consists of pre-contoured metallic bone fixation plates and screws. Used by orthopedic surgeons to stabilize proximal tibial fractures and nonunions. Plates are anatomically shaped to restore original bone structure; variable angle screw technique allows flexible screw configuration for fracture support. System provides mechanical fixation to facilitate bone healing. Device is used in clinical/surgical settings.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including static/fatigue four-point bend testing (ASTM F382), torsion/insertion/pullout strength testing (ASTM F543), fretting corrosion testing (ASTM F897), and biocompatibility assessment (ISO 10993-1).
Technological Characteristics
Plates: Commercially pure titanium grade 4 (ISO 5832-2, ASTM F67). Screws: Titanium alloy Ti6Al4V (ISO 5832-3, ASTM F136). Variable angle locking technology. Sterilization: Validated per ISO 17665-1, AAMI TIR12, ISO TIR17665-2, and ISO 14937 to SAL 10^-6.
Indications for Use
Indicated for patients with proximal tibia fractures and nonunions, including simple, comminuted, lateral/medial wedge, depression, bicondylar, periprosthetic, and associated shaft fractures (AO classifications 41-A2, 41-A3, 41-C1, 41-C2, 41-C3, 42A, 42B).
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
aap LOQTEQ® Proximal Tibia Plate 3.5 System (K132554)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services - USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in two lines, with "FDA U.S. FOOD & DRUG" on the top line and "ADMINISTRATION" on the bottom line. The logo is clean and professional, reflecting the organization's role in regulating food and drugs.
March 21, 2022
Techfit Digital Surgery INC. Leidy Johanna Toro-Gonzalez Regulatory Affairs Specialist 1511 Aviation Center Pkwy. Suite 220H Daytona Beach, Florida 32114
Re: K220199
Trade/Device Name: AFFINITY Proximal Tibia System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: January 19, 2022 Received: January 24, 2022
Dear Leidy Johanna Toro-Gonzalez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220199
Device Name AFFINITY Proximal Tibia System
#### Indications for Use (Describe)
AFFINITY Proximal Tibia System is intended to treat fractures, nonunions of the proximal tibia including simple, comminuted, lateral wedge, depression, medial wedge, bicondylar combination of lateral wedge and depression, periprosthetic, and fractures with associated shaft fractures.
- · Simple metaphyseal fractures (Classification AO 41-A2)
- · Multifragmentary metaphyseal fractures (Classification AO 41-A3)
- · Simple bicondylar fractures (Classification AO41-C1, 41-C2)
- Multifragmentary bicondylar fractures (Classification AO 41-C3)
- · Simple joint, simple metaphyseal fractures (Classification AO 41-C1)
- · Diaphisary fractures (Classification AO 42A and 42B)
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K220199 510(k) Summary 21 CFR 807.92
# Submitter information
| Company name | Techfit Digital Surgery INC. |
|-----------------------------------|---------------------------------------|
| Establishment registration number | 3016560308 |
| Street Address | 1511 Aviation Center Pkwy, Suite 220H |
| City | Daytona Beach, Florida |
| Zip code | FL 32114 |
| Country | United States |
| Phone number | (605) 517-0321 |
| Principal contact person | Leidy Johanna Toro-González |
| Contact title | Regulatory Affairs Specialist |
| Contact e-mail address | leidy.toro@imsampedro.com |
| Additional contact person | Liliana Zuluaga Idárraga |
| Contact title | Technical Director |
| Contact e-mail address | liliana.zuluaga@imsampedro.com |
#### Submission date
The submission date of this Traditional 510(k) submission is the 19th of January 2022.
## Submission information
| Trade name | AFFINITY Proximal Tibia System |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common or Usual name | Plate, Fixation, Bone |
| Classification name | 21 CFR 888.3030 Single/multiple component metallic bone fixation<br>appliances and accessories (Primary)<br>21 CFR 888.3040 Smooth or threaded metallic bone fixation fastener |
| Product code (classification regulation) | HRS/HWC |
| Classification Panel | Orthopedic |
| Device class | Class II |
## Primary Predicate device
The predicate device to which substantial equivalence is claimed to:
| Predicate device: | |
|------------------------------------|---------------------------------------------|
| Trade or proprietary or model name | aap LOQTEQ® Proximal Tibia Plate 3.5 System |
| 510(k) number | K132554 |
| Decision date | 10/03/2013 |
| Product code | HRS |
| Manufacturer | AAP Implantate AG |
| Review Panel | Orthopedic |
## Additional Predicate device 1
| Trade or proprietary or model name | TDM Plate and Screw System |
|------------------------------------|----------------------------|
| 510(k) number | K171808 |
| Decision date | 03/15/2018 |
| Product code | HRS, HWC |
| Manufacturer | TDM Co. Ltd. |
| Review Panel | Orthopedic |
{4}------------------------------------------------
## Additional Predicate device 2
| Trade or proprietary or model name | AFFINITY – Variable Angle Distal Radius System |
|------------------------------------|------------------------------------------------|
| 510(k) number | K191641 |
| Decision date | 09/16/2019 |
| Product code | HRS, HWC |
| Manufacturer | Industrias Médicas Sampedro S.A.S |
| Review Panel | Orthopedic |
## Additional Predicate device 3
| Trade or proprietary or model name | TECHFIT Patient-Specific Maxillofacial System |
|------------------------------------|-----------------------------------------------|
| 510(k) number | K203282 |
| Decision date | 05/19/2021 |
| Product code | JEY |
| Manufacturer | Industrias Médicas Sampedro S.A.S |
| Review Panel | Dental |
#### Additional Predicate device 4
| Trade or proprietary or model name | SYNTHES 3.5mm Cortex Screws |
|------------------------------------|-----------------------------|
| 510(k) number | K043185 |
| Decision date | 02/03/2005 |
| Product code | HWC |
| Manufacturer | SYNTHES (USA) |
| Review Panel | Orthopedic |
#### Additional Predicate device 5
| Trade or proprietary or model name | AXSOS 3 TI Locking Plate System |
|------------------------------------|---------------------------------|
| 510(k) number | K123964 |
| Decision date | 03/28/2013 |
| Product code | HRS |
| Manufacturer | Stryker Trauma AG |
| Review Panel | Orthopedic |
## Device Description
AFFINITY Proximal Tibia System consists of anatomical plates and screws for the placement of the proximal tibial condyles, improving the restoration of the original structure. Similarly, in combination with the variable angle technique, it allows for the placement of screws in different configurations providing appropriate support for the correct healing of fractures.
AFFINITY Proximal Tibia System consists of pre-contoured bone fixation plates and screws. The plates are made from biocompatible commercially pure titanium grade 4 according to ISO 5832-2 and ASTM F67 standard. The screws are made from biocompatible titanium alloy (Ti6Al4V) according to ISO 5832-3 and ASTM F136 standard.
The AFFINITY Proximal Tibia System plates can be fixed with variable angle technique.
{5}------------------------------------------------
## Indications for use
AFFINITY Proximal Tibia System is intended to treat fractures, nonunions of the proximal tibia including simple, comminuted, lateral wedge, depression, medial wedge, bicondylar combination of lateral wedge and depression, periprosthetic, and fractures with associated shaft fractures.
- Simple metaphyseal fractures (Classification AO 41-A2) ●
- Multifragmentary metaphyseal fractures (Classification AO 41-A3)
- Simple bicondylar fractures (Classification A041-C1, 41-C2)
- Multifragmentary bicondylar fractures (Classification AO 41-C3) ●
- Simple joint, simple metaphyseal fractures (Classification AO 41-C1)
- . Diaphisary fractures (Classification AO 42A and 42B)
#### Comparison to the predicate devices
The predicate devices AFFINITY - Variable Angle Distal Radius System (K191641) and TECHFIT Patient-Specific Maxillofacial System (K203182) plates and screws are manufactured in the same materials and using similar manufacturing methods. Also, the AFFINITY - Variable Angle Distal Radius System (K191641) uses the same variable angle technology to subject device.
Both The AFFINITY – Variable Angle Distal Radius System (K191641) and TECHFIT Patient-Specific Maxillofacial System (K203182), were selected as predicate devices to support the fretting corrosion test, the biological tests, the packing test, and chemical characterization tests. The fretting corrosion test, the biological tests, and the chemical characterization tests do not depend on the system's shape but on the type of materials that enter in contact, manufacturing materials, and manufacturing methods.
The TDM Plate and Screw System (K171808) was selected as a predicate device solely for the side-by-side mechanical performance testing with the AFFINITY Proximal Tibia System, according to ASTM F382 "Standard Specification and Test Method for Metallic Bone Plates". Since the TDM Plate and Screw System has the same intended use and manufacturing material, and similar plate design to the subject device. The TDM Plate and Screw System plates dimensions are comparable to the subject device. Both plate thickness and length of this predicate device fall within the range of the subject device.
Predicate devices screws' design share the locking and cortical screws design feature with the subject device. Both The Synthes 3.5 mm Cortex Screws (K043185) and AXSOS 3 TI Locking Plate System (K123964) have longer screws than the subject device. The AFFINITY Proximal Tibia System screws diameters fall within the Synthes 3.5 mm Cortex Screws and AXSOS 3 TI Locking Plate System diameter range as also the lengths from 10mm to 95 mm. and these were selected as predicate devices solely for the side-by-side mechanical performance testing with the AFFINITY Proximal Tibia System, according to ASTM F543 "Standard Specification and Test Methods for Metallic Medical Bone Screws".
The TDM Plate and Screw System (K171808), AFFINITY - Variable Angle Distal Radius System (K191641), TECHFIT Patient-Specific Maxillofacial System (K203182), The Synthes 3.5 mm Cortex Screws (K043185), and AXSOS 3 TI Locking Plate System (K123964) are comparable to the AFFINITY Proximal Tibia System.
{6}------------------------------------------------
# Performance Testing – Non-clinical
- Static and fatique four-point bend testing. Static and fatigue four-point bend tests were performed following the Standard Specification and Test Method for Metallic Bone Plates, indicated in ASTM F382-17 Standard.
- Torsion strength, insertion torque, and pullout strength testing. Torsion strength, insertion torque, and pullout strength tests were performed following the Standard Specification and Test Methods for Metallic Medical Bone Screws, indicated in ASTM F543-17 Standard.
- Fretting corrosion testinq
The fretting corrosion testing was performed following the Standard Test Method for Measuring Fretting Corrosion, indicated in ASTM F897-02 Standard.
- Biocompatibility.
For the assessment of biological endpoints, the procedures and provisions of ISO 10993-1:2018 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process", as well as FDA Guidance "Use of International Standard ISO 10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk Management Process", dated 04 September 2020, were applied.
- . Sterilization.
Sterilization validation was conducted in accordance with international standard ISO 17665-1, AAMI TIR12, ISO TIR17665-2 and ISO 14937 to a sterility Assurance Level (SAL) of 10-°.
#### Performance Testing - Clinical
Clinical testing was not necessary for the substantial equivalence determination.
#### Substantial Equivalence Conclusion
Based on the indications for use, technological characteristics, performance testing, and comparison to predicate devices, the AFFINITY Proximal Tibia System shows to be substantially equivalent to the predicate devices.
- Mechanical testing was performed using FDA consensus standards and laboratory test methods and compared to the predicate device and scientific literature.
- Biocompatibility was demonstrated as per FDA consensus standards.
- Sterilization was demonstrated as per FDA consensus standards and all test method acceptance criteria were met.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.